← Product Code [IYN](/productcode/IYN) · K170374

# SonoBook Series Diagnostic Ultrasound System (K170374)

_Chison Medical Imaging Co., Ltd. · IYN · Jun 6, 2017 · Radiology · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K170374

## Device Facts

- **Applicant:** Chison Medical Imaging Co., Ltd.
- **Product Code:** [IYN](/productcode/IYN.md)
- **Decision Date:** Jun 6, 2017
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 892.1550
- **Device Class:** Class 2
- **Review Panel:** Radiology
- **Attributes:** Pediatric

## Indications for Use

Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows: Fetal, Abdominal, Pediatric, Small Organ (breast, thyroid, testes), Neonatal Cephalic, Adult Cephalic, Cardiac (adult, pediatric), Musculoskeletal (Conventional, Superficial), Peripheral Vascular, Transesophageal, Transrectal, Transvaginal, Urology.

## Device Story

SonoBook Series is a mobile, software-controlled diagnostic ultrasound system. It acquires ultrasound echo data via various probes (linear, convex, phased, volume) and processes signals to display B-Mode, M-Mode, PW/CW Doppler, Color/Power Doppler, TDI, 3D/4D, and elastography. Used in clinical settings by qualified physicians, the system features a mobile console with a keyboard, power supply, and 15-inch color LCD monitor. The device transforms raw echo data into visual images and flow analysis, which the physician interprets to evaluate anatomy and pathology. This aids in clinical decision-making for various diagnostic applications, providing non-invasive imaging benefits to patients.

## Clinical Evidence

No clinical testing was required. Substantial equivalence is supported by bench testing, including electrical safety (IEC 60601-1), electromagnetic compatibility (IEC 60601-1-2), acoustic output (NEMA UD2/UD3), and biocompatibility (ISO 10993-1).

## Technological Characteristics

Mobile console with 15-inch color LCD monitor. Supports linear, convex, phased, and volume transducers. Connectivity includes image file archiving. Software-controlled, Windows-based embedded system. Sterilization/biocompatibility per ISO 10993. Acoustic output complies with NEMA UD2/UD3. Electrical safety per IEC 60601-1 and 60601-2-37.

## Regulatory Identification

An ultrasonic pulsed doppler imaging system is a device that combines the features of continuous wave doppler-effect technology with pulsed-echo effect technology and is intended to determine stationary body tissue characteristics, such as depth or location of tissue interfaces or dynamic tissue characteristics such as velocity of blood or tissue motion. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.

## Predicate Devices

- Q Series Diagnostic Ultrasound System ([K150861](/device/K150861.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the Department of Health & Human Services - USA. The logo is a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right. The profiles are stacked on top of each other, with the top profile being the smallest and the bottom profile being the largest.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

June 6, 2017

Chison Medical Imaging Co., Ltd. % Bob Leiker Leiker Regulatory & Quality 4157 North Del Rey Ave CLOVIS, CA 93619

Re: K170374

Trade/Device Name: SonoBook Series Diagnostic Ultrasound System Regulation Number: 21 CFR 892.1550 Regulation Name: Ultrasonic pulsed doppler imaging system Regulatory Class: II Product Code: IYN, IYO, ITX Dated: April 27, 2017 Received: May 3, 2017

Dear Mr. Bob Leiker:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{1}------------------------------------------------

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours.

Michael D'Hara
For

Robert Ochs. Ph.D. Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration

# Indications for Use

510(k) Number (if known)

K170374

Device Name

SonoBook Series Diagnostic Ultrasound System

Indications for Use (Describe)

The device is a general-purpose ultrasonic instrument intended for use by a qualified physician for evaluation of Fetal , Abdominal, Pediatric, Small Organ (breast, thyroid,testes ), Neonatal Cephalic, Cardiac (adult , pediatric),Musculo-skeletal (Conventional , Superficial) ,Peripheral Vascular,Transesophageal,Transvaginal, Urology.

| Type of Use (Select one or both, as applicable)                                                                                                  |                                                                                                                                                 |
|--------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------|
| <span> <b> <span style="font-size: 16px;">☑</span>                     Prescription Use (Part 21 CFR 801 Subpart D)                 </b> </span> | <span> <b> <span style="font-size: 16px;">☐</span>                     Over-The-Counter Use (21 CFR 801 Subpart C)                 </b> </span> |

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

***DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.***

The burden time for this collection of information is estimated to average 79 hours per response, including the
time to review instructions, search existing data sources, gather and maintain the data needed and complete
and review the collection of information. Send comments regarding this burden estimate or any other aspect
of this information collection, including suggestions for reducing this burden, to:

 Department of Health and Human Services
   
Food and Drug Administration
   
Office of Chief Information Officer
   
Paperwork Reduction Act (PRA) Staff
   
[PRAStaff@fda.hhs.gov](mailto:PRAStaff@fda.hhs.gov)

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

# Diagnostic Ultrasound Indications For Use

## System: SonoBook Series Diagnostic Ultrasound System

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

| Clinical Application      |                                    | Mode of Operation |   |               |               |                  |                  |                   |         |
|---------------------------|------------------------------------|-------------------|---|---------------|---------------|------------------|------------------|-------------------|---------|
| General<br>(Track 1 Only) | Specific<br>(Tracks 1 & 3)         | B                 | M | PW<br>Doppler | CW<br>Doppler | Color<br>Doppler | Power<br>Doppler | Combined<br>Modes | Other   |
| Ophthalmic                | Ophthalmic                         |                   |   |               |               |                  |                  |                   |         |
| Fetal Imaging &<br>Other  | Fetal                              | N                 | N | N             |               | N                | N                | N                 | 3,5,6   |
|                           | Abdominal                          | N                 | N | N             |               | N                | N                | N                 | 3,4,5,6 |
|                           | Intra-operative (Specify)          |                   |   |               |               |                  |                  |                   |         |
|                           | Intra-operative (Neuro)            |                   |   |               |               |                  |                  |                   |         |
|                           | Laparoscopic                       |                   |   |               |               |                  |                  |                   |         |
|                           | Pediatric                          | N                 | N | N             |               | N                | N                | N                 | 4,5,6   |
|                           | Small Organ1 (Specify)             | N                 | N | N             |               | N                | N                | N                 | 4,5,6   |
|                           | Neonatal Cephalic                  | N                 | N | N             | N             | N                | N                | N                 | 6,7     |
|                           | Adult Cephalic                     | N                 | N | N             | N             | N                | N                | N                 | 6,7     |
|                           | Trans-rectal                       | N                 | N | N             |               | N                | N                | N                 | 4,5,6   |
|                           | Trans-vaginal                      | N                 | N | N             |               | N                | N                | N                 | 3,4,5,6 |
|                           | Trans-urethral                     |                   |   |               |               |                  |                  |                   |         |
|                           | Trans-esoph. (non-Card.)           |                   |   |               |               |                  |                  |                   |         |
|                           | Musculo-skeletal<br>(Conventional) | N                 | N | N             |               | N                | N                | N                 | 4,5,6   |
|                           | Musculo-skeletal<br>(Superficial)  | N                 | N | N             |               | N                | N                | N                 | 4,5,6   |
|                           | Other (Urology)                    | N                 | N | N             |               | N                | N                | N                 | 4,5,6   |
| Cardiac                   | Cardiac Adult                      | N                 | N | N             | N             | N                | N                | N                 | 5,6,7   |
|                           | Cardiac Pediatric                  | N                 | N | N             | N             | N                | N                | N                 | 5,6,7   |
|                           | Transesophageal                    | N                 | N | N             | N             | N                | N                | N                 | 6,7     |
| Peripheral Vessel         | Peripheral vessel                  | N                 | N | N             |               | N                | N                | N                 | 5,6     |

N = new indication; P = previously cleared by FDA;

E = added under this appendix

Note : 1. Combined modes are B/M, B/CFM , B/PD, B/PW or CW, B/Color M, B/BC,B/CFM/PW or CW,B/PD or DPD/PW or CW

2. Small Organ: thyroid, testes, breast

- 3. 3D/4D
- 4. Elastography
- 5. Includes guidance of biopsy (2D)
- 6. Fusion Harmonic Imaging
- 7. TDI

 X Prescription Use AND/OR Over-The-Counter Use _ (Part 21 CFR 801 Subpart D) (21 CFR 807 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of In Vitro Diagnostics and Radiological Health (OIR)

{4}------------------------------------------------

#### System: SonoBook Series Diagnostic Ultrasound System Transducer: C3-V

| Clinical Application      | Mode of Operation                  |   |   |               |               |                  |                  |                   |       |
|---------------------------|------------------------------------|---|---|---------------|---------------|------------------|------------------|-------------------|-------|
| General<br>(Track 1 Only) | Specific<br>(Tracks 1 & 3)         | B | M | PW<br>Doppler | CW<br>Doppler | Color<br>Doppler | Power<br>Doppler | Combined<br>Modes | Other |
| Ophthalmic                | Ophthalmic                         |   |   |               |               |                  |                  |                   |       |
| Fetal Imaging &<br>Other  | Fetal                              | N | N | N             | N             | N                | N                | N                 | 5,6   |
| Other                     | Abdominal                          | N | N | N             | N             | N                | N                | N                 | 4,5,6 |
|                           | Intra-operative (Specify)          |   |   |               |               |                  |                  |                   |       |
|                           | Intra-operative (Neuro)            |   |   |               |               |                  |                  |                   |       |
|                           | Laparoscopic                       |   |   |               |               |                  |                  |                   |       |
|                           | Pediatric                          |   |   |               |               |                  |                  |                   |       |
|                           | Small Organ[1] (Specify)           |   |   |               |               |                  |                  |                   |       |
|                           | Neonatal Cephalic                  |   |   |               |               |                  |                  |                   |       |
|                           | Adult Cephalic                     |   |   |               |               |                  |                  |                   |       |
|                           | Trans-rectal                       |   |   |               |               |                  |                  |                   |       |
|                           | Trans-vaginal                      |   |   |               |               |                  |                  |                   |       |
|                           | Trans-urethral                     |   |   |               |               |                  |                  |                   |       |
|                           | Trans-esoph. (non-Card.)           |   |   |               |               |                  |                  |                   |       |
|                           | Musculo-skeletal<br>(Conventional) |   |   |               |               |                  |                  |                   |       |
|                           | Musculo-skeletal<br>(Superficial)  |   |   |               |               |                  |                  |                   |       |
|                           | Other (Urology)                    | N | N | N             | N             | N                | N                | N                 | 4,5,6 |
| Cardiac                   | Cardiac Adult                      |   |   |               |               |                  |                  |                   |       |
|                           | Cardiac Pediatric                  |   |   |               |               |                  |                  |                   |       |
|                           | Transesophageal                    |   |   |               |               |                  |                  |                   |       |
| Peripheral Vessel         | Peripheral vessel                  |   |   |               |               |                  |                  |                   |       |

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

N = new indication: P = previously cleared by FDA; E = added under this appendix

Note : 1. Combined modes are B/M, B/CFM , B/PD, B/PW or CW, B/Color M, B/BC,B/CFM/PW or CW,B/PD or DPD/PW or CW

2. Small Organ: thyroid, testes, breast

3.3D/4D

4. Elastography

5. Includes guidance of biopsy (2D)

6. Fusion Harmonic Imaging

7. TDI

Prescription Use × AND/OR Over-The-Counter Use _ (Part 21 CFR 801 Subpart D) (21 CFR 807 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of In Vitro Diagnostics and Radiological Health (OIR)

{5}------------------------------------------------

### SonoBook Series Diagnostic Ultrasound System System: Transducer: C3M-V

| Clinical Application      | Mode of Operation          |   |   |               |               |                  |                  |                   |       |
|---------------------------|----------------------------|---|---|---------------|---------------|------------------|------------------|-------------------|-------|
| General<br>(Track 1 Only) | Specific<br>(Tracks 1 & 3) | B | M | PW<br>Doppler | CW<br>Doppler | Color<br>Doppler | Power<br>Doppler | Combined<br>Modes | Other |
| Ophthalmic                | Ophthalmic                 |   |   |               |               |                  |                  |                   |       |
| Fetal Imaging &<br>Other  | Fetal                      | N | N | N             | N             | N                | N                | N                 | 5,6   |
|                           | Abdominal                  | N | N | N             | N             | N                | N                | N                 | 4,5,6 |
|                           | Intra-operative (Specify)  |   |   |               |               |                  |                  |                   |       |
|                           | Intra-operative (Neuro)    |   |   |               |               |                  |                  |                   |       |
|                           | Laparoscopic               |   |   |               |               |                  |                  |                   |       |
|                           | Pediatric                  |   |   |               |               |                  |                  |                   |       |
|                           | Small Organ[1] (Specify)   |   |   |               |               |                  |                  |                   |       |
|                           | Neonatal Cephalic          |   |   |               |               |                  |                  |                   |       |
|                           | Adult Cephalic             |   |   |               |               |                  |                  |                   |       |
|                           | Trans-rectal               |   |   |               |               |                  |                  |                   |       |
|                           | Trans-vaginal              |   |   |               |               |                  |                  |                   |       |
|                           | Trans-urethral             |   |   |               |               |                  |                  |                   |       |
|                           | Trans-esoph. (non-Card.)   |   |   |               |               |                  |                  |                   |       |
|                           | Musculo-skeletal           |   |   |               |               |                  |                  |                   |       |
|                           | (Conventional)             |   |   |               |               |                  |                  |                   |       |
|                           | Musculo-skeletal           |   |   |               |               |                  |                  |                   |       |
|                           | (Superficial)              |   |   |               |               |                  |                  |                   |       |
|                           | Other (Urology)            | N | N | N             | N             | N                | N                | N                 | 4,5,6 |
| Cardiac                   | Cardiac Adult              |   |   |               |               |                  |                  |                   |       |
|                           | Cardiac Pediatric          |   |   |               |               |                  |                  |                   |       |
|                           | Transesophageal            |   |   |               |               |                  |                  |                   |       |
| Peripheral Vessel         | Peripheral vessel          |   |   |               |               |                  |                  |                   |       |

### Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

P = previously cleared by FDA; N = new indication;

E = added under this appendix

Note : 1. Combined modes are B/M, B/CFM , B/PD, B/PW or CW, B/Color M, B/BC,B/CFM/PW or CW,B/PD or DPD/PW or CW

2. Small Organ: thyroid, testes, breast

- 3. 3D/4D
- 4. Elastography
- 5. Includes guidance of biopsy (2D)
- 6. Fusion Harmonic Imaging
- 7. TDI

Prescription Use × AND/OR Over-The-Counter Use _ (Part 21 CFR 801 Subpart D) (21 CFR 807 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of In Vitro Diagnostics and Radiological Health (OIR)

{6}------------------------------------------------

### SonoBook Series Diagnostic Ultrasound System System: Transducer: C3T-V

| Clinical Application      |                                    | Mode of Operation |   |               |               |                  |                  |                   |       |
|---------------------------|------------------------------------|-------------------|---|---------------|---------------|------------------|------------------|-------------------|-------|
| General<br>(Track 1 Only) | Specific<br>(Tracks 1 & 3)         | B                 | M | PW<br>Doppler | CW<br>Doppler | Color<br>Doppler | Power<br>Doppler | Combined<br>Modes | Other |
| Ophthalmic                | Ophthalmic                         |                   |   |               |               |                  |                  |                   |       |
| Fetal Imaging &<br>Other  | Fetal                              | N                 | N | N             |               | N                | N                | N                 | 5,6   |
| Other                     | Abdominal                          | N                 | N | N             |               | N                | N                | N                 | 4,5,6 |
|                           | Intra-operative (Specify)          |                   |   |               |               |                  |                  |                   |       |
|                           | Intra-operative (Neuro)            |                   |   |               |               |                  |                  |                   |       |
|                           | Laparoscopic                       |                   |   |               |               |                  |                  |                   |       |
|                           | Pediatric                          |                   |   |               |               |                  |                  |                   |       |
|                           | Small Organ[1] (Specify)           |                   |   |               |               |                  |                  |                   |       |
|                           | Neonatal Cephalic                  |                   |   |               |               |                  |                  |                   |       |
|                           | Adult Cephalic                     |                   |   |               |               |                  |                  |                   |       |
|                           | Trans-rectal                       |                   |   |               |               |                  |                  |                   |       |
|                           | Trans-vaginal                      |                   |   |               |               |                  |                  |                   |       |
|                           | Trans-urethral                     |                   |   |               |               |                  |                  |                   |       |
|                           | Trans-esoph. (non-Card.)           |                   |   |               |               |                  |                  |                   |       |
|                           | Musculo-skeletal<br>(Conventional) |                   |   |               |               |                  |                  |                   |       |
|                           | Musculo-skeletal<br>(Superficial)  |                   |   |               |               |                  |                  |                   |       |
|                           | Other (Urology)                    | N                 | N | N             |               | N                | N                | N                 | 4,5,6 |
|                           | Cardiac                            | Cardiac Adult     |   |               |               |                  |                  |                   |       |
|                           |                                    | Cardiac Pediatric |   |               |               |                  |                  |                   |       |
| Transesophageal           |                                    |                   |   |               |               |                  |                  |                   |       |
| Peripheral Vessel         | Peripheral vessel                  |                   |   |               |               |                  |                  |                   |       |

### Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

P = previously cleared by FDA; N = new indication;

E = added under this appendix

Note : 1. Combined modes are B/M, B/CFM , B/PD, B/PW or CW, B/Color M, B/BC,B/CFM/PW or CW,B/PD or DPD/PW or CW

2. Small Organ: thyroid, testes, breast

- 3. 3D/4D
4. Elastography

- 5. Includes guidance of biopsy (2D)
- 6. Fusion Harmonic Imaging

7. TDI

Prescription Use × AND/OR Over-The-Counter Use _ (Part 21 CFR 801 Subpart D) (21 CFR 807 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

### Concurrence of CDRH, Office of In Vitro Diagnostics and Radiological Health (OIR)

{7}------------------------------------------------

#### System: SonoBook Series Diagnostic Ultrasound System Transducer: L7-V

| Clinical Application      | Mode of Operation                  |   |   |               |               |                  |                  |                   |       |
|---------------------------|------------------------------------|---|---|---------------|---------------|------------------|------------------|-------------------|-------|
| General<br>(Track 1 Only) | Specific<br>(Tracks 1 & 3)         | B | M | PW<br>Doppler | CW<br>Doppler | Color<br>Doppler | Power<br>Doppler | Combined<br>Modes | Other |
| Ophthalmic                | Ophthalmic                         |   |   |               |               |                  |                  |                   |       |
| Fetal Imaging &<br>Other  | Fetal                              |   |   |               |               |                  |                  |                   |       |
|                           | Abdominal                          |   |   |               |               |                  |                  |                   |       |
|                           | Intra-operative (Specify)          |   |   |               |               |                  |                  |                   |       |
|                           | Intra-operative (Neuro)            |   |   |               |               |                  |                  |                   |       |
|                           | Laparoscopic                       |   |   |               |               |                  |                  |                   |       |
|                           | Pediatric                          | N | N | N             |               | N                | N                | N                 | 4,5,6 |
|                           | Small Organ[1] (Specify)           | N | N | N             |               | N                | N                | N                 | 4,5,6 |
|                           | Neonatal Cephalic                  |   |   |               |               |                  |                  |                   |       |
|                           | Adult Cephalic                     |   |   |               |               |                  |                  |                   |       |
|                           | Trans-rectal                       |   |   |               |               |                  |                  |                   |       |
|                           | Trans-vaginal                      |   |   |               |               |                  |                  |                   |       |
|                           | Trans-urethral                     |   |   |               |               |                  |                  |                   |       |
|                           | Trans-esoph. (non-Card.)           |   |   |               |               |                  |                  |                   |       |
|                           | Musculo-skeletal<br>(Conventional) | N | N | N             |               | N                | N                | N                 | 4,5,6 |
|                           | Musculo-skeletal<br>(Superficial)  | N | N | N             |               | N                | N                | N                 | 4,5,6 |
|                           | Other (Urology)                    |   |   |               |               |                  |                  |                   |       |
| Cardiac                   | Cardiac Adult                      |   |   |               |               |                  |                  |                   |       |
|                           | Cardiac Pediatric                  |   |   |               |               |                  |                  |                   |       |
|                           | Transesophageal                    |   |   |               |               |                  |                  |                   |       |
| Peripheral Vessel         | Peripheral vessel                  | N | N | N             |               | N                | N                | N                 | 5,6   |

### Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

N = new indication; P = previously cleared by FDA; E = added under this appendix

Note : 1. Combined modes are B/M, B/CFM , B/PD, B/PW or CW, B/Color M, B/BC,B/CFM/PW or CW,B/PD or DPD/PW or CW

2. Small Organ: thyroid, testes, breast

- 3. 3D/4D
- 4. Elastography
- 5. Includes guidance of biopsy (2D)
- 6. Fusion Harmonic Imaging
- 7. TDI

Prescription Use AND/OR Over-The-Counter Use (Part 21 CFR 801 Subpart D) (21 CFR 807 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of In Vitro Diagnostics and Radiological Health (OIR)

{8}------------------------------------------------

### SonoBook Series Diagnostic Ultrasound System System: Transducer: L7SVA-V

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

| Clinical Application      |                                    | Mode of Operation |   |               |               |                  |                  |                   |       |  |
|---------------------------|------------------------------------|-------------------|---|---------------|---------------|------------------|------------------|-------------------|-------|--|
| General<br>(Track 1 Only) | Specific<br>(Tracks 1 & 3)         | B                 | M | PW<br>Doppler | CW<br>Doppler | Color<br>Doppler | Power<br>Doppler | Combined<br>Modes | Other |  |
| Ophthalmic                | Ophthalmic                         |                   |   |               |               |                  |                  |                   |       |  |
| Fetal Imaging &<br>Other  | Fetal                              |                   |   |               |               |                  |                  |                   |       |  |
|                           | Abdominal                          |                   |   |               |               |                  |                  |                   |       |  |
|                           | Intra-operative (Specify)          |                   |   |               |               |                  |                  |                   |       |  |
|                           | Intra-operative (Neuro)            |                   |   |               |               |                  |                  |                   |       |  |
|                           | Laparoscopic                       |                   |   |               |               |                  |                  |                   |       |  |
|                           | Pediatric                          | N                 | N | N             |               | N                | N                | N                 | 4,5,6 |  |
|                           | Small Organ[1] (Specify)           | N                 | N | N             |               | N                | N                | N                 | 4,5,6 |  |
|                           | Neonatal Cephalic                  |                   |   |               |               |                  |                  |                   |       |  |
|                           | Adult Cephalic                     |                   |   |               |               |                  |                  |                   |       |  |
|                           | Trans-rectal                       |                   |   |               |               |                  |                  |                   |       |  |
|                           | Trans-vaginal                      |                   |   |               |               |                  |                  |                   |       |  |
|                           | Trans-urethral                     |                   |   |               |               |                  |                  |                   |       |  |
|                           | Trans-esoph. (non-Card.)           |                   |   |               |               |                  |                  |                   |       |  |
|                           | Musculo-skeletal<br>(Conventional) | N                 | N | N             |               | N                | N                | N                 | 4,5,6 |  |
|                           | Musculo-skeletal<br>(Superficial)  | N                 | N | N             |               | N                | N                | N                 | 4,5,6 |  |
|                           | Other (Urology)                    |                   |   |               |               |                  |                  |                   |       |  |
| Cardiac                   | Cardiac Adult                      |                   |   |               |               |                  |                  |                   |       |  |
|                           | Cardiac Pediatric                  |                   |   |               |               |                  |                  |                   |       |  |
|                           | Transesophageal                    |                   |   |               |               |                  |                  |                   |       |  |
| Peripheral Vessel         | Peripheral vessel                  | N                 | N | N             |               | N                | N                | N                 | 5,6   |  |

P = previously cleared by FDA; N = new indication; E = added under this appendix

Note : 1. Combined modes are B/M, B/CFM , B/PD, B/PW or CW, B/Color M, B/BC,B/CFM/PW or CW,B/PD or DPD/PW or CW

2. Small Organ: thyroid, testes, breast

3. 3D/4D

4. Elastography

- 5. Includes guidance of biopsy (2D)
6. Fusion Harmonic Imaging

7. TDI

Prescription Use × AND/OR Over-The-Counter Use _ (Part 21 CFR 801 Subpart D) (21 CFR 807 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of In Vitro Diagnostics and Radiological Health (OIR)

{9}------------------------------------------------

### SonoBook Series Diagnostic Ultrasound System System: Transducer: L8Mb-V

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

| Clinical Application      |                                    | Mode of Operation |   |               |               |                  |                  |                   | Other |
|---------------------------|------------------------------------|-------------------|---|---------------|---------------|------------------|------------------|-------------------|-------|
| General<br>(Track 1 Only) | Specific<br>(Tracks 1 & 3)         | B                 | M | PW<br>Doppler | CW<br>Doppler | Color<br>Doppler | Power<br>Doppler | Combined<br>Modes |       |
| Ophthalmic                | Ophthalmic                         |                   |   |               |               |                  |                  |                   |       |
| Fetal Imaging &<br>Other  | Fetal                              |                   |   |               |               |                  |                  |                   |       |
|                           | Abdominal                          |                   |   |               |               |                  |                  |                   |       |
|                           | Intra-operative (Specify)          |                   |   |               |               |                  |                  |                   |       |
|                           | Intra-operative (Neuro)            |                   |   |               |               |                  |                  |                   |       |
|                           | Laparoscopic                       |                   |   |               |               |                  |                  |                   |       |
|                           | Pediatric                          | N                 | N | N             |               | N                | N                | N                 | 4,5,6 |
|                           | Small Organ[1] (Specify)           | N                 | N | N             |               | N                | N                | N                 | 4,5,6 |
|                           | Neonatal Cephalic                  |                   |   |               |               |                  |                  |                   |       |
|                           | Adult Cephalic                     |                   |   |               |               |                  |                  |                   |       |
|                           | Trans-rectal                       |                   |   |               |               |                  |                  |                   |       |
|                           | Trans-vaginal                      |                   |   |               |               |                  |                  |                   |       |
|                           | Trans-urethral                     |                   |   |               |               |                  |                  |                   |       |
|                           | Trans-esoph. (non-Card.)           |                   |   |               |               |                  |                  |                   |       |
|                           | Musculo-skeletal<br>(Conventional) | N                 | N | N             |               | N                | N                | N                 | 4,5,6 |
|                           | Musculo-skeletal<br>(Superficial)  | N                 | N | N             |               | N                | N                | N                 | 4,5,6 |
|                           | Other (Urology)                    |                   |   |               |               |                  |                  |                   |       |
| Cardiac                   | Cardiac Adult                      |                   |   |               |               |                  |                  |                   |       |
|                           | Cardiac Pediatric                  |                   |   |               |               |                  |                  |                   |       |
|                           | Transesophageal                    |                   |   |               |               |                  |                  |                   |       |
| Peripheral Vessel         | Peripheral vessel                  | N                 | N | N             |               | N                | N                | N                 | 5,6   |

P = previously cleared by FDA; N = new indication; E = added under this appendix

Note : 1. Combined modes are B/M, B/CFM , B/PD, B/PW or CW, B/Color M, B/BC,B/CFM/PW or CW,B/PD or DPD/PW or CW

2. Small Organ: thyroid, testes, breast

3. 3D/4D

4. Elastography

- 5. Includes guidance of biopsy (2D)
6. Fusion Harmonic Imaging

7. TDI

Prescription Use × AND/OR Over-The-Counter Use _ (Part 21 CFR 801 Subpart D) (21 CFR 807 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of In Vitro Diagnostics and Radiological Health (OIR)

{10}------------------------------------------------

### System: SonoBook Series Diagnostic Ultrasound System Transducer: L12b-V

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

| Clinical Application      |                                    | Mode of Operation |   |               |               |                  |                  |                   |       |
|---------------------------|------------------------------------|-------------------|---|---------------|---------------|------------------|------------------|-------------------|-------|
| General<br>(Track 1 Only) | Specific<br>(Tracks 1 & 3)         | B                 | M | PW<br>Doppler | CW<br>Doppler | Color<br>Doppler | Power<br>Doppler | Combined<br>Modes | Other |
| Ophthalmic                | Ophthalmic                         |                   |   |               |               |                  |                  |                   |       |
| Fetal Imaging &<br>Other  | Fetal                              |                   |   |               |               |                  |                  |                   |       |
|                           | Abdominal                          |                   |   |               |               |                  |                  |                   |       |
|                           | Intra-operative (Specify)          |                   |   |               |               |                  |                  |                   |       |
|                           | Intra-operative (Neuro)            |                   |   |               |               |                  |                  |                   |       |
|                           | Laparoscopic                       |                   |   |               |               |                  |                  |                   |       |
|                           | Pediatric                          | N                 | N | N             |               | N                | N                | N                 | 4,5,6 |
|                           | Small Organ[1] (Specify)           | N                 | N | N             |               | N                | N                | N                 | 4,5,6 |
|                           | Neonatal Cephalic                  |                   |   |               |               |                  |                  |                   |       |
|                           | Adult Cephalic                     |                   |   |               |               |                  |                  |                   |       |
|                           | Trans-rectal                       |                   |   |               |               |                  |                  |                   |       |
|                           | Trans-vaginal                      |                   |   |               |               |                  |                  |                   |       |
|                           | Trans-urethral                     |                   |   |               |               |                  |                  |                   |       |
|                           | Trans-esoph. (non-Card.)           |                   |   |               |               |                  |                  |                   |       |
|                           | Musculo-skeletal<br>(Conventional) | N                 | N | N             |               | N                | N                | N                 | 4,5,6 |
|                           | Musculo-skeletal<br>(Superficial)  | N                 | N | N             |               | N                | N                | N                 | 4,5,6 |
|                           | Other (Urology)                    |                   |   |               |               |                  |                  |                   |       |
| Cardiac                   | Cardiac Adult                      |                   |   |               |               |                  |                  |                   |       |
|                           | Cardiac Pediatric                  |                   |   |               |               |                  |                  |                   |       |
|                           | Transesophageal                    |                   |   |               |               |                  |                  |                   |       |
| Peripheral Vessel         | Peripheral vessel                  | N                 | N | N             |               | N                | N                | N                 | 5,6   |

P = previously cleared by FDA; N = new indication; E = added under this appendix

Note : 1. Combined modes are B/M, B/CFM , B/PD, B/PW or CW, B/Color M, B/BC,B/CFM/PW or CW,B/PD or DPD/PW or CW

2. Small Organ: thyroid, testes, breast

3. 3D/4D

4. Elastography

- 5. Includes guidance of biopsy (2D)
- 6. Fusion Harmonic Imaging

7. TDI

Over-The-Counter Use _ Prescription Use _____________________________________________________________________________________________________________________________________________________________ × AND/OR (Part 21 CFR 801 Subpart D) (21 CFR 807 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

### Concurrence of CDRH, Office of In Vitro Diagnostics and Radiological Health (OIR)

{11}------------------------------------------------

#### System: SonoBook Series Diagnostic Ultrasound System Transducer: L8M-V

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

| Clinical Application      |                                    | Mode of Operation |   |               |               |                  |                  |                   |       |
|---------------------------|------------------------------------|-------------------|---|---------------|---------------|------------------|------------------|-------------------|-------|
| General<br>(Track 1 Only) | Specific<br>(Tracks 1 & 3)         | B                 | M | PW<br>Doppler | CW<br>Doppler | Color<br>Doppler | Power<br>Doppler | Combined<br>Modes | Other |
| Ophthalmic                | Ophthalmic                         |                   |   |               |               |                  |                  |                   |       |
| Fetal Imaging &<br>Other  | Fetal                              |                   |   |               |               |                  |                  |                   |       |
|                           | Abdominal                          |                   |   |               |               |                  |                  |                   |       |
|                           | Intra-operative (Specify)          |                   |   |               |               |                  |                  |                   |       |
|                           | Intra-operative (Neuro)            |                   |   |               |               |                  |                  |                   |       |
|                           | Laparoscopic                       |                   |   |               |               |                  |                  |                   |       |
|                           | Pediatric                          | N                 | N | N             |               | N                | N                | N                 | 4,5,6 |
|                           | Small Organ[1] (Specify)           | N                 | N | N             |               | N                | N                | N                 | 4,5,6 |
|                           | Neonatal Cephalic                  |                   |   |               |               |                  |                  |                   |       |
|                           | Adult Cephalic                     |                   |   |               |               |                  |                  |                   |       |
|                           | Trans-rectal                       |                   |   |               |               |                  |                  |                   |       |
|                           | Trans-vaginal                      |                   |   |               |               |                  |                  |                   |       |
|                           | Trans-urethral                     |                   |   |               |               |                  |                  |                   |       |
|                           | Trans-esoph. (non-Card.)           |                   |   |               |               |                  |                  |                   |       |
|                           | Musculo-skeletal<br>(Conventional) | N                 | N | N             |               | N                | N                | N                 | 4,5,6 |
|                           | Musculo-skeletal<br>(Superficial)  | N                 | N | N             |               | N                | N                | N                 | 4,5,6 |
|                           | Other (Urology)                    |                   |   |               |               |                  |                  |                   |       |
| Cardiac                   | Cardiac Adult                      |                   |   |               |               |                  |                  |                   |       |
|                           | Cardiac Pediatric                  |                   |   |               |               |                  |                  |                   |       |
|                           | Transesophageal                    |                   |   |               |               |                  |                  |                   |       |
| Peripheral Vessel         | Peripheral vessel                  | N                 | N | N             |               | N                | N                | N                 | 5, 6  |

P = previously cleared by FDA; N = new indication; E = added under this appendix

Note : 1. Combined modes are B/M, B/CFM , B/PD, B/PW or CW, B/Color M, B/BC,B/CFM/PW or CW,B/PD or DPD/PW or CW

2. Small Organ: thyroid, testes, breast

3. 3D/4D

4. Elastography

- 5. Includes guidance of biopsy (2D)
6. Fusion Harmonic Imaging

7. TDI

Prescription Use × AND/OR Over-The-Counter Use (Part 21 CFR 801 Subpart D) (21 CFR 807 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of In Vitro Diagnostics and Radiological Health (OIR)

{12}------------------------------------------------

### SonoBook Series Diagnostic Ultrasound System System: Transducer: L12-V

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

| Clinical Application      |                                    | Mode of Operation |   |               |               |                  |                  |                   | Other |
|---------------------------|------------------------------------|-------------------|---|---------------|---------------|------------------|------------------|-------------------|-------|
| General<br>(Track 1 Only) | Specific<br>(Tracks 1 & 3)         | B                 | M | PW<br>Doppler | CW<br>Doppler | Color<br>Doppler | Power<br>Doppler | Combined<br>Modes |       |
| Ophthalmic                | Ophthalmic                         |                   |   |               |               |                  |                  |                   |       |
| Fetal Imaging &<br>Other  | Fetal                              |                   |   |               |               |                  |                  |                   |       |
|                           | Abdominal                          |                   |   |               |               |                  |                  |                   |       |
|                           | Intra-operative (Specify)          |                   |   |               |               |                  |                  |                   |       |
|                           | Intra-operative (Neuro)            |                   |   |               |               |                  |                  |                   |       |
|                           | Laparoscopic                       |                   |   |               |               |                  |                  |                   |       |
|                           | Pediatric                          | N                 | N | N             |               | N                | N                | N                 | 4,5,6 |
|                           | Small Organ[1] (Specify)           | N                 | N | N             |               | N                | N                | N                 | 4,5,6 |
|                           | Neonatal Cephalic                  |                   |   |               |               |                  |                  |                   |       |
|                           | Adult Cephalic                     |                   |   |               |               |                  |                  |                   |       |
|                           | Trans-rectal                       |                   |   |               |               |                  |                  |                   |       |
|                           | Trans-vaginal                      |                   |   |               |               |                  |                  |                   |       |
|                           | Trans-urethral                     |                   |   |               |               |                  |                  |                   |       |
|                           | Trans-esoph. (non-Card.)           |                   |   |               |               |                  |                  |                   |       |
|                           | Musculo-skeletal<br>(Conventional) | N                 | N | N             |               | N                | N                | N                 | 4,5,6 |
|                           | Musculo-skeletal<br>(Superficial)  | N                 | N | N             |               | N                | N                | N                 | 4,5,6 |
|                           | Other (Urology)                    |                   |   |               |               |                  |                  |                   |       |
| Cardiac                   | Cardiac Adult                      |                   |   |               |               |                  |                  |                   |       |
|                           | Cardiac Pediatric                  |                   |   |               |               |                  |                  |                   |       |
|                           | Transesophageal                    |                   |   |               |               |                  |                  |                   |       |
| Peripheral Vessel         | Peripheral vessel                  | N                 | N | N             |               | N                | N                | N                 | 5,6   |

P = previously cleared by FDA; N = new indication; E = added under this appendix

Note : 1. Combined modes are B/M, B/CFM , B/PD, B/PW or CW, B/Color M, B/BC,B/CFM/PW or CW,B/PD or DPD/PW or CW

2. Small Organ: thyroid, testes, breast

3. 3D/4D

4. Elastography

- 5. Includes guidance of biopsy (2D)
6. Fusion Harmonic Imaging

7. TDI

Prescription Use × AND/OR Over-The-Counter Use _ (Part 21 CFR 801 Subpart D) (21 CFR 807 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of In Vitro Diagnostics and Radiological Health (OIR)

{13}------------------------------------------------

### SonoBook Series Diagnostic Ultrasound System System: Transducer: L12M-V

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

| Clinical Application      |                                    | Mode of Operation |   |               |               |                  |                  |                   |       |  |
|---------------------------|------------------------------------|-------------------|---|---------------|---------------|------------------|------------------|-------------------|-------|--|
| General<br>(Track 1 Only) | Specific<br>(Tracks 1 & 3)         | B                 | M | PW<br>Doppler | CW<br>Doppler | Color<br>Doppler | Power<br>Doppler | Combined<br>Modes | Other |  |
| Ophthalmic                | Ophthalmic                         |                   |   |               |               |                  |                  |                   |       |  |
| Fetal Imaging &<br>Other  | Fetal                              |                   |   |               |               |                  |                  |                   |       |  |
|                           | Abdominal                          |                   |   |               |               |                  |                  |                   |       |  |
|                           | Intra-operative (Specify)          |                   |   |               |               |                  |                  |                   |       |  |
|                           | Intra-operative (Neuro)            |                   |   |               |               |                  |                  |                   |       |  |
|                           | Laparoscopic                       |                   |   |               |               |                  |                  |                   |       |  |
|                           | Pediatric                          | N                 | N | N             |               | N                | N                | N                 | 4,5,6 |  |
|                           | Small Organ[1] (Specify)           | N                 | N | N             |               | N                | N                | N                 | 4,5,6 |  |
|                           | Neonatal Cephalic                  |                   |   |               |               |                  |                  |                   |       |  |
|                           | Adult Cephalic                     |                   |   |               |               |                  |                  |                   |       |  |
|                           | Trans-rectal                       |                   |   |               |               |                  |                  |                   |       |  |
|                           | Trans-vaginal                      |                   |   |               |               |                  |                  |                   |       |  |
|                           | Trans-urethral                     |                   |   |               |               |                  |                  |                   |       |  |
|                           | Trans-esoph. (non-Card.)           |                   |   |               |               |                  |                  |                   |       |  |
|                           | Musculo-skeletal<br>(Conventional) | N                 | N | N             |               | N                | N                | N                 | 4,5,6 |  |
|                           | Musculo-skeletal<br>(Superficial)  | N                 | N | N             |               | N                | N                | N                 | 4,5,6 |  |
|                           | Other (Urology)                    |                   |   |               |               |                  |                  |                   |       |  |
| Cardiac                   | Cardiac Adult                      |                   |   |               |               |                  |                  |                   |       |  |
|                           | Cardiac Pediatric                  |                   |   |               |               |                  |                  |                   |       |  |
|                           | Transesophageal                    |                   |   |               |               |                  |                  |                   |       |  |
| Peripheral Vessel         | Peripheral vessel                  | N                 | N | N             |               | N                | N                | N                 | 5,6   |  |

P = previously cleared by FDA; N = new indication; E = added under this appendix

Note : 1. Combined modes are B/M, B/CFM , B/PD, B/PW or CW, B/Color M, B/BC,B/CFM/PW or CW,B/PD or DPD/PW or CW

2. Small Organ: thyroid, testes, breast

3. 3D/4D

4. Elastography

- 5. Includes guidance of biopsy (2D)
6. Fusion Harmonic Imaging

7. TDI

Prescription Use × AND/OR Over-The-Counter Use (Part 21 CFR 801 Subpart D) (21 CFR 807 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of In Vitro Diagnostics and Radiological Health (OIR)

{14}------------------------------------------------

### SonoBook Series Diagnostic Ultrasound System System: Transducer: L10i-V

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

| Clinical Application     |                                    | Mode of Operation |   |            |            |               |               |                | Other |
|--------------------------|------------------------------------|-------------------|---|------------|------------|---------------|---------------|----------------|-------|
| General                  | Specific                           | B                 | M | PW Doppler | CW Doppler | Color Doppler | Power Doppler | Combined Modes |       |
| (Track 1 Only)           | (Tracks 1 & 3)                     |                   |   |            |            |               |               |                |       |
| Ophthalmic               | Ophthalmic                         |                   |   |            |            |               |               |                |       |
| Fetal Imaging &<br>Other | Fetal                              |                   |   |            |            |               |               |                |       |
|                          | Abdominal                          |                   |   |            |            |               |               |                |       |
|                          | Intra-operative (Specify)          |                   |   |            |            |               |               |                |       |
|                          | Intra-operative (Neuro)            |                   |   |            |            |               |               |                |       |
|                          | Laparoscopic                       |                   |   |            |            |               |               |                |       |
|                          | Pediatric                          |                   |   |            |            |               |               |                |       |
|                          | Small Organ[1] (Specify)           | N                 | N | N          |            | N             | N             | N              | 4,5,6 |
|                          | Neonatal Cephalic                  |                   |   |            |            |               |               |                |       |
|                          | Adult Cephalic                     |                   |   |            |            |               |               |                |       |
|                          | Trans-rectal                       |                   |   |            |            |               |               |                |       |
|                          | Trans-vaginal                      |                   |   |            |            |               |               |                |       |
|                          | Trans-urethral                     |                   |   |            |            |               |               |                |       |
|                          | Trans-esoph. (non-Card.)           |                   |   |            |            |               |               |                |       |
|                          | Musculo-skeletal<br>(Conventional) | N                 | N | N          |            | N             | N             | N              | 4,5,6 |
|                          | Musculo-skeletal<br>(Superficial)  | N                 | N | N          |            | N             | N             | N              | 4,5,6 |
|                          | Other (Urology)                    |                   |   |            |            |               |               |                |       |
| Cardiac                  | Cardiac Adult                      |                   |   |            |            |               |               |                |       |
|                          | Cardiac Pediatric                  |                   |   |            |            |               |               |                |       |
|                          | Transesophageal                    |                   |   |            |            |               |               |                |       |
| Peripheral Vessel        | Peripheral vessel                  | N                 | N | N          |            | N             | N             | N              | 5,6   |

P = previously cleared by FDA; N = new indication; E = added under this appendix

Note : 1. Combined modes are B/M, B/CFM , B/PD, B/PW or CW, B/Color M, B/BC,B/CFM/PW or CW,B/PD or DPD/PW or CW

2. Small Organ: thyroid, testes, breast

3. 3D/4D

4. Elastography

- 5. Includes guidance of biopsy (2D)
6. Fusion Harmonic Imaging

7. TDI

Prescription Use × AND/OR Over-The-Counter Use _ (Part 21 CFR 801 Subpart D) (21 CFR 807 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of In Vitro Diagnostics and Radiological Health (OIR)

{15}------------------------------------------------

### SonoBook Series Diagnostic Ultrasound System System: Transducer: R7-V

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

| Clinical Application      |                                    | Mode of Operation |   |               |               |                  |                  |                   | Other |
|---------------------------|------------------------------------|-------------------|---|---------------|---------------|------------------|------------------|-------------------|-------|
| General<br>(Track 1 Only) | Specific<br>(Tracks 1 & 3)         | B                 | M | PW<br>Doppler | CW<br>Doppler | Color<br>Doppler | Power<br>Doppler | Combined<br>Modes |       |
| Ophthalmic                | Ophthalmic                         |                   |   |               |               |                  |                  |                   |       |
| Fetal Imaging &<br>Other  | Fetal                              |                   |   |               |               |                  |                  |                   |       |
|                           | Abdominal                          |                   |   |               |               |                  |                  |                   |       |
|                           | Intra-operative (Specify)          |                   |   |               |               |                  |                  |                   |       |
|                           | Intra-operative (Neuro)            |                   |   |               |               |                  |                  |                   |       |
|                           | Laparoscopic                       |                   |   |               |               |                  |                  |                   |       |
|                           | Pediatric                          |                   |   |               |               |                  |                  |                   |       |
|                           | Small Organ[1] (Specify)           |                   |   |               |               |                  |                  |                   |       |
|                           | Neonatal Cephalic                  |                   |   |               |               |                  |                  |                   |       |
|                           | Adult Cephalic                     |                   |   |               |               |                  |                  |                   |       |
|                           | Trans-rectal                       | N                 | N | N             |               | N                | N                | N                 | 4,5,6 |
|                           | Trans-vaginal                      |                   |   |               |               |                  |                  |                   |       |
|                           | Trans-urethral                     |                   |   |               |               |                  |                  |                   |       |
|                           | Trans-esoph. (non-Card.)           |                   |   |               |               |                  |                  |                   |       |
|                           | Musculo-skeletal<br>(Conventional) |                   |   |               |               |                  |                  |                   |       |
|                           | Musculo-skeletal<br>(Superficial)  |                   |   |               |               |                  |                  |                   |       |
|                           | Other (Urology)                    | N                 | N | N             |               | N                | N                | N                 | 5,6   |
| Cardiac                   | Cardiac Adult                      |                   |   |               |               |                  |                  |                   |       |
|                           | Cardiac Pediatric                  |                   |   |               |               |                  |                  |                   |       |
|                           | Transesophageal                    |                   |   |               |               |                  |                  |                   |       |
| Peripheral Vessel         | Peripheral vessel                  |                   |   |               |               |                  |                  |                   |       |

P = previously cleared by FDA; N = new indication;

E = added under this appendix

Note : 1. Combined modes are B/M, B/CFM , B/PD, B/PW or CW, B/Color M, B/BC,B/CFM/PW or CW,B/PD or DPD/PW or CW

2. Small Organ: thyroid, testes, breast

3. 3D/4D

4. Elastography

- 5. Includes guidance of biopsy (2D)
6. Fusion Harmonic Imaging

7. TDI

Prescription Use メ AND/OR Over-The-Counter Use _ (Part 21 CFR 801 Subpart D) (21 CFR 807 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of In Vitro Diagnostics and Radiological Health (OIR)

{16}------------------------------------------------

### SonoBook Series Diagnostic Ultrasound System System: Transducer: L8M5-V

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

| Clinical Application      |                                    | Mode of Operation |   |               |               |                  |                  |                   |       |  |
|---------------------------|------------------------------------|-------------------|---|---------------|---------------|------------------|------------------|-------------------|-------|--|
| General<br>(Track 1 Only) | Specific<br>(Tracks 1 & 3)         | B                 | M | PW<br>Doppler | CW<br>Doppler | Color<br>Doppler | Power<br>Doppler | Combined<br>Modes | Other |  |
| Ophthalmic                | Ophthalmic                         |                   |   |               |               |                  |                  |                   |       |  |
| Fetal Imaging &<br>Other  | Fetal                              |                   |   |               |               |                  |                  |                   |       |  |
|                           | Abdominal                          |                   |   |               |               |                  |                  |                   |       |  |
|                           | Intra-operative (Specify)          |                   |   |               |               |                  |                  |                   |       |  |
|                           | Intra-operative (Neuro)            |                   |   |               |               |                  |                  |                   |       |  |
|                           | Laparoscopic                       |                   |   |               |               |                  |                  |                   |       |  |
|                           | Pediatric                          | N                 | N | N             |               | N                | N                | N                 | 4,5,6 |  |
|                           | Small Organ [1] (Specify)          | N                 | N | N             |               | N                | N                | N                 | 4,5,6 |  |
|                           | Neonatal Cephalic                  |                   |   |               |               |                  |                  |                   |       |  |
|                           | Adult Cephalic                     |                   |   |               |               |                  |                  |                   |       |  |
|                           | Trans-rectal                       |                   |   |               |               |                  |                  |                   |       |  |
|                           | Trans-vaginal                      |                   |   |               |               |                  |                  |                   |       |  |
|                           | Trans-urethral                     |                   |   |               |               |                  |                  |                   |       |  |
|                           | Trans-esoph. (non-Card.)           |                   |   |               |               |                  |                  |                   |       |  |
|                           | Musculo-skeletal<br>(Conventional) | N                 | N | N             |               | N                | N                | N                 | 4,5,6 |  |
|                           | Musculo-skeletal<br>(Superficial)  | N                 | N | N             |               | N                | N                | N                 | 4,5,6 |  |
|                           | Other (Urology)                    |                   |   |               |               |                  |                  |                   |       |  |
| Cardiac                   | Cardiac Adult                      |                   |   |               |               |                  |                  |…

---

**Source:** [https://fda.innolitics.com/device/K170374](https://fda.innolitics.com/device/K170374)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
