Custom Stable(TM) Rigid Gas Permeable Scleral Contact Lens
K170335 · Valley Contax, Inc. · HQD · Mar 24, 2017 · Ophthalmic
Device Facts
Record ID
K170335
Device Name
Custom Stable(TM) Rigid Gas Permeable Scleral Contact Lens
Applicant
Valley Contax, Inc.
Product Code
HQD · Ophthalmic
Decision Date
Mar 24, 2017
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 886.5916
Device Class
Class 2
Attributes
Therapeutic, Real-World Evidence
Real-World Evidence
Submission
Device
Sponsor
RWD Sources
RWE Use Summary
Key Tags
K170335 · Mar 24, 2017
Custom Stable(TM) Rigid Gas Permeable Scleral Contact Lens
Valley Contax, Inc.
Clinical case reports
A series of clinical case reports were provided to demonstrate the performance of the Custom Stable Rigid Gas Permeable Scleral Contact Lens when prescribed for the proposed indications for use in a clinical setting.
The Custom Stable(TM) Rigid Gas Permeable Scleral Contact Lenses for daily wear are indicated for use for the management of multiple ocular conditions, such as, degenerations that lead to an irregular corneal shape (e.g. keratoconus. keratoglobus, pellucid marginal degeneration, Salzmann's Nodular Degeneration), dystrophy, granular corneal dystrophy, Lattice Corneal Dystrophy), post-surgery (e.g. corneal transplant, LASIK, radial keratotomy), and corneal scarring. The lens may also be prescribed for the management of ocular surface diseases (e.g. dry eye syndrome, Keratoconjunctivitis Sicca (Graft vs Host Disease, Sjogren's syndrome, Filamentary Keratitis), limbal stem cell deficiency, epidermal ocular disorders, neurotrophic keratitis, and corneal exposure/lagophthalmos). When prescribed for therapeutic use, the Custom Stable RGP Scleral Lenses is also indicated for correction of refractive error in persons with myopia, hyperopia or presbyopia. Eyecare practitioners may prescribe the lenses for frequent/planned replacement wear, with cleaning, disinfection and scheduled replacement. When prescribed for frequent wear, the lens may be cleaned and disinfected using a chemical (not heat) lens care system.
Device Story
Large diameter rigid gas permeable (RGP) scleral contact lens; vaults over cornea; rests on conjunctiva overlying sclera. Lathe-cut from hydrophobic fluoro-silicone acrylate materials (roflufocon D or E). Used by eyecare practitioners for daily wear; frequent/planned replacement. Provides expanded tear reservoir; protects ocular surface; corrects refractive error. Clinical benefit: management of irregular corneal shapes and ocular surface diseases. Shipped dry or wet in multipurpose solution.
Clinical Evidence
No new clinical trials conducted. Safety and effectiveness supported by reference to predicate device (K033594) and a series of clinical case reports demonstrating performance for proposed indications.
Technological Characteristics
Materials: Roflufocon D and Roflufocon E (Group 3 Fluoro Silicone Acrylate). Lathe-cut manufacturing. Dimensions: 14.8-17.8 mm diameter, 0.20-0.40 mm center thickness. Refractive index ~1.43. Water content <1%. UV blocking >98% UVB, >95% UVA. Oxygen permeability (Dk) 100-125 x 10^-11. ANSI Z80.20 compliant.
Indications for Use
Indicated for patients with irregular corneal shape (keratoconus, keratoglobus, pellucid marginal degeneration, Salzmann's Nodular Degeneration, dystrophies, post-surgical corneal conditions, corneal scarring) or ocular surface diseases (dry eye, Keratoconjunctivitis Sicca, limbal stem cell deficiency, epidermal ocular disorders, neurotrophic keratitis, corneal exposure/lagophthalmos). Also indicated for refractive error correction (myopia, hyperopia, presbyopia) in therapeutic contexts.
Regulatory Classification
Identification
A rigid gas permeable contact lens is a device intended to be worn directly against the cornea of the eye to correct vision conditions. The device is made of various materials, such as cellulose acetate butyrate, polyacrylate-silicone, or silicone elastomers, whose main polymer molecules generally do not absorb or attract water.
Optimum GP (roflufocon A,B,C,D&E) Daily Wear Contact Lens (K033594)
Submission Summary (Full Text)
{1}------------------------------------------------
Image /page/0/Picture/2 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, stacked on top of each other.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
March 24, 2017
Valley Contax, Inc. % Mr. Bret Andre Principal Consultant EyeReg Consulting, Inc. 6119 Canter Ln. West Linn. OR 97068
Re: K170335
Trade/Device Name: Custom Stable TM Rigid Gas Permeable Scleral Contact Lens Regulation Number: 21 CFR 886.5916 Regulation Name: Rigid Gas Permeable Contact Lens Regulatory Class: Class II Product Code: HOD Dated: January 30, 2017 Received: February 2, 2017
Dear Mr. Andre:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{2}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies.
You must comply with all the Act's requirements, including, but not limited to: registration and listing
(21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours.
# Denise L. Hampton -S
for Malvina B. Eydelman, M.D. Director Division of Ophthalmic, and Ear, Nose and Throat Devices Office of Device Evaluation Center of Devices and Radiological Health
Enclosure
{3}------------------------------------------------
#### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
## Indications for Use
510(k) Number (if known)
### K170335
Device Name
Custom Stable(TM) Rigid Gas Permeable Scleral Contact Lenses
#### Indications for Use (Describe)
The Custom Stable(TM) Rigid Gas Permeable Scleral Contact Lenses for daily wear are indicated for use for the management of multiple ocular conditions, such as, degenerations that lead to an irregular corneal shape (e.g. keratoconus. keratoglobus, pellucid marginal degeneration, Salzmann's Nodular Degeneration), dystrophy, granular corneal dystrophy, Lattice Corneal Dystrophy), post-surgery (e.g. corneal transplant, LASIK, radial keratotomy), and corneal scarring. The lens may also be prescribed for the management of ocular surface diseases (e.g. dry eye syndrome, Keratoconjunctivitis Sicca (Graft vs Host Disease, Sjogren's syndrome, Filamentary Keratitis), limbal stem cell deficiency, epidermal ocular disorders, neurotrophic keratitis, and corneal exposure/lagophthalmos). When prescribed for therapeutic use, the Custom Stable RGP Scleral Lenses is also indicated for correction of refractive error in persons with myopia, hyperopia or presbyopia.
Eyecare practitioners may prescribe the lenses for frequent/planned replacement wear, with cleaning, disinfection and scheduled replacement. When prescribed for frequent wear, the lens may be cleaned and disinfected using a chemical (not heat) lens care system.
Type of Use (Select one or both, as applicable) Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C)
### PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.
### FOR FDA USE ONLY
Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)
Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement below.
{4}------------------------------------------------
This section applies only to requirements of the Paperwork Reduction Act of 1995.
## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{5}------------------------------------------------
# 510 (k) SUMMARY OF SAFETY AND EFFECTIVENESS
This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92.
The assigned 510(k) number is:
# Applicant information:
| Date Prepared: | January 30th, 2017 |
|---------------------------|----------------------------------------------|
| Name: | Valley Contax, Inc. |
| Address | 200 South Mill St. Springfield, Oregon 97477 |
| Contact Person: | Josh Adams |
| | Vice President |
| Phone number: | (541) 744-9393 |
| Consultant/Correspondent: | EyeReg Consulting, Inc. Bret Andre |
| Phone number | (503) 372-5226 |
| Fax number | (503) 419-4475 |
## Device Information:
| Device Classification: | Class II |
|------------------------|-------------------------------------------------------------|
| Classification Number: | HQD |
| Classification Name: | Lenses, Rigid Gas Permeable, Daily Wear |
| Trade Name: | Custom StableTM Rigid Gas Permeable<br>Scleral Contact Lens |
{6}------------------------------------------------
# Predicate Devices:
The Custom Stable "1" Rigid Gas Permeable Scleral Contact Lens is substantially equivalent to the following predicate device(s)
- "BostonSight PD Prosthetic Device" . - Primary Predicate Manufactured by Boston Foundation for Sight 510(k) number; K161461
- "Optimum GP (roflufocon A,B,C.D&E) Daily Wear Contact Lens" . - Reference Predicate Manufactured by Contamac Ltd. 510(k) number; K033594
# Device Description:
The Custom Stable "11 Rigid Gas Permeable Scleral Contact Lens for daily wear is a large diameter rigid gas permeable contact lens design that vaults over the cornea and rests on the conjunctiva overlying the sclera. The Custom Stable "" Rigid Gas Permeable Scleral Contact Lens is lathe cut from one of the following hydrophobic, fluoro-silicone acrylate materials:
- · roflufocon D (supplied by Contamac Ltd.)
- · roflufocon E (supplied by Contamac Ltd.)
The physical properties of the Custom Stable" Rigid Gas Permeable Scleral Contact Lens are as follows:
| | ROFLUFOCON D | ROFLUFOCON E |
|----------------------------------------------------------------------------------------------------------|------------------------------------------------------------------|------------------------------------------------------------------|
| Refractive Index | 1.4333 | 1.4332 |
| Light Transmission (clear) | >97% | >97% |
| Light Transmission (tinted) | >90% | >90% |
| Water Content | <1% | <1% |
| Dynamic Contact Angle (Receding) | 3° | 6° |
| Specific Gravity | 1.166 | 1.155 |
| Modulus | 697 MPa | 77 MPa |
| Shore D Hardness | 75 | 77 |
| Oxygen Permeability<br>(Dk) | $100 \times 10^{-11}$ (cm²/sec) (ml O₂/ml x mm Hg @ 35°C) | $125 \times 10^{-11}$ (cm²/sec) (ml O₂/ml x mm Hg @ 35°C) |
| ISO/FATT Method | | |
| contain one or more of the following<br>color additives conforming to:<br>21 CFR Part 73 & 74, Subpart D | D & C Green No. 6, FD & C Red<br>No. 17,<br>CI Solvent Yellow 18 | D & C Green No. 6, FD & C Red<br>No. 17,<br>CI Solvent Yellow 18 |
| UV Light Blocking<br>(UVB - 280nm – 315nm; UVA 316nm – 380nm) | >98% UVB<br>>95% UVA | >98% UVB<br>>95% UVA |
{7}------------------------------------------------
The parameters for the Custom Stable™ Rigid Gas Permeable Scleral Contact Lens are as follows:
- Chord Diameter: ન્ન્સ
- Center Thickness: 米
- 米 Base Curve:
- 米 Spherical Powers:
14.8 mm to 17.8 mm 0.20 mm to 0.40 mm 6.6 mm to 11.0 mm -30.00 Diopters to +30.00 Diopters (0.125 Diopter steps)
The Custom Stable"" Rigid Gas Permeable Scleral Contact Lens may be shipped "dry" or "wet". The primary container for shipping the Custom Stable™ Rigid Gas Permeable Scleral Contact Lens is the Bonasse Flat Bed Soaking Case, Model SC 106-with 510(k) clearance under K991206. When shipped "wet", The Custom Stable™ Rigid Gas Permeable Scleral Contact Lens is packaged and shipped in the Menicon Unique pH multipurpose solution.
# Indication for Use:
The Custom Stable "11 Rigid Gas Permeable Scleral Contact Lenses for daily wear are indicated for use for the management of multiple ocular conditions, such as, degenerations that lead to an irregular corneal shape (e.g. keratoconus, keratoglobus, pellucid marginal degeneration, Salzmann's Nodular Degeneration), dystrophies (e.g. Cogan's dystrophy, granular corneal dystrophy, Lattice Corneal Dystrophy), post-surgery (e.g. corneal transplant, LASIK, radial keratotomy), and corneal scarring. The lens may also be prescribed for the management of ocular surface diseases (e.g. dry eye syndrome, Keratoconjunctivitis Sicca (Graft vs Host Disease, Sjogren's syndrome, Filamentary Keratitis), limbal stem cell deficiency, epidermal ocular disorders, neurotrophic keratitis, and corneal exposure/lagophthalmos). When prescribed for therapeutic use, the Custom Stable RGP Scleral Lenses is also indicated for correction of refractive error in persons with myopia, hyperopia or presbyopia.
Eyecare practitioners may prescribe the lenses for frequent/planned replacement wear, with cleaning, disinfection and scheduled replacement. When prescribed for frequent/planned replacement wear, the lens may be cleaned and disinfected using a chemical (not heat) lens care system.
# Description of Testing:
- ~ Non-Clinical Studies ~
Non-clinical testing to demonstrate the safety and effectiveness of contact lenses manufactured from roflufocon D and roflufocon E has been addressed by reference to 510(k) K033594-Optimum GP (roflufocon A.B.C.D&E) Daily Wear Contact Lens (the predicate device).
Additionally, the following testing was performed on the finished Custom Stable™ Rigid Gas Permeable Scleral Contact Lenses:
Bench Testing-manufacturing verification testing was conducted to demonstrate the ability of Valley Contac, Inc. to manufacture lenses, on a repeatable basis, from supplied lens blanks to a variety of prescribed parameters. All lenses were manufactured to established finished product specifications within the ANSI Z80.20 tolerance.
{8}------------------------------------------------
Bioburden Testing-bioburden testing conducted on rigid gas permeable lenses manufactured at Valley Contax, Inc. demonstrated that the colony forming units (CFU) per lens was less than 1, which is within the established acceptance criteria of less than 100 CFU per lens.
- ~ Clinical Studies ~
Clinical testing to demonstrate the safety and effectiveness of contact lenses manufactured from roflufocon D and roflufocon E has been addressed by reference to 510(k) K033594-Optimum GP (roflufocon A,B,C,D&E) Daily Wear Contact Lens (the predicate device).
Additionally, a series of clinical case reports demonstrate the performance of the Custom Stable" Rigid Gas Permeable Scleral Contact Lens when prescribed for the proposed indications for use.
~ Conclusions Drawn from Testing ~
Testing presented in this submission for the Custom Stable™ Rigid Gas Permeable Scleral Contact Lenses demonstrate no significant differences from the predicate devices-supporting the substantial equivalence claim.
# Substantial Equivalence:
The Custom Stable "" Rigid Gas Permeable Scleral Contact Lens is substantially equivalent to the predicate device(s), and does not raise different questions of safety and effectiveness.
The Custom Stable "" Rigid Gas Permeable Scleral Contact Lens is substantially equivalent to the BostonSight PD Prosthetic Device (predicate device) in the following key areas:
- Components/Materials/Formulation (roflufocon D & E contact lens materials) ●
- FDA classification and group number (Class 2, Group #3 Fluoro Silicone Acrylate)
- Lathe cut manufacturing process
- Large diameter (scleral) design
- Actions and intended use
- . Therapeutic indications for use
{9}------------------------------------------------
The following table depicts the characteristics of the Custom Stable™ Rigid Gas Permeable Scleral Contact Lens, as well as the predicate device(s).
| | Custom Stable™ Rigid Gas<br>Permeable Scleral Contact Lens | BostonSight PD Prosthetic<br>Device<br>K161461 | Optimum GP (roflufocon<br>A,B,C,D&E) Daily Wear<br>Contact Lens<br>K033594 |
|----------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | Subject Device | Predicate Device | Predicate Device |
| Indication for Use | Indicated for use for the management of<br>multiple ocular conditions, such as,<br>degenerations that lead to an irregular corneal<br>shape (e.g. keratoconus, keratoglobus,<br>pellucid marginal degeneration, Salzmann's<br>Nodular Degeneration), dystrophies (e.g.<br>Cogan's dystrophy, granular corneal<br>dystrophy, Lattice Corneal Dystrophy), post-<br>surgery (e.g. corneal transplant, LASIK,<br>radial keratotomy), and corneal scarring. The<br>lens may also be prescribed for the<br>management of ocular surface diseases (e.g.<br>dry eye syndrome, Keratoconjunctivitis Sicca<br>(Graft vs Host Disease, Sjogren's syndrome,<br>Filamentary Keratitis), limbal stem cell<br>deficiency, epidermal ocular disorders,<br>neurotrophic keratitis, and corneal<br>exposure/lagophthalmos). When prescribed<br>for therapeutic use, the Custom Stable RGP<br>Scleral Lenses is also indicated for correction<br>of refractive error in persons with myopia,<br>hyperopia or presbyopia. | Indicated for therapeutic use in eyes<br>with ocular surface disease from dry<br>eye (e.g. ocular Graft-versus-Host<br>disease, Sjögren's syndrome, dry eye<br>syndrome), limbal stem cell deficiency<br>(e.g.Stevens-Johnson syndrome,<br>chemical and thermal burns, radiation),<br>disorders of the skin (e.g. atopy,<br>ectodermal dysplasia), neurotrophic<br>keratitis (e.g. Herpes simplex, Herpes<br>zoster, Familial Dysautonomia), and<br>corneal exposure (e.g. anatomic,<br>paralytic) that might benefit from the<br>presence of an expanded tear reservoir<br>and protection against an adverse<br>environment. When prescribed for<br>therapeutic use for distorted cornea or<br>ocular surface disease, the BostonSight<br>PD Prosthetic Device may incidentally<br>provide correction of refractive error. | May be prescribed<br>in otherwise non-diseased<br>eyes that require<br>a rigid gas permeable lens for<br>the management<br>of irregular corneal conditions<br>such as; keratoconus, pellucid<br>marginal degeneration<br>of following penetrating<br>keratoplasty or following<br>refractive (e.g. LASIK)<br>surgery. |
| Device and<br>Classification | Class II<br>Lenses, Rigid Gas Permeable, Daily<br>Wear<br>HQD | Class II<br>Lenses, Rigid Gas Permeable,<br>Daily Wear<br>HQD | Class II<br>Lenses, Rigid Gas Permeable.<br>Daily Wear<br>HQD |
| Production Method | Lathe-cut | Lathe-cut | Lathe-cut |
| USAN | Roflufocon D<br>Roflufocon E | Roflufocon D<br>Roflufocon E<br>Oprifocon A<br>Hexafocon B | Roflufocon A<br>Roflufocon B<br>Roflufocon C<br>Roflufocon D<br>Roflufocon E |
| FDA Group # | Group # 3 Fluoro Silicone Acrylate | Group # 3 Fluoro Silicone<br>Acrylate | Group # 3 Fluoro Silicone<br>Acrylate |
| Water Content | <1% | <1% | <1% |
| UV Absorber/Blocker<br>available | YES | YES | YES |
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.