← Product Code [OLO](/productcode/OLO) · K170937

# VIPER PRIME navigated inserter (K170937)

_Medos International SARL · OLO · Sep 14, 2017 · Neurology · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K170937

## Device Facts

- **Applicant:** Medos International SARL
- **Product Code:** [OLO](/productcode/OLO.md)
- **Decision Date:** Sep 14, 2017
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 882.4560
- **Device Class:** Class 2
- **Review Panel:** Neurology
- **Attributes:** Therapeutic

## Indications for Use

The VIPER PRIME navigated inserter is a navigated instrument for insertion of VIPER PRIME screws in open or percutaneous procedures. The VIPER PRIME navigated inserter is inspinal surgical procedures, in which: · use of the VIPER System is indicated, · use of stereotactic surgery may be appropriate, and · where reference to a rigid anatomical structure, such as the pelvis or a vertebrae can be identified relative to the acquired image (CT, MR, 2D fluoroscopic image or 3D fluoroscopic image reconstruction) and/or an image data based model of the anatomy using a navigation system which includes universal tracking arrays supplied by the navigation manufacturer. These procedures include but are not limited to spinal fusion. The VIPER PRIME navigated inserter requires manual calibration.

## Device Story

Reusable manual screwdriver for insertion of VIPER PRIME screws during open or percutaneous spinal surgery; features attachment sites for third-party universal tracking arrays; requires manual calibration with navigation systems. Device used by surgeons in OR; provides physical interface for screw placement while tracking position relative to patient anatomy via external navigation system. Benefits include enhanced precision in screw placement during spinal fusion procedures.

## Clinical Evidence

Bench testing only. Non-clinical sawbones testing confirmed device performance, including assembly with third-party tracking arrays, manual calibration, and navigated screw insertion. Final screw position verified via secondary imaging modality.

## Technological Characteristics

Manual screwdriver; materials: stainless steel, titanium alloy (Ti-6Al-4V), and polyphenylsulfone (RADEL). Features attachment sites for third-party universal tracking arrays. Requires manual calibration. Standalone mechanical instrument used in conjunction with external navigation systems.

## Regulatory Identification

A stereotaxic instrument is a device consisting of a rigid frame with a calibrated guide mechanism for precisely positioning probes or other devices within a patient's brain, spinal cord, or other part of the nervous system.

## Predicate Devices

- Synthes Navigable Pedicle Preparation Instruments ([K122211](/device/K122211.md))

## Reference Devices

- Stryker SpineMap® 3D Navigation System ([K141941](/device/K141941.md))
- Medtronic StealthStation System ([K133444](/device/K133444.md))
- Brainlab ([K070106](/device/K070106.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

### DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized image of three human profiles facing right, with a wing-like shape extending above them.

Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

Medos International, SARL % Laura Bleyendaal Senior Regulatory Affairs Specialist DePuy Synthes 325 Paramount Drive Raynham, Massachusetts 02767

September 14, 2017

Re: K170937

Trade/Device Name: VIPER PRIME navigated inserter Regulation Number: 21 CFR 882.4560 Regulation Name: Stereotaxic Instrument Regulatory Class: Class II Product Code: OLO Dated: May 5, 2017 Received: May 8, 2017

Dear Laura Bleyendaal:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices. good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device

{1}------------------------------------------------

related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education (DICE) at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education (DICE) at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely,

# Mark N. Melkerson -S

Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

# Indications for Use

510(k) Number (if known) K170937

Device Name VIPER PRIME™ navigated inserter

Indications for Use (Describe)

The VIPER PRIME navigated inserter is a navigated instrument for insertion of VIPER PRIME screws in open or percutaneous procedures. The VIPER PRIME navigated inserter is inspinal surgical procedures, in which:

· use of the VIPER System is indicated,

· use of stereotactic surgery may be appropriate, and

· where reference to a rigid anatomical structure, such as the pelvis or a vertebrae can be identified relative to the acquired image (CT, MR, 2D fluoroscopic image or 3D fluoroscopic image reconstruction) and/or an image data based model of the anatomy using a navigation system which includes universal tracking arrays supplied by the navigation manufacturer. These procedures include but are not limited to spinal fusion. The VIPER PRIME navigated inserter requires manual calibration.

| Type of Use (Select one or both, as applicable) |  |
|-------------------------------------------------|--|
|-------------------------------------------------|--|

X Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

# 510(k) Summary

B.

C.

### A. Submitter Information

| 510(k) Sponsor:                            | Medos International, SARL                                                     |
|--------------------------------------------|-------------------------------------------------------------------------------|
| Contact Person:                            | Laura Bleyendaal<br>DePuy Synthes<br>325 Paramount Drive<br>Raynham, MA 02767 |
| Telephone number:<br>Fax number:<br>Email: | (508) 828-3267<br>(508) 828-3797<br>Lbleyend@its.jnj.com                      |
| Date Prepared                              | September 13, 2017                                                            |
| Device Name                                |                                                                               |
| Trade/Proprietary Name:                    | VIPER PRIME™ navigated inserter                                               |
| Common/Usual Name:                         | Orthopedic stereotaxic instrument                                             |
| Device Classification<br>and Regulation:   | Class II per 21 CFR § 882.4560                                                |

Classification Product and Panel OLO; Orthopedic Code:

### D. Predicate Device Name

Primary Predicate Device: Synthes Navigable Pedicle Preparation Instruments (K122211)

Reference Devices: Stryker SpineMap® 3D Navigation System (K141941) Medtronic StealthStation『System (K133444) Brainlab (K070106)

{4}------------------------------------------------

#### E. Device Description

The VIPER PRIME™ navigated inserter is a reusable manual screwdriver for insertion of the VIPER PRIME screws of the VIPER System in open and percutaneous procedures. The VIPER PRIME navigated inserter also features attachment sites for universal tracking arrays supplied by the navigation manufacturer to enable use with the respective spine navigation system. The VIPER PRIME navigated inserter must be manually calibrated with the third-party navigation system.

### F. Indications for Use

The VIPER PRIME navigated inserter is a navigated instrument for insertion of VIPER PRIME screws in open or percutaneous procedures. The VIPER PRIME navigated inserter is indicated for use in spinal surgical procedures, in which:

- use of the VIPER System is indicated,
- use of stereotactic surgery may be appropriate, and
- where reference to a rigid anatomical structure, such as the pelvis or a ● vertebrae can be identified relative to the acquired image (CT, MR, 2D fluoroscopic image or 3D fluoroscopic image reconstruction) and/or an image data based model of the anatomy using a navigation system which includes universal tracking arrays supplied by the navigation manufacturer.

These procedures include but are not limited to spinal fusion. The VIPER PRIME navigated inserter requires manual calibration.

### G. Summary of Similarities and Differences in Technological Characteristics, Performance and Intended Use

The technological characteristics, including materials, design and performance of the VIPER PRIME navigated inserter are consistent with those of the primary predicate device. In brief, both the primary predicate and subject devices are indicated for use in open or percutaneous spinal surgical procedures in which the use of stereotactic surgery may be appropriate. The Synthes Navigable Pedicle Preparation instruments include awls, probes, taps, screwdrivers, a dilator, a guide tube and a pedicle access cannula while the VIPER PRIME navigated inserter is a screwdriver. The Synthes Navigable Pedicle Preparation instruments are manufactured from surgical grade metals, silicone and polyetheretherketone (PEEK), while the VIPER PRIME navigated inserter is manufactured from stainless steel, titanium alloy (Ti-6Al-4V) and polyphenylsulfone (RADEL"). Both the Synthes Navigable Pedicle Preparation instruments and VIPER PRIME navigated inserter feature attachment sites for universal tracking arrays supplied by the navigation manufacturer to enable use with the respective spine navigation system. The VIPER PRIME navigated inserter requires manual calibration and similarly, the Synthes

{5}------------------------------------------------

Navigable Pedicle Preparation instruments include instruments which may be manually calibrated with the spine navigation systems.

#### H. Materials

The VIPER PRIME navigated inserter is manufactured from stainless steel, titanium alloy (Ti-6Al-4V) and RADEL .

### l. Performance Data

Non-clinical sawbones testing was conducted to confirm device performance for the intended use. The testing consisted of assembly of the VIPER PRIME navigated inserter with the third party universal tracking arrays, manual calibration and navigated insertion of VIPER PRIME screws in a sawbones model. Final screw position in the software was confirmed using a second imaging modality.

### J. Conclusion

The indications for use and intended use of the VIPER PRIME navigated inserter are consistent with those of the primary predicate device. The technological characteristics of the VIPER PRIME navigated inserter in terms of design, materials and performance are consistent with those of the predicate device. The VIPER PRIME navigated inserter is substantially equivalent to the primary predicate device.

---

**Source:** [https://fda.innolitics.com/device/K170937](https://fda.innolitics.com/device/K170937)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
