K170662 · Intrinity Global Limited · FLL · Nov 22, 2017 · General Hospital
Device Facts
Record ID
K170662
Device Name
Non Contact Infrared Forehead Thermometer
Applicant
Intrinity Global Limited
Product Code
FLL · General Hospital
Decision Date
Nov 22, 2017
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.2910
Device Class
Class 2
Attributes
Pediatric
Indications for Use
Non Contact Infrared Forehead Thermometer is a non-sterile, reusable, handheld device. It can be used by consumers in homecare environment and doctors in clinic as reference. It is intended for measuring human body temperature of all ranges of people by detecting infrared heat from the forehead.
Device Story
Handheld, non-sterile, reusable infrared thermometer; measures human body temperature by detecting infrared heat emitted from forehead. Device components: IR sensor, measurement buttons (human/object), battery compartment, buzzer, Celsius/Fahrenheit toggle, LCD, and enclosure. Operated by consumers or clinicians; user points device at forehead (1cm distance) to initiate measurement. Device processes infrared radiation input to calculate temperature; displays result on LCD. Includes fever alert, memory storage (16 sets), and auto-off functionality. Provides non-invasive temperature monitoring to assist in clinical decision-making.
Clinical Evidence
Clinical performance testing conducted per ASTM E1965-98 (2009) standards. Results confirmed compliance with clinical accuracy requirements for infrared thermometers. Additional non-clinical bench testing included electrical safety (IEC 60601-1), EMC (IEC 60601-1-2), and biocompatibility (ISO 10993-5, ISO 10993-10).
Indicated for measuring human body temperature in people of all ages via forehead infrared heat detection. Suitable for use by consumers in homecare settings and by clinicians in medical facilities.
Regulatory Classification
Identification
A clinical electronic thermometer is a device used to measure the body temperature of a patient by means of a transducer coupled with an electronic signal amplification, conditioning, and display unit. The transducer may be in a detachable probe with or without a disposable cover.
Special Controls
(1) Device is not a clinical thermometer with telethermographic functions;
(2) Device is not a clinical thermometer with continuous temperature measurement functions; and
(3) Appropriate analysis and testing (such as that outlined in the currently FDA-recognized editions, as appropriate, of ISO 80601-2-56, “Medical electrical equipment—Part 2-56: Particular requirements for basic safety and essential performance of clinical thermometers for body temperature measurement,” or ASTM E1965, “Standard Specification for Infrared Thermometers for Intermittent Determination of Patient Temperature,” or ASTM E1112, “Standard Specification for Electronic Thermometer for Intermittent Determination of Patient Temperature,” or ASTM E1104, “Standard Specification for Clinical Thermometer Probe Covers and Sheaths”) must validate specifications and performance of the device.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which consists of the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG" in blue, with the word "ADMINISTRATION" underneath.
November 22, 2017
Intrinity Global Limited % Elly Xu Consultant Shenzhen Joyantech Consulting Co., Ltd NO. 55 Shizhou middle road, Nanshan District Shenzhen, 518000 China
Re: K170662
Trade/Device Name: Non Contact Infrared Forehead Thermometer Regulation Number: 21 CFR 880.2910 Regulation Name: Clinical Electronic Thermometer Regulatory Class: Class II Product Code: FLL Dated: October 15, 2017 Received: October 26, 2017
Dear Elly Xu:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
Tina Kiang
-s
Tina Kiang, Ph.D. Acting Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K170662
Device Name
Non Contact Infrared Forehead Thermometer
Indications for Use (Describe)
Non Contact Infrared Forehead Thermometer is a non-sterile, reusable, handheld device. It can be used by consumers in homecare environment and doctors in clinic as reference. It is intended for measuring human body temperature of all ranges of people by detecting infrared heat from the forehead.
Prescription Use (Part 21 CFR 801 Subpart D)
|X | Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# Version:A/0
# VOL_05: 001 510(k) Summary
# 5.1 Administrative Information
| Date of Summary<br>prepared | Nov, 14, 2017 |
|----------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Manufacturer<br>information | Intrinity Global Limited<br>Unit 1707, Floor 17, Westley Square, 48 Hoi Yuen<br>Road, Kwun Tong, Hong Kong |
| | Contact person: Alex Kwok<br>Phone: 852-35231874<br>Fax: 852-37484128<br>E-mail: alex.kwok@intrinity.com.hk |
| Submission<br>Correspondent<br>Image: logo<br>卓远天成 | Shenzhen Joyantech Consulting Co., Ltd.<br>Address: Room 1122, International Mayors<br>Communication Centre, NO. 55 Shizhou middle road<br>Nanshan District, Shenzhen<br>Contact person: Ms. Elly Xu; Mr. Field Fu<br>E-Mail: elly@cefda.com; cefda13485@163.com |
| Establishment | |
## 5.2 Device Information
registration number
| Type of 510(k)<br>submission: | Traditional |
|-------------------------------|-------------------------------------------|
| Trade Name: | Non Contact Infrared Forehead Thermometer |
| Model: | TVT-200, TVT-200 PLUS |
| Classification name: | thermometer, electronic, clinical |
| Review Panel: | General Hospital |
| Product Code: | FLL |
| Device Class: | II |
{4}------------------------------------------------
Subject product: Non Contact Infrared forehead Thermometer
880.2910 Requlation Number:
#### 5.3 Predicate Device Information
KAZ USA, Inc. Sponsor: No Touch + Forehead Thermometer Device: K134043 510(K) Number:
#### 5.4 Device Description
Intrinity Global Limited
The thermometer measures the temperatures of people by detecting the energy.
The external probe plays an important role in the measuring process.
As soon as the thermometer is placed near the body and the radiation sensor is activated, the measurement will be taken instantly by detection of the infrared heat.
The thermometer includes two models: TVT-200, TVT-200 PLUS. The TVT-200 model includes four colors: pink, grey, orange and purple. While TVT-200 PLUS has only orange color.
The Non Contact Infrared Forehead Thermometer is composed by an IR sensor, a Human temperature measurement button, an object temperature button, Battery compartment, Buzzer, ℃ /°F button, a LCD and an enclosure.
The functions of TVT-200 and TVT-200 PLUS are the same, except the Back
Light. The detailed functions see below table:
| | Functions | | | | | | | |
|-----------------|-----------|-------|-----------|-----|--------|-------------|--------------------|------------|
| Models | Memory | °C/°F | Scan Mode | LCD | Buzzer | Fever alert | Power off/Auto Off | Back Light |
| TVT-200 | ● | ● | ● | ● | ● | ● | ● | ● |
| TVT-200<br>PLUS | ● | ● | ● | ● | ● | ● | ● | ○ |
# 5.5 Intended Use/ Indications for Use
Non Contact Infrared Forehead Thermometer is a non-sterile, reusable, handheld device. It can be used by consumers in homecare environment and doctors in
{5}------------------------------------------------
clinic as reference. It is intended for measuring human body temperature of all ranges of people by detecting infrared heat from the forehead.
# 5.6 Technological characteristics of the subject device compared to the predicate device
| Items | Predicate Device<br>(K134043), KAZ | Subject Device | Remarks |
|------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------|
| Indications for<br>use | The No Touch + Forehead<br>Thermometer (Model<br>NTF3000US) is a non-<br>sterile, reusable clinical<br>thermometer intended for<br>the intermittent<br>determination of human<br>body temperature in a touch<br>and no touch on the centre<br>of the forehead as the<br>measurement site on people<br>of all ages. | Non Contact Infrared<br>Forehead Thermometer is a<br>non-sterile, reusable,<br>handheld device. It can be<br>used by consumers in<br>homecare environment and<br>doctors in clinic as reference.<br>It is intended for measuring<br>human body temperature of<br>all ranges of people by<br>detecting infrared heat from<br>the forehead. | Equivalent:<br>intended<br>use<br>remains<br>unchanged |
| Measurement<br>method | Infrared radiation detection | Infrared radiation detection | Same |
| Measurement<br>mode | Forehead measure mode | Forehead measurement<br>mode | Same |
| Measuring<br>range | Forehead temperature<br>mode:<br>34.4°C - 42.2°C (93.9°F -<br>108°F) | Forehead measurement<br>mode:<br>32° to 43°C (89.6°F to<br>109.4°F) | Similar<br>(Note01) |
| Display<br>resolution | 0.1°C (0.1°F) | 0.1°C/0.1°F | Same |
| C/F switchable | Yes | Yes | Same |
| Measuring<br>accuracy | Forehead temperature<br>mode: ±0.2°C (0.4°F) | Forehead measurement<br>mode:<br>±0.2°C (0.4°F) | Same |
| Display | LCD display | LCD display | Same |
| Measurement<br>distance | 0-5cm | 1cm | Similar<br>(Note02) |
| Memory | Not available. | 16 sets. | (Note03) |
| Items | Predicate Device<br>(K134043), KAZ | Subject Device | Remarks |
| Power source | Two AA batteries | One 1.5V AAA alkaline<br>battery | Similar<br>(Note04) |
| Low battery<br>indication | Yes | Yes | Same |
| Waterproof | No, IP20 | No, IP22 | Similar<br>(Note05) |
| Operating<br>condition | 15 °C – 40 °C (59 °F –<br>104 °F), 15–95% non<br>condensing | 15°C~ 40°C (59°F~ 104°F);<br>≤95% RH | Similar<br>(Note06) |
| Patient contact<br>materials | metals and resins | ABS with colorants (pink,<br>grey, orange and purple),<br>Glass and Metal | Different<br>(Note07) |
| Cleaning/<br>disinfection | The thermometer casing<br>and the measuring probe<br>are cleaned and disinfected<br>by alcohol (70% Isopropyl). | The thermometer casing and<br>the measuring probe are<br>cleaned and disinfected by<br>70% alcohol. | Different<br>(Note08) |
| Biocompatibility | Comply with ISO 10993-<br>5:2009, ISO 10993-10:2010 | Comply with ISO 10993-<br>5:2009, ISO 10993-10:2010 | Same |
| Electric Safety<br>and EMC | IEC 60601-1 3rd edition:<br>2005, IEC 60601-1-2: 2007,<br>IEC 60601-1-11: 2010 | IEC 60601-1: 2005+CORR.1<br>(2006)+ CORR.2 (2007), IEC<br>60601-1-2: 2014, IEC 60601-<br>1-11: 2010, ISO 80601-2-56:<br>2009. | Similar |
| Performance | ASTM E1965-98 (2009) | ASTM E1965-98 (2009), ISO<br>80601-2-56: 2009 | Similar |
{6}------------------------------------------------
Subject product: Non Contact Infrared forehead Thermometer
Version:A/0
#### Note 01:
The measuring range of subject device meet the minimum rated output range of clinical thermometer requirement, from 35°C to 42°C, which is stated in standard ISO 80601-2-56. The differences between the predicate device and subject device will not affect the safety and effectiveness of the subject device.
#### Note 02:
The difference in measurement distance will not affect the safety and effectiveness of the subject device. And the use method are described detailed in IFU.
{7}------------------------------------------------
## Note 03:
The memory capacity will not affect the safety and effectiveness of the subject device.
### Note 04:
The main difference between the batteries of the Predicate and the subject device is the size, which will not affect the safety and effectiveness. Furthermore, the electric safety of the battery has been validated with the subject device according to the IEC 60601-1. The differences between the predicate device and subject device will not affect the safety and effectiveness of the subject device.
### Note 05:
The difference in waterproof will not affect the safety and effectiveness of the subject device, which has passed IEC 60601-1 and IEC 60601-1-11 safety test. The differences between the predicate device and subject device will not affect the safety and effectiveness of the subject device.
# Note 06:
The operating condition of subject device has passed the safety test, and the Instructions for Use in VOL 13 Labeling provides the operating condition, so the difference between the operating conditions of subject device and predicate device will not affect the safety and effectiveness of subject device.
#### Note 07:
The patient contact materials including colorants (pink, grey, orange and purple) have passed the ISO 10993-1:2009, ISO 10993-5:2009, ISO 10993-10:2010 Biocompatibility Test. The differences between the predicate device and subject device will not affect the safety and effectiveness of the subject device.
# Note 08:
The cleaning has been validated according to the standards ASTM E2314 -03(2014). The differences between the predicate device and subject device will not affect the safety and effectiveness of the subject device.
{8}------------------------------------------------
The subject device and the predicate device have the same intended use and similar technological characteristics, they both use infrared radiation detection method to detect human body forehead temperature. Their design is compact, small and light-weight. They are same in measuring accuracy, and similar in measuring range. Thus, the subject device is substantially equivalent to the predicate devices.
# 5.7 Brief discussion of nonclinical tests
| Tests | Test Standards | Results |
|-------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------|
| Electric Safety | IEC 60601-1:2005+A1:2012, Medical electrical<br>equipment - Part 1: General requirements for<br>basic safety and essential performance;<br>IEC 60601-1-11: 2010, Medical electrical<br>equipment –Part 1-11: General requirements for<br>basic safety and essential performance –<br>Collateral Standard: Requirements for medical<br>electrical equipment and medical electrical<br>systems used in the home healthcare<br>environment | Pass |
| EMC | IEC 60601-1-2:2014, Medical electrical<br>equipment - Part 1-2: General requirements for<br>basic safety and essential performance -<br>Collateral Standard: Electromagnetic compatibility<br>- Requirements and tests; | Pass |
| Bench performance | ISO 80601-2-56 First Edition 2009-10-01, Medical<br>electrical equipment - Part 2-56: Particular<br>requirements for basic safety and essential<br>performance of clinical thermometers for body<br>temperature measurement | Pass |
Nonclinical tests of the Infrared Thermometer are listed as below table:
{9}------------------------------------------------
#### Subject product: Non Contact Infrared forehead Thermometer
Version:A/0
| Tests | Test Standards | Results |
|-----------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------|---------|
| | ASTM E1965-98 (Reapproved 2009): Standard<br>Specification for Infrared Thermometers for<br>Intermittent Determination of Patient Temperature | |
| Clinical accuracy | ASTM E1965-98 (Reapproved 2009): Standard<br>Specification for Infrared Thermometers for<br>Intermittent Determination of Patient Temperature | Pass |
| Biological Evaluation | AAMI ANSI ISO 10993-1:2009/(R)2013,<br>Biological evaluation of medical devices - Part 1:<br>Evaluation and testing within a risk management<br>process | Pass |
| In vitro Cytotoxicity | ISO 10993-5:2009, Biological evaluation of<br>medical devices - Part 5: Tests for in vitro<br>cytotoxicity | Pass |
| irritation and skin sensitization | ISO 10993-10: 2010, Biological evaluation of<br>medical devices - Part 10: Tests for irritation and<br>skin sensitization | Pass |
#### 5.8 Brief discussion of clinical tests
The clinical performance test protocol and data analysis is conducted as the requirement of ASTM E1965-98 (2009). The test report showed the clinical performance of the subject device complied with the requirement of ASTM E1965-98 (2009).
#### 5.9 Conclusions
Based on the above information, we conclude that the subject device, Non Contact Infrared Forehead Thermometer, is substantially equivalent to the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.