K170810 · Cryofos Medical GmbH · GEH · Sep 27, 2017 · General, Plastic Surgery
Device Facts
Record ID
K170810
Device Name
CRYOFOS
Applicant
Cryofos Medical GmbH
Product Code
GEH · General, Plastic Surgery
Decision Date
Sep 27, 2017
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4350
Device Class
Class 2
Attributes
Therapeutic, Real-World Evidence
Real-World Evidence
Submission
Device
Sponsor
RWD Sources
RWE Use Summary
Key Tags
K170810 · Sep 27, 2017
CRYOFOS
Cryofos Medical GmbH
Published clinical literature (clinical studies of cryotherapy using hyperbaric CO2 gas)
The sponsor conducted a literature review of existing clinical studies to demonstrate the efficacy of hyperbaric CO2 cryotherapy for pain, swelling, inflammation, and hematoma reduction across various clinical indications.
Literature review; Clinical efficacy; Cryotherapy; Hyperbaric CO2
Clinical Evidence
Study Design
Population
Comparator
Key Endpoints
Literature review of hyperbaric CO2 cryotherapy studies; Literature review of individual and randomized clinical studies
Patients with tendinopathies, shoulder/elbow ailments, facial injuries, hemophilia hemorrhage, Sudeck's atrophy, hand/foot surgery, lumbago, and rheumatic myopathies
Not applicable for this study
Reduction of edema, hematoma, inflammation, and pain; post-operative mobility; reduction in medication use
Indications for Use
The CRYOFOS and Accessories indicated for use when cold therapy is indicated for the temporary reduction of pain, swelling, inflammation, and hematoma from minor surgical procedures, minor sprains or other minor sports injuries, and as an adjunct to rehabilitative treatment (e.g., intermittent cold with stretch).
Device Story
Topical refrigerant system; delivers compressed medical-grade CO2 gas via pistol-grip handpiece; creates micro-crystalline dry-ice mist through natural gas expansion. Used in clinical settings by healthcare providers. Operator applies spray in circular/panning motion to treatment site (approx. 7cm distance) for 30-60 seconds. Device features infrared skin temperature monitoring; automatic CO2 beam shutdown if skin temperature drops below -1°C for 1 second to prevent necrosis. Provides rapid cooling (thermal shock) to 2-4°C. Benefits include reduced pain, swelling, and inflammation; potential reduction in medication use; accelerated post-operative mobility; shortened hospital stays.
Clinical Evidence
No clinical data generated for this device. Efficacy supported by literature review of hyperbaric CO2 cryotherapy across various conditions (tendinopathies, sports injuries, post-surgical recovery). Studies indicate reduced edema, hematoma, and inflammation, leading to decreased medication reliance and improved patient mobility.
Technological Characteristics
Topical CO2 cryosurgical unit. Components: pistol grip handpiece, console, rechargeable battery, CO2 cylinder. Principle: expansion cooling of liquid CO2 to form micro-crystalline dry-ice mist. Features: infrared skin temperature sensor, automatic shutdown logic. Electrical safety: IEC 60601-1; EMC: IEC 60601-1-2.
Indications for Use
Indicated for patients requiring temporary reduction of pain, swelling, inflammation, and hematoma resulting from minor surgical procedures, minor sprains, or minor sports injuries, and as an adjunct to rehabilitative treatment.
Regulatory Classification
Identification
(1) Cryosurgical unit with a liquid nitrogen cooled cryoprobe and accessories. A cryosurgical unit with a liquid nitrogen cooled cryoprobe and accessories is a device intended to destroy tissue during surgical procedures by applying extreme cold. (2) Cryosurgical unit with a nitrous oxide cooled cryoprobe and accessories. A cryosurgical unit with a nitrous oxide cooled cryoprobe and accessories is a device intended to destroy tissue during surgical procedures, including urological applications, by applying extreme cold. (3) Cryosurgical unit with a carbon dioxide cooled cryoprobe or a carbon dioxide dry ice applicator and accessories. A cryosurgical unit with a carbon dioxide cooled cryoprobe or a carbon dioxide dry ice applicator and accessories is a device intended to destroy tissue during surgical procedures by applying extreme cold. The device is intended to treat disease conditions such as tumors, skin cancers, acne scars, or hemangiomas (benign tumors consisting of newly formed blood vessels) and various benign or malignant gynecological conditions affecting vulvar, vaginal, or cervical tissue. The device is not intended for urological applications.
{0}------------------------------------------------
Image /page/0/Picture/2 description: The image shows the address for the Food and Drug Administration. The address is 10903 New Hampshire Avenue, Document Control Center - WO66-G609, Silver Spring, MD 20993-0002. The text is left-aligned and in a simple font.
September 27, 2017
CRYOFOS Medical Gmbh % Nick Stocks Vice President TensSource LLC 724 Bald Cypress Place Tampa, Florida 33614
Re: K170810
Trade/Device Name: CRYOFOS and Accessories Regulation Number: 21 CFR 878.4350 Regulation Name: Cryosurgical unit and accessories Regulatory Class: Class II Product Code: GEH, MLY Dated: July 12, 2017 Received: August 1, 2017
Dear Nick Stocks:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Actinclude requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR
{1}------------------------------------------------
Part 807); labeling (21 CFR Part 801); medical device reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education (DICE) at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education (DICE) at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely.
Image /page/1/Picture/9 description: The image shows the name "Jennifer R. Stevenson -S3" in a large, bold font. The name is stacked vertically, with "Jennifer R." on the top line and "Stevenson -S3" on the bottom line. The background of the image is a light blue color, and the letters are black.
For Binita S. Ashar, M.D., M.B.A., F.A.C.S. Director Division of Surgical Devices
Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K170810
Device Name CRYOFOS and Accessories
Indications for Use (Describe)
The CRYOFOS and Accessories indicated for use when cold therapy is indicated for the temporary reduction of pain, swelling, inflammation, and hematoma from minor surgical procedures, minor sports injuries, and as an adjunct to rehabilitative treatment (e.g., intermittent cold with stretch).
| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------|
|-------------------------------------------------|
| <div> <span> <span style="font-size:16px">☒</span> Prescription Use (Part 21 CFR 801 Subpart D) </span> </div> |
|-------------------------------------------------------------------------------------------------------------------------|
| <div> <span> <span style="font-size:16px">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) </span> </div> |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the word "CRYOFOS" in a stylized font. The letters "CRYO" are in green, and the letters "FOS" are in blue. The letters are stacked vertically, with "CRYO" on top of "FOS". The font is sans-serif and appears to be bold.
# 510(k) SUMMARY
In accordance with the requirements of the Safe Medical Device Act of 1990 and 21 CFR 807.92, CRYOFOS Medical is hereby submitting the 510(k) Summary of Safety and Effectiveness for 510(k) Number __K170810__
Date Prepared: February 1, 2017
## A. Submitter
TensSource 724 Bald Cypress Place Tampa, FL 33614 Establishment Registration Number: TBD
## B. Company Contact
Nick Stocks Vice President Tel: (813) 889-0000 Fax: (813) 527-9949
### C. Device Name
| Trade Name: | CRYOFOS and Accessories |
|------------------------|---------------------------------|
| Classification Names: | Unit, Cryosurgical, Accessories |
| Proposed Class/Device: | Class II |
| Product Codes: | GEH, MLY |
| Regulation : | 878.4350 |
### D. Intended Use / Indications
The CRYOFOS and Accessories indicated for use when cold therapy is indicated for the temporary reduction of pain, swelling, inflammation, and hematoma from minor surgical procedures, minor sprains or other minor sports injuries, and as an adjunct to rehabilitative treatment (e.g., intermittent cold with stretch).
### E. Predicate/Legally Marketed Devices
| Device Name: | CRYOTRON 2 Cryotherapy Device |
|---------------|-------------------------------|
| Company Name: | Cryonic Medical North America |
| 510(k) #: | K030281 |
{4}------------------------------------------------
# F. Device Description
The CRYOFOS and Accessories is a topical refrigerant that acts as a vapocoolant. The CRYOFOS system consists of a pistol grip handpiece, console for controlling operations, a rechargeable battery (Panasonic R127R2PG), and a cvlinder of compressed medical grade Carbon Dioxide (CO2) gas (sold separately). The CRYOFOS has a user interface that displays treatment activation. The CRYOFOS system delivers a topical spray of compressed medical grade (CO2) that is pure, dry bacteriostatic and non-explosive via the pistol grip handpiece. The device uses the natural expansion of liquid CO2 in the atmosphere to create a cold spray of microcrystals delivered under pressure. The expansion cooling of the treatment gas forms a micro-crystalline dry-ice mist with which the cooling effect required for treatment is achieved. The end-user uses a gradual circular or panning motion with the handpiece to apply the CO2 to the treatment site for 30-60 seconds at a distance of approximately 7cm. Rapid cooling (thermal shock) of the skin occurs at temperatures of 2-4°C in 30 sec when the spray sublimates (pass directly from solid / ice phase to gas phase) as it contacts the skin.
# G. Testing
Verification and validation testing was completed to confirm that the CRYOFOS and Accessories provides equivalent results in comparison to the predicate device, CRYOTRON 2 Cryotherapy Device (K030281). Verification and validation testing included IEC 60601-1 electrical safety testing and IEC 60601-1-2 EMC testing to confirm device safety, software verification and validations to confirm software interaction and performance, as well as performance / benchmark testing using the CRYOFOS and Accessories as compared with the CRYOTRON 2 Cryotherapy device. The performance / benchmark testing confirmed the temperature of 2-4℃ in 30 sec is reached by both devices, the treatment time to reach thermal shock is 30 sec to 1 minute in both devices, the lowest temperature is -78.5℃ in both devices (due to physical law of CO2 gas), the pistol grip design is more stable with the subject device, and temperature detection is improved with the subject device such that the CO2 beam shuts down if the temperature of the skin is less than -1℃ for 1 second.
Clinical efficacy was determined through an extensive literature review of studies. The publications were presented here to show that it is possible to demonstrate very good to good results with cryotherapy using hyperbaric CO2 gas. The individual and partly randomized studies prove that the patient, which is always in the foreground, can benefit enormously from the treatment. The examinations refer to different ailment and treatment areas and it is therefore possible to give a comprehensive statement that the therapy is proven to be effective. Studies were carried out on tendinopathies, shoulder and elbow ailments, facial injuries, hemorrhage in hemophilia, Sudeck's atrophy, hand surgery, foot surgery, lumbago, rheumatic myopathies and much more. This covers the entire range of therapies required. Overall results for this form of therapy were excellent, confirmed by rapid healing of edemas, hematomas, and inflammation and therefore also pain. In this way, it was possible to reduce or even completely avoid the administration of
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the word "CRYOFOS" in a stylized font. The letters "CRYO" are in a light green color, while the letters "FOS" are in a blue color. The letters are stacked closely together, creating a compact and modern design.
accompanying medications against pain and antibiotic therapies. This ultimately also leads to the goal, i.e. of achieving faster post-operative mobility of individual joints and much earlier overall mobility of the patient. Hospital stays can thus be shortened and the patients can be discharged earlier in order to carry out follow-up rehabilitation. The CRYOFOS met all manufacturing and software specifications. No additional performance or clinical testing was conducted.
# H. Substantial Equivalence
The CRYOFOS and Accessories are substantially equivalent in design, manufacturing materials, intended use, principles of operation, and technical characteristics to the CRYOTRON 2 Cryotherapy Device (K030281) and raises no new issues of safety or effectiveness. The key differences are the CRYOFOS and Accessories include visible warnings as safety features, a more robust pistol handerip design with press kev, focused infrared measurement of skin temperature during treatment, and improved necrosis protection where that the CO2 beam shuts down if the temperature of the skin is detected to be less than -1°C for 1 second to limit burning of the skin. Treatment can be resumed again after a break of 2 seconds.
# I. Conclusion
The CRYOFOS System uses the same identical Indications for Use as the predicate (K030281) device. The CRYOFOS System is very similar to the predicate device in terms of technological characteristics, mode of operation and design features. Based on the verification and validating testing regarding bench performance and the clinical literature provided in the submission: it has been determined that the CRYOFOS System performed as well as the predicate (K030281) device in terms of safety, effectiveness and performance. The bench performance testing also demonstrated that the technological differences between the CRYOFOS System and the predicate device raise no new concerns of safety or effectiveness and is substantially equivalent to predicate (K030281).
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.