DIGITAL RADIOGRAPHY CXDI-710C WIRELESS, DIGITAL RADIOGRAPHY CXDI-810C WIRELESS
K170332 · Canon, Inc. · MQB · Mar 1, 2017 · Radiology
Device Facts
Record ID
K170332
Device Name
DIGITAL RADIOGRAPHY CXDI-710C WIRELESS, DIGITAL RADIOGRAPHY CXDI-810C WIRELESS
Applicant
Canon, Inc.
Product Code
MQB · Radiology
Decision Date
Mar 1, 2017
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.1680
Device Class
Class 2
Indications for Use
The DIGITAL RADIOGRAPHY CXDI-710C Wireless and CXDI-810C Wireless provides digital image capture for conventional film/screen radiographic examinations. This device is intended to capture, for display, radiographic images of human anatomy, and to replace radiographic film/screen systems in all general purpose diagnostic procedures. This device is not intended for mammography applications.
Device Story
Solid state X-ray detectors (CXDI-710C/810C) capture X-ray photons using a CsI(Tl) scintillator; converts photons to visible light; light illuminates photo-detector array to generate electrical signals; signals converted to digital values. Used in general radiography; operated by clinicians. Features wireless or wired communication to control console; includes internal nonvolatile memory for temporary image storage; supports Non-Generator Connection Mode (detects irradiation without direct electrical connection). Reduced weight compared to predicates; improved ingress protection. Output displayed on monitors for diagnostic review by healthcare providers to inform clinical decision-making.
Clinical Evidence
Bench testing only. Included verification/validation testing to internal functional specifications, software validation (moderate Level of Concern), and non-clinical image comparisons between subject and predicate devices. Compliance demonstrated with IEC 60601-1, 60601-1-6, and 60601-2-54 standards.
Indicated for digital image capture for conventional film/screen radiographic examinations of human anatomy in general purpose diagnostic procedures. Not for mammography.
Regulatory Classification
Identification
A stationary x-ray system is a permanently installed diagnostic system intended to generate and control x-rays for examination of various anatomical regions. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
Special Controls
*Classification.* Class II (special controls). A radiographic contrast tray or radiology diagnostic kit intended for use with a stationary x-ray system only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo is a circular emblem with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, with flowing lines suggesting movement or connection.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
March 1, 2017
Canon, Inc. % Ms. Diane Rutherford Submissions Manager Ken Block Consulting 1201 Richardson Drive, Suite 160 RICHARDSON TX 75080
Re: K170332
Trade/Device Name: Digital Radiography CXDI-710C Wireless Digital Radiography CXDI-810C Wireless Regulation Number: 21 CFR 892.1680 Regulation Name: Stationary x-ray system Regulatory Class: II Product Code: MOB Dated: January 6, 2017 Received: February 2, 2017
Dear Ms. Rutherford:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{1}------------------------------------------------
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Michael D'Hara
For
Robert Ochs, Ph.D. Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
#### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
## Indications for Use
Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement below.
510(k) Number (if known)
K170332
Device Name DIGITAL RADIOGRAPHY CXDI-710C Wireless DIGITAL RADIOGRAPHY CXDI-810C Wireless
#### Indications for Use (Describe)
The DIGITAL RADIOGRAPHY CXDI-710C Wireless and CXDI-810C Wireless provides digital image capture for conventional film/screen radiographic examinations. This device is intended to capture, for display, radiographic images of human anatomy, and to replace radiographic film/screen systems in all general purpose diagnostic procedures. This device is not intended for mammography applications.
Type of Use (Select one or both, as applicable)
2 Prescription Use (Part 21 CFR 801 Subpart D)
_ Over-The-Counter Use (21 CFR 801 Subpart C)
### PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.
#### FOR FDA USE ONLY
Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the word "Canon" in a bold, red font. The letters are slightly blurred, giving the image a soft focus. The word is presented against a plain white background, which makes the red color stand out.
# 5. 510(k) SUMMARY
| Submitter: | Canon, Inc. - Medical Equipment Group<br>9-1, Imaikami-cho, Nakahara-ku<br>Kawasaki, Kanagawa 211-8501 Japan | |
|----------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Contact Person: | Mr. Shinji Mori<br>Manager<br>TEL: 81-3-3758-2111<br>FAX: 81-44-739-6493<br>mori.shinji@canon.co.jp | |
| Date Prepared: | January 26, 2017 | |
| Proposed Device | Manufacturer:<br>Trade Name:<br>Common Name:<br>Classification Name:<br>Product Code /<br>Regulatory Standard: | Canon<br>DIGITAL RADIOGRAPHY CXDI-710C Wireless<br>DIGITAL RADIOGRAPHY CXDI-810C Wireless<br>Solid State X-ray Imager<br>Stationary X-ray system<br>MQB<br>892.1680 Stationary X-ray System |
| Predicate Device: | Clearance:<br>Manufacturer:<br>Trade Name:<br>Common Name:<br>Classification Name:<br>Product Code /<br>Regulatory Standard: | K131106 dated July 03, 2013<br>Canon<br>DIGITAL RADIOGRAPHY CXDI-701C Wireless<br>DIGITAL RADIOGRAPHY CXDI-801C Wireless<br>Solid State X-ray Imager<br>Stationary X-ray system<br>MQB<br>892.1680 Stationary X-ray System |
| Device Description: | The two models of detectors included in this submission are solid state x-ray imagers.<br>Model CXDI-710C Wireless has an approximate imaging area of 35.0 x 42.6 cm,<br>while model CXDI-810C Wireless has an approximate imaging area of 35.0 x 27.4 cm.<br><br>For both models, the detector intercepts x-ray photons and the scintillator emits<br>visible spectrum photons that illuminate an array of photo-detectors that create<br>electrical signals. After the electrical signals are generated, the signals are converted<br>to digital values and the images will be displayed on monitors. The digital value can<br>be communicated to the operator console via wiring connection or wireless.<br><br>For the proposed models, temporary image storage is now possible and the detector<br>weight has been reduced from that of the predicates. The proposed models have<br>increased protection against ingress, continue to include the Non-Generator<br>Connection Mode (detection of x-ray irradiation without direct electrical connection<br>to the x-ray generator) and are compatible with the Scatter Correction feature. | |
| Indications for Use: | The DIGITAL RADIOGRAPHY CXDI-710C Wireless and CXDI-810C Wireless<br>provide digital image capture for conventional film/screen radiographic<br>examinations. These devices are intended to capture, for display, radiographic images<br>of human anatomy, and to replace radiographic film/screen systems in all general<br>purpose diagnostic procedures. These devices are not intended for mammography<br>applications. | |
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the word "Canon" in red font. The font is bold and slightly italicized. The word is easily readable and takes up most of the image.
## 5. 510(k) SUMMARY (continued)
Summary of Technological Characteristics: Comparisons with the predicate devices show the technological characteristics of the proposed DIGITAL RADIOGRAPHY CXDI-710C Wireless and DIGITAL RADIOGRAPHY CXDI-810C Wireless devices to be substantially equivalent to the predicate devices. The proposed devices are functionally identical to the predicate devices.
The major difference between the proposed and predicate devices is that the proposed devices have a reduction in weight and can save captured in an internal nonvolatile memory of the detector. No other technological changes have been made to the proposed devices.
| | New Devices: K170332<br>CXDI-710C / 810C Wireless | Predicate Device: K131106<br>CXDI-701C / 801C Wireless |
|------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Indication for Use | The DIGITAL RADIOGRAPHY CXDI-710C<br>Wireless and CXDI-810C Wireless provides<br>digital image capture for conventional<br>film/screen radiographic examinations.<br>This device is intended to capture, for<br>display, radiographic images of human<br>anatomy, and to replace radiographic<br>film/screen systems in all general purpose<br>diagnostic procedures. This device is not<br>intended for mammography applications. | The DIGITAL RADIOGRAPHY CXDI-701G<br>Wireless, CXDI-701C Wireless, CXDI-801G<br>Wireless and CXDI-801C Wireless provide<br>digital image capture for conventional<br>film/screen radiographic examinations.<br>These devices are intended to replace<br>radiographic film/screen systems in all<br>general purpose diagnostic procedures.<br>These devices are not intended for<br>mammography applications. |
| Application | General Radiography | General Radiography |
| Scintillator | CsI(TI)<br>[Cesium lodide doped with Thallium] | CsI(TI)<br>[Cesium lodide doped with Thallium] |
| Pixel Pitch | 125 µm | 125µm |
| Pixels | 710C: 2,800 x 3,408 (~ 9.5 mil)<br>810C: 2,800 x 2,192 (~ 6.1 mil) | 701C: 2,800 x 3,408 (~ 9.5 mil)<br>801C: 2,192 x 2,800 (~ 6.1 mil) |
| External<br>Dimensions | 710C: 384 x 460 x 15.7 mm<br>810C: 307.5 x 384 x 15.7 mm | 701C: 384 x 460 x 15 mm<br>801C: 307 x 384 x 15 mm |
| Weight | 710C: 2.3 kg<br>810C: 1.8 kg | 701C: 3.3 kg<br>801C: 2.3 kg |
| Spatial Resolution | 35% [MTF@2lp/mm] | 35% [MTF@2lp/mm] |
| Control SW | CXDI Control Software<br>Added Standalone function | CXDI Control Software |
| Device FW | PCA-FE-710<br>Added Standalone function<br>Added wireless channels with DFS and<br>TPC functions | PCA-FE-701 |
| Wireless Functions | Communication between Detector and:<br>Multi Box<br>Control PC | Communication between Detector and:<br>X-ray I/F Unit<br>Control PC |
The User's and Installation Manuals provide detailed instructions and information for safe and effective use of the device and users are expected to adhere to the instructions and other information. The User's Manual explains how to use the detector and other equipment. Connected medical equipment, such as X-ray generators, must comply with IEC 60601-1. Before using the product, be sure to read the manual thoroughly in order to utilize it more effectively.
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the word "Canon" in a bold, red font. The letters are slightly stylized, with a slight curve to the top of the "C" and a distinctive shape to the "n". The word is presented against a white background, making the red color stand out.
Summary of Non-Clinical / Test Data:
Tests were performed on the models which demonstrated that the device is safe and effective, performs comparably to the predicate device(s), and is substantially equivalent to the predicate device(s). Tests included verification/validation testing to internal functional specifications (including software) and non-clinical image comparisons involving flat panel display images taken with the new device and the predicate device(s). Documentation was provided demonstrating compliance of the CXDI-710C Wireless and CXDI-810C Wireless to all FDA requirements stated in Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices for a moderate LOC, including results of verification/validation plus traceability of verification/validation tests to software requirements and software risk hazards. Other FDA guidance documents used in development include Radio Frequency Wireless Technology in Medical Devices and Content of Premarket Submissions for Management of Cybersecurity in Medical Devices.
Documentation was provided demonstrating that the CXDI-710C / 810C Wireless complies with the FDA requirements stated in Guidance for the Submission of 510(k)'s for Solid State X-ray Imaging Devices. The evaluations of the CXDI-710C / 810C Wireless compared to the CXDI-701C / 801C Wireless, show the CXDI-710C / 810C Wireless to be equivalent to the CXDI-701C / 801C Wireless.
Testing confirmed that the CXDI-710C Wireless and CXDI-810C Wireless complies with the U.S. Performance Standard for radiographic equipment and with relevant voluntary safety standards for Electrical safety and Electromagnetic Compatibility testing, specifically IEC standards 60601-1. 60601-1-6, and 60601-2-54.
Together, these verification/validation activities successfully demonstrated that the CXDI-710C Wireless and CXDI-810C Wireless correctly performs as designed, has been validated for its intended use, and raises no new questions regarding either safety or effectiveness when compared to the predicate device(s). Therefore, the verification/validation_testing_conducted_supports_a_determination_of_substantial equivalence for the CXDI-710C Wireless and CXDI-810C Wireless device.
Conclusion: Canon, Inc. – Medical Equipment Group considers the DIGITAL RADIOGRAPHY CXDI-710C Wireless and DIGITAL RADIOGRAPHY CXDI-810C Wireless devices to be substantially equivalent to the predicate devices listed above. This conclusion is based on the similarities in primary intended use, principles of operation, functional design, and established medical use.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.