Signia Four-Bay Smart Charger (to be used with Signia Stapler)
K170348 · Covidien · GDW · Mar 2, 2017 · General, Plastic Surgery
Device Facts
Record ID
K170348
Device Name
Signia Four-Bay Smart Charger (to be used with Signia Stapler)
Applicant
Covidien
Product Code
GDW · General, Plastic Surgery
Decision Date
Mar 2, 2017
Decision
SESE
Submission Type
Special
Regulation
21 CFR 878.4750
Device Class
Class 2
Attributes
Pediatric
Indications for Use
The Signia™ Stapler, when used with Endo GIA™ single-use reloads, Endo GIA™ single-use reloads with Tri-Staple™ Technology, Tri-Staple™ 2.0 single-use reloads and Signia™ loading units with Tri-Staple™ 2.0 single-use cartridges, has applications in abdominal, gynecological, pediatric, and thoracic surgery for resection, and creation of anastomosis. It may be used for transection of liver substance, hepatic vasculature, and biliary structures and for transection and resection of the pancreas. The Signia™ Stapler, when used with Endo GIA™ curved tip single use reloads or Tri-Staple™ 2.0 curved tip single-use reloads, can be used to blunt dissect or separate target tissue from other certain tissue. The Signia™ Stapler, when used with Endo GIA™ single use Radial Reloads with Tri-Staple™ Technology, has applications in open or minimally invasive general abdominal, gynecologic, pediatric and thoracic surgery for resection and transection of tissue and creation of anastomosis, as well as application deep in the pelvis, i.e., low anterior resection. It may be used for transection of liver substance, hepatic vasculature and biliary structures and for transection and resection of the pancreas. The Signia™ Stapler, when used with Endo GIA™ single use reinforced reloads with Tri-Staple™ Technology preloaded with polyglycolic acid staple line reinforcement or Tri Staple™ 2.0 single use reinforced reloaded with polyglycolic acid staple line reinforcement, has applications in abdominal, gynecologic, pediatric and thoracic surgery for resection, transection of tissue and creation of anastomosis. It may be used for transection of liver substance, hepatic vasculature and biliary structures, and for transection and resection of the pancreas.
Device Story
Signia Four-Bay Smart Charger is an accessory for the battery-powered, microprocessor-controlled Signia Stapler. The charger provides four charging stations for stapler power handles; features a touch-screen LCD display and data management interface for user convenience. It regulates voltage and current for battery charging. Used by medical professionals in clinical settings to maintain stapler power. The device does not alter the stapling mechanism or surgical function of the Signia Stapler; it serves solely to charge and manage the power handles.
Clinical Evidence
No clinical data. Substantial equivalence demonstrated via non-clinical performance testing, including software verification/validation (IEC 62304), electrical safety (ANSI/AAMI ES 60601-1, IEC 60601-1), EMC testing (IEC 60601-1-2), and usability evaluation (IEC 60601-1-6, IEC 62366-1).
Technological Characteristics
Accessory charger for battery-powered surgical stapler. Features: four charging stations, touch-screen LCD, data ports. Power: voltage and current regulation for battery chemistry. Standards: IEC 62304 (software), ANSI/AAMI ES 60601-1/IEC 60601-1 (electrical safety), IEC 60601-1-2 (EMC), IEC 60601-1-6/IEC 62366-1 (usability).
Indications for Use
Indicated for patients undergoing abdominal, gynecological, pediatric, or thoracic surgery requiring tissue resection, transection, or anastomosis. Includes liver, hepatic vasculature, biliary structures, and pancreas procedures. Also indicated for blunt dissection/separation of target tissue and pelvic procedures (e.g., low anterior resection).
Regulatory Classification
Identification
An implantable staple is a staple-like device intended to connect internal tissues to aid healing. It is not absorbable.
{0}------------------------------------------------
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
March 2, 2017
Covidien Ms. Katherine Choi Principal Regulatory Affairs Specialist 60 Middletown Avenue North Haven, Connecticut 06473
Re: K170348
Trade/Device Name: Signia Four-Bay Smart Charger (to Be Used With Signia Stapler) Regulation Number: 21 CFR 878.4750 Regulation Name: Implantable staple Regulatory Class: Class II Product Code: GDW Dated: February 2, 2017 Received: February 3, 2017
Dear Ms. Choi:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device
{1}------------------------------------------------
related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely,
# David Krause -S
for Binita S. Ashar, M.D., M.B.A., F.A.C.S. Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K170348
#### Device Name
Signia™ Four-Bay Smart Charger
#### Indications for Use (Describe)
The Signia™ Stapler, when used with Endo GIA™ single-use reloads, Endo GIA™ single-use reloads with Tri-Staple™ Technology, Tri-Staple™ 2.0 single-use reloads and Signia™ loading units with Tri-Staple™ 2.0 single-use cartridges, has applications in abdominal, gynecological, pediatric, and thoracic surgery for resection, and creation of anastomosis. It may be used for transection of liver substance, hepatic vasculature, and biliary structures and for transection and resection of the pancreas.
The Signia™ Stapler, when used with Endo GIA™ curved tip single use reloads or Tri-Staple™ 2.0 curved tip single-use reloads, can be used to blunt dissect or separate target tissue from other certain tissue.
The Signia™ Stapler, when used with Endo GIA™ single use Radial Reloads with Tri-Staple™ Technology, has applications in open or minimally invasive general abdominal, gynecologic, pediatric and thoracic surgery for resection and transection of tissue and creation of anastomosis, as well as application deep in the pelvis, i.e., low anterior resection. It may be used for transection of liver substance, hepatic vasculature and biliary structures and for transection and resection of the pancreas.
The Signia™ Stapler, when used with Endo GIA™ single use reinforced reloads with Tri-Staple™ Technology preloaded with polyglycolic acid staple line reinforcement or Tri Staple™ 2.0 single use reinforced reloaded with polyglycolic acid staple line reinforcement, has applications in abdominal, gynecologic, pediatric and thoracic surgery for resection, transection of tissue and creation of anastomosis. It may be used for transection of liver substance, hepatic vasculature and biliary structures, and for transection and resection of the pancreas.
| Type of Use (Select one or both, as applicable) | |
|--------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------|
| <div> <span> <span style="font-size:16px">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) </span> </div> | <div> <span> <span style="font-size:16px">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) </span> </div> |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) Summary of Safety and Effectiveness
## DATE PREPARED:
February 2, 2017
#### SUBMITTER:
Covidien 60 Middletown Avenue North Haven, CT 06473 USA
#### CONTACT PERSON:
Katherine Y. Choi, RAC Principal Regulatory Affairs Specialist Telephone: (203) 492-8412 Fax: (203) 492-5029
#### IDENTIFICATION OF DEVICE:
Signia™ Four-Bay Smart Charger (to be used with Signia™ Stapler) Proprietary/Trade Name: Staples, Implantable Classification Name: 21 CFR 878.4750 Requiation Number: Product Code: GDW Class II Device Class: General and Plastic Surgery Review Panel: Surgical Stapler Common Name:
#### PREDICATE DEVICE:
Proprietary/Trade Name: Signia™ Stapler K160176 (April 26, 2016) 510(k) Number: Classification Name: Staples, Implantable 21 CFR 878.4750 Requlation Number: Product Code: GDW Device Class: Class II General and Plastic Surgery Review Panel: Surgical Stapler Common Name:
# DEVICE DESCRIPTION:
The Signia™ Stapler is a battery powered microprocessor controlled surgical stapler that provides push-button powered maneuverability and firing of compatible Covidien stapling reloads. The Signia™ Stapler is compatible with Endo GJA™ single-use reloads, Endo GIA™ single-use reloads with Tri-Staple™ Technology, Tri-Staple™ 2.0 single use reloads, and Signia™ loading units with Tri-Staple™ Otape - roomnology, The Signia™ Stapler when used with the abovementioned family of Endo GIA™ reloads is a surgical device for stapling and cutting tissues. The Signia™ Stapler is intended to be used by medical professionals qualified in the transportation, preparation, cleaning, sterilization, and use of surgical devices. All of which stay unchanged when compared to the predicate device.
The Signia™ Stapler is composed of the Signia™ Power Handle, Signia™ Power Shell, and Signia™ The Signia - System accessories include the Signia™ Reusable Insertion Guide, Signia™ Manual Retraction Tool, Signia™ Single Bay Charger, Signia™ Sterilization Tray, and this submission introduces a new accessory, Signia™ Four-Bay Smart Charger. The Smart Charger's primary purpose introduces a new as the Signia Single-Bay Charger in K160176, but the difference is that Smart
{4}------------------------------------------------
Charger offers enhanced features such as four charging stations, a touch-screen LCD display, and data management interface in order to improve user convenience.
## INTENDED USE/INDICATIONS FOR USE:
The Signia™ Stapler, when used with Endo GIA™ single-use reloads, Endo GIA™ single-use reloads Trie Signia - Olapic, When assumments 2.0 single-use reloads and Signia™ loading units with Tri-Willi Theolape - " Connology, " The Suplications in abdominal, gynecological, pediatric, and thoracic Otapic - 2.0 Single doo carthagoo, had creation of anastomosis. It may be used for transection and Surgery for research, transculative, and biliary structures and for transection and resection of the pancreas.
The Signia™ Stapler, when used with Endo GIA™ curved tip single use reloads or Tri-Staple™ 2.0 The Sightia - Stapler, when asou with Endo One - Carrou appearate target tissue from other certain tissue.
The Signia™ Stapler, when used with Endo GIA™ single use Radial Reloads with Tri-Staple™ The Olgina - Oldpler, whon acount minimally invasive general abdominal, gynecologic, pediatric recimology, nua upplication of tissue and creation of tissue and creation of anastomosis, as well as and thorable surgery for roosenen and atterior resection. It may be used for transection and resection applivation deep in the porner, for and biliary structures and for transection and resection of the pancreas.
The Signia™ Stapler, when used with Endo GIA™ single use reinforced reloads with Tri-Staple™ The olgha - Oldplar, with polyglycolic acid staple line reinforcement or Tri-Staple™ 2.0 single use reinforced reloads preloaded with polyglycolic acid staple line reinforcement, has applications in romoroco "proloado" proloaded" in d therecic surgery for resection, transection of tissue and creation abounting, gynoongly pound for transection and resection of liver substance, hepatic vasculature and biliary structures, and for transection and resection of the pancreas.
# SUMMARY OF TECHNOLOGICAL CHARACTERISTICS:
The Signia™ Four-Bay Smart Charger is designed for use as an accessory with the Signia™ stapler, The Signia - Tour Day offical Jana Jana touch-screen LCD display screen. Its purpose who houded four power handle. The touchscreen LCD is used to indicate the power handle's is to onalye the organia - power hanigdia™ Single Bay Charger in K160176, the Signia™ Four-Bay Smart Charger employs the smart charging mechanism that provides necessary voltage and current regulation for the given battery chemistry.
#### SUBSTANTIAL EQUIVALENCE:
The Signia™ Stapler used with the Signia™ Four-Bay Smart Charger is substantially equivalent to the TTIe Signia - Olapics asou with the Signia - the Signia " Single-Bay Charger (K160176) since adding a new charger does not affect the stapling technologies employed. Also, when compared to the Signia™ new onarger, the key difference is the Smart Charger offers enhanced features such as four charging stations, touch-screen LCD display and data ports. Other product specifications are very similar.
#### SUMMARY OF STUDIES:
Non-clinical performance data – Tests performed to demonstrate continued conformance with the FDA recognized standards includes:
- 1. Software Verification & Validation per IEC 62304
- 2. Electrical Safety Testing per ANSI/AAMI ES 60601-1 & IEC 60601-1
- 3. EMC Testing per IEC 60601-1-2
- 4. Usability Evaluation per IEC 60601-1-6 & IEC 62366-1
{5}------------------------------------------------
Clinical performance data -- No clinical study has been performed. The substantial equivalence has been demonstrated by non-clinical studies.
#### CONCLUSION:
Based upon the supporting data summarized above, we concluded that Signia" with the new Based upon the Supporting add Summanized as safe and effective as the legally marketed device accessory, Sigma - Tour-Day Omart Online of safety and effectiveness than the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.