Alinity c ICT Sample Diluent
K170320 · Abbott Laboratories · CGZ · Oct 24, 2017 · Clinical Chemistry
Device Facts
| Record ID | K170320 |
| Device Name | Alinity c ICT Sample Diluent |
| Applicant | Abbott Laboratories |
| Product Code | CGZ · Clinical Chemistry |
| Decision Date | Oct 24, 2017 |
| Decision | SESE |
| Submission Type | Traditional |
| Regulation | 21 CFR 862.1170 |
| Device Class | Class 2 |
Intended Use
The Alinity c ICT (Integrated Chip Technology) is used for the quantitation of sodium, potassium, and chloride in human serum, plasma, or urine on the Alinity c analyzer. Sodium measurements are used in the diagnosis and treatment of aldosteronism (excessive secretion of the hormone aldosterone), diabetes insipidus (chronic excretion of large amounts of dilute urine, accompanied by extreme thirst), adrenal hypertension, Addison's disease (caused by destruction of the adrenal glands), dehydration, inappropriate antidiuretic hormone secretion, or other diseases involving electrolyte imbalance. Potassium measurements are used to monitor electrolyte balance in the diagnosis and treatment of diseases conditions characterized by low or high blood potassium levels. Chloride measurements are used in the diagnosis and treatment of electrolyte and metabolic disorders such as cystic fibrosis and diabetic acidosis.
Device Story
Alinity c ICT Sample Diluent is a reagent used on the Alinity c analyzer for indirect potentiometric measurement of sodium, potassium, and chloride ions in human serum, plasma, or urine. The device utilizes ion-selective electrodes (ISE) with ion-specific membranes; an electrical potential develops across the membranes relative to a reference electrode, following the Nernst equation. The analyzer compares this voltage against stored calibrator values to calculate ion concentration. Used in clinical laboratory settings by trained technicians. Output is a quantitative electrolyte concentration value, which clinicians use to diagnose and monitor metabolic and electrolyte disorders. The device benefits patients by providing accurate electrolyte status for clinical decision-making.
Clinical Evidence
Bench testing only. Precision evaluated per CLSI EP05-A2; linearity per EP06-A; interference per EP07-A2; method comparison per EP09-A3. Method comparison against ARCHITECT analyzer showed high correlation (r=1.00) for all analytes across serum and urine. No clinical data presented.
Technological Characteristics
Indirect ion-selective electrode (ISE) potentiometric sensing. Reagent 1 contains buffer. Container: Polypropylene with black colorant; closure: HDPE. Connectivity: Integrated with Alinity c analyzer. Software: Embedded firmware for signal processing and Nernst equation calculation.
Indications for Use
Indicated for quantitation of sodium, potassium, and chloride in human serum, plasma, or urine for diagnosis/treatment of electrolyte/metabolic disorders including aldosteronism, diabetes insipidus, adrenal hypertension, Addison's disease, dehydration, inappropriate ADH secretion, cystic fibrosis, and diabetic acidosis.
Regulatory Classification
Identification
A chloride test system is a device intended to measure the level of chloride in plasma, serum, sweat, and urine. Chloride measurements are used in the diagnosis and treatment of electrolyte and metabolic disorders such as cystic fibrosis and diabetic acidosis.
Predicate Devices
- ICT (Na+, K+, Cl-) Sample Diluent (k980367)
Related Devices
- K222438 — A-LYTE® Integrated Multisensor (IMT Na K Cl) · Siemens Healthcare Diagnostics, Inc. · Jul 25, 2023
- K990036 — CARESIDE LYTES, CARESIDE ANALLYZER MODEL 1000 · Careside, Inc. · Mar 18, 1999
- K051646 — ISE DILUENT, 1N KCL, INTERNAL REFERENCE AND ISE STANDARDS LOW AND HIGH · Pointe Scientific, Inc., · Oct 7, 2005
- K070531 — ISE MODULE AND BAR CODE READER ADDITION FOR THE DATAPRO CLINICAL CHEMISTRY ANALYZER · Thermo Fisher Scientific · Sep 4, 2007
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, with flowing lines beneath them.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002 October 24, 2017
ABBOTT LABORATORIES ZAMAN KHAN ASSOCIATE DIRECTOR, REGULATORY AFFAIRS DEPT. 09AA. BLDG. CP01-3 100 ABBOTT PARK ROAD ABBOTT PARK IL 60064
Re: K170320
Trade/Device Name: Alinity c ICT Sample Diluent Regulation Number: 21 CFR 862.1170 Regulation Name: Chloride test system Regulatory Class: II Product Code: CGZ, CEM, JGS Dated: September 12, 2017 Received: September 13, 2017
Dear Zaman Khan:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{1}------------------------------------------------
If you desire specific advice for your device on our labeling regulations (21 CFR Parts 801 and 809), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
# Kellie B. Kelm -S
for Courtney H. Lias, Ph.D. Director Division of Chemistry and Toxicology Devices Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) k170320
Device Name Alinity c ICT Sample Diluent
#### Indications for Use (Describe)
The Alinity c ICT (Integrated Chip Technology) is used for the quantitation of sodium, and chloride in human serum, plasma, or urine on the Alinity c analyzer.
Sodium measurements are used in the diagnosis and treatment of aldosteronism (excessive secretion of the hormone aldosterone), diabetes insipidus (chronic excretion of large amounts of dilute urine, accompanied by extreme thirst), adrenal hypertension, Addison's disease (caused by destruction of the adrenal glands), dehydration, inappropriate antidiuretic hormone secretion, or other diseases involving electrolyte imbalance.
Potassium measurements are used to monitor electrolyte balance in the diagnosis and treatment of diseases conditions characterized by low or high blood potassium levels.
Chloride measurements are used in the diagnosis and treatment of electrolyte and metabolic disorders such as cystic fibrosis and diabetic acidosis.
| Type of Use (Select one or both, as applicable) | |
|----------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------|
| <span style="font-family: sans-serif;">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) | <span style="font-family: sans-serif;">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
## k170320 510(k) Summary (Summary of Safety and Effectiveness)
This summary of the 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92.
# I. Applicant Name
Abbott Laboratories Dept. 09AA, Bldg CP1 100 Abbott Park Road Abbott Park, IL 60064
Primary contact person for all communications:
Zaman Khan ADD, Associate Director, Regulatory Affairs Phone: (224) 668-2002 Fax: (224) 667-4836 Email: Zaman.Khan@abbott.com
Secondary contact person for all communications:
Amy Ghering, Ph.D. ADD, Associate Director, Regulatory Affairs Phone: (224) 668-6934 Fax: (224) 667-4836 Email: Amy.Ghering@abbott.com
Date Summary Prepared: January 31, 2017
Date Summary Revised: October 19, 2017
{4}------------------------------------------------
# II. Device Name
# Alinity c ICT Sample Diluent
Device Classification: Class II Classification Name: Electrode, ion-specific, chloride/potassium/sodium Governing Regulation: 862.1170/862.1600/862.1665 Product Code: CGZ/CEM/JGS
# III. Predicate Device
Reagent
ICT (Na*, K†, Cl´) Sample Diluent (k980367), List No. (LN) 2P32 for use on Abbott ARCHITECT c8000 and Abbott AEROSET
# IV. Description of Device
# A. Alinity c ICT Sample Diluent
#### Kit Contents
Volumes (mL) listed in the table below indicate the volume per cartridge.
| Component | 07P5320 |
|------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Tests per cartridge | તે તે જેવી સવલતો પ્રાપ્ય થયેલી છે. આ ગામનાં પ્રાથમિક શાળા, પંચાયતઘર, આંગણવાડી તેમ જ દૂધની ડેરી જેવી સવલતો પ્રાપ્ય થયેલી છે. આ ગામનાં પ્રાથમિક શાળા, પંચાયતઘર, આંગણવાડી તેમ જ |
| Number of cartridges per kit | 10 |
| Test per kit | 9350 |
| Reagent 1 (R1) | 68.2 mL |
| Reagent | Active Ingredient | Concentration |
|-----------|-------------------|---------------|
| Reagent 1 | Buffer | NA |
{5}------------------------------------------------
# B. Calibrators
### 1. The Alinity c ICT Serum Calibrator contains:
5 Bottles (2.9 mL each) of Alinity c ICT Serum Calibrators. Calibrators L and H contain Bovine Serum Albumin, Sodium chloride, Potassium chloride, and Sodium nitrate.
| | Concentration | | |
|------------|---------------|------------|------------|
| Calibrator | Sodium | Potassium | Chloride |
| L | 120 mmol/L | 3.4 mmol/L | 80 mmol/L |
| H | 160 mmol/L | 8.0 mmol/L | 120 mmol/L |
Calibrators L and H are manufactured gravimetrically using ACS grade sodium chloride, potassium chloride, and sodium nitrate. The concentrations of sodium and potassium are determined using flame photometry calibrated against NIST Standard Reference Material. The concentration of chloride is determined using titration with silver calibrated against NIST Standard Reference Material (NIST - National Institute of Standards and Technology).
# 2. Alinity c ICT Urine Calibrator
5 Bottles (2.9 mL each) of Alinity c ICT Urine Calibrators. Calibrators L and H contain sodium chloride, potassium chloride, and sodium nitrate.
| | Concentration | | |
|-------------|---------------|-------------|------------|
| Calibrator | Sodium | Potassium | Chloride |
| Urine Cal L | 50 mmol/L | 9.0 mmol/L | 50 mmol/L |
| Urine Cal H | 180 mmol/L | 90.0 mmol/L | 180 mmol/L |
Calibrators L and H are manufactured gravimetrically using ACS grade sodium chloride, potassium chloride, and sodium nitrate. The concentrations of sodium, potassium, and chloride are determined using reference method (flame photometry, titration using silver) calibrated against NIST SRM 918 and NIST SRM 919 (NIST - National Institute of Standards and Technology).
{6}------------------------------------------------
#### C. Principles of the Procedure
Ion-selective electrodes (ISE) for sodium, potassium, and chloride utilize membranes selective to each of these ions. An electrical potential (voltage) is developed across the membranes between the reference and measuring electrodes in accordance with the Nernst equation. The voltage is compared to previously determined calibrator voltages and converted into ion concentration.
Methodology: Ion-selective electrode diluted (Indirect)
#### V. Intended Use of the Device
The Alinity c ICT (Integrated Chip Technology) is used for the quantitation of sodium, potassium, and chloride in human serum, plasma, or urine on the Alinity c analyzer.
Sodium measurements are used in the diagnosis and treatment of aldosteronism (excessive secretion of the hormone aldosterone), diabetes insipidus (chronic excretion of large amounts of dilute urine, accompanied by extreme thirst), adrenal hypertension, Addison's disease (caused by destruction of the adrenal glands), dehydration, inappropriate antidiuretic hormone secretion, or other diseases involving electrolyte imbalance.
Potassium measurements are used to monitor electrolyte balance in the diagnosis and treatment of diseases conditions characterized by low or high blood potassium levels.
Chloride measurements are used in the diagnosis and treatment of electrolyte and metabolic disorders such as cystic fibrosis and diabetic acidosis.
#### VI. Comparison of Technological Characteristics
The Alinity c ICT (Integrated Chip Technology) is used for the quantitation of sodium, potassium, and chloride in human serum, plasma, or urine on the Alinity c System.
A comparison of the candidate assay (Alinity c ICT Sample Diluent, List No. 07P53) and the predicate assay (ICT (Na+, K+, Cl ) Sample Diluent, List No. 2P32 on the ARCHITECT System) is presented in Table 1 and Table 2 starting on page 5.
{7}------------------------------------------------
| Table 1: Assay Similarities | | |
|----------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------|
| Characteristics | Candidate Assay<br>Alinity c ICT Sample Diluent (LN 07P53) | Predicate Assay (k980367)<br>ICT (Na+, K+, Cl-) Sample Diluent Assay on<br>the AEROSET/ARCHITECT c8000 System<br>(LN 2P32) |
| | | Technical Characteristics |
| Reagent<br>Formulation | Active Ingredient: Buffer | Same bulk as Alinity c. |
| Analyte<br>Measured | Sodium, potassium, and chloride. | Same |
| Intended<br>Use/Indications<br>for Use | The Alinity c ICT (Integrated Chip Technology) is used for the quantitation of<br>sodium, potassium, and chloride in human serum, plasma, or urine. | Same |
| Assay Principle | Ion-selective electrode diluted (Indirect). | Same |
| Detection of<br>Analyte | Potentiometric. | Same |
| Specimen Type | Human serum, plasma or urine. | Same |
| Performance Characteristics | | |
| Assay Range | <b>Serum/Plasma</b><br>Sodium: 100 to 200 mmol/L.<br>Potassium 1.0 to 10.0 mmol/L.<br>Chloride: 50 to 150 mmol/L.<br><b>Urine</b><br>Sodium: 20 to 400 mmol/L.<br>Potassium: 1.0 to 300.0 mmol/L.<br>Chloride: 20 to 300 mmol/L. | Same |
| Measuring<br>Interval | <b>Serum/Plasma</b><br>The measuring interval for the sodium assay is 100 to 200 mmol/L.<br>The measuring interval for the potassium assay is 1.0 to 10.0 mmol/L.<br>The measuring interval for the chloride assay is 50 to 150 mmol/L.<br><b>Urine</b><br>The measuring interval for the sodium assay is 20 to 400 mmol/L.<br>The measuring interval for the potassium assay is 1.0 to 300.0 mmol/L.<br>The measuring interval for the chloride assay is 20 to 300 mmol/L. | Same |
| Characteristics | Candidate Assay<br>Alinity c ICT Sample Diluent (LN 07P53) | Predicate Assay (k980367)<br>ICT (Na+, K+, Cl-) Sample Diluent Assay on<br>the AEROSET/ARCHITECT c8000 System<br>(LN 2P32) |
| Tube Type<br>Equivalency | Serum<br>Serum tubes (with or without gel barrier). | Same |
| | Plasma<br>Collection tubes<br>Acceptable anticoagulants are:<br>Lithium heparin (with or without gel barrier)<br>Sodium heparin (full draw) | |
| Use of<br>Calibrators | Yes | Same |
| | Use of Controls | Yes, commercially available controls |
# Table 1: Assay Similarities
{8}------------------------------------------------
# Table 2: Assay Differences
| Assay<br>Characteristics | Candidate Assay<br>Alinity c ICT Sample Diluent (LN 07P53) | Predicate Assay (k980367)<br>ICT (Na $+$ , K $+$ , Cl $-$ ) Sample Diluent Assay on<br>the AEROSET/ARCHITECT c8000 System<br>(LN 2P32) |
|---------------------------------------|------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------|
| Technical Characteristics | | |
| Reagent<br>Container | Polypropylene<br>Black colorant | High Density Polyethylene<br>Natural color (Contains no colorant) |
| Closure<br>Material<br>(contact only) | High Density Polyethylene<br>Black color | F217 cap liner<br>Polyethylene Foam between Low-Density<br>Polyethylene liners<br>Green color. |
{9}------------------------------------------------
# VII. Summary of Nonclinical Performance
Within-Laboratory Precision (20-Day)
Alinity c ICT Sample Diluent Sodium assay – Serum Samples
Precision was performed based on guidance from the Clinical and Laboratory Standards Institute (CLSI) document EP05-A2.
A summary of results is presented below:
| Sample | Control<br>Lot | n | Mean<br>(mmol/L) | Within-Run<br>(Repeatability) | | Within-Laboratory (Total)a | |
|--------------------|----------------|-----|------------------|-------------------------------|-----|----------------------------|--------------------|
| | | | | SD | %CV | SD<br>(Rangeb) | %CV<br>(Rangeb) |
| Control<br>Level 1 | 1 | 249 | 125 | 0.5 | 0.4 | 0.9<br>(0.8 - 1.0) | 0.7<br>(0.6-0.8) |
| | 2 | 249 | 125 | 0.5 | 0.4 | 0.9<br>(0.8 - 0.9) | 0.7<br>(0.6 - 0.8) |
| Control<br>Level 2 | 1 | 252 | 144 | 0.5 | 0.4 | 0.9<br>(0.8-1.0) | 0.6<br>(0.6-0.7) |
| | 2 | 249 | 144 | 0.5 | 0.4 | 0.9<br>(0.8-1.0) | 0.6<br>(0.5-0.7) |
| Control<br>Level 3 | 1 | 249 | 161 | 0.5 | 0.3 | 1.0<br>(0.9-1.1) | 0.6<br>(0.6-0.7) |
| | 2 | 249 | 161 | 0.6 | 0.4 | 1.1<br>(0.9-1.2) | 0.7<br>(0.6-0.7) |
| Panel A | N/A | 496 | 112 | 0.4 | 0.4 | 0.8<br>(0.6-0.9) | 0.7<br>(0.6-0.8) |
| Panel B | N/A | 498 | 190 | 0.8 | 0.4 | 1.4<br>(1.2-1.6) | 0.7<br>(0.6-0.9) |
4 Includes within-run, between-run, and between-day variability.
b Maximum and minimum SD or %CV for each reagent lot and instrument combination.
The precision of the Alinity c ICT Sample Diluent Sodium assay was considered acceptable if the within-laboratory (total) imprecision (within-run, between-run, and between-day) was ≤ 1.5 % CV for serum samples targeted between 131.0 to 153.0 mmol/L.
{10}------------------------------------------------
The Alinity c ICT Sample Diluent Sodium assay demonstrated acceptable precision.
Alinity c ICT Sample Diluent Potassium assay – Serum Samples
Precision was performed based on guidance from the Clinical and Laboratory Standards Institute (CLSI) document EP05-A2.
A summary of results is presented below:
| | | | | | Within-Run<br>(Repeatability) | | Within-Laboratory (Total)a | |
|--------------------|----------------|-----|------------------|------|-------------------------------|---------------------|----------------------------|--|
| Sample | Control<br>Lot | n | Mean<br>(mmol/L) | SD | %CV | SD<br>(Rangeb) | %CV<br>(Rangeb) | |
| Control<br>Level 1 | 1 | 246 | 2.8 | 0.03 | 0.9 | 0.04<br>(0.04-0.04) | 1.4<br>(1.3-1.5) | |
| | 2 | 246 | 2.8 | 0.01 | 0.3 | 0.01<br>(0.00-0.02) | 0.4<br>(0.0-0.6) | |
| Control<br>Level 2 | 1 | 246 | 4.0 | 0.02 | 0.6 | 0.03<br>(0.03-0.03) | 0.8<br>(0.8-0.9) | |
| | 2 | 246 | 4.0 | 0.01 | 0.3 | 0.02<br>(0.02-0.02) | 0.5<br>(0.5-0.5) | |
| Control<br>Level 3 | 1 | 246 | 6.8 | 0.03 | 0.4 | 0.05<br>(0.04-0.05) | 0.7<br>(0.6-0.7) | |
| | 2 | 246 | 6.8 | 0.03 | 0.5 | 0.05<br>(0.04-0.05) | 0.7<br>(0.6-0.7) | |
| Panel A | N/A | 491 | 1.6 | 0.02 | 1.1 | 0.03<br>(0.00-0.04) | 1.7<br>(0.0-2.7) | |
| Panel B | N/A | 492 | 9.4 | 0.04 | 0.4 | 0.06<br>(0.05-0.07) | 0.7<br>(0.5-0.8) | |
4 Includes within-run, between-run, and between-day variability.
b Maximum and minimum SD or %CV for each reagent lot and instrument combination.
The precision Alinity c ICT Sample Diluent Potassium assay was considered acceptable if the within-laboratory (total) imprecision (within-run, between-run, and between-day) was ≤ 2.7 % CV for serum samples targeted between 4.0 to 6.0 mmol/L.
The Alinity c ICT Sample Diluent Potassium assay demonstrated acceptable precision.
{11}------------------------------------------------
# Alinity c ICT Sample Diluent Chloride assay – Serum Samples
Precision was performed based on guidance from the Clinical and Laboratory Standards Institute (CLSI) document EP05-A2.
| Sample | Control<br>Lot | n | Mean<br>(mmol/L) | Within-Run<br>(Repeatability) | | Within-Laboratory (Total)a | |
|--------------------|----------------|-----|------------------|-------------------------------|-----|----------------------------|------------------|
| | | | | SD | %CV | SD<br>(Rangeb) | %CV<br>(Rangeb) |
| Control<br>Level 1 | 1 | 243 | 85 | 0.5 | 0.6 | 0.7<br>(0.6-0.8) | 0.8<br>(0.7-0.9) |
| | 2 | 243 | 84 | 0.4 | 0.5 | 0.6<br>(0.5-0.7) | 0.7<br>(0.5-0.9) |
| Control<br>Level 2 | 1 | 243 | 95 | 0.4 | 0.4 | 0.6<br>(0.6-0.7) | 0.7<br>(0.6-0.8) |
| | 2 | 243 | 94 | 0.4 | 0.5 | 0.6<br>(0.5-0.7) | 0.6<br>(0.5-0.7) |
| Control<br>Level 3 | 1 | 243 | 110 | 0.5 | 0.5 | 0.8<br>(0.6-0.9) | 0.7<br>(0.5-0.9) |
| | 2 | 243 | 109 | 0.5 | 0.5 | 0.7<br>(0.5-0.9) | 0.7<br>(0.5-0.8) |
| Panel A | N/A | 486 | 55 | 0.3 | 0.6 | 0.6<br>(0.4-0.6) | 1.0<br>(0.8-1.2) |
| Panel B | N/A | 485 | 132 | 0.6 | 0.4 | 1.0<br>(0.7-1.3) | 0.8<br>(0.5-0.9) |
A summary of results is presented below:
4 Includes within-run, between-run, and between-day variability.
b Maximum and minimum SD or %CV for each reagent lot and instrument combination.
The precision Alinity c ICT Sample Diluent Chloride assay was considered acceptable if the within-laboratory (total) imprecision (within-run, between-run, and between-day) was ≤ 2.0 % CV for serum samples targeted between 89.0 to 99.0 mmol/L.
The Alinity c ICT Sample Diluent Chloride assay demonstrated acceptable precision.
{12}------------------------------------------------
# Alinity c ICT Sample Diluent Sodium assay – Urine Samples
Precision was performed based on guidance from the Clinical and Laboratory Standards Institute (CLSI) document EP05-A2.
| Sample | Control<br>Lot | n | Mean<br>(mmol/L) | Within-Run<br>(Repeatability) | | Within-Laboratory (Total)a | |
|--------------------|----------------|-----|------------------|-------------------------------|-----|----------------------------|------------------|
| | | | | SD | %CV | SD<br>(Rangeb) | %CV<br>(Rangeb) |
| Control<br>Level 1 | 1 | 240 | 92 | 0.6 | 0.7 | 1.0<br>(0.9-1.1) | 1.1<br>(1.0-1.2) |
| | 2 | 240 | 92 | 0.5 | 0.5 | 0.7<br>(0.7-0.8) | 0.8<br>(0.7-0.9) |
| Control<br>Level 2 | 1 | 240 | 161 | 0.7 | 0.4 | 1.0<br>(0.9-1.1) | 0.6<br>(0.6-0.7) |
| | 2 | 240 | 161 | 0.6 | 0.4 | 1.5<br>(1.0-1.8) | 0.9<br>(0.7-1.1) |
| Panel A | N/A | 480 | 21 | 0.5 | 2.3 | 0.6<br>(0.5-0.8) | 2.9<br>(2.3-3.7) |
| Panel B | N/A | 480 | 383 | 1.5 | 0.4 | 3.9<br>(2.5-4.7) | 1.0<br>(0.7-1.2) |
A summary of results is presented below:
4 Includes within-run, between-run, and between-day variability.
b Maximum and minimum SD or %CV for each reagent lot and instrument combination.
The precision Alinity c ICT Sample Diluent Sodium assay was considered acceptable if the within-laboratory (total) imprecision (within-run, between-run, and between-day) was ≤ 3.0 % CV for urine samples targeted between 79.0 to 181.0 mmol/L.
The Alinity c ICT Sample Diluent Sodium assay demonstrated acceptable precision.
{13}------------------------------------------------
# Alinity c ICT Sample Diluent Potassium assay – Urine Samples
Precision was performed based on guidance from the Clinical and Laboratory Standards Institute (CLSI) document EP05-A2.
| | | | | Within-Run<br>(Repeatability) | | Within-Laboratory (Total)a | |
|--------------------|----------------|-----|------------------|-------------------------------|------|----------------------------|---------------------|
| Sample | Control<br>Lot | n | Mean<br>(mmol/L) | SD | %CV | SD<br>(Rangeb) | %CV<br>(Rangeb) |
| Control<br>Level 1 | 1 | 240 | 16.6 | 0.06 | 0.3 | 0.13<br>(0.13-0.14) | 0.8<br>(0.8-0.9) |
| | 2 | 240 | 16.6 | 0.05 | 0.3 | 0.12<br>(0.11-0.12) | 0.7<br>(0.7-0.7) |
| Control<br>Level 2 | 1 | 240 | 58.1 | 0.17 | 0.3 | 0.34<br>(0.29-0.39) | 0.6<br>(0.5-0.7) |
| | 2 | 240 | 58.2 | 0.17 | 0.3 | 0.49<br>(0.29-0.62) | 0.8<br>(0.5-1.1) |
| Panel A | N/A | 480 | 1.7 | 0.02 | 1.4 | 0.04<br>(0.04-0.04) | 2.4<br>(2.3-2.6) |
| | Panel B | N/A | 480 | 127.7 | 0.35 | 0.3 | 0.68<br>(0.50-0.83) |
| Panel C | | N/A | 479 | 284.5 | 0.82 | 0.3 | 1.91<br>(1.65-2.12) |
A summary of results is presented below:
4 Includes within-run, between-run, and between-day variability.
b Maximum and minimum SD or %CV for each reagent lot and instrument combination.
The precision Alinity c ICT Sample Diluent Potassium assay was considered acceptable if the within-laboratory (total) imprecision (within-run, between-run, and between-day) was ≤ 3 % CV for urine samples targeted between 31.0 to 84.0 mmol/L.
The Alinity c ICT Sample Diluent Potassium assay demonstrated acceptable precision.
{14}------------------------------------------------
# Alinity c ICT Sample Diluent Chloride assay – Urine Samples
Precision was performed based on guidance from the Clinical and Laboratory Standards Institute (CLSI) document EP05-A2.
| Sample | Control<br>Lot | n | Mean<br>(mmol/L) | Within-Run<br>(Repeatability) | | Within-Laboratory (Total)a | |
|--------------------|----------------|-----|------------------|-------------------------------|-----|----------------------------|------------------|
| | | | | SD | %CV | SD<br>(Rangeb) | %CV<br>(Rangeb) |
| Control<br>Level 1 | 1 | 240 | 103 | 0.5 | 0.4 | 0.9<br>(0.8-1.0) | 0.9<br>(0.8-0.9) |
| | 2 | 240 | 103 | 0.5 | 0.5 | 0.8<br>(0.8-0.8) | 0.8<br>(0.8-0.8) |
| Control<br>Level 2 | 1 | 240 | 193 | 0.7 | 0.4 | 1.2<br>(1.1-1.3) | 0.6<br>(0.6-0.7) |
| | 2 | 240 | 193 | 0.7 | 0.4 | 1.9<br>(1.3-2.4) | 1.0<br>(0.7-1.2) |
| Panel A | N/A | 479 | 24 | 0.3 | 1.1 | 0.4<br>(0.1-0.5) | 1.6<br>(0.4-2.2) |
| Panel B | N/A | 480 | 273 | 0.9 | 0.3 | 2.2<br>(1.9-2.3) | 0.8<br>(0.7-0.8) |
A summary of results is presented below:
4 Includes within-run, between-run, and between-day variability.
b Maximum and minimum SD or %CV for each reagent lot and instrument combination.
The precision Alinity c ICT Sample Diluent Chloride assay was considered acceptable if the within-laboratory (total) imprecision (within-run, between-run, and between-day) was ≤ 1.8 % CV for urine samples targeted between 79.0 to 218.0 mmol/L.
The Alinity c ICT Sample Diluent Chloride assay demonstrated acceptable precision
{15}------------------------------------------------
# Linearity
Alinity c ICT Sample Diluent Sodium, Potassium, Chloride assays – Serum Samples
Linearity was evaluated based on guidance from the Clinical and Laboratory Standards Institute (CLSI) document EP06-A.
The mean observed linear range concentrations ranged from 27 to 259 mmol/L for the sodium sample set.
The mean observed linear range concentrations ranged from 0.5 to 14.3 mmol/L for the potassium sample set.
The mean observed linear range concentrations ranged from 22 to 172 mmol/L for the chloride sample set.
Alinity c ICT Sample Diluent Sodium, Potassium, Chloride assays – Urine Samples
Linearity was evaluated based on guidance from the Clinical and Laboratory Standards Institute (CLSI) document EP06-A.
The mean observed linear range concentrations ranged from 1 to 474 mmol/L for the sodium sample set.
The mean observed linear range concentrations ranged from 0.1 to 366.0 mmol/L for the potassium sample set.
The mean observed linear range concentrations ranged from 5 to 346 mmol/L for the chloride sample set.
{16}------------------------------------------------
# Measuring Interval
# Serum
- The measuring interval of the Alinity c ICT Sodium assay is 100 to . 200 mmol/L.
- The measuring interval of the Alinity c ICT Potassium assay is 1.0 to . 10.0 mmol/L.
- The measuring interval of the Alinity c ICT Chloride assay is 50 to 150 mmol/L. .
# Urine
- . The measuring interval of the Alinity c ICT Sodium assay is 20 to 400 mmol/L.
- The measuring interval of the Alinity c ICT Potassium assay is 1.0 to 300.0 mmol/L.
- . The measuring interval of the Alinity c ICT Chloride assay is 20 to 300 mmol/L.
The measuring interval is defined as the range of values in mmol/L which meets the limits of acceptable performance for linearity, imprecision, and bias. The inputs to the measuring interval include imprecision and linearity.
{17}------------------------------------------------
# Interference
Alinity c ICT Sample Diluent Sodium assay Serum Samples
Potential interference was evaluated based on guidance from the Clinical Laboratory and Standards Institute (CLSI) document EP07-A2.
For sodium serum a bias of > 2% was considered significant interference.
The Alinity c ICT Sample Diluent Sodium assay using the serum application is not susceptible to interference effects from the following interferents at the interferent levels listed in the table below:
| Interferent | Interferent Level |
|------------------------|-------------------|
| Unconjugated Bilirubin | ≤ 60 mg/dL |
| Conjugated Bilirubin | ≤ 30 mg/dL |
| Hemoglobin | ≤ 500 mg/dL |
| Triglycerides | ≤ 2000 mg/dL |
| Ascorbic Acid | ≤ 6 mg/dL |
| Acetaminophen | ≤ 20 mg/dL |
| Ibuprofen | ≤ 50 mg/dL |
| Acetylcysteine | ≤ 167 mg/dL |
| Acetylsalicylic Acid | ≤ 66 mg/dL |
| Benzalkonium Chloride | ≤ 1 mg/dL |
{18}------------------------------------------------
# Alinity c ICT Sample Diluent Potassium assay – Serum Samples
Potential interference was evaluated based on guidance from the Clinical Laboratory and Standards Institute (CLSI) document EP07-A2.
For potassium serum a bias of > 10% was considered significant interference.
The Alinity c ICT Sample Diluent Potassium assay using the serum application is not susceptible to interference effects from the following interferents at the interferent levels listed in the table below:
| Interferent | Interferent Level |
|------------------------|-------------------|
| Unconjugated Bilirubin | ≤ 60 mg/dL |
| Conjugated Bilirubin | ≤ 60 mg/dL |
| Hemoglobin | ≤ 100 mg/dL |
| Triglycerides | ≤ 2000 mg/dL |
| Ascorbic Acid | ≤ 6 mg/dL |
| Acetaminophen | ≤ 20 mg/dL |
| Ibuprofen | ≤ 50 mg/dL |
| Acetylcysteine | ≤ 167 mg/dL |
| Acetylsalicylic Acid | ≤ 66 mg/dL |
| Sodium Salicylate | ≤ 70 mg/dL |
| Benzalkonium Chloride | ≤ 5 mg/dL |
{19}------------------------------------------------
# Alinity c ICT Sample Diluent Chloride assay – Serum Samples
Potential interference was evaluated based on guidance from the Clinical Laboratory and Standards Institute (CLSI) document EP07-A2.
For chloride serum a bias of > 10% was considered significant interference.
The Alinity c ICT Sample Diluent Chloride assay using the serum application is not susceptible to interference effects from the following interferents at the interferent levels listed in the table below:
| Interferent | Interferent Level |
|------------------------|-------------------|
| Unconjugated Bilirubin | ≤ 60 mg/dL |
| Conjugated Bilirubin | ≤ 20 mg/dL |
| Hemoglobin | ≤ 1000 mg/dL |
| Triglycerides | ≤ 2000 mg/dL |
| Ascorbic Acid | ≤ 6 mg/dL |
| Acetaminophen | ≤ 20 mg/dL |
| Ibuprofen | ≤ 50 mg/dL |
| Acetylcysteine | ≤ 16.7 mg/dL |
| Acetylsalicylic Acid | ≤ 66 mg/dL |
| Sodium Salicylate | ≤ 70 mg/dL |
| Benzalkonium Chloride | ≤ 10 mg/dL |
| Lithium Bromide | ≤ 20 mg/dL |
| Lithium Iodide | ≤ 25.4 mg/dL |
| Sodium Azide | ≤ 325 mg/dL |
{20}------------------------------------------------
Potential interference was evaluated based on guidance from the Clinical Laboratory and Standards Institute (CLSI) document EP07-A2.
For sodium urine a bias of > 10% was considered significant interference.
The Alinity c ICT Sample Diluent Sodium assay using the urine application is not susceptible to interference effects from the following interferents at the interferent levels listed in the table below:
| Interferent | Interferent Level | | |
|--------------------------|-------------------|--|--|
| Protein | < 50 mg/dL | | |
| Glucose | < 1000 mg/dL | | |
| Ascorbate | < 200 mg/dL | | |
| 8.5 N Acetic Acid | < 6.25 mL/dL | | |
| Boric Acid | < 250 mg/dL | | |
| 6 N Hydrochloric Acid | < 2.5 mL/dL | | |
| 6 N Nitric Acid | < 5.0 mL/dL | | |
| Acetaminophen | < 20 mg/dL | | |
| Ibuprofen | < 50 mg/dL | | |
| Acetylcysteine | < 167 mg/dL | | |
| Conjugated Bilirubin | < 60 mg/dL | | |
| Hemoglobin | < 1000 mg/dL | | |
| pH (range) | 3.52 to 8.58 | | |
| Specific Gravity (range) | 1.004 to 1.027 | | |
{21}------------------------------------------------
Potential interference was evaluated based on guidance from the Clinical Laboratory and Standards Institute (CLSI) document EP07-A2.
For potassium urine a bias of > 10% was considered significant interference.
The Alinity c ICT Sample Diluent Potassium assay using the urine application is not susceptible to interference effects from the following interferents at the interferent levels listed in the table below:
| Interferent | Interferent Level |
|--------------------------|-------------------|
| Protein | ≤ 50 mg/dL |
| Glucose | ≤ 1000 mg/dL |
| Ascorbate | ≤ 200 mg/dL |
| 8.5 N Acetic Acid | ≤ 6.25 mL/dL |
| Boric Acid | ≤ 250 mg/dL |
| 6 N Hydrochloric Acid | ≤ 2.5 mL/dL |
| 6 N Nitric Acid | ≤ 5.0 mL/dL |
| Sodium Oxalate | ≤ 60 mg/dL |
| Sodium Carbonate | ≤ 1.25 g/dL |
| Sodium Fluoride | ≤ 400 mg/dL |
| Acetaminophen | ≤ 20 mg/dL |
| Ibuprofen | ≤ 50 mg/dL |
| Acetylcysteine | ≤ 167 mg/dL |
| Conjugated Bilirubin | ≤ 60 mg/dL |
| Hemoglobin | ≤ 500 mg/dL |
| pH (range) | 3.58 to 8.03 |
| Specific Gravity (range) | 1.010 to 1.025 |
{22}------------------------------------------------
# Alinity c ICT Sample Diluent Chloride assay – Urine Samples
Potential interference was evaluated based on guidance from the Clinical Laboratory and Standards Institute (CLSI) document EP07-A2.
For potassium urine a bias of > 10% was considered significant interference.
The Alinity c ICT Sample Diluent Chloride assay using the urine application is not susceptible to interference effects from the following interferents at the interferent levels listed in the table below:
| Interferent | Interferent Level |
|--------------------------|-------------------|
| Protein | ≤ 50 mg/dL |
| Glucose | ≤ 1000 mg/dL |
| Ascorbate | ≤ 200 mg/dL |
| 8.5 N Acetic Acid | ≤ 6.25 mL/dL |
| Boric Acid | ≤ 250 mg/dL |
| 6 N Nitric Acid | ≤ 5.0 mL/dL |
| Sodium Carbonate | ≤ 1.25 g/dL |
| Sodium Fluoride | ≤ 400 mg/dL |
| Sodium Oxalate | ≤ 60 mg/dL |
| Acetaminophen | ≤ 20 mg/dL |
| Ibuprofen | ≤ 50 mg/dL |
| Acetylcysteine | ≤ 16.7 mg/dL |
| Conjugated Bilirubin | ≤ 60 mg/dL |
| Hemoglobin | ≤ 1000 mg/dL |
| pH (range) | 3.52 to 7.97 |
| Specific Gravity (range) | 1.006 to 1.033 |
{23}------------------------------------------------
# Method Comparison
Alinity c ICT Sample Diluent Sodium, Potassium and Chloride assay – Serum and Urine Samples
The method comparison study was performed based on guidance from the Clinical and Laboratory Standards Institute (CLSI) document EP09-A3.
Human serum and urine specimens that spanned the measuring interval of the assay were evaluated for serum and urine testing respectively.
The results analyzed using Passing-Bablok regression method are summarized in the tables below:
# Sodium
| | Units | n | Correlation<br>Coefficient | Intercept | Slope | Concentration<br>Range |
|--------------------------------------------|--------------|-----|----------------------------|-----------|-------|------------------------|
| Alinity c Sodium<br>vs ARCHITECT<br>Sodium | Serum mmol/L | 141 | 1.00 | -0.50 | 1.00 | 101-197 |
| | Urine mmol/L | 101 | 1.00 | 0.88 | 0.99 | 22-386 |
# Potassium
| | Units | n | Correlation<br>Coefficient | Intercept | Slope | Concentration<br>Range |
|-----------------------------------------------------|--------------|-----|----------------------------|-----------|-------|------------------------|
| Alinity c<br>Potassium<br>vs ARCHITECT<br>Potassium | Serum mmol/L | 122 | 1.00 | 0.00 | 1.00 | 1.3-9.4 |
| | Urine mmol/L | 107 | 1.00 | -1.23 | 1.05 | 4.3-266.1 |
# Chloride
| | Units | n | Correlation<br>Coefficient | Intercept | Slope | Concentration<br>Range |
|------------------------------------------------|--------------|-----|----------------------------|-----------|-------|------------------------|
| Alinity c Chloride<br>vs ARCHITECT<br>Chloride | Serum mmol/L | 120 | 1.00 | 0.00 | 1.00 | 52-148 |
| | Urine mmol/L | 112 | 1.00 | -0.30 | 1.00 | 25-299 |
{24}------------------------------------------------
The method comparison study results for the investigational method, Alinity c ICT Sample Diluent Sodium, Potassium, and Chloride, versus the comparator method, ARCHITECT ICT Sample Diluent Sodium, Potassium, and Chloride, are acceptable.
# Tube Type Equivalency
Alinity c ICT Sample Diluent Sodium, Potassium, and Chloride assay
Tube type equivalency was performed based on guidance from the Clinical Laboratory Standards Institute (CLSI) document EP07-A2 to evaluate whether specific blood collection tube types are suitable for use with the Alinity c ICT Sample Diluent for the Sodium, Potassium, and Chloride assays.
Samples were collected from a minimum of 40 donors and evaluated across tube types.
The following blood collection tube types are acceptable for use with the Alinity c ICT Sample Diluent Sodium, Potassium, and Chloride assays:
- . Serum
- Serum separator ●
- Lithium heparin ●
- Sodium heparin ●
- . Lithium heparin plasma separator
{25}------------------------------------------------
#### VIII. Summary of Clinical Performance
This section does not apply.
# IX. Conclusion Drawn from Nonclinical Laboratory Studies
The similarities and differences between the candidate assay (Alinity c ICT Sample Diluent, List No. 07P53) and the predicate assay (ICT (Na+, K+, Cl ) Sample Diluent, List No. 2P32) are presented in the table on page 5. The minor differences between the candidate ICT sodium, potassium and chloride assays and the predicate ICT sodium, potassium and chloride assays raise no new issues of safety and effectiveness. The performance results presented in this 510(k) demonstrate that the Alinity c ICT Sample Diluent is safe and effective for the stated intended use.
The results presented in this 510(k) premarket notification demonstrate that the Alinity c ICT Sample Diluent is substantially equivalent to the respective predicate device (ICT (Na+, K+, C1 ) Sample Diluent (k980367)).