N Latex Cystatin C; N Protein Standard UY

K171072 · Siemens Healthcare Diagnostics Products GmbH · NDY · May 12, 2017 · Clinical Chemistry

Device Facts

Record IDK171072
Device NameN Latex Cystatin C; N Protein Standard UY
ApplicantSiemens Healthcare Diagnostics Products GmbH
Product CodeNDY · Clinical Chemistry
Decision DateMay 12, 2017
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 862.1225
Device ClassClass 2

Intended Use

N Latex Cystatin C is an in vitro diagnostics kit containing reagents for the quantitative determination of cystatin C in human serum and lithium-heparinized plasma by means of particle-enhanced immunonephelometry using the BN Systems. Cystatin C measurements are used in the diagnosis and treatment of renal diseases. N Protein Standard UY is used for preparing reference curves for the immunonephelometric determination of α1-microglobulin and Cystatin C using the BN Systems.

Device Story

Device is an in vitro diagnostic kit for quantitative cystatin C measurement in human serum/plasma. Principle: particle-enhanced immunonephelometry. Polystyrene particles coated with anti-human cystatin C antibodies aggregate when mixed with patient samples; aggregates scatter light; scattered light intensity is proportional to cystatin C concentration. Used on BN II and BN ProSpec systems in clinical laboratories by trained technicians. Output is a concentration value (mg/L) compared against a reference curve established by N Protein Standard UY. Healthcare providers use results to assess renal function and guide treatment decisions. Benefits include standardized, accurate monitoring of renal disease status.

Clinical Evidence

Bench testing only. Precision studies (n=80 per sample) showed repeatability CVs 1.00-3.85% and within-lab CVs 1.01-3.85%. Linearity confirmed 0.27-10.3 mg/L; LoQ 0.08 mg/L. No high-dose hook effect up to 42.91 mg/L. Method comparison (n=186) against predicate showed high correlation (r=0.988-0.998) and Passing-Bablok regression slopes of 0.981-0.982. Reference interval study (n=203) established range 0.49-1.19 mg/L.

Technological Characteristics

Reagents: polystyrene particles coated with rabbit anti-human cystatin C antibodies. Measurement: particle-enhanced immunonephelometry. Energy: light scattering. Connectivity: BN II and BN ProSpec systems. Calibrator: N Protein Standard UY (lyophilized polygeline with human proteins). Traceability: ERM-DA471/IFCC. Software: embedded system control for nephelometric analysis.

Indications for Use

Indicated for quantitative determination of cystatin C in human serum and lithium-heparinized plasma to aid in the diagnosis and treatment of renal diseases in adult patients.

Regulatory Classification

Identification

A creatinine test system is a device intended to measure creatinine levels in plasma and urine. Creatinine measurements are used in the diagnosis and treatment of renal diseases, in monitoring renal dialysis, and as a calculation basis for measuring other urine analytes.

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ November 7, 2017 Image /page/0/Picture/1 description: The image shows the logos of the Department of Health and Human Services and the Food and Drug Administration (FDA). The Department of Health and Human Services logo is on the left, and the FDA logo is on the right. The FDA logo includes the letters "FDA" in a blue square, followed by the words "U.S. Food & Drug Administration" in blue text. Siemens Healthcare Diagnostics Products GmbH Christine Perkins Regulatory Specialist Emil-von-Behring-Str. 76 Marburg, 35041 De Re: k171072 Trade/Device Name: N Latex Cystatin C; N Protein Standard UY Regulation Number: 21 CFR 862.1225 Regulation Name: Creatinine test system Regulatory Class: Class II Product Code: NDY, JIX Dated: April 6, 2017 Received: April 10, 2017 Dear Christine Perkins: This letter corrects our substantially equivalent letter of May 12, 2017. We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801 and Part 809); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR {1}------------------------------------------------ Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and Part 809), please contact the Division of Industry and Consumer Education (DICE) at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevicesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education (DICE) at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Sincerely, # Kellie B. Kelm -S for Courtney H. Lias, Ph.D. Director Division of Chemistry and Toxicology Devices Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ # Indications for Use 510(k) Number (if known) K171072 #### Device Name N Latex Cystatin C; N Protein Standard UY Indications for Use (Describe) N Latex Cystatin C: N Latex Cystatin C is an in vitro diagnostics kit containing reagents for the quantitative determination of cystatin C in human serum and lithium-heparinized plasma by means of particle-enhanced immunonephelometry using the BN Systems. Cystatin C measurements are used in the diagnosis and treatment of renal diseases. N Protein Standard UY: N Protein Standard UY is used for preparing reference curves for the immunonephelometric determination of a1microglobulin and Cystatin C using the BN Systems. | Type of Use (Select one or both, as applicable) | |----------------------------------------------------------------------------------------------------------------------------------------------| | <span style="font-size: 10pt;"> <span style="font-family: Symbol;">☒</span> Prescription Use (Part 21 CFR 801 Subpart D) </span> | | <span style="font-size: 10pt;"> <span style="font-family: Symbol;">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) </span> | ### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. ### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ # 510(k) Summary per 21 CFR 807.92 This summary of 510(k) safety and effectiveness information is submitted in accordance with the requirements of 21 CFR 807.92 and follows the FDA guidance "The 510(k) Program: Evaluating Substantial Equivalence in Premarket Notifications [510(k)]", issued July 28, 2014. The assigned 510(k) number is: K171072 # 5.1 Submitter Siemens Healthcare Diagnostics Products GmbH Emil-von-Behring-Str. 76 35041 Marburg, Germany | Contact Person: | Christine Perkins | |----------------------|-------------------------------| | Email: | christine.perkins@siemens.com | | Phone: | 302-631-8811 | | Fax: | 302-631-6299 | | Date of Preparation: | May 11, 2017 | # 5.2 Device Information | Trade Name: | N Latex Cystatin C | | |-----------------------|-----------------------------------------|--| | Common or Usual Name: | Test, Cystatin C Creatinine test system | | | Classification Name: | Creatinine test system (21CFR 862.1225) | | | Regulatory Class: | Class II | | | Product Code: | NDY | | | 510(k) Review Panel: | Clinical Chemistry | | | Trade Name: | N Protein Standard UY | | | Common or Usual Name: | Calibrator, Multi-Analyte Calibrator | | | Classification Name: | Calibrator (21 CFR 862.1150) | | sification Name: Class II Regulatory Class: Product Code: الا 510(k) Review Panel: Clinical Chemistry # Calibrator (21 CFR 862.1150) # 5.3 Predicate Device: The predicate device is the Tina-quant Cystatin C Gen.2 Test System, for use on Roche/Hitachi cobas c systems cleared by the FDA under K141143. {4}------------------------------------------------ | Assay: | Tina-quant Cystatin C Gen.2 | |-------------|--------------------------------------------------------| | Calibrator: | C.f.a.s. (Calibrator for automated systems) Cystatin C | # 5.4 Device Description / Test Principle The N Latex Cystatin C is an in vitro diagnostics kit containing reagents for the quantitative determination of cystatin C in human serum and lithium heparinized plasma by means of particle-enhanced immunonephelometry using the BN II and BN ProSpec® Systems. Used in conjunction with other clinical and laboratory findings, N Latex Cystatin C measurements are used as an aid in the diagnosis and treatment of renal diseases. Used in conjunction with the assay reagents, N Protein Standard UY is for use in the establishment of reference curves for the determination of cystatin C on the BN II and BN ProSpec Systems. N Cystatin C Control 1 and 2 are included with N Latex Cystatin C kit and are for use as assayed accuracy controls and precision controls in the determination of cystatin C by immunonephelometry with the BN II and BN ProSpec Systems. The N Latex Cystatin C test system is based upon the principles of particle-enhanced immunonephelometry. Polystyrene particles coated with specific antibodies to human cystatin C are aggregated when mixed with samples containing human cystatin C. These aggregates scatter a beam of light passed through the sample. The intensity of the scattered light is proportional to the concentration of the respective protein in the sample. The result is evaluated by comparison with a standard of known concentration. The N Latex Cystatin C subject of this 510(k) is materially the same as the assay cleared in K041878. The reason for the submission is for a new standardization to ERM-DA471/IFCC. This is being accomplished solely by changing the calibrator value assignment process. # 5.5 Intended Use / Indications for Use # N Latex Cystatin C assay N Latex Cystatin C is an in vitro diagnostics kit containing reagents for the quantitative determination of cystatin C in human serum and lithium-heparinized plasma by means of particle-enhanced immunonephelometry using the BN Systems. Cystatin C measurements are used in the diagnosis and treatment of renal diseases. # N Protein Standard UY N Protein Standard UY is used for preparing reference curves for the immunonephelometric determination of α1-microglobulin and Cystatin C using the BN Systems. {5}------------------------------------------------ # Comparison of Technological Characteristics | | | Predicate Device: | |---------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Attributes | Proposed Device: | Roche Tina-quant Cystatin | | | N Latex Cystatin C assay | C Gen.2 | | | | (K141143) | | Intended Use / Indications<br>for Use | N Latex Cystatin C is an in<br>vitro diagnostics kit containing<br>reagents for the quantitative<br>determination of cystatin C in<br>human serum and lithium-<br>heparinized plasma by means | Tina-quant Cystatin C Gen. 2<br>is an in vitro test for the<br>quantitative determination of<br>cystatin C in human serum<br>and plasma on Roche/Hitachi<br>cobas c systems. | | | of particle-enhanced<br>immunonephelometry using<br>the BN Systems.<br>Cystatin C measurements are<br>used in the diagnosis and | Cystatin C measurements are<br>used as an aid in the<br>diagnosis and treatment of<br>renal diseases. | | | treatment of renal diseases. | | | Analyte | cystatin C | Same | | Assay type/ Technology | Immunonephelometry | Turbidimetry | | Sample Type | Human serum and plasma | Human serum and plasma | | Permissible | Li-Heparin | Li-Heparin, | | anticoagulants | | K2-EDTA, K3-EDTA | | Composition | N Latex Cystatin C consists of<br>a suspension of polystyrene<br>particles coated with<br>approximately 0.016 g/L anti-<br>human cystatin C from rabbits | Reagent 2 is latex particles in<br>qlycine buffer coated with anti-<br>Cystatin C antibodies (rabbit);<br>preservatives and stabilizers. | | Ancillary Reagents | N Cystatin C Supplementary<br>Reagent A contains rabbit<br>immunoglobulin (14 g/L) in<br>buffered solution.<br>N Cystatin C Supplementary<br>Reagent B consists of an<br>aqueous solution of<br>polyethylene glycol sorbitan<br>monolaureate (85 g/L) and<br>polyethylene glycol ether (27<br>g/L). | R1 contains solution of<br>polymers in MOPS-buffered<br>saline; preservative,<br>stabilizers. | | Units | mg/L | Same | | Analytical measuring | 0.27 - 9.4 mg/L | 0.40 - 6.80 mg/L | | range | | 0.40 - 13.6 mg/L (1:2 dilution) | | (Calibrator lot dependent) | | | | Reference Interval | 0.49 - 1.19 mg/L<br>Age 21-72 | Age 21 - 77 0.61 - 0.95 mg/L | {6}------------------------------------------------ | | Proposed Device:<br>N Protein Standard UY | Predicate Device:<br>C.f.a.s. Cystatin C<br>(K141143) | |------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------| | Intended Use | N Protein Standard UY is<br>used for preparing reference<br>curves for the<br>immunonephelometric<br>determination of α1-<br>microglobulin and Cystatin C<br>using the BN Systems. | Calibrator for automated<br>systems Cystatin C is for use<br>in the calibration of<br>quantitative Roche methods<br>on Roche clinical chemistry<br>analyzers. | | Matrix | N Protein Standard UY<br>consists of lyophilized<br>polygeline with urinary<br>proteins of human origin.<br>Contains sodium azide (<1<br>g/L) as a preservative. | C.f.a.s. Cystatin C is based on<br>pooled delipidated human<br>serum enriched with<br>recombinant human Cystatin<br>C produced in E. Coli. | | How supplied | Lyophilized | Liquid | | Analyte Detected | cystatin C,<br>α1-microglobulin* | cystatin C | | Traceability | ERM-DA417/IFCC | Same | *N Protein Standard UY α1-microglobulin analyte was initially cleared in Κ991075 and has not been modified (in terms of composition, value assignment and stability) for the α1-microglobulin analyte since. The differences between the predicate devices and proposed reagents, calibrators and controls do not result in a change to the intended use, the indications for use, or to safety and efficacy when used according to the product labeling. # 5.7 Performance Data Analytical performance studies of N Latex Cystatin C were performed at Siemens Healthcare Diagnostics Products, Marburg Germany on a BN II and a BN ProSpec analvzer. Matrix comparison, interference, specificity and stability claims were not impacted by the new standardization. The claims are unchanged since clearance (K041878). The following performance data are provided in support of a substantial equivalence determination. # 5.7.1 Analytical Performance # 5.7.1.1 Precision Confirmation of precision studies was performed internally on one BN II and one BN ProSpec instrument for twenty days following the scheme of two runs per day, with two {7}------------------------------------------------ replicates per run according to CLSI EP05-A2, Evaluation of Precision Performance of Quantitative Measurement Methods; Approved Guideline-Second Edition. The order of the analysis of parameter samples and quality control samples, between morning and afternoon runs, was varied so as not to add an inherent bias to the study. Precision samples included serum and plasma pools and N Cystatin C Control 1 and 2, spanning the measuring range with sample pools close to the medical decision point (MDP) of 1.1 mg/L. | BN ProSpec Anova | | | | | | | | |------------------|------|----|------------|---------------|------------|--|--| | ID | Unit | n | Mean value | Repeatability | Within- | | | | | | | | CV | device/lab | | | | | | | | [%] | CV | | | | | | | | | 1% | | | | CysC Control 1 | mg/L | 80 | 1.155 | 1.78 | 1.78 | | | | CysC Control 2 | mg/L | 80 | 2.176 | 1.00 | 1.01 | | | | Serum Pool 1 | mg/L | 80 | 0.465 | 3.85 | 3.85 | | | | Serum Pool 2 | mg/L | 80 | 1.142 | 1.39 | 1.41 | | | | Serum Pool 3 | mg/L | 80 | 8.425 | 1.62 | 1.62 | | | | Plasma Pool 1 | mg/L | 80 | 0.481 | 2.38 | 2.65 | | | | Plasma Pool 2 | mg/L | 80 | 1.129 | 2.09 | 2.09 | | | | Plasma Pool 3 | mg/L | 80 | 8.863 | 1.07 | 1.26 | | | | BN II Anova | | | | | | |----------------|------|----|------------|----------------------------|------------------------------------| | ID | Unit | n | Mean value | Repeatability<br>CV<br>[%] | Within-<br>device/lab<br>CV<br>[%] | | CysC Control 1 | mg/L | 80 | 1.070 | 1.93 | 2.24 | | CysC Control 2 | mg/L | 80 | 2.110 | 1.42 | 1.54 | | Serum Pool 1 | mg/L | 80 | 0.445 | 2.54 | 2.62 | | Serum Pool 2 | mg/L | 80 | 1.078 | 1.34 | 1.55 | | Serum Pool 3 | mg/L | 80 | 8.636 | 1.69 | 1.80 | | Plasma Pool 1 | mg/L | 80 | 0.451 | 2.59 | 2.59 | | Plasma Pool 2 | mg/L | 80 | 1.069 | 1.62 | 1.75 | | Plasma Pool 3 | mg/L | 80 | 8.798 | 1.98 | 2.42 | {8}------------------------------------------------ # 5.7.1.2 Measuring Range (Linearity and LoQ) A linearity study was performed according to CLSI EP06-A, Evaluation of the Linearity of Quantitative Measurement Procedures: A Statistical Approach. One N Latex Cystatin C reagent lot, one BN II and one BN ProSpec instrument were used. A concentrated Cystatin C stock solution was used to spike pooled human serum to a Cystatin C concentration between 10 ± 0.5 mg/L which is just above the upper limit of the measuring interval of 9.4 mg/L. This high pool was serially diluted with human serum containing undetectable levels of Cystatin C down to the minimum measuring range. Each dilution was tested in replicates of five using an individual aliquot. The N Latex Cystatin C assay demonstrates acceptable linearity over the range of 0.27 to 10.3 mg/L. Limit of Quantitation (LoQ) testing was performed according to CLSI Guideline EP17-A2: Evaluation of Detection Capability for Clinical Laboratory Measurement Procedures. LoQ was demonstrated to be 0.08 mg/L at < 8% CV. The measuring range of the N Latex Cystatin C assay is 0.27 to 9.4 mg/L. # 5.7.1.3 High Dose Hook Effect (Antigen Excess) A serum sample was spiked with purified cystatin C to obtain a sample with a calculated concentration higher than 28 mg/L. A series of dilutions was prepared using cystatin C deficient serum. Each dilution was run five times. The analyte concentration of the high sample was determined using the mean of at least two manual dilutions that recovered within the measuring range of the assay. The results of the dilutions are compared to the upper limit of the calibration curve. Based on the results, Cystatin C concentrations up to 42.91 mg/L did not exhibit any high dose hook or prozone effects. # 5.7.2 Clinical Studies # 5.7.2.1 Reference Interval Study A reference interval was performed according to CLSI EP28-A3C, ' Defining, Establishing, and Verifying Reference Intervals in the Clinical Laboratory'. Two hundred three (203) serum samples from a US population of apparently healthy adults or patients with diseases not linked to renal disease or failure, between the ages of 21 and 72 and evenly distributed across gender were measured. The data analysis within the 2.5% and 97.5% percentile resulted in a range of Cystatin C from 0.49 to 1.19 mg/L. {9}------------------------------------------------ The sponsor recommends that each facility should determine its own reference intervals since values may vary depending on the population studied. # 5.7.2.2 Method comparison A method comparison study designed according to CLSI EP09-A3: Measurement Procedure Comparison and Bias Estimation Using Patient Samples: Approved Guideline-Third Edition was conducted at one external site in the United States. Serum samples (186) were measured on both the predicate device. Roche Tina-quant Cystatin C Gen.2 assay as well as on the N Latex Cystatin C assay. Results were compared by Passing-Bablok regression analysis. Results met the pre-established acceptance criteria for slope, intercept and Pearson correlation coefficient (r). The following summary of Passing-Bablok regression analysis proves substantial equivalence between the N Latex Cystatin C assay and the predicate device. | Method<br>Comparison<br>(Devices) | Correlation-<br>coefficient<br>(r) | Slope | Intercept | |---------------------------------------------------------------------------------------------|------------------------------------|-------|----------------| | N Latex Cystatin C on BN<br>ProSpec<br>vs.<br>Tina-quant Cystatin C<br>Gen.2 on Roche cobas | 0.988 | 0.982 | -0.078<br>mg/L | | N Latex Cystatin C on BN II<br>vs.<br>Tina-quant Cystatin C<br>Gen.2<br>on Roche cobas | 0.998 | 0.981 | -0.079<br>mg/L | # 5.8 Traceability The calibration of the assay is traceable to the IFCC European Reference Material ERM-DA471/IFCC, certified for cystatin C measurements. # 5.8.1 Calibrator Traceability N Protein Standard UY is traceable to Siemens internal Master Calibrator which is directly traceable to ERM-DA471/IFCC. # 5.9 Conclusion The N Latex Cystatin C assay including reagents and calibrator and the Roche Tinaquant Cystatin C Gen.2 assay and calibrator are substantially equivalent in principle and {10}------------------------------------------------ performance based on intended use and indications for use and supported by the performance testing described in the 510(k) submission.
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...