K171055 · Philips Electronics Nederland B.V. · ONE · Jul 6, 2017 · General, Plastic Surgery
Device Facts
Record ID
K171055
Device Name
Philips BlueControl
Applicant
Philips Electronics Nederland B.V.
Product Code
ONE · General, Plastic Surgery
Decision Date
Jul 6, 2017
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4810
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Philips BlueControl is intended to emit energy in the spectrum, is generally indicated to treat dermatological conditions and specifically indicated to treat mild psoriasis vulgaris.
Device Story
Philips BlueControl is a rechargeable, battery-operated, wearable device for home use by patients to treat mild psoriasis vulgaris. It delivers blue light (452 nm) to the skin via 40 LEDs. The device includes a fixation strap for positioning on the treatment area. Operation involves an on/off button and an infrared sensor that detects skin presence and monitors skin temperature to ensure safety. The device provides visual (LED colors) and audio (beeps) feedback regarding treatment status, battery level, and overheating (auto-shutoff at 51°C or skin >43°C). Treatment consists of a 30-minute session per area, with a regimen of daily treatments for 4 weeks followed by 3 times weekly for 8 weeks. The device benefits patients by providing a wearable, home-based therapeutic option for managing psoriasis symptoms, with clinical evidence showing significant improvement in plaque severity (LPASI scores) compared to control.
Clinical Evidence
Two clinical studies (N=50 total) conducted per ISO 14155. One study (n=23) evaluated blue light treatment (30 min/day) for 12 weeks. Primary endpoint: change from baseline (CfB) of Local Psoriasis Area and Severity Index (LPASI) at week 12. Results showed significant improvement in treated plaques vs. control (ΔCfB: -0.92 ± 1.10, p=0.0005). Mean change from baseline to week 12 was -2.38 in treated plaques vs -1.46 in control. No serious device-related adverse events reported; expected hyperpigmentation observed.
Indicated for adult patients with mild psoriasis vulgaris for treatment of dermatological conditions.
Regulatory Classification
Identification
(1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.
{1}------------------------------------------------
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
July 6, 2017
Philips Electronics Nederland B.V. Ms. Gaozhen Hang Senior Regulatory Affairs Specialist High Tech Campus 5 Eindhoven, 5656 AE The Netherlands
Re: K171055
Trade/Device Name: Philips BlueControl Regulation Number: 21 CFR 878.4810 Regulation Name: Laser Surgical Instrument for Use in General and Plastic Surgery and in Dermatology Regulatory Class: Class II Product Code: ONE Dated: April 7, 2017 Received: April 10, 2017
Dear Ms. Hang:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Actinclude requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device comples with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply
Image /page/0/Picture/11 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo is circular and contains the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. In the center of the circle is a stylized image of three human profiles facing to the right, stacked on top of each other.
{2}------------------------------------------------
with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely.
# Jennifer R. Stevenson -S3
For Binita S. Ashar, M.D., M.B.A., F.A.C.S. Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
# Indications for Use
510(k) Number (if known)
K171055
Device Name Philips BlueControl
Indications for Use (Describe)
The Philips BlueControl is intended to emit energy in the spectrum, is generally indicated to treat dermatological conditions and specifically indicated to treat mild psoriasis vulgaris.
X Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
# 510(k) Summary
The summary of safety and effectiveness information is included in this submission in accordance with the Safe Medical Device Act of 1990 and 21 CFR §807.92.
### I. SUBMITTER
| Philips Electronics Nederland B.V. |
|--------------------------------------|
| High Tech Campus 5 |
| Eindhoven 5656 AE |
| The Netherlands |
| Gaozhen Hang |
| Senior Regulatory Affairs Specialist |
| (302) 358-7026 |
| gaozhen.hang@philips.com |
| |
Date Prepared:April 7, 2017
### II. DEVICE
| Trade Name: | Philips BlueControl |
|-----------------------------|------------------------------------------------------|
| Common/Classification Name: | Powered Light Based Non-Laser Surgical<br>Instrument |
| Regulatory Class: | II |
| Product Code, CFR: | ONE |
| Review Panel: | General and Plastic Surgery (79) |
### PREDICATE DEVICE III.
- 1. Omnilux Clear-U, K081307
- 2. Omnilux Blue, K030883
The predicates have not been subject to a design-related recall.
No reference devices were used in this submission.
#### IV. DEVICE DESCRIPTION
The Philips BlueControl is a rechargeable battery-operated, wearable device for delivery of blue light to treat mild psoriasis vulgaris. It is a prescription device designed for home use.
The Philips BlueControl is sold as a kit, which contains the following items:
- Components:
- o BlueControl Device
- o Fixation Strap (Device Holder and Slings)
- o Power Supply (USB Cable, Adapter and Charger)
- Accessory: ●
{5}------------------------------------------------
- 0 Carrying Case
- Instructions for Use .
### INDICATIONS FOR USE V.
The Philips BlueControl is intended to emit energy in the blue region of the spectrum, is generally indicated to treat dermatological conditions and specifically indicated to treat mild psoriasis vulgaris.
{6}------------------------------------------------
### COMPARISON OF TECHNOLOGICAL CHARACTERISCS WITH THE PREDICATE VI.
Information for predicate devices was obtained from publicly available sources, including the 510(k) Summaries and device instruction manuals.
| Item | Subject Device:<br>Philips BlueControl | Primary Predicate:<br>Omnilux Clear-U<br>K081307 | Secondary Predicate:<br>Omnilux Blue<br>K030883 | Comparison |
|----------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------|
| Energy Source | LED | LED | LED | Same |
| Output<br>Wavelength(s)<br>(nm) | 452+/-7 | Blue: 415 +/- 5<br>Red: 633 +/- 6 | 415 +/- 5 | Similar<br>wavelengths in the<br>blue light<br>spectrum |
| Dose<br>Rate/Intensity<br>(mW/cm²) | 40 +/- 5 | 40 | 40 | Same |
| Average Dose<br>(J/ cm²) | 72 +/- 9 (maximum dose: 90) | 40 (blue)<br>70 (red) | 48 (dosage range: 1-150) | The average<br>treatment dose is<br>within dose range<br>of the predicates. |
| Method of<br>Delivery | Blue light treatment<br>administrated through the lower<br>housing. | Blue light treatment<br>administrated through<br>treatment head. | Blue light treatment<br>administrated through<br>treatment head. | Same |
| Treatment<br>Regimen | A full course of the treatment<br>consists of 4 weeks of daily<br>treatment 5-7 times a week<br>followed by 8 weeks of<br>treatment 3 times a week. A<br>single treatment takes<br>approximately 30 minutes with<br>or without interruptions. | Alternate treatments of blue<br>and red light twice a week<br>for 20 minutes per<br>treatment for 4 weeks.<br>Forearm test for 20 minutes<br>before starting treatment<br>regime. | One treatment consists of<br>20 minutes at standard<br>dose. The full treatment<br>course consists of 8<br>treatments. | Similar in<br>treatment time per<br>day. |
| Item | Subject Device:<br>Philips BlueControl | Primary Predicate:<br>Omnilux Clear-U<br>K081307 | Secondary Predicate:<br>Omnilux Blue<br>K030883 | Comparison |
| Treatment<br>Control(s) | On/off button puts device in<br>standby mode.<br>Infrared sensor on the treatment<br>face measures the skin<br>temperature. It senses the<br>presence of the skin and<br>activates LEDs. | Three position switch: Off-<br>Blue- Red | The control unit consists of<br>an LCD and keyboard<br>together with the control<br>electronics. The user<br>interface software allows<br>the operator to access and<br>control all device functions. | All devices have<br>on/off switches.<br>The sensor<br>provides<br>equivalent or<br>better safety<br>assurance. |
| Treatment<br>Indicator(s) | Visual signal:<br>-flashing blue- waiting for skin<br>detection<br>-solid blue -treatment is in<br>progress;<br>- flashing blue slowly after<br>recovery from skin overheated<br>(exceeds 43 °C) until the skin<br>cool off.<br>-solid orange- device is<br>overheated: exceeds 51 °C<br>(device switches off);<br>-flashing green-charging<br>-solid green-fully charged<br>-flashing and solid red- battery<br>empty<br>Audio signal:<br>-one (1) beep- connected to<br>power supply<br>- a two (2) tones beep- the<br>treatment starts<br>- a two (2) tones beep- the<br>treatment ends | None | Unknown | The treatment<br>light and audio<br>indicators provide<br>equivalent or<br>better safety<br>assurance<br>comparing to<br>Omnilux Clear-U. |
| Item | Subject Device:<br>Philips BlueControl | Primary Predicate:<br>Omnilux Clear-U<br>K081307 | Secondary Predicate:<br>Omnilux Blue<br>K030883 | Comparison |
| | -a long 3-second beep- device<br>overheated (device switches<br>off)<br>-three (3) beeps- battery empty | | | |
| Treatment<br>Timer | Yes- 30 minutes treatment per<br>area after which device<br>automatically switches off. | None | Unknown | The treatment<br>timer provides<br>equivalent or<br>better safety<br>assurance<br>comparing to<br>Omnilux Clear-U. |
| Treatment<br>Surface Area<br>(cm²) | 38 | 28.7 | 525 (LED head active<br>dimensions) | Comparable in<br>size with Omnilux<br>Clear-U. |
| Max<br>Treatment<br>Temperature<br>(°C) | 43 (skin)<br>51 (device) | 39-43 | Unknown | Similar in<br>temperature limit<br>with Omnilux<br>Clear-U |
| Patient<br>Contacting<br>Materials | Clear polycarbonate LED cover<br>OEKO-TEX®-certified fabric<br>device holder and straps | Rigid ABS | NA | Biocompatible<br>materials used in<br>both the<br>BlueControl and<br>Omnilux Clear-U. |
| Power Supply | 3.7 V/rated capacity 6.1Wh<br>lithium-ion rechargeable<br>battery.<br>AC charger:<br>Input: 100-240V at 50-60 Hz | "A separate universal,<br>power supply converts main<br>AC power to the DC power<br>required" | Wall power | The differences in<br>power supply<br>provide equivalent<br>or better safety<br>assurance. |
| Item | Subject Device:<br>Philips BlueControl | Primary Predicate:<br>Omnilux Clear-U<br>K081307 | Secondary Predicate:<br>Omnilux Blue<br>K030883 | Comparison |
| | Output: 5V DC and 1A/7W | | | |
| Dimensions<br>(height x<br>width x depth<br>in cm) | 12.5 x 7.72 x 2.28 | 12 x 6 x 4 | 37 x 18 x 49 | Comparable in<br>size with Omnilux<br>Clear-U |
| Weight | 5.9 oz. (without fixation strap)<br>7.8 oz. (with fixation strap) | Less than 1 pound | 26.4 lb. | Similar in weight<br>with Omnilux<br>Clear-U |
| Treatment<br>Head Design | 40 Blue LEDs | 30 Blue LEDs | Number of LEDs not<br>available | Similar in design<br>with Omnilux<br>Clear-U |
| Principles of<br>Operation | Wearable | Handheld | Desktop | All devices are<br>positioned over<br>the specific skin<br>treatment area to<br>deliver light<br>energy. |
{7}------------------------------------------------
{8}------------------------------------------------
{9}------------------------------------------------
The indications for use statement for the Philips BlueControl is not identical to the predicate devices; however the differences in the disease treated do not alter the intended therapeutic use of the device nor do they affect the safety and effectiveness of the device relative to the predicate. The technical of the Philips BlueControl device are either same or substantially equivalent as compared to the predicate devices. There are no technological differences that raise new or different questions of safety or effectiveness.
{10}------------------------------------------------
### VII. PERFORMANCE DATA
The following performance data were provided in support of the substantial equivalence determination.
## Biocompatibility Testing
The skin contacting materials of the device (lower housing, device holder and strap) were evaluated for biocompatibility per ISO 10993-1 (2009): Biological Evaluation of Medical Devices – Part 1: Evaluation and Testing," for devices in contact with intact skin, including cytotoxicity, dermal sensitization, and dermal irritation.
# Electric Safety and Electromagnetic Compatibility Testing
The Philips BlueControl has been tested and complies with the applicable requirements of the following standards for medical devices used in the home environment:
- . IEC 60601-1:2006 Medical Electrical Equipment - Part 1: General Requirements for Safety, 3rd edition
- . IEC 60601-1-11: 2010 Medical electrical equipment -- Part 1-11: General requirements for basic safety and essential performance -- Collateral standard: Requirements for medical electrical equipment and medical electrical systems used in the home healthcare environment
- IEC 60601-2-57: 2011 Medical electrical equipment: Particular requirements for ● the basic safety and essential performance of non-laser light source equipment intended for therapeutic, diagnostic, monitoring and cosmetic/aesthetic use
- IEC 60601-1-2: 2007 General requirements for basic safety and essential . performance - Collateral Standard: Electromagnetic disturbances - Requirements and tests
# Photobiological Safety Testing
The Philips BlueControl has been tested and complies with IEC 60601-2-57 (2011): Medical Electrical equipment part 2-57: Particular requirements for the basic safety and essential performance of non-laser light source equipment intended for therapeutic, diagnostic, monitoring and cosmetic/aesthetic use, 1* edition and IEC 62471 (2006-07): Photobiological safety of lamps and lamp systems, 1st edition. Those IEC standards incorporate the measurement techniques of the following ANSI IESNA recommended practices:
- RP 27.1:2005 Recommended practice for photobiological safety for lamps and . lamp systems - General requirements
- RP 27.2:2000 Recommended practice for photobiological safety for lamps and . lamp systems - Measurement techniques
- . RP-27.3:2007 Recommended practice for photobiological safety for lamps and lamp systems - Risk group classification and labeling
{11}------------------------------------------------
### Other Bench Testing
Thermal safety and reliability testing has also been conducted to verify the risk controls implemented for the Philips BlueControl are effective to mitigate the hazards identified.
## Software Verification and Validation Testing
Software documentation consistent with moderate level of concern was submitted in this 510(k). System validation testing presented in this 510(k) demonstrated that all software requirement specifications are met and all software hazards have been mitigated to acceptable risk levels.
# Clinical Studies
To support the proposed indication for use, two clinical studies of N=50 have been performed in Europe per ISO 14155: Clinical investigation of medical devices for human subjects -- Good clinical practice. In both studies, psoriasis severity was evaluated using the Local Psoriasis Area and Severity Index (LPASI), a modified version of the validated PASI. No serious adverse events were considered related to any of the studied treatment regimen.
One clinical study was conducted on 23 enrolled subjects who received blue light treatment at 200 mW/cm2 setting for their mild plaque psoriasis. The patients were treated with blue light for 30 minutes per plague per treatment for 12 weeks followed by a 4-week follow up period without treatment. During the full course of 12-week treatment, the 30-minute blue light treatment was performed daily (5 to 7 treatments per week) for the first 4 weeks, followed by 8 weeks with 3 treatments per week. The plaque severity was evaluated using the LPASI. The primary endpoint, change from baseline (CfB) of the LPASI at week 12, revealed a significant improvement of the treated compared to the control plaques (△CfB: -0.92 ± 1.10, p = 0.0005, t test; p = 0.0006, Wilcoxon signed-rank test; mean 95% CI –1.38, –0.45). The mean change from baseline to week 12 was -2.38 (-3.02 to -1.73) in the treated plaque and -1.46 (-2.13 to -0.79) in the control plaque. All 16 adverse events observed during the study were evaluated. No serious adverse events were reported. None of the adverse events were device-related or treatment-related. The hyperpigmentation (tanning) of the treated skin was expected and reported by some of the study participants. The tanning effect reduced within a few days after completion of the treatment cvcle. No other device-related adverse effects of the device are observed.
The results of these studies demonstrate the 12-week treatment cycle of BlueControl reduces the severity of mild psoriasis vulgaris.
# VIII. CONCLUSIONS
The Philips BlueControl and its predicates emit energy in the blue region of the light spectrum to treat dermatological conditions. The performance achieved by these devices uses nearly identical power intensities and wavelengths. Both the Philips BlueControl and the primary predicate are intended to be placed directly on the skin to provide the light energy. The primary predicate device is cleared for over-the- counter home use, and the
{12}------------------------------------------------
Philips BlueControl is a prescription home use device. The intended patient population for the Philips BlueControl fall within the intended use of the secondary predicate device, which is generally indicated to treat dermatological conditions for prescription use. To support the specific indication for use, the clinical testing performed using Philips BlueControl demonstrates its effectiveness for treatment of mild psoriasis vulgaris in adult patients.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.