← Product Code [GAM](/productcode/GAM) · K171001

# Monosyn Quick Synthetic Absorbable Surgical Suture (K171001)

_Aesculap, Inc. · GAM · Oct 23, 2017 · General, Plastic Surgery · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K171001

## Device Facts

- **Applicant:** Aesculap, Inc.
- **Product Code:** [GAM](/productcode/GAM.md)
- **Decision Date:** Oct 23, 2017
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 878.4493
- **Device Class:** Class 2
- **Review Panel:** General, Plastic Surgery
- **Attributes:** Therapeutic

## Indications for Use

Monosyn Quick Synthetic Absorbable Surgical Suture is intended for general soft tissue approximation of the skin and mucosa, where only short term wound support (6-7 days) is required. Monosyn Quick suture is not indicated for use in cardiovascular or neurosurgery.

## Device Story

Monosyn Quick is a sterile, absorbable, flexible monofilament surgical suture composed of a synthetic polyglycolic acid-based copolymer (glycolide, ε-caprolactone, and trimethylene carbonate). It is supplied in various diameters (USP 6-0 to 1) and lengths, with or without attached stainless steel needles. Used by clinicians for soft tissue approximation, the device provides short-term wound support, maintaining 60-70% tensile strength at 5 days and 25-35% at 10 days, with absorption essentially complete by 56 days. The suture is intended for skin and mucosal closure. It is packaged in a sterile pouch and sterilized via ethylene oxide.

## Clinical Evidence

Bench testing only. Performance demonstrated via mechanical testing per USP 39 (diameter, tensile strength, needle attachment) and biocompatibility testing per ISO 10993 (cytotoxicity, local effects, irritation, sensitization, systemic toxicity, EO residuals). Accelerated aging data provided. No clinical data included.

## Technological Characteristics

Synthetic absorbable monofilament; material: 72% glycolide, 14% ε-caprolactone, 14% trimethylene carbonate. USP 39 compliant for mechanical properties. Sterilized via Ethylene Oxide (ISO 11135-1). Packaging: PET/aluminum/LDPE or HDPE support in pouches.

## Regulatory Identification

An absorbable poly(glycolide/l-lactide) surgical suture (PGL suture) is an absorbable sterile, flexible strand as prepared and synthesized from homopolymers of glycolide and copolymers made from 90 percent glycolide and 10 percent l-lactide, and is indicated for use in soft tissue approximation. A PGL suture meets United States Pharmacopeia (U.S.P.) requirements as described in the U.S.P. “Monograph for Absorbable Surgical Sutures;” it may be monofilament or multifilament (braided) in form; it may be uncoated or coated; and it may be undyed or dyed with an FDA-approved color additive. Also, the suture may be provided with or without a standard needle attached.

## Special Controls

*Classification.* Class II (special controls). The special control for this device is FDA's “Class II Special Controls Guidance Document: Surgical Sutures; Guidance for Industry and FDA.” See § 878.1(e) for the availability of this guidance document.

## Predicate Devices

- Monosyn Synthetic Absorbable Surgical Suture ([K011375](/device/K011375.md))
- Caprosyn Absorbable Surgical Suture ([K032586](/device/K032586.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular border with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around it. Inside the circle is a stylized image of three human profiles facing right, with flowing lines connecting them.

DEPARTMENT OF HEALTH & HUMAN SERVICES Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

October 23, 2017

Aesculap®, Inc. Ms. Kathy Racosky Senior Regulatory Affairs Specialist 3773 Corporate Parkway Center Valley, Pennsylvania 18034

Re: K171001

Trade/Device Name: Monosyn Quick Synthetic Absorbable Surgical Suture Regulation Number: 21 CFR 878.4493 Regulation Name: Absorbable Poly(Glycolide/L-Lactide) Surgical Suture Regulatory Class: Class II Product Code: GAM Dated: September 12, 2017 Received: September 13, 2017

Dear Ms. Racosky:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 

{1}------------------------------------------------

801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education (DICE) at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education (DICE) at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely.

# David Krause -S

Binita S. Ashar, M.D., M.B.A., F.A.C.S. for Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

apple Synggerics Aps 25 25 27 2017 11 20

Expiration Date: January 31, 2017

See PRA Statement below.

Monosyn

### 510(k) Prematic Metifisetion OF HEALTH AND HUMAN SERVICES Food and Drug Administration

# Indications for Use

510(k) Number (if known)

# K171001

#### Device Name

Aesculap Monosyn Quick Synthetic Absorbable Surgical Suture

Indications for Use (Describe)

Monosyn Quick Synthetic Absorbable Surgical Suture is intended for general soft tissue approximation of the skin and mucosa, where only short term wound support (6-7 days) is required. Monosyn Quick suture is not indicated for use in cardiovascular or neurosurgery.

Type of Use (Select one or both, as applicable)

X Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Page 1 of 4

#### B. 510(k) SUMMARY (as required by 21 CFR 807.92)

Aesculap Monosyn Quick Synthetic Absorbable Surgical Suture September 12, 2017

| COMPANY:             | Aesculap®, Inc.<br>3773 Corporate Parkway<br>Center Valley, PA 18034<br>Establishment Registration Number: 2916714 |
|----------------------|--------------------------------------------------------------------------------------------------------------------|
| CONTACT:             | Kathy A. Racosky<br>610-984-9291 (phone)<br>610-791-6882 (fax)<br>kathy.racosky@aesculap.com                       |
| TRADE NAME:          | Monosyn Quick Synthetic Absorbable Surgical Suture                                                                 |
| COMMON NAME:         | Synthetic Absorbable Surgical Suture                                                                               |
| CLASSIFICATION:      | Class II                                                                                                           |
| CLASSIFICATION NAME: | Absorbable poly(glycolide/l-lactide) surgical suture                                                               |
| REGULATION NUMBER:   | 878.4493                                                                                                           |
| PRODUCT CODE:        | GAM                                                                                                                |

# PURPOSE FOR PREMARKET NOTIFICATION

The purpose for this submission is to gain marketing clearance for the Monosyn Quick Synthetic Absorbable Surgical Suture.

# PREDICATE DEVICE

- Primary predicate: Monosyn Synthetic Absorbable Surgical Suture (K011375) .
- . Reference predicate: Caprosyn Absorbable Surgical Suture (K032586)

### DEVICE DESCRIPTION

Monosyn Quick is an absorbable flexible monofilament suture which is supplied sterile. It is composed of a synthetic polyglycolic acid-based copolymer. The Monosyn Quick suture is undyed and will be offered in diameters ranging from USP size 6-0 through 1. It is uncoated and will be available in a variety of cut lengths with or with out needles attached.

# INDICATIONS FOR USE

Monosyn Quick Synthetic Absorbable Surgical Suture is intended for general soft tissue approximation of the skin and mucosa, where only short term wound support (6-7 days) is required. Monosyn Quick suture is not indicated for use in cardiovascular or neurosurgery.

{4}------------------------------------------------

# TECHNOLOGICAL CHARACTERISTICS(compared to Predicate(s)}

As established in this submission, the Monosyn Quick suture is a synthetic absorbable monofilament surgical suture offered undyed in the same range of diameters and cut lengths that are substantially equivalent to other predicate devices cleared by FDA. The subject device is shown to be substantially equivalent and has the same technological characteristics to its predicate devices through comparison in design, intended use, material composition, function and range of sizes. The device characteristics comparing the Monosyn Quick Synthetic Absorbable Surgical Suture to the predicate devices are summarized below.

|                                                                                  | Aesculap Inc.<br>Monosyn Quick Synthetic<br>Absorbable Surgical<br>Suture                                                                                                                                                                                                                               | Aesculap, Inc.<br>Monosyn Synthetic<br>Absorbable Surgical<br>Suture                                                                                                                                    | United States Surgical<br>Caprosyn Absorbable<br>Surgical Suture                                                                                                                                                                                                 |
|----------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| K#                                                                               | Proposed device                                                                                                                                                                                                                                                                                         | K011375                                                                                                                                                                                                 | K032586                                                                                                                                                                                                                                                          |
| Indications                                                                      | Monosyn Quick Synthetic<br>Absorbable Surgical Suture<br>is intended for general soft<br>tissue approximation of the<br>skin and mucosa, where<br>only short term wound<br>support (6-7 days) is<br>required. Monosyn Quick<br>suture is not indicated for<br>use in cardiovascular or<br>neurosurgery. | Monosyn Synthetic<br>Absorbable Surgical Suture<br>is indicated for use in<br>general soft tissue<br>approximation and/or<br>ligation, but not for use in<br>cardiovascular or<br>neurological surgery. | Caprosyn synthetic<br>absorbable sutures are<br>indicated for use in general<br>soft tissue approximation<br>and/or ligation, including<br>use in ophthalmic<br>procedures, but not for use<br>in cardiovascular or<br>neurological surgery, or<br>microsurgery. |
| Absorption Type                                                                  | Short term absorbable                                                                                                                                                                                                                                                                                   | Mid term absorbable                                                                                                                                                                                     | Short term absorbable                                                                                                                                                                                                                                            |
| Absorption                                                                       | Essentially complete by 56<br>days                                                                                                                                                                                                                                                                      | Essentially completed<br>between 60-90 days                                                                                                                                                             | Essentially complete by 56<br>days                                                                                                                                                                                                                               |
| Remaining Tensile<br>Strength                                                    | 5 days 60-70%<br>10 days 25-35%                                                                                                                                                                                                                                                                         | 14 days 50%                                                                                                                                                                                             | 5 days 60%<br>10 days 20-30%                                                                                                                                                                                                                                     |
| Material                                                                         | 72% glycolide<br>14% ε-caprolactone<br>14% trimethylene carbonate                                                                                                                                                                                                                                       | 72% glycolide<br>14% ε-caprolactone<br>14% trimethylene carbonate                                                                                                                                       | Glycolide, Caprolactone,<br>Trimethylene Carbonate<br>and Lactide                                                                                                                                                                                                |
| Dyed, Un-dyed                                                                    | Undyed                                                                                                                                                                                                                                                                                                  | Un-dyed and Dyed                                                                                                                                                                                        | Un-dyed and Dyed                                                                                                                                                                                                                                                 |
| Structure                                                                        | Monofilament                                                                                                                                                                                                                                                                                            | Monofilament                                                                                                                                                                                            | Monofilament                                                                                                                                                                                                                                                     |
| Size                                                                             | 6-0 through 1 (various<br>lengths) with or w/out<br>needles attached                                                                                                                                                                                                                                    | 5-0 through 1 (various<br>lengths) with or w/out<br>needles attached                                                                                                                                    | 6-0 through 1 (various<br>lengths) with or w/out<br>needles attached                                                                                                                                                                                             |
| Thread length                                                                    | -45 cm to 120 cm                                                                                                                                                                                                                                                                                        | -35 cm to 150 cm<br>-ligature reels of longer<br>length                                                                                                                                                 | -18 in to 60 in<br>-ligature reels of longer<br>length                                                                                                                                                                                                           |
| Physical:<br>- Diameter<br>- Length<br>- Needle Attachment<br>- Tensile Strength | All characteristics meet<br>USP Requirements, except<br>for diameter.                                                                                                                                                                                                                                   | All characteristics meet<br>USP Requirements, except<br>for diameter.                                                                                                                                   | All characteristics meet<br>USP Requirements, except<br>for diameter.                                                                                                                                                                                            |
| Needle material                                                                  | 300 or 400 series stainless<br>steel                                                                                                                                                                                                                                                                    | 300 or 400 series stainless<br>steel                                                                                                                                                                    | unknown                                                                                                                                                                                                                                                          |

3

{5}------------------------------------------------

## Page 3 of 4

|               | Aesculap Inc.<br>Monosyn Quick Synthetic<br>Absorbable Surgical<br>Suture                                                                                                                                                                                                                                                                                                                                                                                     | Aesculap, Inc.<br>Monosyn Synthetic<br>Absorbable Surgical<br>Suture                                                                                                                | United States Surgical<br>Caprosyn Absorbable<br>Surgical Suture |
|---------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------|
| K#            | Pending                                                                                                                                                                                                                                                                                                                                                                                                                                                       | K011375                                                                                                                                                                             | K032589                                                          |
| Packaging     | Coated cardboard support<br>in double peel pouch. Inner<br>pouch composed of PET,<br>aluminum foil and LDPE<br>layers. Medical grade<br>paper and PET/LDPE<br>laminate outer pouch.<br>or<br>Oval HDPE support with a<br>coated cardboard lid in a<br>single pouch composed of<br>polyethylene terephthalate<br>polyester (PETP).<br>or<br>cardboard support fold card<br>in a single pouch composed<br>of polyethylene<br>terephthalate polyester<br>(PETP). | Coated cardboard support<br>in double peel pouch. Inner<br>pouch composed of PET.<br>aluminum foil and LDPE<br>layers. Medical grade<br>paper and PET/LDPE<br>laminate outer pouch. | unknown                                                          |
| Sterilization | Ethylene Oxide (EO)                                                                                                                                                                                                                                                                                                                                                                                                                                           | Ethylene Oxide (EO)                                                                                                                                                                 | unknown                                                          |

# PERFORMANCE DATA

As recommended by the FDA's Class II Special Control Guidance Document for Surgical Sutures, including mechanical testing in accordance to USP 39 for synthetic absorbable suture, biocompatibility testing in accordance to ISO 10993-1, and in vitro as well as in vivo resorption testing has been performed to demonstrate that the Monosyn Quick Synthetic Absorbable Surgical Suture meets current performance requirements for synthetic absorbable sutures unless otherwise labeled, and that Monosyn Quick is substantially equivalent to other predicate devices. The test results demonstrate that the Monosyn Quick complies with the following standards:

USP 39 Monograph for Absorbable Surgical Sutures

USP 39 <861> Sutures - Diameter (where diameter deviates minimally from USP requirements, it will be labeled as such)

USP 39 <881> Tensile Strength

USP 39 <871> Sutures - Needle Attachment

ISO 10993-1:2009, Biological evaluation of medical devices -- Part 1: Evaluation and testing within a risk management process. (Biocompatibility)

ISO 10993-5, Biological evaluation of medical devices -- Part 5: Tests for in vitro cytotoxicity ISO 10993-6 2007, Biological Evaluation Of Medical Devices - Part 6: Tests For Local Effects After Implantation

ISO 10993-7, Biological evaluation of medical devices - Part 7: Ethylene oxide sterilization residuals

ISO 10993-10 2010, Biological Evaluation Of Medical Devices - Part 10: Tests For Irritation And Skin Sensitization

{6}------------------------------------------------

Page 4 of 4

ISO 10993-11 2006, Biological Evaluation Of Medical Devices - Part 11: Tests For Systemic Toxicity

ISO 11135-1:2007, Sterilization of health care products- Ethylene oxide- Part 1: Requirements for development, validation and routing control of a sterilization process for medical devices

The Monosyn Quick Synthetic Absorbable Surgical Suture is blister packed and sterilized by Ethylene Oxide. Accelerated aging data for the Monosyn Quick has been generated to support this submission.

#### CONCLUSION:

Non clinical testing demonstrated that the device is as safe, as effective, and performs as well as the predicate devices.

In addition, residual strength and absorption rate studies were performed. The results of this testing demonstrates that the Monosyn Quick is substantially equivalent to the predicate devices.

---

**Source:** [https://fda.innolitics.com/device/K171001](https://fda.innolitics.com/device/K171001)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).