← Product Code [LLZ](/productcode/LLZ) · K171053

# syngo Dynamics Version VA30 (K171053)

_Siemens Healthcare GmbH · LLZ · Jun 1, 2017 · Radiology · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K171053

## Device Facts

- **Applicant:** Siemens Healthcare GmbH
- **Product Code:** [LLZ](/productcode/LLZ.md)
- **Decision Date:** Jun 1, 2017
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 892.2050
- **Device Class:** Class 2
- **Review Panel:** Radiology
- **Attributes:** Software as a Medical Device, 3rd-Party Reviewed

## Indications for Use

syngo Dynamics is an image and information system intended for acceptance, transfer, display, storage, archive and manipulation of digital medical images, including review, analysis, quantification and reporting. As a Cardiology PACS and information system, syngo Dynamics supports the physician in interpretation and evaluation of examinations within healthcare institutions, in particular, in Cardiology, Obstetrics and Gynecology or other departments. syngo Dynamics is not intended to be used for displaying of digital mammography images for diagnosis in the U.S.

## Device Story

syngo Dynamics VA30 is a client-server PACS for digital medical images (ultrasound, XA, DX, DR, PET, nuclear medicine). Server processes data from imaging modalities; client provides UI for viewing, processing, and reporting. Used by physicians in clinical settings for interpretation, quantification, and reporting. Features include window/leveling, edge enhancement, digital subtraction, and vascular/cardiac measurements. Integrates with EHRs and third-party rules engines to support clinical decision-making. System enables collaborative reporting and critical result communication. Benefits include centralized image management, standardized reporting, and efficient clinical workflow.

## Clinical Evidence

No clinical studies were performed. Evidence consists of non-clinical verification and validation testing, including software testing, risk analysis per ISO 14971, and conformance to DICOM and HL7 standards.

## Technological Characteristics

Client-server software architecture; runs on Windows 2012 R2 Server (server) and Windows 7+ (client). Connectivity via TCP/IP, DICOM, HL7, and HTTP. Supports lossless/lossy compression. Algorithms include window/leveling, edge enhancement, digital subtraction, and pixel size evaluation. Conforms to NEMA PS3, ISO 14971, IEC 62304, IEC 62366-1, and ISO/HL7 21731.

## Regulatory Identification

A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.

## Special Controls

*Classification.* Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).

## Predicate Devices

- syngo Dynamics ([K123922](/device/K123922.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image is a black and white logo for the Department of Health & Human Services - USA. The logo is circular, with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter of the circle. In the center of the circle is a stylized image of three human profiles facing to the right, stacked on top of each other.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

June 1, 2017

Siemens Healthcare GmbH % Mr. Olaf Teichert Third Party Reviewer TUV SUD America Inc. 1775 Old Hwy 8 NW, Suite 104 NEW BRIGHTON MN 55112

Re: K171053

Trade/Device Name: syngo® Dynamics Version VA30 Regulation Number: 21 CFR 892.2050 Regulation Name: Picture archiving and communications system Regulatory Class: II Product Code: LLZ Dated: May 18, 2017 Received: May 25, 2017

Dear Mr. Teichert:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{1}------------------------------------------------

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours.

Michael D. O'Hara
For

Robert Ochs. Ph.D. Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

# Indications for Use

510(k) Number (if known) k171053

Device Name syngo Dynamics Version VA30

#### Indications for Use (Describe)

syngo Dynamics is an image and information system intended for acceptance, transfer, display, storage, archive and manipulation of digital medical images, including review, analysis, quantification and reporting.

As a Cardiology PACS and information system, syngo Dynamics supports the physician in interpretation and evaluation of examinations within healthcare institutions, in particular, in Cardiology. Obstetrics and Gynecology or other departments.

syngo Dynamics is not intended to be used for displaying of digital mammography images for diagnosis in the U.S.

Type of Use (Select one or both, as applicable)

X Prescription Use (Part 21 CFR 801 Subpart D)

| Over-The-Counter Use (21 CFR 801 Subpart C)

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows the logo for Siemens Healthineers. The word "SIEMENS" is written in teal, and the word "Healthineers" is written in orange below it. To the right of the words is a graphic of orange dots arranged in a triangular shape.

# 510(k) Summary

## syngo® Dynamics (Version VA30)

This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR §807.92.

Date prepared: March 27, 2017

### 1. Submitter

Siemens Healthcare GmbH Henkestrasse 127 D-91052 Erlangen Germany

## Establishment Registration Number 3002808157

### 2. Contact Person

Ms. Lauren Bentley Regulatory Affairs Manager Siemens Medical Solutions USA, Inc. 40 Liberty Boulevard Mailcode: 65-3 Malvern, PA 19355 USA Email: lauren.bentley@siemens.com Telephone: +1 (610) 448 - 6104 Fax: +1 (610) 448 – 6557

### 3. Device Name and Classification

| Trade Name:           | syngo® Dynamics Version VA30               |  |
|-----------------------|--------------------------------------------|--|
| Classification Name:  | Picture Archiving and Communication System |  |
| Classification Panel: | Radiology                                  |  |
| CFR Section:          | 21 CFR §892.2050                           |  |
| Device Class:         | Class II                                   |  |
| Product Code:         | LLZ                                        |  |

### 4. Legally Marketed Predicate Device

| Trade Name:       | syngo Dynamics |
|-------------------|----------------|
| 510(k) Clearance: | K123922        |

{4}------------------------------------------------

| Clearance Date:       | January 16, 2013                           |
|-----------------------|--------------------------------------------|
| Classification Name:  | Picture Archiving and Communication System |
| Classification Panel: | Radiology                                  |
| CFR Section:          | 21 CFR §892.2050                           |
| Device Class:         | Class II                                   |
| Product Code:         | LLZ                                        |

#### 5. Device Description

This premarket submission addresses the Siemens syngo® Dynamics Version VA30 Picture Archiving and Communication System.

syngo Dynamics VA30, is a digital image display and reporting system. This system can function as a standalone medical device that includes a DICOM server or as an integrated module within and Electronic Health Record (EHR) System with a DICOM archive that receives images from digital image acquisition devices such as ultrasound and x-ray angiography machines. The syngo Dynamics system provides components that can be used to review, edit and manipulate image data, as well as to generate quantitative data, qualitative date and diagnostic reports. syngo Dynamics VA30 also provides advanced reporting support for cardiology, OB/GYN, MFM (maternal fetal medicine) and vascular ultrasound studies.

syngo Dynamics is a software only medical device. Recommended configurations are defined for the hardware required to run the device. The hardware itself is not seen as a medical device and not in the scope of this 510(k) submission.

syngo Dynamics is based on a client-server architecture. The server processes the data from the connected imaging modalities. The client provides the user interface for interactive image viewing and processing and can be installed on remote, network connected, workstation machines or through industry standard virtualization software.

Since the majority of the data processing is performed by the server, the client can be installed on standard off-the-shelf computers with a variety of monitor types. The quality of displayed images highly depends on the quality and settings of monitors, graphics cards, and graphics drivers. It is the customer's responsibility that client monitors are compatible with graphics cards and graphics drivers. It is also the customer's responsibility to use suitable monitors for diagnostic purposes.

In the United States, monitors (displays) should not be used for diagnosis, unless the monitor (display) has specifically received 510(k) clearance for this purpose.

syngo® Dynamics VA30 and its predicate device have the same fundamental technical characteristics.

{5}------------------------------------------------

#### 6. Intended Use

syngo Dynamics is an image and information system intended for acceptance, transfer. display, storage, archive and manipulation of digital medical images, including review, analysis, quantification and reporting.

As a Cardiology PACS and information system, syngo Dynamics supports the physician in interpretation and evaluation of examinations within healthcare institutions, in particular, in Cardiology, Obstetrics and Gynecology or other departments.

syngo Dynamics is not intended to be used for displaying of digital mammography images for diagnosis in the U.S.

#### 7. Summary of Differences Between the Subject Device and the Predicate Device

The differences between the subject device described in this premarket notification and the predicate device are summarized in the following comparison table:

|                                                                       | Subject Device<br>syngo Dynamics VA30                                                                                                                                                                                                                                                                                                                          | Predicate Device<br>syngo Dynamics VA10A                                                                                                                                                                                                                                                                                                        |
|-----------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Manufacturer                                                          | Siemens Healthcare GmbH                                                                                                                                                                                                                                                                                                                                        | Siemens Medical Solutions USA, Inc.                                                                                                                                                                                                                                                                                                             |
| Software Architecture                                                 | Client-server architecture                                                                                                                                                                                                                                                                                                                                     | Client-server architecture                                                                                                                                                                                                                                                                                                                      |
| Operating System                                                      | Server<br>Windows 2012 R2 Server<br>Standard Edition R2 (64-bit)<br><br>Portal Client<br>Windows 7 SP1 or higher<br>64-bit<br><br>Portal Website Host<br>Windows 2012 R2 Server<br>Standard Edition R2 (64-bit)                                                                                                                                                | Server<br>Windows 2008 R2 Server<br>Standard Edition R2 SP1 or<br>SP2 (64-bit)<br><br>Workplace<br>Microsoft Windows 7 or<br>Windows 7 SP1 or higher<br>· 32-bit or 64-bit<br>· Ultimate, Professional,<br>Enterprise, Ultimate N,<br>Professional N, or<br>Enterprise N<br><br>Portal Website Host<br>Windows Server 2008 R1 32-bit or greater |
| Image Source                                                          | DICOM Ultrasound, XA, DX, DR<br>and Nuclear Medicine, including<br>PET.                                                                                                                                                                                                                                                                                        | DICOM Ultrasound, XA, DX, DR and<br>Nuclear Medicine, including PET.                                                                                                                                                                                                                                                                            |
| Image Display                                                         | Ultrasound, XA, DX,<br>DR, PET and Nuclear Medicine<br>through Corridor4DM                                                                                                                                                                                                                                                                                     | Ultrasound, XA, MR, DX,<br>DR, PET and Nuclear Medicine<br>through Corridor4DM                                                                                                                                                                                                                                                                  |
| Data Export                                                           | DICOM, bmp, avi                                                                                                                                                                                                                                                                                                                                                | DICOM, bmp, avi                                                                                                                                                                                                                                                                                                                                 |
| Image<br>Communication                                                | Within the network, the<br>following communication<br>protocols are used:<br>· TCP/IP: for communication<br>and transport<br>· DICOM and HL7 at                                                                                                                                                                                                                | Within the network, the<br>following communication<br>protocols are used:<br>· TCP/IP: for communication<br>and transport<br>· DICOM and HL7 at                                                                                                                                                                                                 |
|                                                                       | application level<br>- HTTP for communication<br>and transport of thumbnails                                                                                                                                                                                                                                                                                   | application level<br>- HTTP for communication<br>and transport of thumbnails                                                                                                                                                                                                                                                                    |
| Image Data<br>Compression                                             | Lossless compression with<br>compression factor 2 to 3 and<br>lossy compression with higher<br>compression rate.                                                                                                                                                                                                                                               | Lossless compression with<br>compression factor 2 to 3 and<br>lossy compression with higher<br>compression rate.                                                                                                                                                                                                                                |
| Imaging<br>Algorithms                                                 | Window/Leveling, Edge<br>Enhancement, and Digital<br>Subtraction                                                                                                                                                                                                                                                                                               | Window/Leveling, Edge<br>Enhancement, and Digital<br>Subtraction                                                                                                                                                                                                                                                                                |
| Quantitative<br>Algorithms                                            | Pixel Size Evaluation                                                                                                                                                                                                                                                                                                                                          | Pixel Size Evaluation                                                                                                                                                                                                                                                                                                                           |
| Network Access                                                        | Yes                                                                                                                                                                                                                                                                                                                                                            | Yes                                                                                                                                                                                                                                                                                                                                             |
| Analysis                                                              | Yes                                                                                                                                                                                                                                                                                                                                                            | Yes                                                                                                                                                                                                                                                                                                                                             |
| Reporting                                                             | Yes                                                                                                                                                                                                                                                                                                                                                            | Yes                                                                                                                                                                                                                                                                                                                                             |
| Decision Support<br>Interface to Rules<br>Engine                      | Ability to interface with a third party<br>rules engine (BizTalk), where rules<br>are configured by the end<br>customer to determine clinical<br>relevance of selected<br>observations.<br><br>Customers identify and store<br>selected patient data.<br>Orchestrations provides a trigger<br>to pull in previously stored relevant<br>data for a given study. | n/a                                                                                                                                                                                                                                                                                                                                             |
| Send Critical Results                                                 | End user, identified critical Results<br>are sent to the EHR quickly                                                                                                                                                                                                                                                                                           | n/a                                                                                                                                                                                                                                                                                                                                             |
| Multimodality<br>storage and<br>review                                | Yes                                                                                                                                                                                                                                                                                                                                                            | Yes                                                                                                                                                                                                                                                                                                                                             |
| Web Server for<br>images and clips                                    | Yes                                                                                                                                                                                                                                                                                                                                                            | Yes                                                                                                                                                                                                                                                                                                                                             |
| Report upload to<br>Information<br>Systems                            | Yes, through broker or<br>interface engine                                                                                                                                                                                                                                                                                                                     | Yes, through broker or<br>interface engine                                                                                                                                                                                                                                                                                                      |
| DICOM Structured<br>Reporting                                         | Yes                                                                                                                                                                                                                                                                                                                                                            | Yes                                                                                                                                                                                                                                                                                                                                             |
| Export/Import<br>Data Sets via<br>removable media<br>or network means | n/a                                                                                                                                                                                                                                                                                                                                                            | Yes                                                                                                                                                                                                                                                                                                                                             |
| Vascular<br>Quantification                                            | Yes, measurements and<br>calculations                                                                                                                                                                                                                                                                                                                          | Yes, measurements and<br>calculations                                                                                                                                                                                                                                                                                                           |
| Data Mining                                                           | Yes                                                                                                                                                                                                                                                                                                                                                            | Yes                                                                                                                                                                                                                                                                                                                                             |
| Discrete Data<br>Export                                               | Yes                                                                                                                                                                                                                                                                                                                                                            | Yes                                                                                                                                                                                                                                                                                                                                             |
| Cardiac<br>Measurements                                               | Yes                                                                                                                                                                                                                                                                                                                                                            | Yes                                                                                                                                                                                                                                                                                                                                             |
| Interactive<br>graphical<br>documentation for<br>reporting            | Coronary Tree Diagrams are the<br>same in VA30. Congenital Heart<br>Diagrams and Vascular Diagrams<br>have been added in the Common<br>Reporting Component.                                                                                                                                                                                                    | Coronary Tree Diagrams                                                                                                                                                                                                                                                                                                                          |
| Hemodynamic                                                           | Yes                                                                                                                                                                                                                                                                                                                                                            | Yes                                                                                                                                                                                                                                                                                                                                             |
| data import (third                                                    |                                                                                                                                                                                                                                                                                                                                                                |                                                                                                                                                                                                                                                                                                                                                 |
| party vendor)                                                         |                                                                                                                                                                                                                                                                                                                                                                |                                                                                                                                                                                                                                                                                                                                                 |
| Web Reporting                                                         | Yes                                                                                                                                                                                                                                                                                                                                                            | Yes                                                                                                                                                                                                                                                                                                                                             |
| Electronic Health                                                     | Basic report and discrete data                                                                                                                                                                                                                                                                                                                                 | Basic report and discrete data upload to                                                                                                                                                                                                                                                                                                        |
| Record (EHR)                                                          | upload to EHR.                                                                                                                                                                                                                                                                                                                                                 | EHR                                                                                                                                                                                                                                                                                                                                             |
| Interfacinq                                                           |                                                                                                                                                                                                                                                                                                                                                                |                                                                                                                                                                                                                                                                                                                                                 |
|                                                                       | Ability to integrate DICOM Image                                                                                                                                                                                                                                                                                                                               |                                                                                                                                                                                                                                                                                                                                                 |
|                                                                       | Review and Common Reporting                                                                                                                                                                                                                                                                                                                                    |                                                                                                                                                                                                                                                                                                                                                 |
|                                                                       | Component with Cerner Electronic                                                                                                                                                                                                                                                                                                                               |                                                                                                                                                                                                                                                                                                                                                 |
|                                                                       | Health Record System and PACS.                                                                                                                                                                                                                                                                                                                                 |                                                                                                                                                                                                                                                                                                                                                 |
| Collaborative                                                         | Yes                                                                                                                                                                                                                                                                                                                                                            | Yes                                                                                                                                                                                                                                                                                                                                             |
| Reporting                                                             |                                                                                                                                                                                                                                                                                                                                                                |                                                                                                                                                                                                                                                                                                                                                 |
| Mobile Device                                                         | Images can be viewed on mobile                                                                                                                                                                                                                                                                                                                                 | Yes - Non-diagnostic, and                                                                                                                                                                                                                                                                                                                       |
| Support                                                               | devices, but are still non-                                                                                                                                                                                                                                                                                                                                    | read-only. Support Apple                                                                                                                                                                                                                                                                                                                        |
|                                                                       | diagnostic.                                                                                                                                                                                                                                                                                                                                                    | iPhone and iPad.                                                                                                                                                                                                                                                                                                                                |

{6}------------------------------------------------

{7}------------------------------------------------

## 8. Clinical Testing

No clinical studies were carried out for syngo Dynamics VA30. All performance testing was conducted in a non-clinical fashion as part of the verification and validation activities for the medical device.

# 9. Non-Clinical Performance Testing

Non-clinical testing was conducted for the device syngo Dynamics during product development. The modifications described in this Premarket Notification were supported with verification and validation testing.

Siemens claims conformance to the following standards:

- NEMA PS3 Digital Imaging and Communications in Medicine (DICOM)
- ISO 14971:2007
- ANSI/AAMI ES 60601-1. A1. clauses 14.11 and 14.13 ●
- IEC 62304: 2006 ●
- IEC 62366-1:2015 .
- IEC 10918-1:1994 + Technical Corrigendum 1:2005
- . ISO/HL7 21731:2014

### Software Verification and Validation

In accordance with the FDA's Guidance Document "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices" issued on May 11, 2005, documentation is included within this submission for software of a Moderate Level of Concern. Non-clinical Testing was conducted during product development. Evidence provided within this submission demonstrates conformance with special controls for medical devices containing software.

{8}------------------------------------------------

Cybersecurity considerations related to syngo Dynamics are included within this submission. Siemens conforms to cybersecurity requirements by implementing a means to prevent unauthorized access, modification, misuse, denial of use or unauthorized use of information stored, accessed or transferred from a medical device to an external recipient.

A risk analysis, in compliance with ISO 14971:2007, for syngo Dyanamics was conducted and mitigation controls were implemented for identified hazards. Verification and validation testing confirms that all software specifications have been implemented and met the defined acceptance criteria. Further, documentation is provided to support the claim of substantial equivalence.

### 10. Safety and Effectiveness Information

Device labeling contains instructions for use and any necessary cautions and warnings to provide for safe and effective use of this device. Risk management was implemented throughout the development process to control potential hazards.

The device does not come into contact with the patient and is only used by trained professionals. The output of the device is evaluated by clinicians, providing for sufficient review to identify and intervene in the event of a malfunction.

Siemens believes that syngo Dynamics version VA30 is safe and effective as the identified predicate device and does not introduce new safety and effectiveness concerns.

# 11. Conclusion as to Substantial Equivalence

The comparison of intended use, technological characteristics, performance specifications, device hazards as well as verification and validation results demonstrate that synqo Dynamics is safe, effective and performs as well as the predicate device.

In summary, Siemens is of the opinion that syngo Dynamics version VA30 does not introduce any new significant potential safety risks and is substantially equivalent to the predicate device.

---

**Source:** [https://fda.innolitics.com/device/K171053](https://fda.innolitics.com/device/K171053)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
