Vesair 0 degree Cystoscope, Vesair 30 degree Long Cystoscope, Vesair 70 degree Long Cystoscope,Vesair 30 degree Cystoscope,Vesair 70 degree Cystoscope
K170862 · Solace Therapeutics, Inc. · FAJ · Dec 19, 2017 · Gastroenterology, Urology
Device Facts
| Record ID | K170862 |
| Device Name | Vesair 0 degree Cystoscope, Vesair 30 degree Long Cystoscope, Vesair 70 degree Long Cystoscope,Vesair 30 degree Cystoscope,Vesair 70 degree Cystoscope |
| Applicant | Solace Therapeutics, Inc. |
| Product Code | FAJ · Gastroenterology, Urology |
| Decision Date | Dec 19, 2017 |
| Decision | SESE |
| Submission Type | Traditional |
| Regulation | 21 CFR 876.1500 |
| Device Class | Class 2 |
Intended Use
Vesair Cystoscopes are indicated to provide the user with the means for endoscopic diagnostic and therapeutic surgical procedures. Examples for the devices include the visualization and manipulation of anatomy as the surgeon deems appropriate. Vesair Cystoscopes are intended to be used in general urological surgery through a minimally invasive approach by utilizing natural orifices to access the surgical site.
Device Story
Vesair Cystoscopes are rigid, reusable endoscopes used for visualization during minimally invasive urological procedures. The device consists of an insertion tube (4mm diameter) with fiber optics for light transmission and a rigid lens system for image transmission. It requires connection to an external light source, light guide, video camera, monitor, and printer. The device is operated by a surgeon in a clinical setting. Light is transmitted from the source to the surgical site; images are captured via the lens system and displayed on a monitor for the surgeon. This visualization assists the surgeon in navigating anatomy and performing therapeutic interventions. The device is provided non-sterile and requires steam sterilization before use.
Clinical Evidence
Bench testing only. Testing included reprocessing validation, biocompatibility, electrical and thermal safety, shelf-life, and optical image quality. All tests met required specifications.
Technological Characteristics
Rigid, reusable endoscope; 4mm outer diameter; 304 stainless steel patient-contacting surfaces; fiber optic light transmission; rigid lens image transmission; 0°, 30°, 70° direction of view; 105° field of view; steam sterilization; no working channel.
Indications for Use
Indicated for patients undergoing endoscopic diagnostic and therapeutic surgical procedures in general urological surgery via natural orifices.
Regulatory Classification
Identification
An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.
Special Controls
*Classification* —(1)*Class II (special controls).* The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.
Predicate Devices
- Schoelly Cystoscopes (K150158)
Reference Devices
- Henke Sass Wolf Arthroscope (K080560)
Related Devices
- K093725 — CYSTOSCOPE, MODELS CE0004-C, CE006-C, CE3004-C, CE3027-C, CE7004-C, CE1204-C, CE1227-C · Canada Endoscope Corporation · Aug 30, 2010
- K992983 — BAHO AUTOCLAVABLE CYSTOSCOPE · Innovative Endoscopy Components, LLC · Oct 7, 1999
- K243653 — KARL STORZ HOPKINS Telescopes for Urology · Karl Storz SE & CO. KG · Aug 13, 2025
- K171336 — Cystoscope 4.0mm autoclavable 0° WL 302mm; Hysteroscope 4.0mm autoclavable 0° WL 302mm; Cysto-Urethroscope Sheath 19 Fr.; Bridge for 17-25 Fr. Cysto-Urethroscope Sheath · Henke-Sass, Wolf GmbH · Dec 7, 2017
- K191357 — Flexible HD Cysto-Urethroscope System · KARL STORZ Endoscopy-America, Inc. · Sep 18, 2019
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logos of the Department of Health & Human Services and the U.S. Food & Drug Administration (FDA). The Department of Health & Human Services logo is on the left, and the FDA logo is on the right. The FDA logo includes the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.
December 19, 2017
Solace Therapeutics, Inc. Scott Blood Vice President, Quality Assurance and Regulatory Affairs 135 Newbury Street Framingham, MA 01701
Re: K170862
> Trade/Device Name: Vesair 0 degree Cystoscope, Vesair 30 degree Long Cystoscope, Vesair 70 degree Long Cystoscope, Vesair 30 degree Cystoscope, Vesair 70 degree Cystoscope Regulation Number: 21 CFR§ 876.1500 Regulation Name: Endoscope and Accessories Regulatory Class: II Product Code: FAJ Dated: November 3, 2017 Received: November 6, 2017
Dear Scott Blood:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies.
{1}------------------------------------------------
You must comply with all the Act's requirements. including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Image /page/1/Picture/6 description: The image shows the name "Benjamin R. Fisher -S" in a large, bold font. The text is black and appears to be centered on a white background. The letters are clearly legible, and the overall impression is clean and professional.
Benjamin R. Fisher, Ph.D. Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K170862
#### Device Name
Vesair 0 degree Cystoscope, Vesair 30 degree Long Cystoscope, Vesair 70 degree Long Cystoscope, Vesair 30 degree Cystoscope, Vesair 70 degree Cystoscope
#### Indications for Use (Describe)
Vesair Cystoscopes are indicated to provide the user with the means for endoscopic diagnostic and therapeutic surgical procedures. Examples for the devices include the visualization and manipulation of anatomy as the surgeon deems appropriate. Vesair Cystoscopes are intended to be used in general urological surgery through a minimally invasive approach by utilizing natural orifices to access the surgical site.
Type of Use (Select one or both, as applicable)
| <span></span> | <span></span> |
|----------------------------------------------|--------------------------------------------|
| Prescription Use (Part 21 CFR 801 Subpart D) | <div style="display:inline-block;">☑</div> |
| Over-The-Counter Use (21 CFR 801 Subpart C) | <div style="display:inline-block;">☐</div> |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for Solace Therapeutics. The logo consists of a purple circle with white lines inside, followed by the word "Solace" in black, with a registered trademark symbol next to it. Below "Solace" is the word "THERAPEUTICS" in purple, with a sans-serif font.
# 510(k) Summary
| 1. Sponsor Name | |
|----------------------|--------------------------------------------------------------------|
| Submitter's Name: | Solace Therapeutics |
| Address: | 135 Newbury St, Framingham, MA 01701 |
| Phone: | (508) 283-1200 |
| Fax: | (508) 283-1199 |
| Contact Person: | Scott Blood, Vice President Quality Assurance & Regulatory Affairs |
| Date of Preparation: | December 18, 2017 |
#### 2. Device Information
| Trade Name: | Vesair 0 degree Cystoscope, Vesair 30 degree Long Cystoscope, Vesair 70 degree Long Cystoscope, Vesair 30 degree Cystoscope, Vesair 70 degree Cystoscope |
|--------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------|
| Common Name: | Cystoscope |
| Regulation Number: | 21 CFR 876.1500 |
| Regulation Name: | Endoscope and Accessories |
| Class: | II |
| Product Code: | FAJ |
### 3. Predicate Device
Schoelly Cystoscopes K150158
#### 4. Device Description
The Vesair Cystoscopes described in this submission are rigid reusable endoscopes for visualization of the operating site during cystoscopic minimally invasive procedures in conjunction with a commercially available light guide, light source, video camera, monitor, and printer. Light that is created by an external light source is transmitted from the endoscope's light guide connector through the endoscope itself to the tip via a fiber optic system. Images are transferred the other way back through a rigid lens system.
Technical parameters of the Vesair Cystoscopes that characterize the optical view are the Direction of View (0°, 30°, 70°) and the Field of View (105°). The image can be displayed by a camera/monitor system which can be connected to the endoscope eyepiece. The Vesair Cystoscopes have the same diameter (4mm), and are available in different lengths of the insertion tube (0° - 280mm length, 30° - 185mm or 278mm length, 70° - 185mm or 281mm length). None of the cystoscope models have a working channel. Like other currently marketed rigid cystoscopes, all models have patient-contacting outer surfaces mainly made from metal (304 stainless steel) and incorporate fiber optics for light transmission and rigid lenses for image transmission.
{4}------------------------------------------------
Each cystoscope model is provided non-sterile and requires cleaning and steam sterilization prior to use and between subsequent uses.
## 5. Intended Use
Vesair Cystoscopes are indicated to provide the user with the means for endoscopic diagnostic and therapeutic surgical procedures. Examples for the devices include the visualization and manipulation of anatomy as the surgeon deems appropriate. Vesair Cystoscopes are intended to be used in general urological surgery through a minimally invasive approach by utilizing natural orifices to access the surgical site.
### 6. Comparison of Technological Characteristics
The Vesair Cystoscopic Sheath is substantially equivalent to the Schoelly Cystoscope per Table 1:
| Feature/Specification | Proposed Vesair Cystoscope | Schoelly Cystoscope<br>K150158 |
|--------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Patient-Contacting<br>Material | Stainless Steel | Stainless Steel |
| Sterilization | Steam | Steam |
| Outer Diameter | 4mm | Available in 2.9mm and 4mm<br>sizes |
| Working length | 0° - 280mm<br>30° -185mm or 278mm<br>70° -185mm or 281mm | Available in 280mm, 298mm,<br>300mm, 302mm,303mm |
| Working Channel | Not available | Not available |
| Direction of View | Available in 0°,30°,70° | Available in<br>0°,5°,12°,25°,30°,70° |
| Field of View | 1020 | 70° - 85° |
| Indications for Use | Vesair Cystoscopes are indicated to<br>provide the user with the means for<br>endoscopic diagnostic and therapeutic<br>surgical procedures. Examples for the use<br>of the devices include the visualization<br>and manipulation of anatomy as the<br>surgeon deems appropriate. Vesair<br>Cystoscopes are intended to be used in<br>general urological surgery through a<br>minimally invasive approach by utilizing<br>natural orifices to access the surgical site. | The Schoelly Cystoscopes/Hysteroscopes<br>and Accessories arc indicated to provide<br>the user with the means for endoscopic<br>diagnostic and therapeutic surgical<br>procedures. Examples for the use of the<br>devices include the visualization and<br>manipulation of anatomy as the surgeon<br>deems appropriate. The Schoelly<br>Cystoscopes/Hysteroscopes and<br>Accessories are intended to be used in<br>general urological and gynecological<br>surgery through a minimally invasive<br>approach by utilizing natural orifices to |
Table 1 - Comparison of Proposed Vesair Cystoscope to Predicate Device
{5}------------------------------------------------
| Feature/Specification | Proposed Vesair Cystoscope | Schoelly Cystoscope K150158 |
|-----------------------|----------------------------|-----------------------------|
| | | access the surgical site. |
#### 7. Reference Device
These same models have already been cleared for marketing by FDA (K080560) for use in arthroscopic procedures. The proposed Vesair Cystoscope is the exact same device as the Henke Sass Wolf Arthroscope, except for the working length and Indications for Use.
| Feature/Specification | Proposed Vesair Cystoscope | Henke Sass Wolf Arthroscope<br>K080560 |
|--------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Patient-Contacting<br>Material | Stainless Steel | Stainless Steel |
| Sterilization | Steam | Steam |
| Outer Diameter | 4mm | 4mm |
| Working length | 0° - 280mm<br>30° -185mm or 278mm<br>70° -185mm or 281mm | 185mm |
| Working Channel | Not available | Not available |
| Direction of View | Available in 0°,30°,70° | Available in 0°,30°,70° |
| Field of View | 105° | 105° |
| Indications for Use | Vesair Cystoscopes are indicated to provide the user with the means for endoscopic diagnostic and therapeutic surgical procedures. Examples for the use of the devices include the visualization and manipulation of anatomy as the surgeon deems appropriate. Vesair Cystoscopes are intended to be used in general urological surgery through a minimally invasive approach by utilizing natural orifices to access the surgical site. | The HSW Arthroscope and accessories is a tubular endoscopic device with accessory devices which attach to the Arthroscope and is intended to examine and / or perform surgery on the interior of a joint. Arthroscopic minimal invasive procedures are performed in the hip, knee, shoulder, wrist (carpal tunnel syndrome), temporal-mandibular joint, ankle, elbow and feet (plantar fascia release). |
Table 2 - Comparison of Proposed Vesair Cystoscope to Reference Device
### 8. Performance Testing
Performance tests on the bench were performed on the Vesair Cystoscope. The following were performed:
- Reprocessing validation study ●
- . Biocompatibility Testing
- Electrical and Thermal Safety Testing
{6}------------------------------------------------
- Shelf-life study
- Optical image quality testing
All bench testing has been performed and all components, subassemblies and/or full devices and systems have met the required specifications for the completed tests.
#### 9. Summary
Solace Therapeutics, Inc. has demonstrated that the Vesair Cystoscope is substantially equivalent to its listed predicate devices.