Bi-Ostetic Bioactive Glass Foam

K170917 · Berkeley Advanced Biomaterials · MQV · Oct 30, 2017 · Orthopedic

Device Facts

Record IDK170917
Device NameBi-Ostetic Bioactive Glass Foam
ApplicantBerkeley Advanced Biomaterials
Product CodeMQV · Orthopedic
Decision DateOct 30, 2017
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 888.3045
Device ClassClass 2
AttributesTherapeutic

Intended Use

Bi-Ostetic Bioactive Glass Foam is indicated for use in bony voids or gaps of the skeletal system (i.e. extremities, pelvis and posterolateral spine). These osseous defects are surgically created or result from traumatic injury to the bone, and are not intrinsic to the stability of the bony structure. The graft resorbs and is replaced by the growth of new bone during the healing process. Bi-Ostetic Bioactive Glass Foam must be used with autogenous bone marrow aspirate and autograft in the posterolateral spine.

Device Story

Bone void filler; collagen matrix mineralized with hydroxyapatite (HA), tri-calcium phosphate (TCP), and 45S5 bioactive glass granules. Used in orthopedic surgery to fill osseous defects. Surgeon hydrates graft with sterile water, blood, or bone marrow aspirate; mixes with autograft bone for spine applications. Provides osteoconductive scaffold for new bone growth; graft resorbs and remodels into bone over time. Used in clinical settings by orthopedic surgeons. Benefits patient by facilitating bone healing in non-structural voids.

Clinical Evidence

No clinical data. Evidence based on non-clinical bench testing and animal models. Biocompatibility confirmed per ISO 10993. Bioactivity demonstrated via apatite layer formation in simulated body fluid. Endotoxin levels verified per ANSI/AAMI ST72 and ISO 10993-11. Animal studies compared performance to FormaGraft in spine and femoral models, showing substantial equivalence.

Technological Characteristics

Collagen matrix with 60 wt% HA, 40 wt% TCP, and 20 wt% 45S5 bioactive glass granules. Available as granules, strips, or putty. Sterilization per ISO 11137. Characterization via FTIR, XRD, ICP-MS, and SEM. Non-pyrogenic per ISO 10993-11.

Indications for Use

Indicated for patients with surgically created or traumatic bony voids or gaps in the skeletal system (extremities, pelvis, posterolateral spine) not intrinsic to structural stability. Must be used with autogenous bone marrow aspirate and autograft in posterolateral spine.

Regulatory Classification

Identification

A resorbable calcium salt bone void filler device is a resorbable implant intended to fill bony voids or gaps of the extremities, spine, and pelvis that are caused by trauma or surgery and are not intrinsic to the stability of the bony structure.

Special Controls

*Classification.* Class II (special controls). The special control for this device is the FDA guidance document entitled “Class II Special Controls Guidance: Resorbable Calcium Salt Bone Void Filler Device; Guidance for Industry and FDA.” See § 888.1(e) of this chapter for the availability of this guidance.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue. Berkeley Advanced Biomaterials François Génin, Ph.D. Chief Executive Officer 901 Gravson Street. Suite 101 Berkeley, California 94710 October 30, 2017 Re: K170917 Trade/Device Name: Bi-Ostetic Bioactive Glass Foam Regulation Number: 21 CFR 888.3045 Regulation Name: Resorbable calcium salt bone void filler device Regulatory Class: Class II Product Code: MQV Dated: September 25, 2017 Received: September 29, 2017 Dear Dr. Génin: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. {1}------------------------------------------------ If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education (DICE) at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevicesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education (DICE) at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Sincerely, Mark N. Melkerson -S Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement below. # Indications for Use 510(k) Number (if known) K170917 Device Name Bi-Ostetic Bioactive Glass Foam #### Indications for Use (Describe) Bi-Ostetic Bioactive Glass Foam is indicated for use in bony voids or gaps of the skeletal system (i.e. extremities, pelvis and posterolateral spine). These osseous defects are surgically created or result from traumatic injury to the bone, and are not intrinsic to the stability of the bony structure. The graft resorbs and is replaced by the growth of new bone during the healing process. Bi-Ostetic Bioactive Glass Foam must be used with autogenous bone marrow aspirate and autograft in the posterolateral spine. | Type of Use (Select one or both, as applicable) | | |----------------------------------------------------------------------------|---------------------------------------------------------------------------| | <div> <span>☑</span> Prescription Use (Part 21 CFR 801 Subpart D) </div> | <div> <span>☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) </div> | #### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ ## BERKELEY ADVANCED BIOMATERIALS 901 Grayson Street, Suite 101, Berkeley, CA 94710, USA Tel: (510) 883 0500; Fax: (510) 883 0511 http://www.ostetic.com Image /page/3/Picture/2 description: The image shows a logo consisting of a circular arrangement of blue dots. The dots vary in size, with the largest dots concentrated in the center and the smaller dots positioned towards the outer edges. The overall effect is a gradient or radial pattern, creating a visually dynamic and modern design. # 510(K) Summary - Bi-Ostetic Bioactive Glass Foam In accordance with the Food and Drug Admisnistration Rule to implement provisions of the Safe Medical Devices Act of 1990 and in conformance with 21 CFR 807, this information serves as a Summary of Safety and Effectiveness for the use of the device. | Submitted By: | Berkeley Advanced Biomaterials | |--------------------------|-------------------------------------------------| | Date: | 27 March 2017 | | Contact Person: | François Génin, Ph.D. | | Position: | Chief Executive Officer | | Contact Information | Phone: 510-883-0500; Fax: 510-883-0511 | | Proprietary Name: | Bi-Ostetic Bioactive Glass Foam | | Regulation Name: | Resorbable Calcium Salt Bone Void Filler Device | | Regulation Number: | 888.3045 | | Classification: | Class II | | Device Code/ Panel Code: | Orthopedics/87/MQV | ## DEVICE INFORMATION ### A. INTENDED USE Bi-Ostetic Bioactive Glass Foam is indicated for use in bony voids or gaps of the skeletal system (i.e. extremities, pelvis and posterolateral spine). These osseous defects are surgically created or result from traumatic injury to the bone, and are not intrinsic to the stability of the bony structure. The graft resorbs and is replaced by the growth of new bone during the healing process. Bi-Ostetic Bioactive Glass Foam must be used with autogenous bone marrow aspirate and autograft in the posterolateral spine. ### B. DEVICE DESCRIPTION The device is a bone void filler consisting of a collagen matrix mineralized with hydroxyapatite (HA) - tri-calcium phosphate (TCP) granules and 45S5 bioactive glass granules. The bovine fibrillar collagen component is biocompatible. The 45S5 glass particles are bioactive. The device is available in the form of granules, strips and putty. The composition of the granules is 60 wt% HA and 40 wt% TCP. The amount of 45S5 granules in the graft is 20 wt%. When used in the extremities, the granules and strips can be hydrated with sterile water and blood (nominally 1:1). When used in the posterolateral spine, the graft material must be hydrated with bone marrow aspirate (1:1 ratio), then mixed with autograft bone (1:1 ratio). The defect site should be filled as completely as possible. The device provides a scaffold around which new bone can grow. A series of surface reactions on the particles results in the formation of a calcium phosphate layer that is substantially equivalent in composition and structure to the hydroxyapatite found in bone mineral. This apatite layer on the 4555 granules and the hydroxyapatite- tri-calcium phosphate granules provide an osteoconductive scaffold onto which the patient's new bone will grow. During healing, the graft particulate is absorbed and remodeled into new bone. ### C. SUBSTANTIAL EQUIVALENCE INFORMATION The intended use, materials and design features of the device are substantially equivalent to the predicate devices FormaGraft and Bi-Ostetic Foam. Bi-Ostetic Bioactive Glass Foam has the {4}------------------------------------------------ same classification as the predicate devices. It has the same intended use and the same or similar technological characteristics, principles of operation and indications as the predicate devices. FormaGraft, Bi-Ostetic Foam and Bi-Ostetic Bioactive Glass Foam all contain the same HA-TCP granules and similar amounts of bovine collagen. Bi-Ostetic Bioactive Glass is provided in strip and putty forms like the Bi-Ostetic Foam (FormaGraft comes only as strips). Device dosage, packaging and sterilization methods are the same as Bi-Ostetic Foam. Bi-Ostetic Bioactive Glass Foam can be mixed with autogenous bone marrow like FormaGraft (Bi-Ostetic Foam is mixed with blood or saline). The safety and effectiveness of the devices are adequately supported by the substantial equivalence information provided within the Premarket Notification and as demonstrated by comparative evaluation in animal studies. #### D. TECHNOLOGICAL CHARACTERISTICS AND SUBSTANTIAL EQUIVALENCE The chemistry and specifications of the granules (hydroxyapatite - beta tri-calcium phosphate and 45S5 bioactive glass) in a collagen matrix are subtantially equivalent to the predicate device. Testing of the chemistry of the components was performed by FTIR and X-Ray Diffraction analysis. The purity of the components was measured by ICP-MS. Physical properties were evaluated by Scanning Electron Microscopy (SEM). Sterilization is performed according to ISO 11137 standards. Bi-Ostetic BGF Foam is Bi-Ostetic Foam (K092046) in which part of the HA-TCP granules has been substituted with 45S5 granules. Bi-Ostetic BGF Foam is also similar in characteristics to FormaGraft (K050789). The radiographic, morphometric and histologic performance of the device was compared to that of the predicate. ### E. PERFORMANCE DATA In vivo and in vitro testing of the device in accordance to ISO 10993 Biological Evaluation of Medical Devices has demonstrated that the materials are safe and biocompatible. The device is considered bioactive based on in vitro studies that show apatite layer formation on the surface of the bioactive glass particles following immersion in simulated body fluid (SBF). The device was tested to establish non-pyrogenicity according ISO 10993-11. Limulus Amebocyte Lysate (LAL) batch testing demonstrated that bacterial endotoxin levels were below the standard 0.5EU/mL and 20EU/device in accordance with ANSI/AAMI ST72(2002), and ISO 10993-11. The performance of the device in postero-lateral spine fusion and femoral cancellous defect animal models was compared to the performance of FormaGraft. The results showed substantial equivalence with FormaGraft. Test results have not been correlated to clinical performance. The results are consistent with the literature and previous laboratory results for this model. The conclusions drawn from the nonclinical tests demonstrate that the device is as safe, as effective, and performs as well as or better than the predicate devices.
Innolitics
510(k) Summary
Decision Summary
Classification Order
Enter a record ID and click Load to view the document.
100%