Urbanek Device

K170985 · Tmj Services, LLC · MQC · Sep 25, 2017 · DE

Device Facts

Record IDK170985
Device NameUrbanek Device
ApplicantTmj Services, LLC
Product CodeMQC · DE
Decision DateSep 25, 2017
DecisionSESE
Submission TypeTraditional
Device ClassClass U
AttributesTherapeutic

Indications for Use

The Urbanek Device is for the amelioration of clenching and bruxing associated with TMD and is to aid in the relief of symptoms of TMD/TMJ.

Device Story

Intraoral, maxillary posterior dis-occluding device; custom-fabricated from heat-cured dental resin with metal ball clasps for retention. Used by patients under dentist prescription; worn 24 hours/day (except meals) for set intervals (e.g., 9 weeks). Device sits on palate; prevents posterior maxillary and mandibular teeth (molars/premolars) from occluding in centric or eccentric positions. Disocclusion stabilizes jaw muscles and joints; reduces inflammation; allows healing. Dentist performs follow-up assessments for symptom relief, device comfort, wear, and vertical dimension adjustments.

Clinical Evidence

No clinical data. Bench testing only: material composition meets ISO 20795-1:2008 (Denture base polymers) and has history of safe use in legally marketed dental devices.

Technological Characteristics

Custom-fabricated intraoral appliance; materials: thermoplastic resin (ISO 20795-1:2008 compliant) and ASTM 313 stainless steel. Principle: mechanical disocclusion of posterior teeth. Form factor: maxillary palate-supported. Non-powered.

Indications for Use

Indicated for patients experiencing clenching and bruxing associated with TMD/TMJ to aid in symptom relief.

Predicate Devices

Reference Devices

Submission Summary (Full Text)

{1}------------------------------------------------ ### DEPARTMENT OF HEALTH & HUMAN SERVICES Public Health Service Image /page/0/Picture/2 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, stacked on top of each other. Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002 TMJ Services, LLC % Dawn Norman Executive Vice President MRC-X, LLC 6075 Poplar Ave; Suite 500 Memphis. Tennessee 38119 Re: K170985 Trade/Device Name: Urbanek Device Regulatory Class: Unclassified Product Code: MQC Dated: August 24, 2017 Received: August 25, 2017 Dear Ms. Dawn Norman: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. September 25, 2017 {2}------------------------------------------------ Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education (DICE) at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education (DICE) at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Sincerely, Mary S. Runner -S Tina Kiang, Ph.D. Acting Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {3}------------------------------------------------ # Indications for Use 510(k) Number (if known) K170985 Device Name Urbanek Device Indications for Use (Describe) The Urbanek Device is for the amelioration of clenching and bruxing associated with TMD and is to aid in the relief of symptoms of TMD/TMJ. | Type of Use (Select one or both, as applicable) | | |-------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------| | <div> <span> <span style="font-size: 20px;">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) </span> </div> | <div> <span> <span style="font-size: 20px;">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) </span> </div> | ### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. ### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {4}------------------------------------------------ ## K170985 ## 510(k) Summary | Proprietary Name | Urbanek Device | |--------------------------------|-----------------------------------------------------------------------------------------------------------------------------------| | Date of Submission: | September 25, 2017 | | Company: | TMJ Services<br>ATTN: Anthony P. Urbanek, DDS, MD, MS<br>2009 Mallory Ln<br>Franklin, TN 37067<br>Phone: 615-771-2189 | | Establishment<br>Registration: | N/A | | Primary Contact: | Dawn Norman, MS<br>Exec. Vice President MRC-X, LLC<br>6075 Poplar Avenue, Suite 500<br>Memphis, TN, 38119<br>Phone: 618-604-3064 | | Company/Secondary<br>Contact: | TMJ Services<br>Anthony P. Urbanek, DDS, MD, MS<br>2009 Mallory Ln<br>Franklin, TN 37067<br>Phone: 615-771-2189 | | Trade Name: | Urbanek Device | | Common Name: | Mouthguard, Prescription | | Classification: | Unclassified | | Product Code: | MQC, OCO | | Regulation Number: | Pre-amendment | | Panel: | Dental | {5}------------------------------------------------ ### Predicate Devices: Primary Predicate Device: - . NTI-TSS Inc.: NTI Tension Suppression System – K010876 (Cleared 06/20/2001) Reference Predicate Device: - MigraTherapy, LLC Brux: TMD QuickSplint-K111066 (Cleared 11/22/2011) . Device Description: The Urbanek Device is an intraoral, maxillary posterior dis-occluding device designed to aid in the treatment of clenching and bruxing associated with TMD/TMJ. The purpose of the device is to keep posterior maxillary and mandibular teeth (molars and premolars) from occluding in the centric or eccentric positions. Constructed of heat cured dental resin and two metal ball clasps for retention, the device is customized for each patient (per prescription by a dentist). > The device is designed to be worn for 24 hours per day (except at meals), 7 days per week for a set interval of time (e.g., 9 weeks). At scheduled follow-up visits, patients are assessed for symptom relief and the device inspected for comfort, proper design, excessive wear and vertical dimension for the inter-incisal distance with the device in place. The device may be adjusted for comfort and inter-incisal distance to ensure no posterior tooth or portion of a tooth occludes with the opposing tooth. > This device is effective because it allows the posterior teeth to be separated (disoccluded) for a period of time in the centric relation position. Disocclusion helps to relieve TMJ/TMD symptoms by stabilizing jaw muscles and joints. By stabilizing jaw muscles and joints to prevent overuse, inflammation of tissue within the temporomandibular joint resolves and healing is allowed to occur. - Indications for Use: The Urbanek Device is for the amelioration of clenching and bruxing associated with TMD and is to be used to aid in the relief of symptoms of TMD/TMJ. {6}------------------------------------------------ | Device | Urbanek Device<br>(Subject Device) | NTI-TSS NTI Tension<br>Suppression System<br>(K010876); Primary<br>Predicate | MigraTherapy LLC Brux- TMD<br>QuickSplint (K111066);<br>Reference | |-----------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Product Code | OCO, MQC | OCO | MQC | | Indications for<br>Use | The Urbanek Device is for<br>the amelioration of clenching<br>and bruxing associated with<br>TMD and is to be used to aid<br>in the relief of symptoms of<br>TMD/TMJ. | Treatment of tooth wear<br>from bruxing and<br>clenching<br>Treatment of muscle<br>pain associated with<br>muscle dysfunction<br>Diagnostic treatment<br>planning<br>The NTI-tss protocol is<br>indicated for the<br>prevention of medically<br>diagnosed migraine pain<br>and jaw disorders<br>through the reduction<br>of trigeminally<br>innervated muscular<br>activity | 1. Protection of teeth and<br>restorations from injury due<br>to bruxism or clenching.<br>2. Temporary relief of<br>Temporomandibular Joint<br>Disorder (TMD) and bruxism by<br>reducing muscle tension.<br>3. Temporary treatment of<br>Temporomandibular Disorder<br>(TMD) along with the relief of<br>associated headaches and pains.<br>The intended user includes patients<br>diagnosed with bruxism, patients<br>with headaches and pain related to<br>bruxism, Patients with TMD, patients<br>with headaches and pain related to<br>TMD and patients who may damage<br>teeth or dental restorations from the<br>clenching and grinding related to<br>bruxism activity. | | Material | • Thermoplastic<br>resin<br>• ASTM 313<br>Stainless Steel | • Thermoplastic<br>resins | • Polycarbonate tray<br>• vinyl polysiloxane liner | | Mechanism<br>of action | Disocclusion | Disocclusion | Disocclusion | | Removable? | Yes | Yes | Yes | | Device<br>placement | Palate and anterior<br>maxillary arch | Anterior maxillary or<br>mandibular arch | Anterior maxillary or<br>mandibular arch | | Location of<br>disocclusion | Upper molars and<br>premolars | Lower incisors | Upper/lower incisors from<br>bicuspid to<br>bicuspid | ## Comparison of Subject and Predicate Intraoral Devices {7}------------------------------------------------ | Technological<br>Characteristics: | The material base for the Urbanek Device is FDA cleared and previously<br>demonstrated to be safe for dental use. The Urbanek Device is<br>substantially equivalent to the predicate and reference devices as its<br>primary mode of action, like that of the predicate and reference devices, is<br>to keep the posterior maxillary and mandibular teeth (molars and<br>premolars) from occluding in the centric or eccentric positions. This<br>disocclusion results in the alleviation of pressure on the<br>temporomandibular joints. The primary difference between the proposed<br>device and the predicate(s) is where the device sits within the oral cavity.<br>The Urbanek Device sits on the palate of the mouth while the NTI-TSS NT<br>Tension Suppression System uses the lower incisors and the Brux TMD<br>QuickSplint is an anterior bite plate that covers either the upper or lower<br>front teeth from bicuspid to bicuspid. Although, the positions of the device<br>within the oral cavity are different, the intent in all of the devices is<br>disocclusion. The proposed and predicate devices are all patient specific<br>and reusable for the given patient. | |-----------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Performance Testing: | Performance testing has not been performed as there are no recognized<br>standards to which prescription mouthguards are tested; however, the<br>material composition of the Urbanek Device base has been found in legally<br>marketed devices and found previously safe for dental use since 2001<br>and meets the requirements of ISO 20795-1:2008 Dentistry -- Base<br>polymers -- Part I: Denture base polymers (previously ISO 1567:1999). | | Conclusion: | The subject Urbanek Device has similar indications for use, mode of<br>action, technological characteristics and materials as the predicate<br>devices. The minor differences between the Urbanek device and the<br>predicate devices raise no new questions of safety and effectiveness.<br>Thus, the Urbanek Device is substantially equivalent to the predicate<br>devices. |
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...