MiiS Horus+ Scope DEA 200 is a digital hand-held slit lamp system indicated for non-invasive illumination, magnification, visualization and to record digital photographs and video of anterior segment (including cornea, anterior chamber, and lens) of the human eye and surrounding area.
Device Story
Digital hand-held slit lamp system; uses white LED illumination and high-resolution lens with 5M pixel CMOS sensor to capture images/video of anterior eye segment. Operated by clinicians in clinical settings. Device features 3.5-inch TFT-LCD display for real-time visualization; stores data on SD memory card or transfers via USB to computer; supports AV output to external monitors. Provides high-quality digital records for medical filing; assists in diagnosis and documentation of anterior segment conditions. Battery-powered (Li-ion).
Clinical Evidence
No clinical studies were performed. Evidence is based on bench testing, including software/hardware validation, EMC testing, safety testing, and optical radiation hazard evaluation per ISO 15004-2:2007.
Technological Characteristics
Hand-held slit lamp; Li-ion battery-powered; 5M pixel CMOS sensor; white LED light source; 3.5-inch TFT-LCD display; turret-type slit switching; blue/green filters; SD card/USB connectivity; AV output. Complies with ISO 15004-2:2007 for light safety (Group 2 instrument).
Indications for Use
Indicated for non-invasive illumination, magnification, visualization, and recording of digital photographs and video of the anterior segment (cornea, anterior chamber, lens) and surrounding area of the human eye.
Regulatory Classification
Identification
An AC-powered slitlamp biomicroscope is an AC-powered device that is a microscope intended for use in eye examination that projects into a patient's eye through a control diaphragm a thin, intense beam of light.
Special Controls
*Classification.* Class II (special controls). The device, when it is intended only for the visual examination of the anterior segment of the eye, is classified as Group 1 per FDA-recognized consensus standard ANSI Z80.36, does not provide any quantitative output, and is not intended for screening or automated diagnostic indications, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 886.9.
{1}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food & Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
October 6, 2017
Medimaging Integrated Solution Inc. (miis) % Yi-Ying Chen Engineer 1F, No.7, R&D Rd. II Hsinchu Science Park Hsinchu, Taiwan 30076 (R.O.C.)
Re: K170470
Trade/Device Name: Digital Eye Anterior Camera Regulation Number: 21 CFR 886.1850 Regulation Name: AC-Powered Slitlamp Biomicroscope Regulatory Class: Class II Product Code: HJO, HKI Dated: August 24, 2017 Received: August 29, 2017
Dear Yi-Ying Chen:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies.
{2}------------------------------------------------
You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling
(21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events)
(21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation
(21 CFR Part 820); and if applicable, the electronic product radiation control provisions
(Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education (DICE) at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also. please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education (DICE) at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely.
# Bradley S. Cunningham -S
for Malvina B. Eydelman, M.D. Director Division of Ophthalmic and Ear, Nose and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
# Indications for Use
510(k) Number (if known) K170470
Device Name MiiS Horus+ Scope DEA 200
Indications for Use (Describe)
MilS Horus+ Scope DEA 200 is a digital hand-held slit lamp system indicated for non-invasive illumination, magnification, visualization and to record digital photographs and video of anterior segment (including cornea, anterior chamber, and lens) of the human eye and surrounding area.
| Type of Use (Select one or both, as applicable) | |
|----------------------------------------------------------------------------------|--------------------------------------------------------------------------------|
| <div> <span> Prescription Use (Part 21 CFR 801 Subpart D) </span> </div> | <div> <span> Over-The-Counter Use (21 CFR 801 Subpart C) </span> </div> |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
Image /page/3/Picture/0 description: The image contains a single, large, uppercase letter 'K' in a simple, sans-serif font. The letter is black and stands out against a white background. The letter is centered in the image and takes up most of the space. The image is simple and clear.
Image /page/3/Picture/1 description: The image contains the logo for MiiS. The logo is a circular target with alternating black and white rings. The text "MiiS" is written in blue below the target.
K170470
# 510(k) Summary
Prepared: October 1, 2017
| Submitter/Owner's<br>Name/ Address | Medimaging Integrated Solution Inc. (MiiS)<br>1F, No.7, R&D Rd. II, Hsinchu Science Park, Hsinchu, Taiwan<br>30076 (R.O.C.) |
|------------------------------------|-----------------------------------------------------------------------------------------------------------------------------|
| Contact Person: | Hsu, Chih-Lu (COO) |
| Telephone: Fax: | +886-3-5798860 Ext: 1110 |
| Email Address: | +886-3-5798821<br>luu@miis.com.tw |
{5}------------------------------------------------
Image /page/4/Picture/1 description: The image is a logo with a circular design at the top and the word "MiiS" in blue at the bottom. The circular design consists of concentric rings divided into alternating black and white segments, resembling a target or a stylized eye. The word "MiiS" is written in a bold, sans-serif font, with a blue color that stands out against the white background. The entire logo is enclosed within a rounded square with a thin blue border.
K170470
## Device Identification:
Trade/Device Name MiiS Horus Scope DEA 200 Regulation Number: 21 CFR 886.1850 Regulation Name: AC-powered slit lamp biomicroscope Regulatory Class: Class II Product Code: HJO, HKI
## Predicate Device:
- 1. K063640
Trade/Device Name: Kowa SL-15 Slit Lamp Regulation Number: 21 CFR 886.1850 Regulation Name: AC-powered slit lamp biomicroscope Regulatory Class: Class II Product Code: HJO
- 2. K120982
Trade/Device Name: MiiS Horus Scope DEC 100 Regulation Number: 21 CFR 886.1120 Regulation Name: Ophthalmic Camera Regulatory Class: Class II Product Code: HKI
### Description of Device:
MiiS Horus* Scope DEA 200 is a digital hand-held slit lamp system indicated for non-invasive illumination, magnification, visualize and to record digital photographs and video of anterior segment (including cornea, anterior chamber, and lens) of the human eye and surrounding area. It is a Li-ion battery-powered optical device. It brings more complete medical records about the static photos as well as the dynamic videos. MiiS Horus* Scope DEA 200 has an LED light source with visible white light. The device is designed with
high-resolution lens and 5M pixels CMOS Sensor, faithful rendering color of the anterior segment (including cornea, anterior chamber, and lens) of the human eye and surrounding area. The device can store pictures or videos in SD memory card, or via the USB transfer pictures or videos to a computer. In addition to rendering images in the 3.5-inch full color TFT-LCD, through the AV output, you can connect the device to the big screen (TV, LCD screen) showing the pictures or videos. Below includes a summary of the technical information used in the substantial equivalence comparison. It is more efficient and suitable for many different applications, such as electronic filing.
{6}------------------------------------------------
Image /page/5/Picture/1 description: The image shows a logo for MiiS. The logo is a rounded square with a white background. Inside the square is a circular target-like design with alternating black and white sections. Below the target design, the text "MiiS" is written in blue, with a line underneath it.
# Indications for Use:
MiiS Horus+ Scope DEA 200 is a digital hand-held slit lamp system indicated for noninvasive illumination, magnification, visualization and to record digital photographs and video of anterior segment (including cornea, anterior chamber, and lens) of the human eye and surrounding area.
# Substantial Equivalence Summary
Key technological seminaries and differences between the ICON and the predicates:
| Device | K170470 | K063640<br>Predicate Device | K120982<br>Predicate Device |
|--------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Model name | MiiS Horus+ Scope<br>DEA 200 | KOWA SL-15 | MiiS Horus Scope<br>DEC 100 |
| Intended use | MiiS Horus+ Scope DEA 200 is<br>a digital hand-held slit lamp<br>system indicated for non-<br>invasive illumination,<br>magnification, visualization and<br>to record digital photographs<br>and video of anterior segment<br>(including cornea, anterior<br>chamber, and lens) of the<br>human eye and surrounding<br>area. | KOWA SL-15 is an ophthalmic<br>device indicated for<br>non-invasive illumination,<br>magnification and observation<br>of the human eye. It consists of<br>a hand-held, battery powered<br>slit-lamp biomicroscope with<br>viewing and illumination<br>optical systems and an<br>AC-powered stand. | MiiS Horns Scope DEC 100 is<br>a digital hand-held eye-fundus<br>camera used to record digital<br>photographs and video of<br>fundus of the human eye and<br>surrounding area. |
| Slit Lamp | | | |
| Slit selection | Turret-type switching | Turret-type switching | |
| Slit length | 10 mm Fixed | 12 mm Fixed | |
| Slit Width | $≤$ 0.2, 0.2, 0.5, 2, 5,<br>10 mm | 0.1, 0.2, 0.8 mm and<br>$ ⌀ $ 12mm spot | |
| Slit projection<br>angle | $ \pm $ 45 degree (+/- 5%) | 60 degree for<br>Horizontal | |
| Filter | Blue, Green | Blue | None |
| Digital Camera | | | |
| Picture/Video | Built-in | Supports electronic image<br>with the optional camera<br>connection adapter<br>(C-mount). | Built-in |
| Image sensor | CMOS | CCD | CMOS |
{7}------------------------------------------------
Image /page/6/Picture/1 description: The image shows a logo for Miis. The logo is a circular target with alternating black, white, and gray sections. The word "Miis" is written in blue, bold letters below the target.
| Magnification | 10X / 16X | 10X / 16X | 0.25X | | | | | | | | |
|----------------------------------|----------------------------------------------------------------------------|------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------|--|--|--|--|--|--|--|--|
| Display | Display on the TFT-LCD or<br>connect to and display on<br>computer | Connect to and display on<br>the camera and computer<br>display. | Display on the TFT-LCD or<br>connect to and display on<br>computer | | | | | | | | |
| Image storage | SD card and could be transfer<br>to user's computer | User's computer | SD card and could be transfer<br>to user's computer | | | | | | | | |
| Light Source | | | | | | | | | | | |
| Type | White LED | Halogen Lamp | White LED and Infrared LED | | | | | | | | |
| Maximum<br>Output power | 2.9V, 1.0 W | 7.5V, 15 W | 2.9V, 1.0 W | | | | | | | | |
| Light Intensity<br>Adjustment | Six selection: 0B, 1B, 2B, 3B,<br>4B, 5B<br>From 0 mw /cm2 to 19.12 mw/cm2 | Three selection: Full, 1/2 and<br>1/4 | Sixteen selection:<br>0B, 1B, 2B, 3B, 4B, 5B, 6B, 7B,<br>8B, 9B, 10B, 11B, 12B, 13B,<br>14B, 15B | | | | | | | | |
| | | | White<br>LED Infrared<br>LED Cornea 338.21<br>mw/cm2 3.11<br>mw/cm2 Retina 3.88<br>mw/cm2 0.036<br>mw/cm2 | | | | | | | | |
| | | | | | | | | | | | |
| | | | | | | | | | | | |
| External light | Background Light | None | None | | | | | | | | |
| System | | | | | | | | | | | |
| Power Source | Lithium-ion rechargeable<br>battery | Lithium-ion rechargeable<br>battery | Lithium-ion rechargeable<br>battery | | | | | | | | |
| Appearance<br>Outline<br>Drawing | Image: microscope | Image: microscope | Image: microscope | | | | | | | | |
{8}------------------------------------------------
Image /page/7/Picture/1 description: The image is a logo with a circular target-like design at the top and the word "MiiS" in blue at the bottom. The target design consists of concentric circles divided into alternating black and white quadrants. The word "MiiS" is in a bold, sans-serif font and is underlined with a blue line. The entire logo is set against a white background with a rounded blue border.
## Substantial Equivalence Discussion
### Similarities:
- MiiS Horus scope DEA 200 and the primary predicate device, KOWA SL-15, have the i same intended use. DEA 200 has similar specifications to KOWA SL-15, such as the working distance, slit length, slit width, filters, light weight and magnification function.
- Both MiiS Horus scope DEA 200 and secondary predicate device, MiiS Horus scope -DEC 100 are Medimaging Integrated Solution Inc. (MiiS)'s products. They have same concept of design for image sensor, display function and image storage.
### Differences:
- The Major differences are the Changes in light sources and optical system. Optical radiation hazard evaluation was performed that these changes will not affect the device safety.
## Nonclinical and Clinical Tests
In this submission, the nonclinical tests were conducted, such as software (S/W) and hardware (H/W), EMC, safety and light hazard. The results of this testing indicate that the MiiS Horus+ scope DEA 200 is equivalent to predicate device.
No clinical studies were performed.
## Optical Performance
The MiiS Horus Scope DEA 200 was tested according to 15004-2:2007 to determine acceptable light safety limits for both the illumination and background lights. The test results demonstrate the DEA 200 is in compliance with the of Group 2 instrument requirements provided by the standard.
## Conclusion
Substantial equivalence comparison and bench performance tests support the conclusion of substantial equivalence of MiiS Horus* Scope DEA 200 to the predicate devices (KOWA SL-15 and MiiS Horus Scope DEC 100).
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.