K170681 · Irvine Scientific Sales Co., Inc. · MQL · Jun 8, 2017 · Obstetrics/Gynecology
Device Facts
Record ID
K170681
Device Name
Continuous Single Culture -NX (CSCM-NX)
Applicant
Irvine Scientific Sales Co., Inc.
Product Code
MQL · Obstetrics/Gynecology
Decision Date
Jun 8, 2017
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 884.6180
Device Class
Class 2
Indications for Use
The Continuous Single Culture®-NX (CSCM-NX) is intended for use as a culture medium for human gametes and embryos from fertilization through day 5/6 of development in vitro.
Device Story
CSCM-NX is a single-use embryo culture medium used in clinical IVF laboratories. It supports human gametes and embryos from fertilization through blastulation (day 5/6). The medium contains salts, energy substrates, amino acids, buffering agents, nutrients, and antibiotics. Unlike its predicate, it is supplied without human serum albumin (HSA), which must be added by the clinician before use. The device is aseptically filled into bottles. Clinicians use the medium to provide an optimized environment for embryo development in an incubator. By maintaining specific pH and osmolality, the medium supports embryo viability and development, potentially improving IVF success rates.
Clinical Evidence
Bench testing only. Performance validated via pH, osmolality, sterility (USP <71>), endotoxin (USP <85>), and mouse embryo assay (MEA). MEA results showed ≥80% development to blastocyst stage at 96 hours. Human sperm survival assay (HSSA) showed ≥70% motility at 24 hours. Shelf-life and stability testing confirmed specifications are maintained over time and after repeated bottle opening.
Technological Characteristics
Liquid culture medium containing salts, energy substrates, amino acids, buffering agents, nutrients, and antibiotics. Aseptically filled; SAL 10^-3. Sterility per USP <71>; Endotoxin per USP <85>; Aseptic processing per ISO 13408-1:2008 and ISO 13408-2:2003. pH 7.25-7.54; Osmolality 260-270 mOsm/kg. Supplied in 20 mL and 60 mL bottles. Shelf-life 4 months; stable for 4 weeks after opening at 2-8°C.
Indications for Use
Indicated for use as a culture medium for human gametes and embryos from fertilization through day 5/6 of development in vitro.
Regulatory Classification
Identification
Reproductive media and supplement are products that are used for assisted reproduction procedures. Media include liquid and powder versions of various substances that come in direct physical contact with human gametes or embryos (including water, acid solutions used to treat gametes or embryos, rinsing solutions, sperm separation media, supplements, or oil used to cover the media) for the purposes of preparation, maintenance, transfer or storage. Supplements are specific reagents added to media to enhance specific properties of the media (e.g., proteins, sera, antibiotics, etc.).
Special Controls
*Classification.* Class II (special controls) (mouse embryo assay information, endotoxin testing, sterilization validation, design specifications, labeling requirements, biocompatibility testing, and clinical testing). The device, when it is phosphate-buffered saline used for washing, and short-term handling and manipulation of gametes and embryos; culture oil used as an overlay for culture media containing gametes and embryos; and water for assisted reproduction applications, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 884.9.
{0}------------------------------------------------
Image /page/0/Picture/2 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, with a flowing, abstract design connecting them.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
June 8, 2017
Irvine Scientific Sales Co., Inc. Jayme Yamaguchi-Owens Regulatory Affairs Manager 2511 Daimler Street Santa Ana, CA 92705
Re: K170681
> Trade/Device Name: Continuous Single Culture® -NX (CSCM-NX) Regulation Number: 21 CFR§ 884.6180 Regulation Name: Reproductive Media and Supplements Regulatory Class: II Product Code: MOL Dated: May 18, 2017 Received: May 23, 2017
Dear Jayme Yamaguchi-Owens:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies.
You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely.
Image /page/1/Picture/10 description: The image shows the text "Benjamin R. Fisher -S" in a large, bold, sans-serif font. The text is black and is set against a white background. The letters are evenly spaced and the text is easy to read. The overall impression is clean and professional.
Benjamin R. Fisher, Ph.D. Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
#### Indications for Use
510(k) Number (if known) K170681
Device Name
Continuous Single Culture®-NX (CSCM-NX)
Indications for Use (Describe)
The Continuous Single Culture®-NX (CSCM-NX) is intended for use as a culture medium for human gametes and embryos from fertilization through day 5/6 of development in vitro.
Type of Use (Select one or both, as applicable)
> Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement below.
{3}------------------------------------------------
# 510(k) Summary
#### -General Information on Submitter
| Submitter/Address: | Irvine Scientific Sales Co., Inc.<br>2511 Daimler Street<br>Santa Ana, CA 92705<br>Telephone: 800-437-5706<br>Facsimile: 949-261-6522 |
|--------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Contact Person: | Jayme Yamaguchi-Owens<br>Regulatory Affairs Manager<br>Irvine Scientific<br>2511 Daimler Street<br>Santa Ana, CA 92705<br>Tel: 949-261-7800 ext 254<br>Fax: 949-261-6522<br>Email: jfy@irvinesci.com |
#### II. Date Prepared: June 6, 2017
### III. General Information on Device
| Device Name: | Continuous Single Culture®-NX (CSCM-NX) |
|-----------------------|-----------------------------------------------------------------------------------------------------|
| Common Name: | Embryo Culture Media |
| Classification Name: | Reproductive Media and Supplements (21 CFR 884.6180) |
| Product code: | MQL (Media, Reproductive) |
| Regulatory Class: | II |
| IV. Predicate Device: | Continuous Single Culture Complete (K121572) manufactured by Irvine Scientific Sales Co., Inc. This |
#### V. Description of the Device:
The Continuous Single Culture®-NX (CSCM-NX) is a culture medium using a formulation modified from the predicate device. It is intended for culture of embryos from fertilization to blastulation. This medium consists of salts, energy substrates, amino acids, buffering agents, nutrients supplements, and antibiotics.
related recalls.
predicate device has not been subject to any design
This product is a single-use device supplied in a fill volume of 20 mL and 60 mL. It is aseptically filled into sterilized bottles and has a sterility assurance level (SAL) of 10° The product is tested for pH, osmolality, embryotoxicity, spermtoxicity, endotoxin, and sterility before lot release. The product has a four-month shelf-life and remains stable for four weeks after opening/closing of the bottle when it is stored at 2-8°C.
{4}------------------------------------------------
# VI. Indication for Use:
The Continuous Single Culture®-NX (CSCM-NX) is intended for use as a culture medium for human gametes and embryos from fertilization through day 5/6 of development in vitro.
## VII. Comparison of Intended Use and Technological Characteristics of Subject and Predicate Devices
| Device & Predicate<br>Device(s): | K170681<br>(subject device) | K121572<br>(predicate device) |
|----------------------------------|--------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Indications for Use | Same as predicate device | Continuous Single Culture complete<br>is intended for use as a culture<br>medium for human gametes and<br>embryos from fertilization through<br>day 5/6 of development in vitro. |
| pH (under 5% CO2) | Same as predicate device | 7.25-7.54 |
| Osmolality | Same as predicate device | 260-270 mOsm/kg |
| Formulation | Comparable to predicate device | 39 ingredients |
The subject and predicate devices have the same indication - culture of embryos from fertilization to blastulation. Therefore, they have the same intended use.
The subject and predicate devices have the same pH, osmolality and comparable formulations. Unlike the predicate device, the subject device does not contain human serum albumin (HSA), but needs addition of HSA before use. In addition, the concentrations of certain ingredients are modified in the subject device. The differences noted between the subject and predicate devices are minor and do not raise different questions of safety and effectiveness.
In addition to pH and osmolality, the acceptance criteria for other product characteristics of the subject device, including mouse embryo assay (MEA), human sperm survival assay (HSSA), endotoxin and sterility, are same to those of the predicate device.
### VIII. Summary of Non-clinical Performance Testing
The following studies have been performed to support substantial equivalence to the predicate device:
- pH ●
- Osmolality
- . Aseptic Processing Validation testing per ISO 13408-1:2008 and ISO 13408-2:2003
- Sterility testing per USP <71> ●
- Endotoxin testing per USP <85> ●
- . Mouse embryo assay (MEA)
Continuous Single Culture®-NX (CSCM-NX)
{5}------------------------------------------------
One-cell mouse embryos were exposed to subject devices and cultured at 37°C in an atmosphere containing 5% CO2. The percent of embryos developed to the expanded blastocyst stage at 96 hours were assessed in comparison with the control group.
- . Shelf-life studies (real-time and accelerated) were conducted to ensure that the following acceptance criteria for product characteristics are met at time zero and the end of shelf-life.
- pH See table above *
- * Osmolality - See table above
- * 1-cell MEA - ≥80% developed to the blastocyst stage at 96 hours
- * HSSA - ≥70% of original motility at 24 hours
- Endotoxin ≤0.25 EU/ml (LAL) *
- ★ Sterility - No microbiological growth
- . Stability testing was conducted to ensure that acceptance criteria for the product specifications are met after repeated opening/closing of bottles for four (4) weeks if handled appropriately.
### XI. Conclusion:
The subject and predicate devices have the same intended use. Although there is a difference in formulation between the subject and predicate devices, this difference does not raise different questions of safety or effectiveness. The performance data also demonstrate that the subject device is substantially equivalent to the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.