← Product Code [DTN](/productcode/DTN) · K170583

# Intersept Filtered Cardiotomy Reservoir with Cortiva BioActive Surface (K170583)

_Medtronic, Inc. · DTN · Mar 17, 2017 · Cardiovascular · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K170583

## Device Facts

- **Applicant:** Medtronic, Inc.
- **Product Code:** [DTN](/productcode/DTN.md)
- **Decision Date:** Mar 17, 2017
- **Decision:** SESE
- **Submission Type:** Special
- **Regulation:** 21 CFR 870.4400
- **Device Class:** Class 2
- **Review Panel:** Cardiovascular

## Indications for Use

The Intersept filtered cardiotomy reservoir with Cortive surface is indicated for use in the cardiopulmonary bypass circuit during surgery for: an air/fluid separation chamber a temporary storage reservoir for priming solution and blood the filtration of particulates from bank blood and the storage and filtration of blood recovered from the field by suction the addition of medications or other fluids.

## Device Story

Single-use cardiotomy reservoir; 2,600 mL capacity; 2 LPM max flow rate. Functions as air/fluid separation chamber, temporary storage for priming solution/blood, and filtration unit for bank/field-recovered blood. Contains open-cell polyurethane defoamer and 20-micron microaggregate filter with polyester sleeve. Features eight access ports for suction, pre-defoamer, post-defoamer, and venting. Used in cardiopulmonary bypass circuits during surgery; operated by perfusionists/surgical staff. Cortiva BioActive surface coating minimizes blood-material interaction. Output is processed blood/fluid ready for return to bypass circuit. Benefits patient by removing particulates and air from blood during extracorporeal circulation.

## Clinical Evidence

Bench testing only. Testing included bioactivity, leaching, and coverage for the Cortiva coating process, and biocompatibility, filtration efficiency, and blood trauma for the alternate material formulation. All tests passed.

## Technological Characteristics

Single-use reservoir; 2,600 mL capacity; 2 LPM max flow. Materials: open-cell polyurethane defoamer, 20-micron microaggregate filter, polyester sleeve. Cortiva BioActive surface coating. Eight luer/0.6 cm ID access ports. Sterile, nonpyrogenic fluid path. Mechanical filtration and air/fluid separation principle.

## Regulatory Identification

A cardiopulmonary bypass blood reservoir is a device used in conjunction with short-term extracorporeal circulation devices to hold a reserve supply of blood in the bypass circulation.

## Special Controls

*Classification.* Class II (special controls), except that a reservoir that contains a defoamer or filter is classified into the same class as the defoamer or filter. The device, when it is a cardiopulmonary bypass blood reservoir that does not contain defoamers or blood filters, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 870.9.

## Predicate Devices

- Intersept Filtered Cardiotomy Reservoir ([K151110](/device/K151110.md))
- Intersept Filtered Cardiotomy Reservoir with Medtronic/Carmeda BioActive Surface ([K926226](/device/K926226.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the Department of Health & Human Services - USA. The logo is a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" written around the top half of the circle. The bottom half of the circle contains a stylized image of a bird.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

March 17, 2017

Medtronic, Inc. Gayatri Ghadge Regulatory Affairs Specialist 710 Medtronic Parkway Minneapolis, MN 55432-5604

Re: K170583

Trade/Device Name: Intersept™ Filtered Cardiotomy Reservoir with Cortiva™ BioActive Surface Regulation Number: 21 CFR 870.4400 Regulation Name: Cardiopulmonary Bypass Blood Reservoir Regulatory Class: Class II Product Code: DTN, DTP

Received: February 28, 2017

Dated: February 27, 2017

Dear Ms. Ghadge:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set

{1}------------------------------------------------

forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely,

M.A. Hilleman

for Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

# Indications for Use

# 510(k) Number (if known)

K170583

Device Name

Intersept™ Filtered Cardiotomy Reservoir with Cortiva™ BioActive surface

Indications for Use (Describe)

The Intersept filtered cardiotomy reservoir with Cortive surface is indicated for use in the cardiopulmonary bypass circuit during surgery for:

· an air/fluid separation chamber

- · a temporary storage reservoir for priming solution and blood
- the filtration of particulates from bank blood and the storage and filtration of blood recovered from the field by suction
- · the addition of medications or other fluids.

| Type of Use (Select one or both, as applicable)                                                                |                                                                                                               |
|----------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------|
| <div> <span>                     ☒ Prescription Use (Part 21 CFR 801 Subpart D)                 </span> </div> | <div> <span>                     □ Over-The-Counter Use (21 CFR 801 Subpart C)                 </span> </div> |

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

Page 1 of 1

{3}------------------------------------------------

# 510(k) Summary

| Date Prepared:  | February 27, 2017                                                                                                                                                        |
|-----------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Applicant:      | Medtronic, Inc.<br>Medtronic Perfusion Systems<br>7611 Northland Drive<br>Minneapolis, MN 55428<br>Establishment Registration Number: 2184009                            |
| Contact Person: | Gayatri Ghadge<br>Regulatory Affairs Specialist<br>Medtronic Perfusion Systems<br>Phone: 763.526.9336<br>Fax: 763.367.8361<br>Email: gayatri.a.ghadge@medtronic.com      |
|                 | Alternate Contacts:                                                                                                                                                      |
|                 | Chelsea Pioske<br>Senior Regulatory Affairs Supervisor<br>Medtronic Perfusion Systems<br>Phone: 763.514.9838<br>Fax: 763.367.8360<br>Email: chelsea.pioske@medtronic.com |
|                 | Jake Roeller<br>Senior Regulatory Affairs Manager<br>Medtronic Perfusion Systems<br>Phone: 763.526.0404<br>Fax: 763.367.8360<br>Email: jake.w.roeller@medtronic.com      |

# Device Name and Classification

| Trade Name:             | Intersept™ Filtered Cardiotomy Reservoir with Cortiva™<br>BioActive Surface   |
|-------------------------|-------------------------------------------------------------------------------|
| Classification Name:    | Reservoir, Blood, Cardiopulmonary Bypass; Defoamer,<br>Cardiopulmonary Bypass |
| Regulation Number:      | 21 CFR 870.4400                                                               |
| Product Classification: | Class II                                                                      |
| Product Code:           | DTN, DTP                                                                      |

{4}------------------------------------------------

Name of Predicate Devices: Intersept™ Filtered Cardiotomy Reservoir

Intersept® Filtered Cardiotomy Reservoir with Medtronic/Carmeda® BioActive Surface

## Device Description

The Intersept Filtered Cardiotomy Reservoir with Cortiva BioActive Surface is a single-use, device with a sterile, nonpyrogenic fluid path. The maximum capacity of each reservoir is 2,600 mL. The maximum recommended flow rate is 2 liters per minute (LPM).

Each reservoir has eight luer or 0.6 cm (0.25 in) ID access ports;

- four suction access ports
- two pre-defoamer and ●
- two post-defoamer access ports. One of the post-defoamer access ports is intended for use ● as a vent.

Each Intersept Filtered Cardiotomy Reservoir contains an open cell polyurethane defoamer with a 20-micron microaggregate filter covered with a polyester sleeve.

# Indications for Use

The Intersept filtered cardiotomy reservoir with Cortive surface is indicated for use in the cardiopulmonary bypass circuit during surgery for:

- an air/fluid separation chamber ●
- a temporary storage reservoir for priming solution and blood ●
- the filtration of particulates from bank blood and the storage and filtration of blood ● recovered from the field by suction
- the addition of medications or other fluids

# Contraindications

Use the device only as indicated.

#### Predicate Devices

# K926226 Intersent Filtered Cardiotomy Reservoir with Medtronic/Carmeda BioActive Surface

Premarket Notification submission K926226. for the Intersept Cardiotomy Reservoir with filter (model CB1351), was cleared for market by FDA on June 22, 1993. This submission was for the initial introduction of the Intersept Filtered Cardiotomy Reservoir with Carmeda BioActive Surface into the market.

{5}------------------------------------------------

#### K151110 Intersept Filtered Cardiotomy Reservoir

Premarket Notification submission K151110, for the Intersept Cardiotomy Reservoir with filter (model 1351), was cleared for market by FDA on May 27, 2015. This submission included a proposed alternate material formulation to the inner frames used within the Intersept Filtered Cardiotomy Reservoir, as well as a proposed Instructions for Use/labeling change.

#### Comparison to Predicate Devices

A comparison of the Intersept Filtered Cardiotomy Reservoir with Cortiva BioActive Surface to the predicate device (the Intersept Cardiotomy Reservoir with Filter and Intersept Filtered Cardiotomy Reservoir with Medtronic/Carmeda BioActive Surface) indicates the following similarities:

- Intended Use: The intended use is the same as predicate device.
- Design: The overall design is the same as the predicate. ●
- Materials: The material types used are the same as the predicate.
	- The proposed modification in this submission from the predicate device is to an O alternate material formulation to the inner frame used within the Intersept Filtered Cardiotomy Reservoir.
- Principles of Operation and Technology: The principles of operation are the same as the ● predicate device.
- Performance: The performance is substantially equivalent to the predicate device.
- Coating: The Cortiva coating is the same as the predicate.
	- The proposed modification in the submission from the predicate device includes: O slight alterations to the process steps and solution concentrations of the coating. This includes the alteration to coating process of the defoamer frame assembly inside the reservoir jar; and the filtration sock assembly inside a container.

#### Summary of Testing

Testing has demonstrated that the Intersept Filtered Cardiotomy Reservoir with Cortiva BioActive Surface is substantially equivalent to the predicate.

| Change                                | Verification/Validation | Results |
|---------------------------------------|-------------------------|---------|
| Cortiva Coating Process Change        | Bioactivity             | Pass    |
|                                       | Leaching                | Pass    |
|                                       | Coverage                | Pass    |
| Alternate Material Formulation Change | Biocompatibility        | Pass    |
|                                       | Filtration Efficiency   | Pass    |
|                                       | Blood Trauma            | Pass    |

The following tests were conducted:

{6}------------------------------------------------

## Conclusion

The data included in this submission is sufficient to provide reasonable assurance that the Intersept Filtered Cardiotomy Reservoir with Cortiva BioActive Surface is substantially equivalent to the legally marketed predicate device, the Intersept Filtered Cardiotomy Reservoir with Medtronic/Carmeda BioActive Surface and Intersept Filtered Cardiotomy Reservoir.

---

**Source:** [https://fda.innolitics.com/device/K170583](https://fda.innolitics.com/device/K170583)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
