K170398 · S.I.N. – Sistema DE Implante Nacional S.A. · DZE · Nov 21, 2017 · Dental
Device Facts
Record ID
K170398
Device Name
S.I.N. Dental Implant System
Applicant
S.I.N. – Sistema DE Implante Nacional S.A.
Product Code
DZE · Dental
Decision Date
Nov 21, 2017
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 872.3640
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
S.I.N. Dental Implant System is intended for placement in the maxillary or mandibular arch to provide support for single-unit or multi-unit restorations. When a one-stage surgical approach is applied, the S.I.N. Dental Implant System is intended for immediate loading when good primary stability is achieved and with appropriate occlusal loading. Revolution Compact with a 6 mm length is intended for delayed loading only.
Device Story
S.I.N. Dental Implant System consists of endosseous implants and abutments for dental restoration. Implants are surgically placed in maxillary or mandibular bone to support crowns, bridges, or overdentures. System includes six product lines (Revolution Compact, Micro-Mini Revolution, Strong SW HE, TryOn, Strong SW HI) featuring acid-etched surfaces and various abutment designs (healing, provisional, cemented, UCLA, overdenture, conical, mini). Implants are cylindrical with apical cutting flutes; some include micro-threads. Components are manufactured from commercially pure titanium, titanium alloy, or cobalt-chromium alloy. Used by dental professionals in clinical settings. Output is a stable foundation for prosthetic teeth, restoring chewing function. Benefits include tooth replacement and improved oral function.
Clinical Evidence
Bench testing only. Evidence includes radiation and steam sterilization validation (ISO 11137, ISO 17665), biocompatibility (ISO 10993-1, ISO 10993-5), LAL endotoxin testing, and mechanical shelf-life/seal integrity testing (ASTM F88, F1929, F1980). No clinical data included.
Indicated for patients requiring single-unit or multi-unit dental restorations in the maxillary or mandibular arch. Supports immediate loading when primary stability is achieved; 6 mm Revolution Compact implants are indicated for delayed loading only.
Regulatory Classification
Identification
An endosseous dental implant is a prescription device made of a material such as titanium or titanium alloy that is intended to be surgically placed in the bone of the upper or lower jaw arches to provide support for prosthetic devices, such as artificial teeth, in order to restore a patient's chewing function.
Special Controls
*Classification.* (1) Class II (special controls). The device is classified as class II if it is a root-form endosseous dental implant. The root-form endosseous dental implant is characterized by four geometrically distinct types: Basket, screw, solid cylinder, and hollow cylinder. The guidance document entitled “Class II Special Controls Guidance Document: Root-Form Endosseous Dental Implants and Endosseous Dental Implant Abutments” will serve as the special control. (See § 872.1(e) for the availability of this guidance document.)(2)
*Classification.* Class II (special controls). The device is classified as class II if it is a blade-form endosseous dental implant. The special controls for this device are:(i) The design characteristics of the device must ensure that the geometry and material composition are consistent with the intended use;
(ii) Mechanical performance (fatigue) testing under simulated physiological conditions to demonstrate maximum load (endurance limit) when the device is subjected to compressive and shear loads;
(iii) Corrosion testing under simulated physiological conditions to demonstrate corrosion potential of each metal or alloy, couple potential for an assembled dissimilar metal implant system, and corrosion rate for an assembled dissimilar metal implant system;
(iv) The device must be demonstrated to be biocompatible;
(v) Sterility testing must demonstrate the sterility of the device;
(vi) Performance testing to evaluate the compatibility of the device in a magnetic resonance (MR) environment;
(vii) Labeling must include a clear description of the technological features, how the device should be used in patients, detailed surgical protocol and restoration procedures, relevant precautions and warnings based on the clinical use of the device, and qualifications and training requirements for device users including technicians and clinicians;
(viii) Patient labeling must contain a description of how the device works, how the device is placed, how the patient needs to care for the implant, possible adverse events and how to report any complications; and
(ix) Documented clinical experience must demonstrate safe and effective use and capture any adverse events observed during clinical use.
Predicate Devices
Sistema de Implante Nacional Dental Implant System (K021859)
{1}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo features the letters 'FDA' in a blue square, followed by the words 'U.S. FOOD & DRUG ADMINISTRATION' in blue text.
S.I.N. - Sistema De Implante Nacional S.A. % Linda Schulz Regulatory Affairs PaxMed International, LLC 12264 El Camino Real, Suite 400 San Diego, California 92130
November 21, 2017
Re: K170398
Trade/Device Name: S.I.N. Dental Implant System Regulation Number: 21 CFR 872.3640 Regulation Name: Endosseous Dental Implant Regulatory Class: Class II Product Code: DZE, NHA Dated: October 23, 2017 Received: October 24, 2017
## Dear Linda Schulz:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{2}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
Mary S. Runner -S
for
Tina Kiang, Ph.D. Acting Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control. and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
#### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
## Indications for Use
Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement below.
510(k) Number (if known)
K170398
Device Name
S.I.N. Dental Implant System
Indications for Use (Describe)
S.I.N. Dental Implant System is intended for placement in the maxillary or mandibular arch to provide support for single-unit or multi-unit restorations. When a one-stage surgical approach is applied, the S.I.N. Dental Implant System is intended for immediate loading when good primary stability is achieved and with appropriate occlusal loading. Revolution Compact with a 6 mm length is intended for delayed loading only.
Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
# 510(k) Summary S.I.N. - Sistema de Implante Nacional S.A. S.I.N. Dental Implant System K170398
November 21, 2017
## ADMINISTRATIVE INFORMATION
| Manufacturer Name | S.I.N. - Sistema de Implante Nacional S.A.<br>Avenida Vereador Abel Ferreira, 1100<br>São Paulo, São Paulo<br>03340-000 Brazil<br>Telephone: +55-11-21693000 ext 3236 |
|---------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Official Contact | Dênis Oliveira<br>Quality and Regulatory Manager |
| Representative/Consultant | Linda K. Schulz, BSDH, RDH<br>Floyd G. Larson, MS, MBA<br>PaxMed International, LLC<br>12264 El Camino Real, Suite 400<br>San Diego, CA 92130<br>Telephone: +1 (858) 792-1235<br>Fax: +1 (858) 792-1236<br>Email: lschulz@paxmed.com<br>flarson@paxmed.com |
## DEVICE NAME AND CLASSIFICATION
| Trade/Proprietary Name | S.I.N. Dental Implant System |
|----------------------------|------------------------------------|
| Common Name | Dental implant |
| | Dental implant abutment |
| Classification Name | Implant, endosseous, root form |
| | Endosseous dental implant abutment |
| Classification Regulations | 21 CFR 872.3640, Class II |
| Product Code | DZE, NHA |
| Classification Panel | Dental Products Panel |
| Reviewing Branch | Dental Devices Branch |
## PREDICATE DEVICE INFORMATION
Primary Predicate Sistema de Implante Nacional Dental Implant System K021859 Sistema de Implante Nacional, Ltda.
{5}------------------------------------------------
| Reference Predicates | | |
|----------------------|-------------------------------------------------|--------------------------------------------------------|
| K120414 | OsseoSpeed™ Plus | Astra Tech AB |
| K072570 | NobelActive Multi Unit Abutment | Nobel Biocare AB |
| K072363 | BIOMET 3i NanoTite™ Dental Implants | Biomet 3i, Inc. |
| K092035 | Bicon Implants with a 2.5mm Internal Connection | Bicon, L.L.C. |
| K101945 | Neodent Implant System | JJGC Industria e Comercio de Materiais<br>Dentarios SA |
#### INDICATIONS FOR USE
S.I.N. Dental Implant System is intended for placement in the maxillary or mandibular arch to provide support for single-unit or multi-unit restorations. When a one-stage surgical approach is applied, the S.I.N. Dental Implant System is intended for immediate loading when good primary stability is achieved and with appropriate occlusal loading. Revolution Compact with a 6 mm length is intended for delayed loading only.
#### DEVICE DESCRIPTION
The subject device includes six product lines (Revolution Compact, Micro-Mini Revolution, Strong SW HE, TryOn, Strong SW HI). Five of the product lines have an external hex implant/abutment connection (HE) and one has an internal hex implant/abutment connection (HI). Each product line includes implants with an acid-etched surface and abutments in multiple designs (healing, provisional, cemented, UCLA, overdenture, conical and mini).
Revolution implants (Revolution Compact, Micro-Mini Revolution) are cylindrical implants with apical cutting flutes and are placed with the Revolution Assembler mount. Strong SW HI are cvlindrical implants with apical cutting flutes and micro-threads on the collar. Strong SW HE has and external connection and Strong SW HI has an internal connection. TryOn implants with apical cutting flutes and a machined collar. Subject Device components are made of commercially pure titanium, titanium alloy or cobalt-chromium alloy. All subject device abutments have 0° angulation.
| Summary of Implant Sizes | | | |
|--------------------------|--------------------------|------------------------------|--------------------------|
| | Body<br>Diameter<br>(mm) | Platform<br>Diameter<br>(mm) | Length<br>(mm) |
| Revolution | 4.0 | 4.1 | 7 |
| Revolution Compact | 4.0, 5.0 | 5.0 | 6 |
| Strong SW HE | 3.5 | 3.65 | 7, 8.5, 10, 11.5, 13, 15 |
| | 3.75, 4.0 | 4.1 | 7, 8.5, 10, 11.5, 13, 15 |
| | 4.5 | 4.5 | 8.5, 10, 11.5, 13, 15 |
| | 5.0 | 5.0 | 7, 8.5, 10, 11.5, 13, 15 |
| TryOn | 3.25 | 4.1 | 8.5, 10, 11.5, 13, 15 |
| | 3.75, 4.0 | 4.1 | 7, 8.5, 10, 11.5, 13, 15 |
| | 5.0 | 5.0 | 7, 8.5, 10, 11.5, 13, 15 |
| Strong SW HI | 3.8 | 3.8 | 8.5, 10, 11.5, 13, 15 |
| | 4.5 | 4.5 | 8.5, 10, 11.5, 13, 15 |
| | 5.0 | 5.0 | 8.5, 10, 11.5, 13, 15 |
y of Implant Sizes
{6}------------------------------------------------
## PERFORMANCE DATA
Non-clinical testing data submitted, referenced, or relied upon to demonstrate substantial equivalence included: radiation sterilization validation to an SAL of 10° according to ISO 11137-1, ISO 11137-2; Steam sterilization validation to an SAL of 10° according to ISO 17665-1 and ISO 17665-2; biocompatibility evaluation according to ISO 10993-1 by reference to K051859 and biocompatibility testing according to ISO 10993-5 (cytotoxicity), demonstrating acceptable biocompatibility ; Limulus amebocyte lysate (LAL) endotoxin testing in accordance with FDA Guidance documents Submission and Review of Sterility Information in Premarket Notification (510(k)) Submissions for Devices Labeled as Sterile; shelf life testing seal leaks according to ASTM F88/F88M, seal integrity testing according to ASTM F1929, accelerated age testing according to ASTM F1980, and surface area analysis showing substantial equivalence to a predicate.
No clinical data were included in this submission.
## EQUIVALENCE TO MARKETED DEVICE
The subject device is substantially equivalent in indications and design principles to the predicate devices shown above. Below are summary tables comparing the Indications for Use and the technological characteristics of the subject device and the predicate devices.
| Subject Device | |
|------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| K170398<br>S.I.N Implant System<br>SIN - Sistema de Implante Nacional<br>S.A. | S.I.N. Dental Implant System is intended for placement in the maxillary or mandibular arch to provide<br>support for single-unit or multi-unit restorations. When a one-stage surgical approach is applied, the S.I.N.<br>Dental Implant System is intended for immediate loading when good primary stability is achieved and with<br>appropriate occlusal loading. Revolution Compact with a 6 mm length is intended for delayed loading only. |
| Primary Predicate Device | |
| K051859<br>Sistema de Implante Nacional Dental<br>Implant System<br>SIN - Sistema de Implante Nacional,<br>Ltda. | The Sistema de Implante Nacional Dental Implant System is intended to be surgically placed in the bone of<br>the maxillary and/or mandibular arch to provide support for crowns, bridges or overdentures. Implants may be<br>placed immediately after tooth extraction or following bone healing. Restorations supported by two or more<br>Sistema de Implante Nacional implants may be loaded immediately after implant placement if primary<br>implant stability has been achieved. |
| Reference Predicate Devices | |
| K120414<br>OsseoSpeedTM Plus<br>Astra Tech AB | Implants: The Astra Tech Dental Implants are intended for both one- and two-stage surgical procedures in the<br>following situations and with the following clinical protocols:<br>replacing single and multiple missing teeth in the mandible and maxilla, immediate placement in extraction sites and in situations with a partially or completely healed alveolar ridge, especially indicated for use in soft bone applications where implants with other implant surface treatments may be less effective, immediate loading in all indications, except in single tooth situations on implants shorter than 8 mm, or in soft bone (type IV) where implant stability may be difficult to obtain and immediate loading may not be appropriate. The intended use for OsseoSpeedTM Plus 3.0S is limited to replacement of maxillary lateral incisors and mandibular incisors.<br>Abutments: Astra Tech Implant System Plus abutments are intended to be used in conjunction with Astra<br>Tech Implant System Plus in fully edentulous or partially edentulous maxillary and/or mandibular arches to<br>provide support for crowns, bridges or overdentures.<br>Atlantis Abutments: The AtlantisTM Abutment is intended for use with an endosseous implant to support a<br>prosthetic device in a partially or completely edentulous; patient. It is intended for use to support single and<br>multiple tooth prostheses, in the mandible or maxilla. The prosthesis can be cemented, screw retained or<br>friction fit to the abutment. The abutment screw is intended to secure the abutment to the endosseous implant.<br>The AtlantisTM Crown Abutment in Zirconia is intended for use with an endosseous; implant to function as a |
| | substructure that also serves as the final restoration, in partially or completely edentulous; patients. The<br>prosthesis is screw retained. The abutment screw is intended to secure the crown abutment to the endosseous<br>implant. |
| K072570<br>NobelActive Multi Unit Abutment<br>Nobel Biocare AB | NobelActive Multi Unit Abutment is a pre-manufactured prosthetic component directly<br>connected to the endosseous dental implant and intended for use as an aid in prosthetic<br>rehabilitation. |
Comparison of Indications for Use Statements
{7}------------------------------------------------
| K072363<br>BIOMET 3i NanoTite™ Dental<br>Implants<br>BIOMET 3i, Incorporated | BIOMET 3i dental implants are intended for surgical placement in the upper or lower jaw to provide a means<br>for prosthetic attachment in single tooth restorations and in partially or fully edentulous spans with multiple<br>single teeth utilizing delayed or immediate loading, or as a terminal or intermediary abutment for fixed or<br>removable bridgework, and to retain overdentures.<br>BIOMET 3i NanoTite dental implants are intended for immediate function on single tooth and/or multiple<br>tooth applications when good primary stability is achieved, with appropriate occlusal loading, in order to<br>restore chewing function. |
|------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| K092035 | The Bicon implant is designed for use in edentulous sites in the mandible or maxilla for support of a complete |
| Bicon Implants with a 2.5mm Internal<br>Connection | denture prosthesis, a final or intermediate abutment for fixed bridgework or for partial dentures, or as a single<br>tooth replacement. |
| Bicon, L.L.C. | |
| K101945 | The Neodent Implant System is intended to be surgically placed in the bone of the upper or lower jaw to<br>provide support for prosthetic devices such as artificial teeth, to restore chewing function. It may be used with |
| Neodent Implant System | single-stage or two-stage procedures, for single or multiple unit restorations, and may be loaded immediately<br>when good primary stability is achieved and with appropriate occlusal loading. Multiple tooth applications |
| JJGC Industria e Comercio de Materiais<br>Dentarios SA | may be rigidly splinted. |
{8}------------------------------------------------
# Comparison of Technological Characteristics
| | Subject Device | Primary<br>Predicate Device | Reference Predicate Devices | | | | |
|-------------------------------|--------------------------------------------------|----------------------------------------------------------|---------------------------------------------|------------------------------------|-------------------------------------------|-------------------------------------------------------|-----------------------------------------------------------|
| | | K051859 | K120414 | K072570 | K072363 | K092035 | K101945 |
| | S.I.N. Implant<br>System | Sistema de Implante<br>Nacional<br>Dental Implant System | OsseoSpeed™ Plus | NobelActive Multi<br>Unit Abutment | BIOMET 3i<br>NanoTite™<br>Dental Implants | Bicon Implants with<br>a 2.5mm Internal<br>Connection | Neodent Implant System |
| | S.I.N. - Sistema<br>de Implante<br>Nacional S.A. | S.I.N. - Sistema de<br>Implante Nacional,<br>Ltda. | Astra Tech AB | Nobel Biocare AB | BIOMET 3i, Inc. | Bicon, L.L.C. | JJGC Industria e<br>Comercio de Materiais<br>Dentarios SA |
| DESIGN | | | | | | | |
| Prosthesis Attachment | Screw-retained<br>Cement-retained | Screw-retained<br>Cement-retained | Screw-retained<br>Cement-retained | Screw-retained | Screw-retained<br>Cement-retained | Screw-retained<br>Cement-retained | Screw-retained<br>Cement-retained |
| Restoration | Single-unit<br>Multi-unit | Single-unit<br>Multi-unit | Single-unit<br>Multi-unit | Multi-unit | Single-unit<br>Multi-unit | Single-unit<br>Multi-unit | Single-unit<br>Multi-unit |
| Implant/Abutment<br>Interface | External<br>Internal | External<br>Internal | Internal | Internal | External<br>Internal | Internal | Internal |
| Implant Diameter, mm | 3.25 - 5.0 | 3.25 - 6.3 | 3.0 - 5.4 | NA | 3.25 - 6.0 | 4.0 - 4.5 | 3.3 - 5.0 |
| Implant Length, mm | 6.0 - 18 | 7.0 - 18 | 6.0 - 17 | NA | 8.5 - 18 | 5 - 11 | 7.0 - 19 |
| Abutment Diameter, mm | 3.4 - 5.0 | 3.4 - 5.5 | 3.0 - 5.4 | NP, RP | NA | 4.0 - 6.5 | 3.3 - 5.0 |
| Abutment Angle | 0° | 0° - 19° | 0° - 30° | 0° - 30° | NA | 0° - 25° | 0° - 30° |
| MATERIAL | | | | | | | |
| Implant | CPTi Gr 4 | CPTi Gr 4 | CPTi Gr 4 | NA | CPTi Gr 4 | TI6AL4V | CPTi Gr 4 |
| Surface | Acid-Etched | Acid-Etched | OsseoSpeed™ | NA | NanoTite™ | HA, TPS, Integra Ti,<br>NanoTite™ | Grit blasted and<br>acid etched |
| Abutment | Ti-6Al-4V<br>CoCr | CPTi Gr 2<br>Ti-6Al-4V<br>CoCr<br>Gold alloy | Ti-6Al-4V<br>Zirconia<br>Gold alloy<br>PEEK | Ti-6Al-4V | NA | Ti-6Al-4V | Ti-6Al-4V<br>Zirconia<br>Gold alloy<br>POM |
| Abutment Screw | Ti-6Al-4V | Ti-6Al-4V | Ti-6Al-4V | Ti-6Al-4V | NA | Ti-6Al-4V | Ti-6Al-4V |
{9}------------------------------------------------
510(k) Summary Page 6 of 6
The subject device and predicate devices are substantially equivalent in intended use. Small differences in language in the Indications for Use Statement between the subject device and the primary predicate K051859 do not change the intended use of implant placement and function. Both are implant systems placed in the maxilla or mandible to support single or multi-unit restore chewing function, with possible immediate loading. A comparison of equivalent phrasing for the subject device and primary predicate device is shown in the table below.
| Subject Device | Primary Predicate Device – K051859 |
|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| S.I.N. Dental Implant System is intended for placement<br>in the maxillary or mandibular arch. | The Sistema de Implante Nacional Dental Implant<br>System is intended to be surgically placed in the bone of<br>the maxillary and/or mandibular arch |
| to provide support for single-unit or multi-unit<br>restorations. | to provide support for crowns, bridges or overdentures |
| | Implants may be placed immediately after tooth<br>extraction or following bone healing. |
| When a one-stage surgical approach is applied, the S.I.N.<br>Dental Implant System is intended for immediate loading<br>when good primary stability is achieved and with<br>appropriate occlusal loading. Revolution Compact with a<br>6 mm length is intended for delayed loading only. | Restorations supported by two or more Sistema de<br>Implante Nacional implants may be loaded immediately<br>after implant placement if primary implant stability has<br>been achieved. |
Subject device implants are substantially equivalent to K051859 implants in body, connection design, material and acid etched surface. The short is substantially equivalent to K092035 in size, surface area and function. Subject device abutments are substantially equivalent to K051859 abutments in material, design and function, with the exception of the design of the ball abutment. Subject device abutments are substantially equivalent to K120414 abutments in material, design and function. Subject device multi-unit abutments are substantially equivalent to K072570 abutments in material, design and function.
## CONCLUSION
The subject device and the predicate devices have the same intended use, have similar technological characteristics, and are made of the same or similar materials. The subject device and predicate devices encompass the same range of physical dimensions, including diameter, length and angle of the abutments. The subject and predicate devices are packaged in similar materials and are to be sterilized using similar methods.
The data included in this submission demonstrate substantial equivalence to the predicate devices listed above.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.