K170319 · Applied Medical Technology, Inc. · KNT · Oct 6, 2017 · Gastroenterology, Urology
Device Facts
Record ID
K170319
Device Name
Low Profile Non-Balloon Feeding Device
Applicant
Applied Medical Technology, Inc.
Product Code
KNT · Gastroenterology, Urology
Decision Date
Oct 6, 2017
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 876.5980
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Low Profile Non-Balloon Feeding Device is indicated for use in patients who require long term feeding, are unable to tolerate oral feeding, who are at low risk for aspiration, require gastric decompression and/or medication delivered directly into the stomach through a secured (initial placement) or formed (replacement) stoma.
Device Story
Device provides gastric access for nutrition, medication, and decompression. Consists of external bolster, feeding catheter, and internal retention bolster. Two insertion methods: 'capsule' (water-dissolvable capsule encasing internal bolster, deployed via pull-tab/suture) and 'obturator' (T-handle/ratcheting rod assembly to elongate internal dome for insertion). Removal uses specialized tool assembly to elongate bolster for easier extraction. Used in clinical settings; operated by healthcare professionals. Device features anti-reflux valve and feeding port interlock. Benefits include low-profile design and controlled insertion/removal process. Output is direct gastric access; provider uses device to manage patient nutritional/medication needs.
Clinical Evidence
Bench testing only. No clinical data. Performance testing included stoma pullout force, tip poke-through force, snap arm/T-handle engagement, interlock pullout, capsule rupture time, suture deployment force, ripcord tensile strength, pull tab-to-suture bond strength, liquid leakage, flow rate (water/viscous), tube tensile test, obturator bond strength, duckbill flow/backflow, and obturator push-through force.
Technological Characteristics
Medical-grade silicone construction. Non-balloon retention bolster. Two insertion mechanisms: water-dissolvable capsule or T-handle/ratcheting obturator rod. Includes anti-reflux valve and feeding port interlock. Available in 14, 18, 20, 24 Fr diameters; lengths 1.0 cm to 4.4 cm. Sterile, single-use. No electronic components or software.
Indications for Use
Indicated for patients requiring long-term enteral feeding, gastric decompression, or medication delivery; patients must be unable to tolerate oral feeding and at low risk for aspiration. Applicable for initial placement or replacement via stoma.
Regulatory Classification
Identification
A gastrointestinal tube and accessories is a device that consists of flexible or semi-rigid tubing used for instilling fluids into, withdrawing fluids from, splinting, or suppressing bleeding of the alimentary tract. This device may incorporate an integral inflatable balloon for retention or hemostasis. This generic type of device includes the hemostatic bag, irrigation and aspiration catheter (gastric, colonic, etc.), rectal catheter, sterile infant gavage set, gastrointestinal string and tubes to locate internal bleeding, double lumen tube for intestinal decompression or intubation, feeding tube, gastroenterostomy tube, Levine tube, nasogastric tube, single lumen tube with mercury weight balloon for intestinal intubation or decompression, and gastro-urological irrigation tray (for gastrological use).
Special Controls
*Classification.* (1) Class II (special controls). The barium enema retention catheter and tip with or without a bag that is a gastrointestinal tube and accessory or a gastronomy tube holder accessory is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I (general controls) for the dissolvable nasogastric feed tube guide for the nasogastric tube. The class I device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 876.9.
Predicate Devices
AMT Button Replacement Gastrostomy Device (K980305)
Capsule Monarch Trans shaping Gastrostomy Tube (K043027)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food & Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
October 6, 2017
Applied Medical Technology, Inc. Shraddha Saini Regulatory Affairs Specialist 8006 Katherine Boulevard Brecksville, OH 44141
Re: K170319
> Trade/Device Name: Low Profile Non-Balloon Feeding Device Regulation Number: 21 CFR§ 876.5980 Regulation Name: Gastrointestinal Tube and Accessories Regulatory Class: II Product Code: KNT Dated: September 12, 2017 Received: September 13, 2017
Dear Shraddha Saini:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies.
{1}------------------------------------------------
You must comply with all the Act's requirements. including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education (DICE) at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevicesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education (DICE) at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely,
# Benjamin R. Fisher -S
Benjamin R. Fisher, Ph.D. Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K170319
#### Device Name
Low Profile Non-Balloon Feeding Device.
#### Indications for Use (Describe)
The Low Profile Non-Balloon Feeding Device is indicated for use in patients who require long term feeding, are unable to tolerate oral feeding, who are at low risk for aspiration, require gastric decompression and/or medication delivered directly into the stomach through a secured (initial placement) or formed (replacement) stoma.
| Type of Use (Select one or both, as applicable) |
|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| <div> <span> <svg fill="none" height="16" viewbox="0 0 16 16" width="16" xmlns="http://www.w3.org/2000/svg"> <path d="M13.2929 3.29289C13.6834 2.90237 14.3166 2.90237 14.7071 3.29289C15.0976 3.68342 15.0976 4.31658 14.7071 4.70711L5.70711 13.7071C5.31658 14.0976 4.68342 14.0976 4.29289 13.7071L1.29289 10.7071C0.902369 10.3166 0.902369 9.68342 1.29289 9.29289C1.68342 8.90237 2.31658 8.90237 2.70711 9.29289L5 11.5858L13.2929 3.29289Z" fill="black"></path> </svg> </span> Prescription Use (Part 21 CFR 801 Subpart D) </div> |
| <div> <span> <svg fill="none" height="16" viewbox="0 0 16 16" width="16" xmlns="http://www.w3.org/2000/svg"> <rect height="15" rx="2.5" stroke="black" width="15" x="0.5" y="0.5"></rect> </svg> </span> Over-The-Counter Use (21 CFR 801 Subpart C) </div> |
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
## SECTION - 5
# 510(k) Summary
## Low Profile Non-Balloon Feeding Device
I
| Date Prepared: | September 12, 2017 |
|-----------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Submitter: | Applied Medical Technology, Inc.<br>8006 Katherine Boulevard<br>Brecksville, OH 44141<br>Phone: 440-717-4000<br>Fax: 440-717-4200<br>Contact Person: Shraddha Saini - Regulatory Affairs Specialist<br>Email: Shraddha.Saini@appliedmedical.net |
| Device<br>Information: | Trade Name: Low Profile Non-Balloon Feeding Device<br>Common Name: Gastrointestinal Feeding Tubes<br>Classification Name: Gastrointestinal Tube and Accessories (21 CFR<br>876.5980)<br>Product Code: KNT<br>Regulatory Class: II<br>Review Panel: Gastroenterology and Urology |
| Predicate<br>Device: | AMT Button Replacement Gastrostomy Device (K980305)<br>Capsule Monarch Trans shaping Gastrostomy Tube (K043027) |
| Intended Use: | The Low Profile Non-Balloon Feeding Device is indicated for use in patients<br>who require long term feeding, are unable to tolerate oral feeding, who are at<br>low risk for aspiration, require gastric decompression and/or medication<br>delivered directly into the stomach through a secured (initial placement) or<br>formed (replacement) stoma. |
| Device<br>Description: | The Low Profile Non-Balloon Feeding Device is used to provide nutrition,<br>medication, and decompression access into the stomach through a secured<br>(initial placement) or formed (replacement) stoma. The device may be placed<br>using one of two insertion methods including an 'obturator' insertion method<br>and a 'capsule' insertion method.<br>Capsule Insertion Method: The internal bolster of the device is encased by water<br>dissolvable capsule. The capsule technology provides a low-profile option that<br>is delivered by lubricating the capsule with lubricant, inserting the device<br>through the stoma site, and pulling on a pull tab that pulls a suture through the<br>capsule and deploys the internal dome. Once the capsule is deployed, the<br>internal dome returns to its uncompressed shape and the device is held in place.<br>Obturator Insertion Method: A T-Handle and ratcheting obturator rod (Snap<br>Arm Assembly) is used to extend the internal dome, decreasing the outer profile<br>of the dome. When the dome is fully elongated, the internal dome maintains a |
| | narrow profile that can easily be inserted into the stoma site with proper<br>lubrication. Once inserted into the stoma site, the ratcheting rod and T-handle<br>are removed and device is held in place by the internal bolster.<br>Once placed, both devices provide access to nutrition, medication, and<br>decompression into the stomach via a Mini ONE® feeding port. Both devices<br>are identical after placement regardless of the insertion method chosen.<br>Removal of feeding device can be performed by using a removal tool assembly<br>(Snap Arm and T-handle with removal tool reinforcer). Removal is done by<br>elongating the internal bolster to reduce the outer diameter of the dome,<br>providing for an easier removal process. The Low Profile Non-Balloon Feeding<br>Device will be offered in several different diameters including14 Fr, 18 Fr, 20<br>Fr, and 24 Fr and will be available in stoma lengths ranging from 1.0 cm to 4.4<br>cm. |
| | The Low Profile Non-Balloon Feeding Device is provided sterile for single use<br>only. The molded body of the Non-Balloon feeding device is made of medical<br>silicone. The Low Profile Non-Balloon Feeding Device consists of an external<br>bolster, feeding catheter, and internal retention bolster similar to the predicate<br>devices. The external bolster consists of a feeding port for access to the stomach<br>through the tubing of the device and a strap with a plug to close the feeding port<br>while not in use. In addition, an anti-reflux valve is included in the feeding port<br>area to prevent backflow of stomach contents while not in use. The feeding<br>catheter is inserted into the stomach through a stoma and is held in place with<br>the internal retention bolster. There are two different insertion methods for the<br>proposed device, the 'capsule' and the 'obturator' method. The 'capsule' version<br>includes a tapered capsule enclosing the internal bolster at one end that allows<br>ease of insertion and an obturator rod is pre-loaded in to the shaft prior to use.<br>The 'obturator' version uses a T-Handle and snap arm assembly for controlled<br>insertion and elongation through the device shaft. The Low Profile Non-Balloon<br>Feeding Device and the predicate devices are provided in a number of sizes to<br>accommodate different stoma diameters and lengths. |
| Technological<br>Characteristics: | Design Differences:<br>Method of Insertion: The proposed devices provides two different methods of<br>insertion ('Capsule' or 'Obturator'), whereas the predicate devices are only<br>provided with a single method of insertion. The Capsule Monarch Trans<br>shaping Gastrostomy Tube uses the same Capsule technology of the proposed<br>device but does not offer a Snap arm assembly and T-handle method for<br>insertion. The AMT Button Replacement Gastrostomy Device uses an insertion<br>rod to insert the device, but does not offer the level of control and repeatability<br>for insertion that the proposed device offers with the Snap arm assembly and T-<br>handle components. Both predicate devices don't use removal tools, allowing<br>the proposed device to have a more comfortable means of removal by avoiding a<br>necessary retraction removal. |
| | Interlock and Anti-Reflux Valve: The proposed device incorporates an<br>interlock for its feeding port, whereas the predicate devices do not have<br>interlocks. The proposed device incorporates an anti-reflux valve that is located<br>along the proximal end of the tubing along the external bolster. The Capsule<br>Monarch Trans shaping Gastrostomy Tube does not include an anti- reflux valve<br>and relies only on external plugs to prevent leakage. The AMT Button<br>Replacement Gastrostomy Device includes an anti-reflux valve along the distal |
| | the proposed device. |
| | The overall functionality and safety of the proposed Low Profile Non Balloon<br>Feeding Device remains substantially equivalent to the predicate devices. |
| Biocompatibility<br>Testing: | The patient contacting components of the Low Profile Non-Balloon Feeding<br>Device are in contact with mucosal membrane and/or breached/compromised<br>surfaces for prolonged or permanent exposure. No additional biocompatibility<br>testing was conducted because the patient-contacting components in the final<br>finished device are identical to those in previously marketed devices (in terms<br>of formulation, processing, sterilization, geometry, nature of use, and duration<br>of exposure) and no chemicals were added (e.g., plasticizers, fillers, additives,<br>cleaning agents, mold release agents). |
| Performance<br>Testing: | AMT conducted various performance tests on the components contained within<br>the Low Profile Non-Balloon Feeding Device. Testing found that all<br>components and materials met or exceeded design specifications established by<br>AMT.<br>Bench tests have been carried out on samples of the Low Profile Non-Balloon<br>Feeding Device. The tests carried out included:<br>• Stoma Pullout Force<br>• Tip Poke-Through Force<br>• Snap Arm Body to Rod Attachment Force<br>• Snap Arm to T-Handle Snap Engagement Force<br>• T-Handle Tooth Shear Strength<br>• Interlock Pullout<br>• Time to Capsule Rupture<br>• Suture Deployment Force<br>• Ripcord Tensile Strength<br>• Pull Tab-to-Suture Bond Strength<br>• Liquid Leakage Test<br>• Flow Rate Test (Water and Viscous Fluid)<br>• Tube Tensile Test<br>• Obturator Bond Strength Test<br>• Duckbill Flow and Backflow Test<br>• Obturator Poke-Through Evaluation<br>• Obturator Push-Through Force Test |
| Conclusion: | The Low Profile Non-Balloon Feeding Device is substantially equivalent to the<br>predicate devices cleared under K980305 and K043027 in intended use, design,<br>biocompatibility, performance, and principles of operation. |
Applied Medical Technology, Inc. - 510(k) Submission Low Profile Non Balloon Feeding Device Section 5 – Page 5.1
{4}------------------------------------------------
# SECTION - 5
Applied Medical Technology, Inc. - 510(k) Submission Low Profile Non Balloon Feeding Device Section 5 – Page 5.2
{5}------------------------------------------------
## SECTION - 5
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.