Total knee arthroplasty is intended to provide increased patient mobility and reduced pain by replacing the damaged knee joint articulation in patients where there is evidence of sufficient sound bone to seat and support the components. Total knee replacement may be considered for younger patients if, in the opinion of the surgeon, an unequivocal indication for total knee replacement outweighs the risks associated with the age of the patient, and if limited demands regarding activity and knee joint loading can be assured. This includes severely crippled patients with multiple joint involvement for whom a gain in knee mobility may lead to an expectation of significant improvement in the quality of their lives.
Device Story
ATTUNE Cemented Tibial Base, Fixed Bearing is a total knee prosthesis component; provides tibial fixation in total knee arthroplasty. Device features stem, keel, recessed undercut cement pockets, and grit-blasted surface to enhance cement-to-implant bond. Utilizes central universal locking mechanism to capture tibial insert. Implanted by orthopedic surgeons in clinical settings. Device includes a temporary Base Protector to prevent marring during impaction; protector removed prior to insert placement. Benefits include increased patient mobility and pain reduction through joint replacement.
Clinical Evidence
No clinical data. Substantial equivalence determined via non-clinical bench testing, including fatigue testing (ASTM F1800), intraoperative ROM simulation, pull-off testing with lipid/marrow, surgeon validation, and bacterial endotoxin testing (ANSI AAMI ST-72:2011).
Technological Characteristics
Cast Co-Cr-Mo alloy (ASTM F75). Features stem, keel, undercut cement pockets, and grit-blasted surface. Fixed bearing design with central universal locking mechanism. Sizes 1-10. Non-electronic, mechanical implant.
Indications for Use
Indicated for patients with severely painful and/or disabled knee joints due to osteoarthritis, post-traumatic arthritis, rheumatoid arthritis, or failed previous implants.
Regulatory Classification
Identification
A knee joint patellofemorotibial polymer/metal/polymer semi-constrained cemented prosthesis is a device intended to be implanted to replace a knee joint. The device limits translation and rotation in one or more planes via the geometry of its articulating surfaces. It has no linkage across-the-joint. This generic type of device includes prostheses that have a femoral component made of alloys, such as cobalt-chromium-molybdenum, and a tibial component or components and a retropatellar resurfacing component made of ultra-high molecular weight polyethylene. This generic type of device is limited to those prostheses intended for use with bone cement (§ 888.3027).
{0}------------------------------------------------
Public Health Service
Image /page/0/Picture/2 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" around the perimeter. Inside the circle is an abstract symbol that resembles a stylized caduceus, featuring three human profiles in silhouette.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
June 15, 2017
DePuy (Ireland) Nancy Friddle Associate Director, Regulatory Affairs DePuv Orthopaedics. Inc. 700 Orthopaedic Dr. Warsaw, Indiana 46582
Re: K170806
Trade/Device Name: ATTUNE Cemented Tibial Base, Fixed Bearing Regulation Number: 21 CFR 888.3560 Regulation Name: Knee Joint Patellofemorotibial Polymer/Metal/Polymer Semi-Constrained Cemented Prosthesis Regulatory Class: Class II Product Code: JWH, OIY Dated: March 16, 2017 Received: March 17, 2017
Dear Nancy Friddle:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply
{1}------------------------------------------------
with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely,
Image /page/1/Picture/9 description: The image shows the name "Vincent J. Devlin -S" in a large, sans-serif font. The text is black and appears to be against a white background. The letters are evenly spaced and the text is left-aligned.
for Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K170806
Device Name
ATTUNE Cemented Tibial Base, Fixed Bearing
Indications for Use (Describe)
Candidates for total knee replacement include patients with a severely disabled joint resulting from osteoarthritis, post-traumatic arthritis, rheumatoid arthritis, or a failed previous implant.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------|--|
| | |
X Prescription Use (Part 21 CFR 801 Subpart D)
| | Over-The-Counter Use (21 CFR 801 Subpart C)
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
510(k) Summary (As required by 21 CFR 807.92 and 21 CFR 807.93)
| Submitter Information | |
|------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Name | DePuy Orthopaedics, Inc. |
| Address | 700 Orthopaedic Drive<br>Warsaw, IN 46582 |
| Phone number | (574) 371-4923 |
| Fax number | (574) 371-4987 |
| Establishment | 1818910 |
| Registration Number | |
| Name of contact person | Nancy Friddle |
| Date prepared | March 16, 2016 |
| Name of device | |
| Trade or proprietary name | ATTUNE Cemented Tibial Base, Fixed Bearing |
| Common or usual name | Total Knee Prosthesis |
| Classification name | Knee joint patellofemorotibial polymer/metal/polymer semi-<br>constrained cemented prosthesis |
| Class | II |
| Classification panel | 87 Orthopedics |
| Regulation | 21 CFR 888.3560 |
| Product Code(s) | OIY, JWH |
| Legally marketed<br>device(s) to which<br>equivalence is claimed | Primary Predicate:<br>K101433: DePuy ATTUNE Knee System<br>Reference Predicate:<br>K032151: Sigma Co-Cr Tibial Trays |
| Reason for 510(k)<br>submission | Addition of a new Cemented Tibial Base component to the ATTUNE<br>Knee System |
| Device description | The ATTUNE Cemented Tibial Base, Fixed Bearing (FB) are<br>available in sizes 1-10. The fixation surface incorporates a stem<br>and keel to provide additional stability as well as recessed undercut<br>cement pockets and a grit blasted surface for enhanced cement<br>fixation. The tibial base utilizes a central universal locking<br>mechanism to capture the tibial insert. The ATTUNE Cemented<br>Tibial Base, FB is manufactured from cast Co-Cr-Mo alloy<br>conforming to ASTM F75. Additionally, the ATTUNE Cemented<br>Tibial Base, FB utilizes a previously cleared Base Protector<br>assembled with the device to minimize potential marring of the<br>proximal bearing surface of the tibial base during impaction. After<br>impaction is completed, the Base Protector is then removed and<br>discarded prior to inserting the ATTUNE Fixed Bearing Insert.<br><br>The ATTUNE CR Total Knee is designed to accommodate knee<br>flexion to 150 degrees in those patients able to attain a high degree of<br>knee flexion. |
| | The ATTUNE PS Total Knee is designed to accommodate knee<br>flexion to 145 degrees in those patients able to attain a high degree of<br>knee flexion. |
| Intended use of the<br>device | Total knee arthroplasty is intended to provide increased patient<br>mobility and reduced pain by replacing the damaged knee joint<br>articulation in patients where there is evidence of sufficient sound<br>bone to seat and support the components. Total knee replacement<br>may be considered for younger patients if, in the opinion of the<br>surgeon, an unequivocal indication for total knee replacement<br>outweighs the risks associated with the age of the patient, and if<br>limited demands regarding activity and knee joint loading can be<br>assured. This includes severely crippled patients with multiple joint<br>involvement for whom a gain in knee mobility may lead to an<br>expectation of significant improvement in the quality of their lives. |
| Indications for use | Candidates for total knee replacement include patients with a<br>severely painful and/or severely disabled joint resulting from<br>osteoarthritis, post-traumatic arthritis, rheumatoid arthritis, or a<br>failed previous implant. |
{4}------------------------------------------------
## Summary of Technologies/ Substantial Equivalence Discussion:
The ATTUNE Cemented Tibial Base, FB provides a macro geometric feature and an optimized micro-blast finish which are both intended to aid in fixation of the tibial implant to the bone cement. The ATTUNE Cemented Tibial Base, FB is designed to enhance fixation by improving resistance (relative to the industry) to intra-operative factors which can result in a reduction in cement to implant bond.
The subject ATTUNE Cemented Tibial Base, FB has the same indications, intended used, materials, biocompatibility, packaging configurations, manufacturing methods, and sterilization methods as the previously cleared ATTUNE Cemented Tibial Base (K101433). The subject ATTUNE Cemented Tibial Base, FB is manufactured with undercut cement pockets similar to those previously cleared with the Sigma Co-Cr Trays (K032151).
The testing provided below was used to support equivalence to the predicate devices.
| Non-Clinical Testing: | Functional testing was conducted in compliance with FDA Guidance,<br>Class II Special Controls Guidance Document: Knee Joint<br>Patellofemorotibial and Femorotibial Metal/Polymer Porous-Coated<br>Uncemented Prostheses: Guidance for Industry and FDA, to verify<br>that the implant performance is substantially equivalent to predicate<br>devices for anticipated in-vivo loading via the following tests:<br>Fatigue Testing per ASTM F 1800, Standard Practice for<br>Cyclic Testing of Metal Tibial Tray Components of Total |
|-----------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
|-----------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
{5}------------------------------------------------
| | Knee Joint Replacements |
|-------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | Intraoperative ROM simulation and pull-off testing with lipid<br>● |
| | and marrow |
| | Surgeon validation<br>● |
| | The proposed devices also meet the requirement of<br>Bacterial endotoxin testing as specified in ANSI AAMI ST-<br>●<br>72:2011. |
| Clinical Testing: | Clinical testing was not necessary to determine substantial<br>equivalence between the ATTUNE Cemented Tibial Base, FB and<br>the predicate devices. |
| Conclusions drawn<br>from non-clinical and<br>clinical data | The subject Fixed Bearing ATTUNE Cemented Tibial Base, FB is<br>substantially equivalent to the predicate Fixed Bearing ATTUNE<br>Total Knee System Tibial Bases and Sigma Co-Cr Tibial Trays. |
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.