ChitoPulse 9in / 23 cm, ChitoPulse 12in / 30.5 cm, ChitoPulse 15in / 38 cm

K170958 · Tricol Biomedical, Inc. · DXC · Aug 29, 2017 · Cardiovascular

Device Facts

Record IDK170958
Device NameChitoPulse 9in / 23 cm, ChitoPulse 12in / 30.5 cm, ChitoPulse 15in / 38 cm
ApplicantTricol Biomedical, Inc.
Product CodeDXC · Cardiovascular
Decision DateAug 29, 2017
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 870.4450
Device ClassClass 2
AttributesTherapeutic

Intended Use

ChitoPulse™ Radial Hemostatic Device is intended to promote hemostasis following a catheterization or other puncture into a blood vessel in a patient's radial artery, including arterial or venous line or sheath removal, hemodialysis and in patients on anticoagulation therapy. ChitoPulse™ Hemostatic Device is intended to promote hemostasis following a catheterization or other puncture into a blood vessel in a patient's leg, including dorsalis pedis or tibial blood vessels, arterial or venous line or sheath removal and in patients on anticoagulation therapy.

Device Story

Mechanical compression band with integrated hemostatic chitosan patch; used for radial artery or leg vascular access sites. Operator (trained medical professional) secures band over puncture site using strap fastener; inflates balloon via syringe and luer lock valve to apply localized pressure. Chitosan patch provides adjunct bleeding control and bacterial barrier. Once hemostasis achieved, band removed; chitosan patch remains in situ for up to 24 hours, secured by dressing. Benefits include rapid bleeding control and reduced bacterial penetration risk. Single-use device.

Clinical Evidence

No clinical data provided. Substantial equivalence supported by bench testing, including biocompatibility (ISO 10993), pressure equivalence testing, and bacterial barrier testing (AATCC 100-2004).

Technological Characteristics

Biocompatible PVC strap with hook-and-loop fastener; inflatable balloon for compression; chitosan patch for adjunct hemostasis. Dimensions: 9in, 12in, 15in. Sterilization: Gamma irradiation (SAL 10^-6).

Indications for Use

Indicated for patients undergoing catheterization or vascular puncture (radial artery, leg, dorsalis pedis, or tibial vessels) requiring hemostasis, including arterial/venous line or sheath removal, hemodialysis, and patients on anticoagulation therapy.

Regulatory Classification

Identification

A vascular clamp is a surgical instrument used to occlude a blood vessel temporarily.

Predicate Devices

Reference Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/2 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, stacked on top of each other. Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002 August 29, 2017 Tricol Biomedical. Inc. Máire Ní Beilliú Vice President Regulatory & Quality 720 SW Washington Street, Suite 200 Portland, Oregon 97205 Re: K170958 Trade/Device Name: ChitoPulse 9in/23cm, ChitoPulse 12in/30.5cm, ChitoPulse 15in/38cm Regulation Number: 21 CFR 870.4450 Regulation Name: Vascular Clamp Regulatory Class: Class II Product Code: DXC Dated: Julv 26, 2017 Received: July 31, 2017 Dear Dr. Ní Beilliú: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device {1}------------------------------------------------ related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education (DICE) at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education (DICE) at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Sincerely. Nicole G. Ibrahim -S for Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ ## Indications for Use 510(k) Number (if known) K170958 Device Name HemCon® ChitoPulse™ #### Indications for Use (Describe) ChitoPulse™ Radial Hemostatic Device is intended to promote hemostasis following a catheterization or other puncture into a blood vessel in a patient's radial artery, including arterial or venous line or sheath removal, hemodialysis and in patients on anticoagulation therapy. ChitoPulse™ Hemostatic Device is intended to promote hemostasis following a catheterization or other puncture into a blood vessel in a patient's leg, including dorsalis pedis or tibial blood vessels, arterial or venous line or sheath removal and in patients on anticoagulation therapy. | Type of Use (Select one or both, as applicable) | |-------------------------------------------------| |-------------------------------------------------| X Prescription Use (Part 21 CFR 801 Subpart D) | Over-The-Counter Use (21 CFR 801 Subpart C) ## CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. ## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ 720 SW Washington Street, Suite 200 Portland, OR 97205-3504 USA TEL: +1.503.245.0459 FAX: +1.971.223.1000 WEB: www.tricolbiomedical.com # 1. 510(k) Summary Tricol Biomedical Inc 510(k) Notification ## l. Submitter: | Company Name: | Tricol Biomedical, Inc. | |-----------------------------|----------------------------------------------------------------| | Company Address: | 720 SW Washington Street, Suite 200<br>Portland, OR 97205-3504 | | Contact Person: | Máire E. Ní Beilliú PhD<br>Vice President Regulatory & Quality | | Contact Phone: | (971) 327.5729 | | Contact Fax: | (503) 245.1326 | | Date of Preparation: | 28th August 2017 | | II. Device | | | Trade Name: | HemCon® ChitoPulseTM | | Common Name: | Vascular Compression Device, Hemostat | | Classification Name: | Vascular Clamp (21 CFR 870.4450) | | Product Code: | DXC | | Regulatory Class: | Class II (Performance Standards) | | Classification Panel: | 870 - Cardiovascular Devices | | III. Predicate Device(s): | | | Primary Predicate Device: | Zephyr Vascular Compression Devices (K151363) | | Secondary Predicate Device: | Vasc Band Hemostat (K142359) | | Reference Device(s): | HemCon® Patch PRO (K150916) | #### IV. Description of the Device: ChitoPulse™ is a mechanical compression band incorporating a hemostatic chitosan patch. The compression band and chitosan patch are positioned over the radial access site and secured in place by a strap fastener. Addition of air to the compression band balloon provides rapid bleeding control by localized compression. The chitosan patch provides adjunct bleeding control to the balloon compression and provides a barrier against bacterial penetration by a wide range of gram positive and gram negative organisms. {4}------------------------------------------------ TEL: +1.503.245.0459 FAX: +1.971.223.1000 WEB: www.tricolbiomedical.com Compression is applied or reduced using a syringe and luer lock valve to add or remove air volume to or from the balloon. Once hemostasis has been achieved, compression may be reduced by gradual removal of air from the balloon. ChitoPulse™ is a single use device. Once hemostasis is achieved, the compression band is intended to be removed leaving the hemostatic chitosan patch in place. The hemostatic chitosan patch can be secured in place with a securement dressing for up to 24 hrs. #### V. Indications for Use: #### HemCon® ChitoPulse™ Radial Hemostatic Device ChitoPulse™ is intended to promote hemostasis following a catheterization or other into a blood vessel in a patient's radial artery, including arterial or venous line or sheath removal, hemodialysis and in patients on anticoagulation therapy #### HemCon® ChitoPulse Hemostatic Device ChitoPulse™ is intended to promote hemostasis following a catheterization or other puncture into a blood vessel in a patient's leg, including dorsalis pedis or tibial blood vessels, arterial or venous line or sheath removal and in patients on anticoagulation therapy. ChitoPulse™ Radial Hemostatic Device has the same indications for use as the primary predicate device and same intended use as the secondary predicate device. ChitoPulse™ Hemostatic Device has the same indications for use as the primary predicate device Both the subject and predicate devices have substantially equivalent intended use as compression devices to promote hemostasis following a catheterization or other puncture into a blood vessel in a patients arm or leg, including: radial, dorsalis pedis or tibial blood vessels, arterial or venous line or sheath removal, hemodialysis and in patients on anticoaqulation therapy. Both the subject and predicate devices are intended to be used by trained medical professionals. #### VI. Comparison of Technological Characteristics with the Predicate Devices: Both the subject device and primary predicate device are intended to promote hemostasis following a catheterization or other puncture into a blood vessel in a patients arm or leg, including: radial, brachial, dorsalis pedis or tibial blood vessels, arterial or venous line or sheath removal, hemodialysis and in patients on anticoagulation therapy. Both the subject device and predicate devices are composed of biocompatible PVC straps that are applied around a limb and fastened with hook and loop material. {5}------------------------------------------------ TEL: +1.503.245.0459 FAX: +1.971.223.1000 WEB: www.tricolbiomedical.com Both the subject and predicate devices use inflatable balloons to apply compression to the limb thereby applying pressure to the underlying vasculature. The subject and predicate devices differ in that the subject device has a chitosan patch that is intended to be left in situ for up to 24 hours following removal of the compression band. The chitosan patch acts as an adjunct to the mechanical compression provided by the strap and inflatable balloon. #### VII. Performance Data: To demonstrate substantial performance equivalence to the predicate devices, biocompatibility and bench top testing was performed. Performance testing included: - Biocompatibility testing - Pressure equivalence testing ● - . Bacterial barrier testing (chitosan patch) #### Sterilitv Each ChitoPulse™ is individually packaged in a heat sealed foil pouch and is terminally sterilized by gamma irradiation to a sterility assurance level (SAL) of 10 °. #### Biocompatibility Testing Biocompatibility testing for ChitoPulse™ has been conducted in accordance with the FDA Guidance Document "Use of International Standard ISO 10993, "Biological Evaluation of Medical Devices Part 1: Evaluation and Testing", June 16" 2016 and International Standard ISO 10993-1 "Biological Evaluation of Medical Devices – Part 1: Evaluation and Testing within a Risk Management Process" as recognized bv FDA. Categorization of the device by nature of body contact deduces that ChitoPulse™ is a surface contacting device, device is intended for limited contact (≤ 24hrs). Cytotoxicity, intracutaneous irritation, sensitization, and acute systemic toxicity were performed by contract testing laboratories per ISO 10993 standard protocols. #### Bacterial Barrier Properties The barrier properties of the chitosan patch have been demonstrated in vitro using AATCC Test Method 100-2004, Antibacterial Finishes on Textiles - Evaluation of (Technical Manual of the American Association of Textile Chemists and Colorists). The results of the test demonstrated the chitosan patch is capable of preventing bacterial penetration. #### Performance Testing Benchtop pressure equivalence testing was performed to demonstrate that ChitoPulse™ applies equivalent pressure as the predicate Vasc Band Hemostat device. The tensile strength of the ChitoPulse™ strap was also evaluated and demonstrated to be equivalent to the predicate Vasc Band Hemostat device. {6}------------------------------------------------ ## Clinical Performance Data: No clinical data was required for evaluation of this device. ## Summary: The conclusion drawn from the technological characteristics and non-clinical performance data is that the HemCon® ChitoPulse™ has been found to be substantially equivalent to the predicate devices. Image /page/6/Picture/5 description: The image shows a pattern of rounded rectangles arranged in a circular fashion. There are eight rounded rectangles in total, with four of them being a light gray color and the other four being a lighter, almost white color. The rectangles are oriented diagonally, creating a sense of movement and symmetry in the image. The overall effect is a simple, abstract design.
Innolitics
510(k) Summary
Decision Summary
Classification Order
Enter a record ID and click Load to view the document.
100%