HV Screw System

K170889 · Wrightmedicaltechnologyinc · HWC · Apr 26, 2017 · Orthopedic

Device Facts

Record IDK170889
Device NameHV Screw System
ApplicantWrightmedicaltechnologyinc
Product CodeHWC · Orthopedic
Decision DateApr 26, 2017
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 888.3040
Device ClassClass 2
AttributesTherapeutic

Intended Use

The HV Screw is indicated for fixation of bone fractures or for bone reconstruction. Examples include: - Mono or Bi-Cortical osteotomies in the foot or hand - Distal or Proximal metatarsal or metacarpal osteotomies - Fusion of the first metatarsophalangeal joint and interphalangeal joint - Fixation of osteotomies for Hallux Valgus treatment (such as Scarf, Chevron, etc.) - Akin type osteotomy - Arthrodesis base first metatarsal cuneiform joint to reposition and stabilize metatarsus varus primus - Talar/navicular arthrodesis

Device Story

HV Screw System consists of 3.0mm diameter cannulated titanium alloy screws with self-tapping distal and proximal threads. Used by orthopedic surgeons for bone fixation and reconstruction in foot and hand procedures, including osteotomies and arthrodesis. Device provides mechanical stabilization of bone segments to facilitate healing. Operates as a traditional orthopedic fixation fastener; no electronic or software components.

Clinical Evidence

No clinical data provided. Substantial equivalence supported by non-clinical bench testing including torsional testing, axial pullout force, bending strength, and bacterial endotoxin testing.

Technological Characteristics

Cannulated bone screw; 3.0mm diameter; titanium alloy construction; self-tapping distal and proximal threads. Mechanical fixation device; no energy source or software.

Indications for Use

Indicated for fixation of bone fractures or bone reconstruction in the foot or hand, including osteotomies, arthrodesis, and Hallux Valgus treatment.

Regulatory Classification

Identification

A smooth or threaded metallic bone fixation fastener is a device intended to be implanted that consists of a stiff wire segment or rod made of alloys, such as cobalt-chromium-molybdenum and stainless steel, and that may be smooth on the outside, fully or partially threaded, straight or U-shaped; and may be either blunt pointed, sharp pointed, or have a formed, slotted head on the end. It may be used for fixation of bone fractures, for bone reconstructions, as a guide pin for insertion of other implants, or it may be implanted through the skin so that a pulling force (traction) may be applied to the skeletal system.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002 April 26, 2017 Wright Medical Technology, Inc. Ms. Tara Conrad Regulatory Affairs Specialist II 1023 Cherry Road Memphis. Tennessee 38117 Re: K170889 Trade/Device Name: HV Screw System Regulation Number: 21 CFR 888.3040 Regulation Name: Smooth Or Threaded Metallic Bone Fixation Fastener Regulatory Class: Class II Product Code: HWC Dated: March 15, 2017 Received: March 24, 2017 Dear Ms. Conrad: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device- Image /page/0/Picture/11 description: The image is a circular seal for the U.S. Department of Health & Human Services. The seal features the department's name in a circular arrangement around the edge. In the center of the seal is a stylized image of three faces in profile, stacked on top of each other, creating a sense of depth and unity. {1}------------------------------------------------ related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Sincerely, # Mark N. Melkerson -S Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ ### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration # Indications for Use Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement on last page. # 510(k) Number (if known) K170889 Device Name HV Screw System Indications for Use (Describe) The HV Screw is indicated for fixation of bone fractures or for bone reconstruction. Examples include: - · Mono or Bi-Cortical osteotomies in the foot or hand - · Distal or Proximal metatarsal or metacarpal osteotomies - · Fusion of the first metatarsophalangeal joint and interphalangeal joint - · Fixation of osteotomies for Hallux Valgus treatment (such as Scarf, Chevron, etc.) - Akin type osteotomy - · Arthrodesis base first metatarsal cuneiform joint to reposition and stabilize metatarsus varus primus - · Talar/navicular arthrodesis | Type of Use (Select one or both, as applicable) | | |-------------------------------------------------|--| |-------------------------------------------------|--| X Prescription Use (Part 21 CFR 801 Subpart D) _ Over-The-Counter Use (21 CFR 801 Subpart C) ### PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED. ### FOR FDA USE ONLY Concurrence of Center for Devices and Radiological Health (CDRH) (Signature) {3}------------------------------------------------ K170889 Page 1 of 2 Headquarters Wright Medical Technology, Inc. mohis TN 38117 901 867 997 wmt.com Image /page/3/Picture/4 description: The image contains the logo for Wright. The logo consists of two overlapping trapezoids, one red and one orange, on the left side of the word "WRIGHT" in red, all caps. Below the word "WRIGHT" is the phrase "FOCUSED EXCELLENCE" in a smaller, gray font. # 510(K) SUMMARY OF SAFETY AND EFFECTIVENESS In accordance with the Food and Drug Administration Rule to implement provisions of the Safe Medical Devices Act of 1990 and in conformance with 21 CFR 807.92, this information serves as a Summary of Safety and Effectiveness for the USe of the HV Screw System. | 1. | Submitted By: | Wright Medical Technology, Inc.<br>1023 Cherry Road<br>Memphis, TN 38117 | | |----|-------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------|--| | | 510(k) Number: | K170889 | | | | Date: | April 25, 2017 | | | | Contact Person: | Tara Conrad<br>Regulatory Affairs Specialist II<br>Office<br>(901) 867-4367<br>(901) 867-4190<br>Fax | | | 2. | Proprietary Name:<br>Common Name:<br>Classification Name and Reference: | HV Screw System<br>Cannulated Bone Screw<br>21 CFR 888.3040- Class II- Smooth or threaded | | | | Device Product Code, Device Panel: | metallic bone fixation fastener<br>HWC - Orthopedic | | | 3. | Predicate Device: | K082320- Wright TM Compression Screw | | | 4. | Device Description | | | The HV Screws are cannulated screws offered in 3.0mm diameter, various lengths, and have self-tapping features on both distal and proximal threads. All screws are manufactured from titanium alloy. ### ഹ Intended Use and Indications for Use The HV Screw is indicated for fixation of bone fractures or for bone reconstruction. Examples include: - · Mono or Bi-Cortical osteotomies in the foot or hand - · Distal or Proximal metatarsal or metacarpal osteotomies - Fusion of the first metatarsophalangeal joint and interphalangeal joint - · Fixation of osteotomies for Hallux Valgus treatment (such as Scarf, Chevron, etc.) - · Akin type osteotomy {4}------------------------------------------------ - Arthrodesis base first metatarsal cuneiform joint to reposition and stabilize - metatarsus varus primus - Talar/navicular arthrodesis #### 6. Technological Characteristics Comparison The HV Screw System utilizes similar instrumentation, is made from identical materials, and has identical sterilization methods when compared to the legally marketed predicate device. ### 7. Substantial Equivalence- Non-Clinical Evidence | Validation and/or | Acceptance Value/Criteria | Results | | |--------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------|------------|--| | Verification Method | | | | | Torsional Testing<br>Insertion Torque<br>।<br>Removal Torque<br>।<br>Ultimate Torque<br>। | Ultimate Torque must<br>exceed insertion torque and<br>removal torque by a safety<br>factor | Acceptable | | | Engineering Analysis<br>Axial Pullout Force<br>Bending Strength<br>-<br>Torsional Strength | Comparable to the predicate<br>device design | Acceptable | | | Bacterial Endotoxins<br>Testing | < 20EU/device | Acceptable | | #### 8. Substantial Equivalence- Clinical Evidence - N/A #### 9. Substantial Equivalence- Conclusions The design characteristics of the subject devices do not raise any different types of questions of safety or effectiveness. From the evidence submitted in this 510(k), the subject devices can be expected to perform at least as well as the predicate system and are substantially equivalent.
Innolitics
510(k) Summary
Decision Summary
Classification Order
Enter a record ID and click Load to view the document.
100%