← Product Code [NHI](/productcode/NHI) · K170902

# Sprint System, Single Procedure Kit, Sprint System with Pad II, Single Procedure Kit, Sprint System, Multi-Procedure Kit, Sprint System, Sterile Components Kit, Sprint System, Stimulator Kit (K170902)

_Spr Therapeutics, LLC · NHI · Jun 26, 2017 · Neurology · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K170902

## Device Facts

- **Applicant:** Spr Therapeutics, LLC
- **Product Code:** [NHI](/productcode/NHI.md)
- **Decision Date:** Jun 26, 2017
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 882.5890
- **Device Class:** Class 2
- **Review Panel:** Neurology
- **Attributes:** Therapeutic

## Indications for Use

The SPRINT Peripheral Nerve Stimulation (PNS) System is indicated for up to 60 days in the back and/or extremities for: - Symptomatic relief of chronic, intractable pain, post-surgical and post-traumatic acute pain; - Symptomatic relief of post-traumatic pain; - Symptomatic relief of post-operative pain. The SPRINT PNS System is not intended to treat pain in the craniofacial region.

## Device Story

System consists of percutaneous stainless steel coiled wire electrode inserted via introducer needle near target peripheral nerve; wearable external stimulator delivers electrical stimulation therapy. Used in clinical or home settings; operated by patient/clinician. Device provides symptomatic pain relief; therapy duration up to 60 days. Healthcare provider uses output to manage pain conditions. Benefits include non-pharmacological pain management.

## Clinical Evidence

Clinical data provided for 60-day extension. Studies on chronic, post-surgical, and post-traumatic pain demonstrated safety; reported adverse events (skin irritation, erythema, blisters, skin tears) were non-serious and resolved quickly. A randomized controlled trial for post-stroke shoulder pain failed to show effectiveness. Clinical experience from other electrical stimulation devices supports long-term use.

## Technological Characteristics

Percutaneous electrode: flexible, coiled, stainless steel wire. External stimulator: wearable, battery-powered (25mAhr capacity). System is percutaneous, indwelling for up to 60 days. Electrical stimulation therapy. Software verification and validation performed. Biocompatibility and electromagnetic compatibility testing conducted.

## Regulatory Identification

A transcutaneous electrical nerve stimulator for pain relief is a device used to apply an electrical current to electrodes on a patient's skin to treat pain.

## Predicate Devices

- Smartpatch PNS System ([K161154](/device/K161154.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Public Health Service

Image /page/0/Picture/2 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, with flowing lines representing hair or a cape.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002 June 26, 2017

SPR Therapeutics, LLC Kathryn Stager Manager of Regulatory Affairs 22901 Millcreek Blvd.. Suite 110 Cleveland, Ohio 44122

Re: K170902

Trade/Device Name: Sprint System. Single Procedure Kit. Sprint System with Pad II, Single Procedure Kit, Sprint System, Multi-Procedure Kit, Sprint System, Sterile Components Kit, Sprint System, Stimulator Kit Regulation Number: 21 CFR 882.5890 Regulation Name: Transcutaneous Electrical Nerve Stimulator For Pain Relief Regulatory Class: Class II Product Code: NHI Dated: March 28, 2017 Received: March 28, 2017

Dear Kathryn Stager:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply 

{1}------------------------------------------------

with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely,

William J. Heetderks -S 2017.06.26 13:39:09 -04'00'

for Carlos L. Peña, PhD, MS Director Division of Neurological and Physical Medicine Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

# Indications for Use

## 510(k) Number (if known) K170902

### Device Name

SPRINT Peripheral Nerve Stimulation (PNS) System

Indications for Use (Describe)

The SPRINT Peripheral Nerve Stimulation (PNS) System is indicated for up to 60 days in the back and/or extremities for:

- · Symptomatic relief of chronic, intractable pain, post-surgical and post-traumatic acute pain;
- · Symptomatic relief of post-traumatic pain;
- · Symptomatic relief of post-operative pain.

The SPRINT PNS System is not intended to treat pain in the craniofacial region.

| Type of Use (Select one or both, as applicable) |  |
|-------------------------------------------------|--|
|-------------------------------------------------|--|

X Prescription Use (Part 21 CFR 801 Subpart D)

| Over-The-Counter Use (21 CFR 801 Subpart C)

## CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image contains the logo for SPR Therapeutics. The logo features the letters "SPR" in a large, green, sans-serif font. Below the letters, the words "SPR Therapeutics" are written in a smaller, gray, sans-serif font. Above the letters, there is a gray swoosh design.

# 510(k) Summary

#### 1. SUBMITTER

SPR Therapeutics, LLC 22901 Millcreek Boulevard, Suite 110 Cleveland, OH 44122

216-378-9108 (phone) 216-378-9116 (fax)

Contact Person: Kathryn Stager, MS, RAC, Manager of Regulatory Affairs Telephone: 216-378-9067

Date Prepared: March 28, 2017

#### 2. DEVICE

| Trade/Proprietary Name: | SPRINT PNS System                                                         |
|-------------------------|---------------------------------------------------------------------------|
| Common/Usual Name:      | Peripheral Nerve Stimulator                                               |
| Classification Name:    | Percutaneous Electrical Nerve Stimulation (PENS devices (21 CFR 882.5890) |
| Regulatory Class:       | II                                                                        |
| Product Code:           | NHI                                                                       |

#### 3. PREDICATE DEVICE

Smartpatch PNS System (K161154)*

#### 4. DEVICE DESCRIPTION

The Sprint PNS System is comprised of a percutaneous electrode placed via an introducer needle in proximity to a target peripheral nerve associated with a painful area and a wearable external stimulator that delivers stimulation therapy to the percutaneous electrode. The Sprint PNS System provides peripheral nerve stimulation (PNS) therapy to relieve pain. The percutaneous electrode is a sterile,

<sup>*</sup> The Smartpatch PNS System is presently being marketed under the name Sprint PNS System as reported in K161154/A001. The name Smartpatch is being used in this 510(k) application for clarity.

{4}------------------------------------------------

Image /page/4/Picture/0 description: The image shows the logo for SPR Therapeutics. The logo features the letters "SPR" in a large, green, sans-serif font. Below the letters, the words "SPR Therapeutics" are written in a smaller, gray, sans-serif font. A gray swoosh design curves over the top of the letters.

flexible, coiled, stainless steel wire designed to be percutaneously inserted through the skin via an introducer needle and remain indwelling for the duration of the therapy (up to 60 days).

#### 5. COMPARISON OF INDICATIONS FOR USE WITH THE PREDICATE DEVICE

## Indications Statement:

The SPRINT Peripheral Nerve Stimulation (PNS) System is indicated for up to 60 days in the back and/or extremities for:

- Symptomatic relief of chronic, intractable pain, post-surgical and post-. traumatic acute pain;
- Symptomatic relief of post-traumatic pain; ●
- Symptomatic relief of post-operative pain. ●

The SPRINT PNS System is not intended to treat pain in the craniofacial region.

## Comparison:

This indications statement is identical to the Smartpatch PNS System cleared in K161154 except that the duration of therapy was extended from 30 days to 60 days. Extending the treatment period to an upper limit of 60 days does not result in a new intended use when compared to the predicate device. In contrast to treatment periods longer than 60 days, this extension to an upper limit of 60 days does not raise different safety and effectiveness questions because the device, like the predicate device, is intended to provide treatment for the pain conditions specified in the indications statement (e.g., post-traumatic pain, post-operative pain) using a limited duration of therapy.

### COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATE 6. DEVICE

The purpose of this 510(k) is to obtain clearance for a change in the indication for use statement. The Sprint PNS System has the same technological characteristics as the Smartpatch PNS System. No changes have been made to the physical components of the system since clearance of K161154 other than:

- elimination of the longer lead length (clinical experience has revealed that . it provides no increased utility compared to the shorter lead, as the introducer needle is the same length),
- elimination of the lead connector tapes (clinical experience has revealed . that they provide no increased utility compared to commercially available bandages), and
- addition of a Pad that has a larger battery capacity (25mAhr compared to ●

{5}------------------------------------------------

Image /page/5/Picture/0 description: The image contains the logo for SPR Therapeutics. The logo features the letters "SPR" in a large, green font, with the words "SPR Therapeutics" in a smaller, gray font underneath. A gray swoosh design is above the letters.

14mAhr) to reduce the frequency at which Pads need to be replaced.

These minor modifications to the Smartpatch PNS System are intended to result in a simpler and easier to use product for the end-users (called the Sprint PNS System). They do not raise new questions regarding safety and effectiveness.

#### 7. PERFORMANCE DATA

Nonclinical testing of this device includes biocompatibility testing, electrical testing (safety and electromagnetic compatibility), system performance testing, and software verification and validation.

Clinical data was provided in K170902 to establish safety and effectiveness for extension of use (as per the IFU) of the SPRINT PNS system from 30 to 60 days. Clinical testing on the subject device includes multiple completed and ongoing studies on chronic or intractable pain, post-surgical pain, and post-traumatic pain. These studies have demonstrated safety of the therapy. The adverse events reported in the studies included skin irritation, erythema, a blister, or a mild skin tear. The majority of the adverse events were resolved with little to no intervention and resolved within a few days, and none were classified as serious. A randomized controlled trial failed to show that the Smartpatch was effective for post-stroke shoulder pain.

In addition to the clinical data provided in K170902, clinical experience from other electrical stimulation devices intended for the relief of pain, which deliver the same electrical stimulation therapy, demonstrates that there are no different or increased risks resulting from the long-term use of electrical stimulation for pain. Therefore, extending the treatment period to 60 days also does not raise different questions of effectiveness than the predicate or significantly increase an effectiveness concern.

#### 8. CONCLUSIONS

The Sprint PNS System has been shown to be substantially equivalent to the identified predicate device.

---

**Source:** [https://fda.innolitics.com/device/K170902](https://fda.innolitics.com/device/K170902)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
