For use in ablating a channel in occlusive peripheral vascular disease.
Device Story
DABRA Laser System consists of RA-308 excimer laser source and model 101 catheter. System uses 308nm ultraviolet laser energy to photoablate lesion material in peripheral vasculature. Catheter features optically conducting fluid within FEP tubing, titanium distal tip, and glass window. Physician operates system in catheterization laboratory; inserts catheter into vasculature to target blockage. Laser energy creates lumen through chronic total occlusion, restoring blood flow and enabling subsequent interventional treatments like balloon angioplasty. System benefits patients by treating flow obstructions that cause pain, non-healing wounds, gangrene, and potential amputation.
Clinical Evidence
Clinical study conducted on 50 patients (66 lesions). Primary endpoints included successful device deployment, navigation, lesion crossing, procedure outcome, and safety. Results showed device performed as predicted with no observed adverse events or safety issues. Performance goals were exceeded.
Technological Characteristics
308nm excimer laser source. Catheter materials: FEP tubing, titanium tip, glass window, epoxy adhesive. Conforms to ISO 10993 (biocompatibility), ISO 10555 (catheters), ISO 11137 (gamma sterilization), IEC 60601-1 (electrical safety), IEC 60825-1 (laser safety), and IEC 60601-1-2 (EMC). 1.5mm diameter, 150cm working length. Single-use.
Indications for Use
Indicated for crossing chronic total occlusions in patients with symptomatic infrainguinal lower extremity vascular disease.
Regulatory Classification
Identification
A percutaneous catheter is a device that is introduced into a vein or artery through the skin using a dilator and a sheath (introducer) or guide wire.
{0}------------------------------------------------
Public Health Service
Image /page/0/Picture/2 description: The image is a circular seal for the U.S. Department of Health & Human Services. The seal features the department's emblem, which consists of a stylized caduceus with three intertwined snakes and a staff. The emblem is positioned in the center of the seal. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged around the upper perimeter of the circle.
May 24, 2017
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
Ra Medical Systems, Inc. Mr. Dean Irwin CEO 1930 Kellogg Ave. Carlsbad. CA 92008
Re: K170349
Trade/Device Name: DABRA Laser System (DABRA Laser model RA-308 and DABRA Catheter model 101) Regulation Number: 21 CFR 870.1250 Regulation Name: Percutaneous Catheter Regulatory Class: Class II Product Code: PDU Dated: April 24, 2017 Received: April 24, 2017
Dear Mr. Irwin:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR
{1}------------------------------------------------
Part 807); labeling (21 CFR Part 801); medical device reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely.
Fernando
Aguel-S
for Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K170349
Device Name
DABRA Laser System (DABRA Laser model RA-308 and DABRA Catheter model 101)
Indications for Use (Describe)
The DABRA Laser System is indicated for crossing chronic total occlusions in patients with symptomatic infrainguinal lower extremity vascular disease.
Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
## PREMARKET NOTIFICATION
## TRADITIONAL 510(k) SUMMARY
#### Submitter Information
A. Company Name: Ra Medical Systems, Inc.
- B. Company address: 1930 Kellogg Ave, Carlsbad CA 92008
- C. Company Phone: (760) 804 1648 / (877) 635 1800 Fax: (760) 804 1657
- D. Contact Person: Dean Irwin, Chief Executive Officer
- E. Date Prepared: January 20, 2017
### Device Identification
- A. Device Trade Name: DABRA Laser System (DABRA Laser model RA-308 and DABRA Catheter model 101)
- B. Device Common Name: Laser Catheter, Excimer Laser
- C. Classification Name: Percutaneous catheter
- D. Device Class: Class II (per 870.1250 percutaneous catheter)
E. Device Code: PDU
#### Identification of Predicate Device
Spectranetics CLiRpath Laser Catheters; K040067.
## Identification of Reference Devices
Ra Medical EX-308; K062963
Spectranetics CVX-300; K052514
#### Device Description
Ra Medical Systems is requesting FDA 510(k) clearance for the DABRA Laser (model RA-308) and DABRATM Catheter (model 101) as a system, collectively the DABRA Laser System™.
The Ra Medical Systems' DABRA Laser System™ is composed of a laser light source and catheter consisting of optically conducting fluid encased in medical grade tubing enclosed by an optical window on either end. The tip and the fluid conduct the ultraviolet laser energy from the laser light source to the distal tip of the catheter. The laser light is generated by a 308nm excimer source. The catheter is connected to the laser for the procedure, and then inserted into the patient's vasculature, allowing the physician to target the laser energy to a blockage or lesion. Patient contacting parts of the device are the distal tip (titanium), the distal tip window (glass), the glue for the distal tip (epoxy), and the catheter tube (FEP). These parts are limited (<24 hours) blood contact. The laser energy photoablates the lesion material creating a lumen that permits blood flow, and allows access for other interventional treatment devices, such as balloons. The system is designed to be used in a catheterization laboratory. The candidates for this type of laser treatment are people who have blockages in their leg arteries that completely obstruct flow (chronic total occlusions). Flow obstruction of this type causes pain, wounds that do not heal, gangrene and ultimately limb amputation.
{4}------------------------------------------------
## Intended Use (807.92(a)(5))
For use in ablating a channel in occlusive peripheral vascular disease.
### Indications for Use
The DABRA Laser System is indicated for crossing chronic total occlusions in patients with symptomatic infrainguinal lower extremity vascular disease.
## Comparison to Predicate Devices
The DABRA Laser™ and DABRA Catheter™ use the same materials (plastics, glass, metal), the same light source and energy (308nm excimer at 15ml), the same constructions, and the same interventional techniques (step-by-step) to operate as the predicate device. The differences between the devices include a more compact excimer source and an optically conducting material as opposed to a continuous glass element. The materials that define the laser output as the materials that contact the patient are all either identical, or are substantially equivalent in regards to the interaction of the device with the patient or the operator. The DABRA Catheter™ is 1.5mm in diameter whereas the predicate device is for a line of different sized catheters ranging from .9mm to 2.5mm. The device that is closest in diameter is the CLiRpath 1.4mm device, which was the subject of comparison bench tests. The DABRA Catheter™ has a working length of 150cm.
The RA-308 light source for the device is nearly identical to the EX-308, with different delivery device connections and software. The RA-308 Laser is identical to the EX-308 in fit, form, function, energy type, materials, and use.
The Spectranetics CVX-300 reference device also has an 80Hz repetition rate and continuous operation.
The DABRA Laser™ and DABRA Catheter™ were bench tested using a wide variety of protocols. These protocols also included side-by-side performance tests. The direct comparison tests included:
- a. Energy output
- b. Output fluence in mJ/mm2 applied to the target
- c. Beam divergence and beam profile
- d. Energy transmission in test models
- e. Ablation though non-biological samples to demonstrate hole size and features
- f. Ablation through biological materials (in vitro porcine)
- g. Vasculature maneuverability in test models
- h. Insertion and retraction force, at time of manufacture and aged 2 years
Other bench tests on sterilized catheters (where applicable) include:
- a. Pull testing, at time of manufacture and aged 2 years
- b. Corrosion resistance
{5}------------------------------------------------
- c. Patent artery perforation testing
- d. Kink testing, at time of manufacture and aged 2 years
- Torque testing, at time of manufacture and aged 2 years e.
- f. Radio opaque tip testing
- g. Catheter fluid leak analysis
- h. Package integrity validation
- i. Process indicator validation
- Dimensional testing, at time of manufacture and aged 2 years i.
- k. Particulate investigation
- Simulated use testing l.
Other testing to show conformance to various standards include:
- IEC 60601-1 Electrical safety a.
- b. IEC 60825-1 Safety of laser products
- c. IEC 60601-2-22 Basic safety and essential performance of laser equipment
- d. IEC 60601-1-2 Electromagnetic compatibility
- e. 21 CFR 1010 and 1040 Electronic products and light-emitting products
- f. ISO 10555 Sterile single-use intravascular catheters
- g. IEC 62366 Usability engineering
#### Biocompatibility and Sterilization
The Ra Medical DABRA Catheters™ are manufactured from materials and components that are commonly used in other catheters already marketed. All of the materials conform to ISO 10993, Biological Evaluation of Medical Devices. The following biocompatibility testing has been performed:
- C3a complement activation ●
- Cytotoxicity
- Guinea pig maximization sensitization
- Direct and indirect hemolysis
- Intracutaneous reactivity
- Partial thromboplastin time
- Material-mediated Pyrogenicity
- SC5b-9 complement activation
- Acute systemic toxicity
- In vivo human thrombus evaluation ●
Ra Medical conducts and maintains valid gamma radiation sterilization processes in conformance with ISO 11137 Sterilization of Health Care Products - Radiation. The packaging for the DABRA Catheters™ has been initially validated, and each is visually inspected prior to delivery to finished goods. A shelf life of 1 year was validated based on successful packaging integrity and product performance testing using accelerated aged and real time aged device samples.
{6}------------------------------------------------
Each catheter is inspected and 100% tested for dimensions and functionality. It has been validated for integrity to conform to ISO 10555-1 Sterile Single-Use Intravascular Catheters.
### Clinical Results
The device was used on 50 patients (66 lesions) in two studies to determine that the device is substantially equivalent to the predicate in terms of procedure outcome, adverse events, and physician use. The performance goal was exceeded.
- a. the device deployed properly and predictably from the packaging,
- b. the device navigated the vasculature easily,
- c. the device crossed the lesions with the predicted manner, time and ease,
- d. the outcome of the procedure was as predicted,
- e. the follow-up on the patients revealed no undesired effects,
- f. there were no observed adverse events or safety issues according to the study protocol.
#### Conclusion
The analysis, testing, and clinical study establish the substantial equivalence between the Ra Medical Systems DABRA Laser™ and DABRA Catheter™ and the predicate device, and demonstrate that the device performs equivalently to the referenced predicate and exceeded the performance goal.
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.