K170353 · HORIBA ABX SAS · GKZ · Nov 1, 2017 · Hematology
Device Facts
Record ID
K170353
Device Name
ABX MICROS ES 60 OT and ABX MICROS ES 60 CT
Applicant
HORIBA ABX SAS
Product Code
GKZ · Hematology
Decision Date
Nov 1, 2017
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 864.5220
Device Class
Class 2
Attributes
Real-World Evidence, Pediatric
Real-World Evidence
Submission
Device
Sponsor
RWD Sources
RWE Use Summary
Key Tags
K170353 · Nov 1, 2017
ABX MICROS ES 60 OT and ABX MICROS ES 60 CT
HORIBA ABX SAS
Clinical laboratory venous whole blood specimens; Published pediatric reference ranges from US hospitals
Retrospective clinical specimens were used to evaluate the performance and bias of the device in a pediatric population (1 month to 21 years) compared to a predicate device. Published reference ranges were used to confirm the transference of pediatric reference intervals.
Pediatric patients between 1 month and 21 years of age; Sample Size: 137; Number of Sites: 3
Sysmex XN Series, XN-10
Bias estimation at medical decision points
Reference interval verification
Normal pediatric patients; Sample Size: 69; Number of Sites: 1
Not applicable for this study
Transference of pediatric reference ranges
Indications for Use
The ABX MICROS ES 60 (OT and CT models) is a quantitative multi-parameter, automated hematology analyzer for in vitro diagnostic use in clinical laboratories to identify and enumerate the following parameters: WBC, RBC, HGB, HCT, MCV, MCH, MCHC, RDW, PLT, MPV, LYM%, LYM#, MON%, MON#, GRA%, GRA#, in K2EDTA and K3EDTA anticoagulated venous whole blood samples of adult patients and pediatric patients ≥ 1 month of age.
Device Story
ABX MICROS ES 60 is an automated hematology analyzer; performs quantitative analysis of venous whole blood samples (K2EDTA/K3EDTA). Inputs: venous blood samples; processed via impedance (RBC, PLT, WBC, DIFF) and spectrophotometry (HGB). Outputs: CBC and leukocyte differential parameters (WBC, RBC, HGB, HCT, MCV, MCH, MCHC, RDW, PLT, MPV, LYM, MON, GRA). Used in clinical laboratories; operated by trained personnel. Manual sample identification via barcode scanner. Provides quantitative data to clinicians for hematological assessment; aids in patient diagnosis and monitoring. Available in Open Tube (OT) and Closed Tube (CT) models; CT features cap-piercing mechanism. Benefits include automated, efficient cell counting and differential analysis for adult and pediatric (≥ 1 month) populations.
Clinical Evidence
Clinical evaluation included 137 venous whole blood specimens from pediatric patients (1 month to 21 years) across three sites. Samples analyzed in duplicate on ABX MICROS ES 60 and compared against Sysmex XN-10. Bias was estimated at medical decision points; acceptance criteria met for all measurands except RDW (due to calculation differences). Reference interval verification performed on 69 normal pediatric samples confirmed results within published ranges.
Technological Characteristics
Automated hematology analyzer; dimensions 43x36x36 cm. Measurement principles: Impedance (RBC, PLT, WBC, DIFF) and Spectrophotometry (HGB). Sample volume: 10 µL. Throughput: 60 (OT) or 50 (CT) samples/hour. Connectivity: manual barcode scanner. Software: embedded firmware (no changes from K141161).
Indications for Use
Indicated for quantitative multi-parameter hematology analysis in clinical laboratories using K2EDTA or K3EDTA anticoagulated venous whole blood. Patient population includes adults and pediatric patients ≥ 1 month of age.
Regulatory Classification
Identification
An automated differential cell counter is a device used to identify one or more of the formed elements of the blood. The device may also have the capability to flag, count, or classify immature or abnormal hematopoietic cells of the blood, bone marrow, or other body fluids. These devices may combine an electronic particle counting method, optical method, or a flow cytometric method utilizing monoclonal CD (cluster designation) markers. The device includes accessory CD markers.
Special Controls
*Classification.* Class II (special controls). The special control for this device is the FDA document entitled “Class II Special Controls Guidance Document: Premarket Notifications for Automated Differential Cell Counters for Immature or Abnormal Blood Cells; Final Guidance for Industry and FDA.”
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
November 01, 2017
HORIBA ABX SAS Caroline Ferrer Regulatory Affairs Manager Parc Euromedecine Rue du Caducee BP 7290 34184 Montpellier Cedex 4 - France
Re: K170353
Trade/Device Name: ABX MICROS ES 60 OT and ABX MICROS ES 60 CT Regulation Number: 21 CFR 864.5220 Regulation Name: Automated differential cell counter Regulatory Class: Class II Product Code: GKZ Dated: September 28, 2017 Received: October 2, 2017
Dear Caroline Ferrer:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801 and Part 809); medical device reporting of medical device-related adverse events) (21 CFR
{1}------------------------------------------------
803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
# Leonthena R. Carrington -S
Lea Carrington Director Division of Immunology and Hematology Devices Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
### Indications for Use
510(k) Number (if known) k170353
### Device Name
ABX MICROS ES 60 OT (Open Tube model) ABX MICROS ES 60 CT (Close Tube model)
### Indications for Use (Describe)
The ABX MICROS ES 60 (OT and CT models) is a quantitative multi-parameter, automated hematology analyzer for in vitro diagnostic use in clinical laboratories to identify and enumerate the following parameters: WBC, RBC, HGB, HCT, MCV, MCH, MCHC, RDW, PLT, MPV, LYM%, LYM#, MON%, MON#, GRA%, GRA#, in K2EDTA and K3EDTA anticoagulated venous whole blood samples of adult patients and pediatric patients ≥ 1 month of age.
| Type of Use (Select one or both, as applicable) | <span> <span style="font-size: 16px;">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) </span> <span> ☐ Over-The-Counter Use (21 CFR 801 Subpart C) </span> |
|-------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
|-------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
***DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.***
The burden time for this collection of information is estimated to average 79 hours per response, including the
time to review instructions, search existing data sources, gather and maintain the data needed and complete
and review the collection of information. Send comments regarding this burden estimate or any other aspect
of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
*PRAStaff@fda.hhs.gov*"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
### Premarket Notification [510(k)] Summary
This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92.
The assigned 510(k) number is : k170353
1.0 Submitted bv : Company: HORIBA ABX SAS HORIBA MEDICAL Parc Euromédecine Rue du Caducée - BP 7290 34184 Montpellier cedex 4 FRANCE
Contact Person: Caroline FERRER Telephone (France): +33 467 141843 Email: caroline.ferrer@horiba.com
Date Submitted : 2.0
28th September 2017
#### 3.0 Device Name and Classification : ABX MICROS ES 60 Trade/Proprietary Name:
### Classification:
Device: Counter, differential cell Panel: 81 Hematology Regulation number: 864.5220 Product Code: GKZ Class: 2
### 4.0 System description : 4.1 Device Description
The ABX MICROS ES 60 is an automated hematology analyzer developed by HORIBA Medical and cleared under k141161.
No modification (hardware or software) to the ABX MICROS ES 60 cleared device (k141161) was made to support the pediatric claim addition.
Performance verification has been done to support the pediatric claim addition.
{4}------------------------------------------------
The corresponding User Instructions for Use was updated to include this additional claim to the intended use.
### 4.2 Principles of Operation
The ABX MICROS ES 60 performs hematology analyses using the following methods:
- RBC / PLT: Impedance ●
- WBC: Impedance ●
- HGB: Spectrophotometry
- DIFF: Impedance ●
The Principles of Operation are unchanged.
### 4.3 Modes of Operation
Two models of ABX MICROS ES 60 are available.
The Closed Tube (CT) instrument: it has a cap-piercing mechanism. The blood collection tube ● is placed directly in the analyzer without removing the cap.
. The Open Tube (OT) instrument: the cap form the blood collection tube must be removed before analyzing the sample.
### 4.4 Specimen Identification
Specimen identification is by manual sample identification with the use of a hand held barcode scanner.
### 4.5 Calibration
Calibration is a procedure that is performed during specific situations such as installation, maintenance or service intervention. It is performed by a HORIBA ABX SAS representative. It ensures that the precision and accuracy of the analyzer are acceptable, so that accurate measurements are performed by the analyzer.
There is no specific calibrator for pediatric addition.
{5}------------------------------------------------
### 4.6 Quality Control
Quality control allows the user to monitor a set of analyses based on known sample values and ranges over a period of several months. Statistical computation performed on these populations allows the extraction of qualitative information related to the stability of the instrument.
The manufacturer's instructions are to be followed for material and frequency of quality control analysis.
There is no specific control for pediatric addition.
### 4.7 Software
FDA has reviewed HORIBA ABX SAS's Hazard Analysis and Software Development processes for this line of product types during the 510(k) file k141161 review.
No change was made to the Hazard Analysis and Software Development processes relative to the pediatric use addition.
#### 5.0 Intended use
### 5.1 Indications for Use :
The ABX MICROS ES 60 (OT and CT models) is a quantitative multi-parameter, automated hematology analyzer for in vitro diagnostic use in clinical laboratories to identify and enumerate the following parameters: WBC, RBC, HGB, HCT, MCV, MCH, MCHC, RDW, PLT, MPV, LYM%, LYM#, MON%, MON#, GRA%, GRA#, in K2EDTA and K3EDTA anticoagulated venous whole blood samples of adult patients and pediatric patients > 1 month of age.
### 5.2 Special Conditions for Use Statements:
For prescription use only
{6}------------------------------------------------
#### 6.0 Comparison with the original device :
# Similarities
| Item | ABX MICROS ES 60<br>Device with additional pediatric<br>intended use | ABX MICROS ES 60 (K141161)<br>Previous Device |
|-------------------------------------------|-------------------------------------------------------------------------------------------------------------------|-----------------------------------------------|
| Manufacturer | HORIBA ABX SAS | HORIBA ABX SAS |
| Type of product | Automated blood cell counter | Automated blood cell counter |
| Parameters | Complete Blood Cell Count (CBC)<br>WBC, RBC, HGB, HCT, MCV,<br>MCH, MCHC, RDW, PLT, MPV | Complete Blood Cell Count (CBC)<br>Same |
| | Leukocyte Differential (DIFF):<br>LYM (# and %), MON (# and %),<br>GRA (# and %) | Leukocyte Differential (DIFF):<br>Same |
| Sample types | K2 and K3EDTA anti-coagulated<br>whole blood | Same |
| Principles of<br>measurements | RBC / PLT: Impedance<br>WBC / DIFF: Impedance<br>HGB: Spectrophotometry<br>RDW, MCV, MCH, MCHC:<br>Calculation | Same |
| Reagents | ABX Minidil LMG<br>ABX Minilyse LMG or ABX<br>Alphalyse 360<br>ABX Miniclean<br>ABX Minipack LMG<br>ABX Minoclair | Same |
| Calibrators | ABX Minocal | Same |
| Controls | ABX Minotrol 16 (3 levels) | Same |
| Throughput | OT / CT models: 60 / 50 samples per<br>hour | Same |
| Dimensions<br>(Height x Width x<br>Depth) | 43 cm x 36 cm x 36 cm | Same |
{7}------------------------------------------------
## Differences
| Item | ABX MICROS ES 60<br>Device with additional pediatric<br>intended use | ABX MICROS ES 60 (K141161)<br>Previous Device |
|--------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------|
| Intended Use | The ABX MICROS ES 60 (OT and CT<br>models) is a quantitative multi-<br>parameter, automated hematology<br>analyzer for in vitro diagnostic use in<br>clinical laboratories to identify and<br>enumerate the following parameters:<br>WBC, RBC, HGB, HCT, MCV, MCH,<br>MCHC, RDW, PLT, MPV, LYM%,<br>LYM#, MON%, MON#, GRA%,<br>GRA#, in K2EDTA and K3EDTA<br>anticoagulated venous whole blood<br>samples of adult patients and pediatric<br>patients ≥ 1 month of age. | Same except:<br>- sample type was venous only<br>- intended population was adult<br>patients only. |
{8}------------------------------------------------
#### 7.0 Substantial Equivalence Information :
The following tables show the similarities and differences between the candidate device and its predicate device identified below.
### 7.1 Predicate Device Name and 510(k) number:
For the validation of the pediatric use on the ABX MICROS ES 60 the following predicate device was used during the clinical evaluation:
| Candidate device | Predicate device | Manufacturer | Predicate 510(k) number |
|------------------|-------------------------|--------------|-------------------------|
| ABX MICROS ES 60 | Sysmex XN Series, XN-10 | SYSMEX | K112605 |
# 7.2 Comparison with predicate Device : Similarities
| Item | ABX MICROS ES 60 (k141161)<br>Candidate device | Sysmex XN Series, XN-10<br>(k112605)<br>Predicate Device |
|-------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Manufacturer | HORIBA ABX SAS | Sysmex |
| Type of product | Automated blood cell counter | Automated blood cell counter |
| Parameters | Complete Blood Cell Count (CBC)<br>WBC, RBC, HGB, HCT, MCV,<br>MCH, MCHC, RDW, PLT, MPV<br>Leukocyte Differential (DIFF):<br>LYM (% and #) | Complete Blood Cell Count (CBC)<br>WBC, RBC, HGB, HCT, MCV,<br>MCH, MCHC, RDW-CV, PLT,<br>MPV<br>Leukocyte Differential (DIFF):<br>LYM (# and %), |
| Sample types | K2 and K3EDTA anti-coagulated<br>venous samples | K2 and K3EDTA anti-coagulated<br>whole blood |
| Principles of<br>measurements | RBC / PLT: Impedance<br>HGB: Spectrophotometry | RBC: Impedance (Sheath Flow)<br>HGB: Spectrophotometry |
| Item | ABX MICROS ES 60 (k141161) Candidate device | Sysmex XN Series, XN-10 (k112605) Predicate Device |
| Intended Use | The ABX MICROS ES 60 (OT and CT models) is a quantitative multi-parameter, automated hematology analyzer for in vitro diagnostic use in clinical laboratories to identify and enumerate the following parameters: WBC, RBC, HGB, HCT, MCV, MCH, MCHC, RDW, PLT, MPV, LYM%, LYM#, MON%, MON#, GRA%, GRA#, in K2EDTA and K3EDTA anticoagulated venous whole blood samples of adult patients and pediatric patients ≥ 1 month of age. | The Sysmex® XN-10 and XN-20 modules are quantitative multi-parameter automated hematology analyzers intended for in vitro diagnostic use in screening patient populations found in clinical laboratories. The XN-Series modules classify and enumerate the following parameters: WBC, RBC, HGB, HCT, MCV, MCH, MCHC, RDW-CV, RDW-SD, PLT, NEUT%/#, LYMPH%/#, MONO%/#, EO%/#, BASO%/#, NRBC%/#, IG%/#, , MPV, RET%/#, IRF, IPF, RET-He |
| Parameters | Leukocyte Differential (DIFF):<br>MON (% and #)<br>GRA (% and #)<br><br>RDW-SD, NRBC%/#, IG%/#, , MPV, RET%/#, IRF, IPF, RET-He: parameters not available | Leukocyte Differential (DIFF):<br>NEUT%/#<br>ΜΟΝΟ%/#,<br>EO%/#,<br>BASO%/#<br><br>RDW-SD, NRBC%/#, IG%/#, , MPV, RET%/#, IRF, IPF, RET-He |
| Minimum<br>Sample<br>Volume | 50 µL | Sampler Mode – 88 µL<br>Manual (Closed Cap) Mode – 88 µL<br>Manual (Open Cap) Mode – 88 µL<br>Dilution Mode – 70 µL |
| Specimen<br>sample volume | 10 µL | |
| Dimensions | OT / CT models:<br>43 x 36 x 36 cm | 85.5 cm x 64.5 cm x 75.5 cm |
| Weight | OT model: 14 kg<br>CT model: 17 kg | 57 kg |
| Throughput | OT / CT models: 60 / 50 samples per hour | 100 samples/hour maximum depending on mode used. |
| Modes of<br>Operation | Open Tube Mode<br>Closed Tube Mode | Sampler Analysis Mode<br>Manual Closed Analysis Mode<br>Pre-dilute Analysis Mode Dilute sample 1:7 |
| Principles of Measurement | | |
| WBC, WBC<br>Differential | Impedance | WBC: Radio-frequency (RF) / Direct-current (DC) Detection Method Flow |
| Item | ABX MICROS ES 60 (k141161)<br>Candidate device | Sysmex XN Series, XN-10 (k112605)<br>Predicate Device |
| (LYM, MON,<br>GRA) | | Cytometry Methods using a Semiconductor Laser<br>Particle characterization and identification is<br>based on detection of forward scatter,<br>fluorescence and adaptive cluster analysis.<br>The XN-Series analyzers automatically<br>classify cells from whole blood |
| Calibrator | ABX Minocal | XN CAL (XN-10/XN-20 Calibrator)<br>XN CAL PF (Platelet F Calibrator) |
| Quality<br>Controls | ABX Minotrol 16<br>(3 levels) | XN-Check - 3 Levels |
| Reagents | ABX Minidil LMG (Diluent)<br>ABX Minilyse LMG or ABX Alphalyse<br>360 (Lyse)<br>ABX Miniclean (Cleaner)<br>ABX Minoclair (Concentrated cleaning<br>reagent)<br>ABX Minipack LMG (Reagent Pack) | CELLPACKTM DCL (Diluent)<br>CELLPACKTM DFL (Diluent)<br>LYSERCELL WNR (Lyse) LYSERCELL<br>WDF (Lyse) LYSERCELL WPC (Lyse)<br>FLUOROCELL WNR (Stain)<br>FLUOROCELL WDF (Stain)<br>FLUOROCELL RET (Stain)<br>FLUOROCELL PLT (Stain)<br>FLUOROCELL WPC (Stain) |
{9}------------------------------------------------
# 7.3 Comparison with predicate Device : Differences
{10}------------------------------------------------
#### 8.0 Special Control/Guidance Document Referenced :
### 8.1 Standards Followed
- CLSI EP09-A3: Measurement Procedure Comparison and Bias Estimation Using ● Patient Samples - 2013
- CLSI EP28-A3: Defining, Establishing and Verifying Reference Intervals in the ● Clinical Laboratory - 2008
### 8.2 FDA Guidances Followed
- Guidance for Industry and FDA Staff : Format for Traditional and Abbreviated 510(k)s ● - 2005
- Final Guidance for Industry and FDA: Class II Special Controls Guidance Document: Premarket Notifications for Automated Differential Cell Counters for Immature or Abnormal Blood Cells - 2001
{11}------------------------------------------------
#### 9.0 Summary of Performance Data :
### 9.1 Comparability with Predicate Device
A total of 137 venous whole blood specimens (collected in K2EDTA) from pediatric patients between 1 month and 21 years of age were collected and analyzed at two test sites in the US and one test site in France. Different instruments and operators were used at each site.
Each of the samples was analyzed in duplicate on the ABX MICROS ES 60 and once on the predicate XN Series, XN-10 from SYSMEX.
Bias was estimated at various data points, particularly around medical decision points, for each parameter. Bias was estimated for one replicate. Acceptance criteria were met for all measurands at all levels, except for RDW, due to the difference of calculation methods on the two analyzers. These findings support the claim that the ABX MICROS ES 60 candidate device and the XN Series, XN-10 predicate device are substantially equivalent, and demonstrate acceptable levels of bias for pediatric samples.
### 9.2 Reference Interval
69 (37 female and 32 male) normal pediatric samples (whole blood samples collected in K2EDTA) were analyzed, when possible in duplicate, on the ABX MICROS ES 60 at one test site in the US. Results were compared to a pediatric reference range obtained from several US hospitals published ranges. Results were contained in the pediatric reference range, confirming the transference of the pediatric reference range below for the ABX MICROS ES 60:
| Combined Pediatric | | MALE | | | FEMALE | | |
|----------------------------|-----------|--------------|--------------|--------------|--------------|--------------|--------------|
| Reference Range parameters | | 1 m - 2 y | >2y - 12 y | >12 y - 21 y | 1 m - 2 y | >2y - 12 y | >12 y - 21 y |
| WBC | (10³/mm³) | 5.0 - 20.0 | 3.4 - 17.0 | 3.5 - 13.0 | 5.0 - 20.0 | 3.4 - 17.0 | 3.5 - 13.0 |
| RBC | (10⁶/mm³) | 2.7 - 6.0 | 3.9 - 5.5 | 4.0 - 5.7 | 2.7 - 6.0 | 3.8 - 5.5 | 3.7 - 5.2 |
| HGB | g/dL | 8.9 - 18.0 | 10.2 - 15.5 | 11.0 - 18.0 | 9.0 - 18.0 | 10.2 - 15.5 | 10.6 - 16.0 |
| HCT | % | 28.0 - 55.0 | 31.0 - 45.0 | 33.9 - 49.0 | 27.7 - 55.0 | 31.0 - 45.0 | 32.9 - 49.0 |
| MCV | μm³ | 70 - 123 | 70 - 95 | 77 - 102 | 70 - 123 | 70 - 95 | 76.9 - 102 |
| RDW | % | 11.0 - 16.5 | 11.0 - 16.0 | 11.0 - 15.6 | 11.0 - 16.5 | 11.0 - 16.0 | 11.0 - 15.5 |
| MCH | pg | 22.7 - 36.0 | 23.0 - 33.0 | 25.0 - 35.0 | 24.4 - 35 | 23.7 - 33.0 | 24.8 - 35.0 |
| MCHC | % | 28 - 37 | 30 - 37 | 31 - 37 | 28 - 36 | 30 - 37 | 31 - 37 |
| PLT | (10³/mm³) | 140 - 562 | 140 - 450 | 140 - 450 | 140 - 597 | 140 - 450 | 140 - 450 |
| MPV | μm³ | 6.5 - 12.4 | 6.5 - 12.4 | 6.5 - 12.4 | 6.5 - 12.4 | 6.5 - 12.4 | 6.5 - 12.4 |
| LYM# | (10³/mm³) | 1.50 - 16.50 | 0.97 - 10.50 | 0.85 - 6.50 | 1.50 - 16.50 | 1.16 - 10.50 | 0.90 - 6.50 |
| LYM% | % | 26 - 86 | 15 - 75 | 12 - 53 | 27 - 87 | 17 - 75 | 13 - 50 |
| MON# | (10³/mm³) | 0.00 - 1.40 | 0.00 - 1.00 | 0.00 - 0.90 | 0.00 - 1.40 | 0.00 - 1.00 | 0.00 - 0.90 |
| MON% | % | 0 - 14 | 0 - 12 | 0 - 13 | 0 - 13 | 0 - 12 | 0 - 13 |
| GRA# | (10³/mm³) | 0.84 - 10.1 | 1.50 - 9.45 | 1.50 - 8.80 | 1.01 - 10.45 | 1.50 - 9.45 | 1.50 - 8.80 |
| GRA% | % | 10 - 80 | 12 - 85 | 31 - 87 | 9 - 86 | 12 - 81 | 31 - 87 |
These intervals are given in the labeling of the ABX MICROS ES 60.
However, expected values will vary with sample population and/or geographical location. HORIBA highly recommends that each laboratory establishes its own normal ranges based upon its local population.
{12}------------------------------------------------
### 10.0 Proposed Labeling :
The labeling is written as per the recommendations given in standard EN18113-2. It takes into account the requirements of 21 CFR Part 809.10.
### 11.0 Conclusion :
As per 21CFR Part §807.92(b)(3), the nonclinical tests demonstrate that the ABX MICROS ES 60 device is as safe, as effective, and performs as well as or better than the predicate device for pediatric population (≥1 month of age).
The submitted information in this premarket notification is complete and supports a substantial equivalence decision on pediatric use addition.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.