Reprocessed HARMONIC FOCUS Shears + Adaptive Tissue Technology
K170456 · Stryker Sustainability Solutions · NLQ · Apr 5, 2017 · SU
Device Facts
| Record ID | K170456 |
| Device Name | Reprocessed HARMONIC FOCUS Shears + Adaptive Tissue Technology |
| Applicant | Stryker Sustainability Solutions |
| Product Code | NLQ · SU |
| Decision Date | Apr 5, 2017 |
| Decision | SESE |
| Submission Type | Traditional |
| Device Class | Class U |
| Attributes | Therapeutic, Pediatric |
Indications for Use
Reprocessed HARMONIC FOCUS Shears + Adaptive Tissue Technology are indicated for soft tissue incisions when bleeding control and minimal thermal injury are desired. The instrument can be used as an adjunct to or substitute for electrosurgery, lasers, and steel scalpels in general, otorhinolaryngologic (ENT), plastic, pediatric, gynecologic, urologic, exposure to orthopedic structures (such as spine and joint space) and other open procedures.
Device Story
Reprocessed ultrasonic surgical instrument; used for soft tissue incision and coagulation. Input: electrical power from Generator G11; device contains EEPROM chip for communication with generator. Operation: ultrasonic energy converts electrical power to mechanical vibration at blade; Adaptive Tissue Technology algorithms adjust generator output based on tissue conditions. Output: mechanical cutting and thermal coagulation of vessels up to 5mm. Used in OR by surgeons for open procedures. Benefits: provides hemostasis with minimal thermal injury; serves as adjunct/substitute for electrosurgery, lasers, or scalpels. Reprocessing involves cleaning, decontamination, and functional testing of components.
Clinical Evidence
Bench testing and preclinical animal studies (acute and chronic survival). Bench tests: biocompatibility, sterilization validation, electrical safety (IEC 60601-1, IEC 60601-2-2), EMC (IEC 60601-1-2), thermal spread, transection time, burst pressure, and device reliability. Animal studies evaluated hemostasis and thermal spread on vessels 1-5mm in diameter. Results demonstrate performance equivalent to the predicate.
Technological Characteristics
Ultrasonic shears with curved blade and clamp arm. Includes EEPROM memory chip for generator communication. Materials: standard medical-grade components (reprocessed). Energy: ultrasonic vibration powered by external generator. Connectivity: proprietary interface with Generator G11. Sterilization: validated reprocessing cycle. Software: embedded firmware/algorithms for Adaptive Tissue Technology.
Indications for Use
Indicated for soft tissue incisions in patients undergoing general, ENT, plastic, pediatric, gynecologic, urologic, or orthopedic open procedures where bleeding control and minimal thermal injury are required. Suitable for cutting and coagulating vessels up to 5mm in diameter.
Regulatory Classification
Identification
Single-use reprocessed versions of the ultrasonic devices used in surgical procedures either for fragmentation, emulsification and aspiration of soft tissue and hard tissue or for ligation of vessels (under product code LFL). Reprocessing validation data for this device type must be included in a 510(k) submission.. (70 FR 56911, available at https://www.govinfo.gov/content/pkg/FR-2005-09-29/pdf/05-19510.pdf).
Predicate Devices
- HARMONIC FOCUS® Shears + Adaptive Tissue Technology (K133314)
Submission Summary (Full Text)
{0}------------------------------------------------
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
April 5, 2017
Stryker Sustainability Solutions Ms. Chelsea Cullen Sr. Regulatory Affairs Specialist 1810 W Drake Drive Tempe, Arizona 85283
Re: K170456
Trade/Device Name: Reprocessed HARMONIC FOCUS Shears + Adaptive Tissue Technology Regulatory Class: Unclassified Product Code: NLQ Dated: February 13, 2017 Received: February 15, 2017
Dear Ms. Cullen:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in
Image /page/0/Picture/11 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, with flowing lines above them.
{1}------------------------------------------------
the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely,
# Jennifer R. Stevenson -S
For Binita S. Ashar, M.D., M.B.A., F.A.C.S. Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Device Models Subject to Clearance:
| Model<br>Number | Device Name/Description | Device Length | Active Blade<br>Length |
|-----------------|--------------------------------------------------------|---------------|------------------------|
| HAR9F | HARMONIC FOCUS® Shears + Adaptive<br>Tissue Technology | 9cm | 16mm |
{3}------------------------------------------------
Form Approved: OMB No. 0910-0120
Expiration Date: January 31, 2017
See PRA Statement below.
#### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
## Indications for Use
510(k) Number (if known)
K170456
Device Name
Reprocessed HARMONIC FOCUS® Shears + Adaptive Tissue Technology
Indications for Use (Describe)
The Reprocessed HARMONIC FOCUS® Shears + Adaptive Tissue Technology are indicated for soft tissue incisions when bleeding control and minimal thermal injury are desired. The instrument can be used as an adjunct to or substitute for electrosurgery, lasers, and steel scalpels in general, otorhinolaryngologic (ENT), plastic, pediatric, gynecologic, urologic, exposure to orthopedic structures (such as spine and joint space) and other open procedures.
| Type of Use ( <i>Select one or both, as applicable</i> ) | |
|----------------------------------------------------------|--|
|----------------------------------------------------------|--|
> Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
Page 1 of 1
{4}------------------------------------------------
# K170456
## SECTION 5: 510(k) SUMMARY
#### Submitter:
Stryker Sustainability Solutions 1810 W. Drake Drive Tempe, Arizona 85283
Contact:
Chelsea Cullen Sr. Regulatory Affairs Specialist 888-888-3433x5506 (o) 480-763-5310 (f) chelsea.cullen@stryker.com
Date of Preparation: February 13, 2017
#### Name of Device:
| Trade/Proprietary Name: | Reprocessed HARMONIC FOCUS® Shears + Adaptive<br>Tissue Technology |
|-----------------------------|--------------------------------------------------------------------------------|
| Model Number: | HAR9F |
| Common Name: | Scalpel, Ultrasonic, Reprocessed |
| Classification Information: | Class: Unclassified<br>Unclassified Reason: Pre-Amendment<br>Product Code: NLQ |
#### Predicate Devices:
| Model<br>Number | 510(k)<br>Number | 510(k) Title | Original<br>Manufacturer |
|-----------------|------------------|--------------------------------------------------------|--------------------------|
| HAR9F | K133314 | HARMONIC FOCUS® Shears + Adaptive Tissue<br>Technology | Ethicon Endo-<br>Surgery |
### Device Description:
Reprocessed HARMONIC FOCUS® Shears + Adaptive Tissue Technology are a sterile, single patient use instrument consisting of a soft grip scissor handle housing assembly with two hand controls (MIN for minimum power level and MAX for maximum power level). The instrument has a curved blade and clamp arm with Teflon pad. The instrument is 9 cm in length with a 16 mm active blade length. The Reprocessed HARMONIC FOCUS®+ Shears instrument allows for the cutting and coagulation of vessels up to and including 5 mm in diameter.
{5}------------------------------------------------
Each Reprocessed HARMONIC FOCUS®+ Shears instrument is packaged with one sterile, single patient use, disposable Torque Wrench.
### Intended Use:
Reprocessed HARMONIC FOCUS® Shears + Adaptive Tissue Technology are indicated for soft tissue incisions when bleeding control and minimal thermal injury are desired. The instrument can be used as an adjunct to or substitute for electrosurgery, lasers, and steel scalpels in general, otorhinolarynqologic (ENT), plastic, pediatric, gynecologic, urologic, exposure to orthopedic structures (such as spine and joint space) and other open procedures.
### Summary of Technological Characteristics:
The Reprocessed HARMONIC FOCUS Shears + Adaptive Tissue Technology use an EEPROM memory chip that stores device identification, usage tracking, and operating parameters for use by the Generator G11 that provides power for the HARMONIC FOCUS Shears + Adaptive Tissue Technology. Adaptive Tissue Technology refers to the advanced algorithms that allow the generator to communicate with the device and properly respond to the changing tissue conditions. The design, materials, and intended use of the reprocessed device are equivalent to the predicate device. The mechanism of action of the reprocessed device is identical to the predicate device in that the same standard mechanical design, materials, and sizes is utilized. The differences between the predicate device and the reprocessed device include a replacement Torque Wrench and replacement tissue pad. There are no changes to the claims, intended use, clinical applications, patient population, performance specifications, or method of operation. In addition, Stryker Sustainability Solutions' reprocessing of the HARMONIC FOCUS® Shears + Adaptive Tissue Technology includes removal of adherent visible soil and decontamination. Each individual device is tested for appropriate function of its components prior to packaging and labeling operations.
#### Performance Data:
Bench and laboratory testing was conducted to demonstrate performance (safety and effectiveness) of Reprocessed HARMONIC FOCUS® Shears + Adaptive Tissue Technology. This included the following tests:
- Biocompatibility
- . Validation of Reprocessing
- . Sterilization Validation
- Functional Performance Tests ●
- Electrical Safety Testing
- Electromagnetic Compatibility Testing
- . Packaging Validation
The functional performance testing involved electrical safety and electromagnetic compatibility testing in accordance with IEC 60601-1, IEC 60601-1-2, and IEC 60601-2-2. and verification/comparative testing (to the predicate device). The bench testing involved evaluation of the device's performance and ability to seal and divide vessels up
{6}------------------------------------------------
to 5mm, including: thermal spread, transection time, burst pressure, device functionality, and device reliability.
Additionally, preclinical laboratory evaluations in an animal model were performed, which included acute and chronic survival studies. The studies were done to evaluate thermal spread and the ability to achieve hemostasis of the reprocessed device. The results of the evaluations demonstrate that the Reprocessed HARMONIC FOCUS Shears + Adaptive Tissue Technology effectively cut and coagulated vessels 1 to 5mm in diameter.
The performance testing demonstrates that reprocessed devices are as safe and effective as the predicate and operate as originally intended.
#### Conclusion:
The results of bench testing and preclinical laboratory evaluations demonstrate that the Reprocessed HARMONIC FOCUS® Shears + Adaptive Tissue Technology are at least as safe and effective and perform as well as the identified leqally marketed predicate device as described herein.