← Product Code [LPH](/productcode/LPH) · K170444

# PROCOTYL® PRIME Acetabular Cup System (K170444)

_Microport Orthopedics, Inc. · LPH · Jun 26, 2017 · Orthopedic · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K170444

## Device Facts

- **Applicant:** Microport Orthopedics, Inc.
- **Product Code:** [LPH](/productcode/LPH.md)
- **Decision Date:** Jun 26, 2017
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 888.3358
- **Device Class:** Class 2
- **Review Panel:** Orthopedic
- **Attributes:** Therapeutic

## Indications for Use

The PROCOTYL® PRIME Acetabular Cup System is intended for use in total hip arthroplasty for reduction or relief of pain and/or improved hip function in skeletally mature patients with the following conditions: 1) non-inflammatory degenerative joint disease such as osteoarthritis, avascular necrosis, ankylosis, protrusio acetabuli, and painful hip dysplasia; 2) inflammatory degenerative joint disease such as rheumatoid arthritis; 3) correction of functional deformity; and, 4) revision procedures where other treatments or devices have failed. Shells with BIOFOAM® metal foam coating are intended only for uncemented arthroplasty.

## Device Story

PROCOTYL® PRIME Acetabular Cup System is a modular hip implant consisting of titanium alloy acetabular shells (with BIOFOAM® coating), A-CLASS® highly crosslinked UHMWPE liners, and optional titanium alloy cancellous bone screws. Used in total hip arthroplasty to replace diseased or damaged hip joints; intended for reduction of pain and restoration of function. Implanted by orthopedic surgeons in a clinical/hospital setting. The system is modular, allowing surgeons to select shell/liner configurations based on patient anatomy. The device provides a stable, articulating surface for the femoral head component. Benefits include improved hip function and pain relief for patients with degenerative or inflammatory joint conditions. The system is compatible with various previously cleared MicroPort femoral components and instrumentation.

## Clinical Evidence

No clinical data provided. Substantial equivalence supported by nonclinical bench testing, including ASTM 1820 liner lock testing (push-out, lever-out, torque-out), deformation and frictional torque testing, long-term axial cyclic fatigue testing, wear rate and particle analysis, range of motion studies, and ASTM F543 bone screw testing.

## Technological Characteristics

Materials: Titanium alloy (shells/screws), A-CLASS® highly crosslinked UHMWPE (liners). Coating: BIOFOAM® metal foam. Dimensions: Shells 42-68mm OD; Liners 22-44mm ID; Screws 6.5mm diameter, 15-80mm length. Sterilization: Ethylene Oxide. Connectivity: None (mechanical implant).

## Regulatory Identification

A hip joint metal/polymer/metal semi-constrained porous-coated uncemented prosthesis is a device intended to be implanted to replace a hip joint. The device limits translation and rotation in one or more planes via the geometry of its articulating surfaces. It has no linkage across the joint. This generic type of device has a femoral component made of a cobalt-chromium-molybdenum (Co-Cr-Mo) alloy or a titanium-aluminum-vanadium (Ti-6Al-4V) alloy and an acetabular component composed of an ultra-high molecular weight polyethylene articulating bearing surface fixed in a metal shell made of Co-Cr-Mo or Ti-6Al-4V. The femoral stem and acetabular shell have a porous coating made of, in the case of Co-Cr-Mo substrates, beads of the same alloy, and in the case of Ti-6Al-4V substrates, fibers of commercially pure titanium or Ti-6Al-4V alloy. The porous coating has a volume porosity between 30 and 70 percent, an average pore size between 100 and 1,000 microns, interconnecting porosity, and a porous coating thickness between 500 and 1,500 microns. The generic type of device has a design to achieve biological fixation to bone without the use of bone cement.

## Predicate Devices

- LINEAGE® Acetabular System ([K002149](/device/K002149.md), [K052026](/device/K052026.md))
- Cancellous Bone Screw ([K864626](/device/K864626.md))
- DYNASTY® Acetabular System ([K082924](/device/K082924.md), [K002149](/device/K002149.md), [K061547](/device/K061547.md), [K070785](/device/K070785.md))
- Zimmer Biomet Trilogy Acetabular Cup System ([K934765](/device/K934765.md), [K953490](/device/K953490.md))
- Depuy Pinnacle 100 Shell ([K090998](/device/K090998.md))

## Reference Devices

- [K111699](/device/K111699.md)
- [K971429](/device/K971429.md)
- [K043099](/device/K043099.md)
- [K041114](/device/K041114.md)
- [K072656](/device/K072656.md)
- [K122382](/device/K122382.md)
- [K130376](/device/K130376.md)
- [K140043](/device/K140043.md)
- [K142119](/device/K142119.md)
- [K893685](/device/K893685.md)
- [K920593](/device/K920593.md)
- [K925512](/device/K925512.md)
- [K932222](/device/K932222.md)
- [K021349](/device/K021349.md)
- [K004043](/device/K004043.md)
- [K051348](/device/K051348.md)
- PMA [P030027](/device/P030027.md)
- [K003016](/device/K003016.md)
- [K012091](/device/K012091.md)
- [K021346](/device/K021346.md)
- [K041586](/device/K041586.md)
- [K051995](/device/K051995.md)
- [K052915](/device/K052915.md)
- [K053588](/device/K053588.md)
- [K060358](/device/K060358.md)
- [K080663](/device/K080663.md)
- [K081090](/device/K081090.md)
- [K091423](/device/K091423.md)
- [K100866](/device/K100866.md)
- [K110399](/device/K110399.md)
- [K111698](/device/K111698.md)
- [K111910](/device/K111910.md)
- [K112080](/device/K112080.md)
- [K112150](/device/K112150.md)
- [K121221](/device/K121221.md)
- [K123434](/device/K123434.md)
- [K123688](/device/K123688.md)
- [K130984](/device/K130984.md)
- [K140676](/device/K140676.md)
- [K141235](/device/K141235.md)
- [K150133](/device/K150133.md)
- [K150302](/device/K150302.md)

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/2 description: The image is a seal for the U.S. Department of Health & Human Services. The seal is circular and contains the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the top half of the circle. The bottom half of the circle contains a stylized image of three human profiles facing to the right.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

MicroPort Orthopedics Inc. Kaitlin Grove Regulatory Affairs Specialist II 5677 Airline Road Arlington, Tennessee 38002

June 26, 2017

Re: K170444

Trade/Device Name: PROCOTYL® PRIME Acetabular Cup System Regulation Number: 21 CFR 888.3358 Regulation Name: Hip Joint Metal/Polymer/Metal Semi-Constrained Porous-Coated Uncemented Prosthesis Regulatory Class: Class II Product Code: LPH, LZO, HWC Dated: Mav 24, 2017 Received: May 25, 2017

Dear Kaitlin Grove:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR

{1}------------------------------------------------

Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely.

# Mark N. Melkerson -S

Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Expiration Date: January 31, 2017

See PRA Statement below.

Form Approved: OMB No. 0910-0120

# Indications for Use

510(k) Number (if known)

#### Device Name

PROCOTYL® PRIME Acetabular Cup System

Indications for Use (Describe)

The PROCOTYL® PRIME Acetabular Cup System is intended for use in total hip arthroplasty for reduction or relief of pain and/or improved hip function in skeletally mature patients with the following conditions:

1) non-inflammatory degenerative joint disease such as osteoarthritis, avascular necrosis, ankylosis, protrusio acetabuli. and painful hip dysplasia;

2) inflammatory degenerative joint disease such as rheumatoid arthritis;

3) correction of functional deformity; and,

4) revision procedures where other treatments or devices have failed.

Shells with BIOFOAM® metal foam coating are intended only for uncemented arthroplasty.

X Prescription Use (Part 21 CFR 801 Subpart D)

| Over-The-Counter Use (21 CFR 801 Subpart C)

## CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/1 description: The image contains the logo for MicroPort Orthopedics. The logo consists of a blue square with rounded corners, inside of which is a red diamond with a white star in the center. To the right of the square is the text "MicroPort" in blue, with "Orthopedics" in red below it. Underneath the entire logo is the text "MicroPort Orthopedics Inc." in black.

PROCOTYL® PRIME Acetabular Cup System Traditional 510(k) 510(k) Summary

# 510(k) Summary

In accordance with the Food and Drug Administration Rule to implement provisions of the Safe Medical Devices Act of 1990 and in conformance with 21 CFR 807.92, this information serves as a Summary of Substantial Equivalence for the use of the PROCOTYL® PRIME Acetabular Cup System.

| Submitted by:                           | MicroPort Orthopedics Inc.<br>5677 Airline Rd, Arlington TN, 38002<br>Phone: 866-872-0211<br>Fax: 855-446-2247                   |
|-----------------------------------------|----------------------------------------------------------------------------------------------------------------------------------|
| Date:                                   | June 22, 2017                                                                                                                    |
| Contact Person:                         | Kaitlin Grove<br>Regulatory Affairs Specialist II                                                                                |
| Proprietary Name:                       | PROCOTYL® PRIME Acetabular Cup System                                                                                            |
| Common Name:                            | acetabular shell, acetabular liner and bone screw                                                                                |
| Classification Name and<br>Reference:   | 21 CFR 888.3358 - Hip joint metal/polymer/metal semi-<br>constrained porous-coated uncemented prosthesis - Class<br>II           |
|                                         | 21 CFR 888.3353 - Hip joint metal/ceramic/polymer<br>semi-constrained cemented or nonporous uncemented<br>prosthesis - Class II  |
|                                         | 21 CFR 888.3040 - Smooth or threaded metallic bone<br>fixation fastener - Class II                                               |
| Subject Product Code and Panel<br>Code: | Orthopedics/87/LPH, LZO, HWC                                                                                                     |
| Predicate Device:                       | LINEAGE® Acetabular System (K002149  – Shell)<br>LINEAGE® Acetabular System (K052026 - Liner)<br>Cancellous Bone Screw (K864626) |
|                                         | DYNASTY® Acetabular System (K082924 Shells;<br>K002149, K061547, K070785, K082924 Liners)                                        |
|                                         | Zimmer Biomet Trilogy Acetabular Cup System<br>(K934765, K953490)                                                                |
|                                         | Depuy Pinnacle 100 Shell (K090998)                                                                                               |

{4}------------------------------------------------

Image /page/4/Picture/1 description: The image shows the text "Pg.2/6". The text is in a simple, sans-serif font and is positioned in the center of the image. The text likely refers to a page number within a larger document.

Image /page/4/Picture/2 description: The image contains the logo for MicroPort Orthopedics. The logo consists of a blue rounded square with a white circle inside, containing a red diamond shape. To the right of the icon is the text "MicroPort" in blue, with "Orthopedics" underneath in red. Below the logo is the text "MicroPort Orthopedics Inc." in black.

# DEVICE INFORMATION

# A. Intended Use

The PROCOTYL® PRIME Acetabular Cup System is intended for use in total hip arthroplasty for reduction or relief of pain and/or improved hip function in skeletally mature patients with the following conditions:

- 1. non-inflammatory degenerative joint disease such as osteoarthritis, avascular necrosis, ankylosis, protrusio acetabuli, and painful hip dysplasia;
- 2. inflammatory degenerative joint disease such as rheumatoid arthritis;
- 3. correction of functional deformity; and,
- 4. revision procedures where other treatments or devices have failed

Shells with BIOFOAM® metal foam coating are intended only for uncemented arthroplasty.

Similar to the explanation that was given in K111699 for MicroPort hip stems, these indications were stated as five (5) items (cleared with K971429, K002149 and K043099) and were eventually consolidated to four (4) items (as first cleared with hip stem K041114). The current indications are applied to all MicroPort polyethylene total acetabular systems and have been cleared in K052026, K061547, K070785, K072656, K082924, K122382, K130376, K140043.

# B. Device Description

The PROCOTYL® PRIME Acetabular Cup System includes acetabular shells, acetabular liners, and optional cancellous bone screws. The design features are summarized below:

- . Acetabular Shells
	- o Manufactured from titanium alloy
	- BIOFOAM® coated o
	- Available in Solid or Quad configurations o
	- Outer Diameter sizes 42mm to 68mm in 2mm increments o
- Acetabular Liners ●
	- Manufactured from A-CLASS® highly crosslinked ultra high molecular о weight polvethylene)
	- Available in Standard. Lipped Face-changing or Lateralized о configurations
	- Inner Diameter sizes 22mm to 44mm o
- Bone Screws ●
	- Manufactured from titanium alloy o
	- 6.5mm diameter о
	- 0 Available in lengths 15mm to 80mm in 5mm increments

{5}------------------------------------------------

Pg.3/6

Image /page/5/Picture/1 description: The image contains the logo for MicroPort Orthopedics. The logo consists of a blue square with rounded corners, inside of which is a white circle with a red diamond shape in the center. To the right of the square is the text "MicroPort" in blue, with "Orthopedics" in red underneath. Below the logo is the text "MicroPort Orthopedics Inc." in black.

# C. Substantial Equivalence Information

The indications for use of the PROCOTYL® PRIME Acetabular Cup System are identical to those for the predicate devices. The identical highly crosslinked UHMWPE material was characterized for the predicate liner in K052026. The design features and materials are substantially equivalent to those of the predicate devices. The fundamental scientific technology of the subject devices has not changed relative to the predicate devices. The safety and effectiveness of the subject devices are adequately supported by the substantial equivalence information, materials information, and analysis data provide within this Premarket Notification.

# D. Nonclinical Testing

The lock detail of the subject PROCOTYL® PRIME Acetabular Cup System was evaluated by testing Push-out, Lever-out and Torque-out of the Liner per ASTM 1820.

The subject System underwent deformation and frictional torque testing to evaluate the properties under pinch-loading conditions.

The subject System completed long term axial cyclic loading to evaluate fatigue resistance properties.

The subject System underwent wear testing to determine the wear rate and a wear particle analysis was performed on the wear debris.

Additionally, a range of motion study was completed for the PROCOTYL® PRIME Acetabular Cup System.

Finally, per ASTM F543, the properties of the cancellous bone screws were examined.

# E. Clinical Testing

Clinical data was not provided for the subject devices

# F. Biocompatibility

The intended patient contact and materials used in the subject implant devices are identical to those of the predicate devices. Furthermore, that the materials and processing methods of the subject device are identical to predicates K002149, K082924, K052026, and K864626. No subject implants use colorants. Therefore, biocompatibility testing was not completed on the subject devices.

The Limulus Amebocyte Lysate (LAL) test was used to determine the subject devices meets the specified pyrogen limit specification of ≤ 20 EU/device.

For all associated instruments that were not previously cleared, Master File references were provided for raw material biocompatibility and manufacturing process effects on biocompatibility were evaluated.

{6}------------------------------------------------

Image /page/6/Picture/2 description: The image contains the logo for MicroPort Orthopedics. The logo consists of a blue square with a white circle inside, and a red diamond inside the white circle. To the right of the square is the text "MicroPort" in blue, with "Orthopedics" in red below it. Below the logo is the text "MicroPort Orthopedics Inc." in black.

PROCOTYL® PRIME Acetabular Cup System Traditional 510(k) 510(k) Summary

# G. Component and Accessory Compatibility

Tables 1 and 2 show the compatibility of the subject device with previously cleared MicroPort Orthopedics products.

| Table 1: PROCOTYL® PRIME Acetabular Cup System Compatibility |  |  |
|--------------------------------------------------------------|--|--|
|--------------------------------------------------------------|--|--|

| 510(k)                                                               | Intended Combinations                                  |
|----------------------------------------------------------------------|--------------------------------------------------------|
| Subject Liners Compatibility with Modular Femoral Heads (up to 44mm) |                                                        |
| K893685                                                              | Ceramic-Polyethylene with ID 28-36mm**                 |
| K920593                                                              | Ceramic-Polyethylene, OD 28mm                          |
| K925512                                                              | Ceramic-Polyethylene, OD 28mm                          |
| K932222                                                              | Metal-Polyethylene, OD 28mm XXL                        |
| K002149                                                              | Metal-Polyethylene with ID 22-32mm                     |
| K021349*                                                             | Metal-Polyethylene with ID 38-56mm                     |
| K004043*                                                             | Metal-Polyethylene with ID 28-36mm                     |
| K051348*                                                             | Metal-Polyethylene with ID 38-56mm                     |
| K072656                                                              | Ceramic-Polyethylene with ID 38-46mm with neck sleeves |
| K130376                                                              | Ceramic-Polyethylene with ID 32-40mm                   |
| K140043                                                              | Ceramic-Polyethylene with ID 28mm                      |
| Subject Shell Compatibility with Bone Screws                         |                                                        |
| K864626                                                              | Cancellous Bone Screws                                 |
| Subject Screws Compatibility with Acetabular Shells                  |                                                        |
| K002149                                                              | LINEAGE® System Shells                                 |
| K061547                                                              | DYNASTY® PC Shells                                     |
| K070785                                                              | DYNASTY® PC Shells                                     |
| K082924                                                              | DYNASTY® BIOFOAM® Shells                               |
| K122382                                                              | DYNASTY® BIOFOAM® Shells                               |
| K142119                                                              | PROCOTYL® I/O Shells                                   |

* Metal femoral heads in K021349, K004043 and K051348 were originally cleared for use with metal-metal bearings, and later cleared for compatibility with DYNASTY® A-CLASS® polyethylene liners in K070785.

**36mm Forte Ceramic heads were originally cleared for use in PMA P030027, and later cleared for compatibility with Procotyl L / O Acetabular System in K142119.

{7}------------------------------------------------

Pg.5/6

Image /page/7/Picture/1 description: The image contains the logo for MicroPort Orthopedics. The logo consists of a blue square with rounded corners, inside of which is a white circle containing a red diamond shape. To the right of the square is the text "MicroPort" in blue, with "Orthopedics" in red underneath. Below the logo is the text "MicroPort Orthopedics Inc."

| 510(k)  | Device Name                                      |
|---------|--------------------------------------------------|
| K003016 | PRO-FEMUR R                                      |
| K012091 | PRO-FEMUR                                        |
| K021346 | STEM HIP REPLACEMENT SYSTEM                      |
| K041114 | PROFEMUR TAPERED HIP STEM                        |
| K041586 | PROFEMUR S HIP STEM                              |
| K051995 | PROFEMUR RENAISSANCE HIP STEM                    |
| K052915 | PROFEMUR XTR HIP STEM                            |
| K053588 | PROFEMUR LX HIP STEM                             |
| K060358 | PROFEMUR TL HIP STEM                             |
| K080663 | PROFEMUR LX REVISION 5/8 COATED HIP STEM         |
| K081090 | PROFEMUR LX 5/8 COATED HIP STEM                  |
| K091423 | PROFEMUR HIP SYSTEM MODULAR NECKS                |
| K100866 | PROFEMUR HIP SYSTEM MODULAR                      |
| K110399 | GLADIATOR PLASMA CLASSIC HIP STEM                |
| K111698 | PROFEMUR(R) E CEMENTLESS HIP STEM                |
| K111699 | PROFEMUR(R) Z TITANIUM PLASMA SPRAYED HIP STEM   |
| K111910 | GLADIATOR HIP STEM                               |
| K112080 | PRESERVE HIP STEM                                |
| K112150 | PROFEMUR GLADIATOR HA HIP STEM                   |
| K121221 | PROFEMUR Z REVISION HIP STEM                     |
| K123434 | PROFEMUR Z CLASSIC STEM                          |
| K123688 | PROFEMUR TL CLASSIC STEM                         |
| K130984 | PROFEMUR RENAISSANCE CLASSIC STEM                |
| K140676 | PROFEMUR TL CLASSIC LONG NECK HIP STEMS          |
| K141235 | PROFEMUR RENAISSANCE CLASSIC LONG NECK HIP STEMS |
| K150133 | PROFEMUR PRESERVE SIZE 1-3 HIP STEMS             |
| K150302 | PROFEMUR PRESERVE CLASSIC STEM                   |

Table 2: Compatible Femoral Components, Including 510(k) Information.

The PROCOTYL® PRIME Acetabular Cup System instrumentation includes reamers, trial shells, trial liners, impactors and impactor handles, extraction instruments, and screw instruments. The PROCOTYL® PRIME Acetabular Cup System may be used with alternative surgical approaches submitted in K122382; related instrumentation is included in the subject biocompatibility and surgical technique information.

{8}------------------------------------------------

Image /page/8/Picture/1 description: The image shows the logo for MicroPort Orthopedics. The logo consists of a blue square with rounded corners, inside of which is a white circle with a red diamond shape in the center. To the right of the square is the text "MicroPort" in blue, with "Orthopedics" in red below it. Below the entire logo is the text "MicroPort Orthopedics Inc." in black.

# H. Sterilization Residuals

Summary residuals results related to Ethylene Oxide sterilization were provided previously in K140043. Worst-case device was selected based on product, density, size, and worst case for Ethylene Oxide sterilization. Selected devices were LINEAGE® Acetabular Poly Liner (K002149), LINEAGE® Ceramic Head (K893685).

# I. Conclusions

The substantial equivalence of the PROCOTYL® PRIME Acetabular Cup System is adequately supported by the substantial equivalence information, materials information, and nonclinical testing data provided within this premarket notification.

---

**Source:** [https://fda.innolitics.com/device/K170444](https://fda.innolitics.com/device/K170444)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
