Ion PGM Dx System

K170299 · Life Technologies Corporation · PFF · Jun 22, 2017 · Immunology

Device Facts

Record IDK170299
Device NameIon PGM Dx System
ApplicantLife Technologies Corporation
Product CodePFF · Immunology
Decision DateJun 22, 2017
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 862.2265
Device ClassClass 2
AttributesReal-World Evidence

Real-World Evidence

SubmissionDeviceSponsorRWD SourcesRWE Use SummaryKey Tags
K170299 · Jun 22, 2017Ion PGM Dx SystemLife Technologies CorporationFFPE clinical tumor samples (resection and core needle biopsy)Clinical FFPE samples were used to evaluate the accuracy (PPA/NPA) of the Ion PGM Dx System by comparing results against validated reference detection methods (NGS assay and ROS1 FISH).FFPE; Clinical samples; Accuracy study; Concordance

Clinical Evidence

Study DesignPopulationComparatorKey Endpoints
Accuracy study using FFPE tumor samples; Retrospective concordance study290 FFPE tumor samples; Sample Size: 290Validated NGS assay and ROS1 FISH reference testPositive percent agreement (PPA) and negative percent agreement (NPA)

Indications for Use

The Ion PGM™ Dx Instrument System is composed of a sequencing instrument that measures the hydrogen ions that are generated during the incorporation of nucleotides in the DNA sequencing reaction, and the ancillary instrumentation necessary for sample processing. This instrument system is used in conjunction with the instrument-specific Ion PGM™ Dx Library Kit, Ion OneTouch™ Dx Template Kit, Ion PGM™ Dx Sequencing Kit, and Ion 318™ Dx Chip Kit, and data analysis software. The Ion PGM™ Dx Instrument System is intended for targeted sequencing of human genomic DNA (gDNA) from peripheral whole-blood samples and DNA and RNA extracted from formalin-fixed, paraffin-embedded (FFPE) samples. The Ion PGM™ Dx Instrument System is not intended for whole genome or de novo sequencing.

Device Story

System performs targeted DNA/RNA sequencing; measures hydrogen ions released during nucleotide incorporation. Components: Ion OneTouch Dx/ES Dx instruments, PGM Dx sequencer, Torrent Server, Torrent Suite Dx software, and specific library/template/sequencing/chip kits. Used in clinical laboratories by trained professionals. Input: gDNA from whole blood or DNA/RNA from FFPE tissue. Workflow: sample processing, library preparation, sequencing, and automated data analysis. Output: qualitative variant calls (SNVs, MNVs, insertions, deletions, fusions). Healthcare providers use results to identify genetic variants; aids clinical decision-making regarding cancer-related genes. Benefits: enables targeted genomic analysis from diverse sample types.

Clinical Evidence

Bench testing only. Accuracy evaluated using 290 FFPE tumor samples against validated NGS/FISH methods; PPA 96.9-98.5% and NPA 98.4-100% (excluding no calls). Reproducibility assessed across 4 sites/operators/instruments; repeatability 100% for DNA variants, 87.5% for RNA. Tissue/DNA/RNA input studies confirmed performance at specified concentrations (10 ng). Interfering substances (paraffin, xylene, ethanol, hemoglobin, protease) showed 100% concordance with controls. Cross-contamination rate 0% for DNA, 1.25% for RNA.

Technological Characteristics

Sequencing instrument measuring hydrogen ions (pH-based) during nucleotide incorporation. Includes ancillary sample processing instrumentation. Connectivity: Torrent Server/Suite Dx software. Software: Torrent Suite Dx. Form factor: Benchtop system. Sterilization: N/A (laboratory equipment).

Indications for Use

Indicated for targeted sequencing of human genomic DNA from peripheral whole blood and DNA/RNA from FFPE tissue samples. Not for whole genome or de novo sequencing. Prescription use only.

Regulatory Classification

Identification

A high throughput genomic sequence analyzer for clinical use is an analytical instrument system intended to generate, measure and sort signals in order to analyze nucleic acid sequences in a clinical sample. The device may include a signal reader unit; reagent handling, dedicated instrument control, and other hardware components; raw data storage mechanisms; data acquisition software; and software to process detected signals.

Special Controls

*Classification.* Class II (special controls). The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9. The special controls for this device are:(1) The labeling for the instrument system must reference legally marketed pre-analytical and analytical reagents to be used with the instrument system and include or reference legally marketed analytical software that includes sequence alignment and variant calling functions, to be used with the instrument system. (2) The labeling for the instrument system must include a description of the following information: (i) The specimen type(s) validated as an appropriate source of nucleic acid for this instrument. (ii) The type(s) of nucleic acids ( *e.g.,* germline DNA, tumor DNA) validated with this instrument.(iii) The type(s) of sequence variations ( *e.g.* single nucleotide variants, insertions, deletions) validated with this instrument.(iv) The type(s) of sequencing ( *e.g.,* targeted sequencing) validated with this instrument.(v) The appropriate read depth for the sensitivity claimed and validation information supporting those claims. (vi) The nucleic acid extraction method(s) validated for use with the instrument system. (vii) Limitations must specify the types of sequence variations that the instrument cannot detect with the claimed accuracy and precision ( *e.g.,* insertions or deletions larger than a certain size, translocations).(viii) Performance characteristics of the instrument system must include: (A) Reproducibility data generated using multiple instruments and multiple operators, and at multiple sites. Samples tested must include all claimed specimen types, nucleic acid types, sequence variation types, and types of sequencing. Variants queried shall be located in varying sequence context ( *e.g.,* different chromosomes, GC-rich regions). Device results shall be compared to reference sequence data with high confidence.(B) Accuracy data for all claimed specimen types and nucleic acid types generated by testing a panel of well characterized samples to query all claimed sequence variation types, types of sequencing, and sequences located in varying sequence context ( *e.g.,* different chromosomes, GC-rich regions). The well-characterized sample panel shall include samples from at least two sources that have highly confident sequence based on well-validated sequencing methods. At least one reference source shall have sequence generated independently of the manufacturer with respect to technology and analysis. Percent agreement and percent disagreement with the reference sequences must be described for all regions queried by the instrument.(C) If applicable, data describing endogenous or exogenous substances that may interfere with the instrument system. (D) If applicable, data demonstrating the ability of the system to consistently generate an accurate result for a given sample across different indexing primer combinations. (ix) The upper and lower limit of input nucleic acid that will achieve the claimed accuracy and reproducibility. Data supporting such claims must also be summarized.

Predicate Devices

Submission Summary (Full Text)

{0} 1 # 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY INSTRUMENT ONLY TEMPLATE A. 510(k) Number: k170299 B. Purpose for Submission: Expansion of the intended use of the device to include DNA and RNA from FFPE samples. C. Manufacturer and Instrument Name: Life Technologies Corporation Ion PGM Dx System D. Type of Test or Tests Performed: High-throughput DNA sequencing E. System Descriptions: 1. Device Description: The Ion PGM Dx System is a high throughput DNA sequence analyzer for clinical use. The Ion PGM Dx System consists of the Ion OneTouch Dx Instrument, the Ion OneTouch ES Dx Instrument, the Ion OneTouch Rack Kit, the Ion PGM Dx Chip Minifuge, the Ion PGM Dx Sequencer, the Ion PGM Wireless Scanner, the DynaMag Dx Kit – Tube and Plate, the Ion Torrent Server and the Torrent Suite Dx Software. It is for use with the Ion PGM Dx Library Kit, the Ion OneTouch Dx Template Kit, the Ion PGM Dx Sequencing Kit, and the Ion 318 Dx Chip Kit. The end-user inputs extracted DNA to be sequenced and uses an established assay or provides the Analyte Specific Reagents (ASRs) to develop a sequencing assay that targets their sequence of interest. 2. Principles of Operation: DNA from whole blood samples is isolated using a commercially available extraction method, or DNA and RNA from FFPE tissue samples are isolated and the RNA is reverse transcribed into cDNA. The DNA and/or cDNA is made into amplicon libraries via polymerase chain reaction (PCR) to specifically amplify the regions intended for sequencing while also adding indexing sequences (barcodes) to the amplified products. Employing emulsion PCR, each sample library is templated onto Ion PGM Dx Ion Sphere Particles (ISPs) using the Ion {1} OneTouch Dx Instrument. Templated ISPs are enriched from non-templated ISPs and loaded onto an Ion 318 Dx Chip for sequencing on the Ion PGM Dx Sequencer. The sequencing reaction measures hydrogen ions that are generated during the incorporation of nucleotides into the nascent strand complementary to the template sequence. The resulting signal is translated into base calls which are assembled into reads, which are strings of nucleotide bases in the order found in the original library molecules. Reads are mapped to a reference sequence and mapped reads are assessed at specific nucleotide locations to identify variation from the sequence information in the reference sequence from which variant calls are determined. 3. Modes of Operation: Does the applicant’s device contain the ability to transmit data to a computer, webserver, or mobile device? Yes ☐ or No ☑ Does the applicant’s device transmit data to a computer, webserver, or mobile device using wireless transmission? Yes ☐ or No ☑ 4. Specimen Identification: Sample barcode adapters are used in all runs to assign a unique nucleic acid barcode to each sample DNA allowing the ability to pool up to sixteen (16) barcoded sample libraries into a single sequencing run. 5. Specimen Sampling and Handling: The Ion PGM Dx specimen is a library (or multiple pooled libraries) derived from genomic DNA extracted from peripheral whole blood or DNA or RNA from FFPE samples that is generated and sequenced through the following steps: nucleic acid is extracted from a sample, quantified and qualified, and used to make an indexed barcode library; the library is processed to remove remaining library preparation reagents (e.g. unused primers), normalized as necessary to ensure that each library is equally represented in the pooled samples, and templated onto beads; the templated beads are added onto a sequencing chip and sequenced. 6. Calibration: There is no end-user calibration of the system. During the installation of the Ion PGM Dx System a company representative performs an Installation Qualification/ Operational Qualification/ Performance Qualification (IQ/OQ/PQ) protocol to ensure that software has been installed properly on all items (IQ), the instruments have been calibrated properly (OQ), and performance of the system is adequate (PQ). This includes an actual 2 {2} run on the system and uses all instruments included in the Ion PGM Dx System; the run comprises a control for the Ion OneTouch Dx Instrument, and another control for the Ion PGM Dx Sequencer. Two controls are needed in case one fails so the company representative can troubleshoot at which step the sequencing run is failing. These assessments can only be performed by a company representative and are required after major software updates. The system cannot be used by customers for any other runs until a PQ run has been successfully performed. 7. Quality Control: A double-stranded 221 bp oligonucleotide (CF-1 control fragment) synthesized with the appropriate ends needed for amplification is spiked into every templating reaction setup. One region of the CF-1 oligonucleotide contains a unique sequence to differentiate it from library amplicons. CF-1 is used as a run quality control to ensure that templating and sequencing was performed correctly. A quality control check is run on results of CF-1 sequencing and must be passed before results are released to a user. 8. Software: FDA has reviewed applicant’s Hazard Analysis and Software Development processes for this line of product types: Yes ☐ X ☐ or No ☐ F. Regulatory Information: 1. Regulation section: 21 CFR 862.2265 2. Classification: Class II (special controls). The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to 862.9. 3. Product code: PFF - High throughput DNA sequence analyzer 4. Panel: Toxicology (91) {3} G. Intended Use: 1. Indication(s) for Use: The Ion PGM Dx Instrument System is composed of a sequencing instrument that measures the hydrogen ions that are generated during the incorporation of nucleotides in the DNA sequencing reaction, and the ancillary instrumentation necessary for sample processing. This instrument system is used in conjunction with the instrument-specific Ion PGM Dx Library Kit, Ion OneTouch Dx Template Kit, Ion PGM Dx Sequencing Kit, and Ion 318 Dx Chip Kit, and data analysis software. The Ion PGM Dx Instrument System is intended for targeted sequencing of human genomic DNA (gDNA) from peripheral whole-blood samples and DNA and RNA extracted from formalin-fixed, paraffin-embedded (FFPE) samples. The Ion PGM Dx Instrument System is not intended for whole genome or de novo sequencing. 2. Special Conditions for Use Statement(s): For in vitro diagnostic use. For prescription use only. The instrument was previously legally marketed for use with genomic DNA from whole blood samples under 21 CFR 862.2265; special conditions for use described here are specific to RNA and genomic DNA from FFPE samples only. Special conditions for use of the instrument with genomic DNA from whole blood are included in the device labeling. Special Conditions statement for performance derived from a representative assay using RNA and genomic DNA from FFPE: - The Ion PGM Dx System has been validated to deliver the following using a representative assay: - Sequencing output > 0.7 gigabases - Reads > 3 million - Read length up to 141 base pairs - The system has been validated for the detection of single nucleotide variants, multi-nucleotide variants and up to 18 bp deletions. - The system is designed to deliver qualitative results. - As with any hybridization-based workflow, underlying polymorphisms or mutations in primer binding regions can affect the regions being sequenced and, consequently, the ability to make calls. - The minimal coverage required to call a SNV, MNV or deletion variant is ≥ 347x. The minimal coverage required to call a fusion variant is ≥ 41X. - The Ion PGM Dx System can be used only with the Ion PGM Dx Library Kit, Ion OneTouch Dx Template Kit, Ion PGM Dx Sequencing Kit, and the Ion 318 Dx Chip Kit. 4 {4} H. Substantial Equivalence Information: 1. Predicate Device Name(s) and 510(k) numbers: Illumina MiSeqDx Platform, k123989 (DEN130011) 2. Comparison with Predicate Device: | Similarities and Differences | | | | --- | --- | --- | | Item | Device Ion PGM Dx System (k170299) | Predicate MiSeqDx Platform (k123989) | | Intended Use | For targeted sequencing of human genomic DNA from peripheral whole blood samples and for DNA and RNA extracted from formalin-fixed, paraffin-embedded (FFPE) samples | For targeted sequencing of human genomic DNA from peripheral whole blood samples | | Environment of Use | Clinical Laboratories | Same | | Software | Combined functions software for separate use in IVD and research use only modes | Same | | Specimen Type | Whole blood or formalin fixed paraffin embedded (FFPE) samples | Whole blood samples | | Input Sample | Genomic DNA and cDNA | Genomic DNA | | Technology | Measurement of hydrogen ions that are generated during the incorporation of nucleotides during a DNA sequencing reaction | Measurement of fluorescent signals from labeled nucleotides incorporated during a DNA sequencing reaction | | Software | Combined functions software for separate use in IVD and research use only modes | Combined functions software for separate use in IVD and research use only modes | I. Special Control/Guidance Document Referenced (if applicable): November 19, 2013, De Novo Classification Order for k123989 (DEN130011) {5} J. Performance Characteristics: The instrument was previously legally marketed for use with genomic DNA from whole blood samples under 21 CFR 862.2265. The performance information provided here is specific to RNA and genomic DNA from FFPE samples only. Performance characteristics of the instrument with genomic DNA from whole blood are included in the device labeling. The following table is a summary of the sample type distribution used to characterize instrument performance with the indicated variants in nucleic acid extracted from FFPE samples using a representative assay: | Type of Variant | Number of variants detected by the representative assay | Number of samples tested for detection by sample type | | | | --- | --- | --- | --- | --- | | | | Plasmid/FFPE Sample Blend | FFPE Cell Line or FFPE Cell Line Blend | FFPE Clinical Sample | | MNV | 9 | 9 | 0 | 2 | | SNV | 326 | 329 | 8 | 113 | | 3-bp deletion | 4 | 6 | 0 | 0 | | 6-bp deletion | 4 | 8 | 0 | 0 | | 9-bp deletion | 4 | 8 | 0 | 1 | | 12-bp deletion | 7 | 7 | 0 | 0 | | 15-bp deletion | 7 | 10 | 3 | 23 | | 18-bp deletion | 7 | 8 | 0 | 11 | 1. Analytical Performance: a. Accuracy: A study was performed to demonstrate positive percent agreement (PPA) and negative percent agreement (NPA) concordance between detection of somatic variants in FFPE tumor samples using a representative next-generation sequencing (NGS) assay and detection using validated reference methods. The following reference detection methods were used: - A validated NGS assay to detect SNV and deletion hotspot variants to detect SNV and deletion hotspot variants - FISH/IHC testing to detect RNA fusions {6} Variants detected by the representative assay for which the reference method testing failed and did not yield a valid result were not included in the calculations. Accuracy data was analyzed by the following: each variant location, bins (or categories of variants, i.e., RNA fusions, simple and complex single nucleotide variants (SNVs) and deletions), and each FFPE sample. The results are shown in the following tables: Positive Percent Agreement (PPA) | PPA Measure | Excluding no calls | | Including no calls | | | --- | --- | --- | --- | --- | | | Percent agreement | 95% CI | Percent agreement | 95% CI | | Variant | 98.5% (195/198) | 95.6%, 99.7% | 98.5% (195/198) | (95.6%, 99.7%) | | Bin | 97.2% (176/181) | 93.7%, 99.1% | 97.2% (176/181) | (93.7%, 99.1%) | | Sample | 96.9% (158/163) | 93.0%, 99.0% | 96.9% (158/163) | (93.0%, 99.0%) | Negative Percent Agreement (NPA) | NPA Measure | Excluding no calls | | Including no calls | | | --- | --- | --- | --- | --- | | | Percent agreement | 95% CI | Percent agreement | 95% CI | | Variant | 100.0% (118,155/118,159) | 99.99%, 100.0% | 96.8% (118,155/122,012) | (96.7%, 96.9%) | | Bin | 99.8% (942/944) | 99.2%, 100.0% | 70.0% (657/939) | (66.9%, 72.9%) | | Sample | 98.4% (124/126) | 94.4%, 99.8% | 23.4% (29/124) | (16.3%, 31.8%) | {7} Overall Percent Agreement (OPA) | OPA Measure | Excluding no calls | | Including no calls | | | --- | --- | --- | --- | --- | | | Percent agreement | 95% CI | Percent agreement | 95% CI | | Variant | 100% (118,350/118,357) | 99.99%, 100.0% | 96.8% (118,350/122,210) | (96.74%, 96.94%) | | Bin | 99.4% (1,118/1,125) | 98.72%, 99.75% | 74.4% (833/1,120) | (71.71%, 76.91%) | | Sample | 97.6% (282/289) | 95.07%, 99.02% | 65.2% (187/287) | (59.34%, 70.66%) | b. Precision/Reproducibility: Two studies were conducted to assess reproducibility: Sample Reproducibility Two wild-type (WT) and ten variant-positive samples were evaluated at 4 testing sites, each with 4 Ion PGM Dx systems and 4 operators, to determine the reproducibility and repeatability of detection for multiple variants using a representative assay. Each sample was tested 8 times at each site, for a total of 32 replicates per sample, or 768 sample sequencing results (12 samples × 32 replicates × 2 library types (RNA and DNA)). The call rate, no call rate, positive call rate, negative call rate, and within-run repeatability were computed at each variant location of interest. Including no calls and excluding known positive variant locations, the negative call rate at each clinical variant location for all samples was 100%. Including no calls, all positive call rates from positive variant locations were >84%. Excluding no calls and combining data across all study samples, the estimate of repeatability was 100% for DNA variants and 87.5% for the RNA variants. The lower limit of the 95% two-sided confidence interval (CI) for repeatability exceeded 96% at all variant locations. Including no calls from the data, the estimate of repeatability was 100% at 218 out of 605 variant locations, 94–99.9% at 186 out of 605 variant locations, and 71.6–93.9% at 184 out of 605 variant locations. Including no calls, the lower limit of the 95% two-sided confidence interval for repeatability exceeded 64.6% at all variant locations. The lower limit of the 95% two-sided confidence interval (CI) for repeatability exceeded 96% at all variant locations. {8} Sample reproducibility results are shown in the two tables below (data from one study split into two tables): | Sample | Variant identification | Variant location | # of valid sample results (N) | # of positive calls (A) | # of negative calls (B) | # of No Calls (C) | Positive call rate + 95% CI | | | --- | --- | --- | --- | --- | --- | --- | --- | --- | | | | | | | | | Including no calls (A/N) | Excluding no calls (A/(A+B)) | | B | COSM6223 | EGRF Exon19del | 32 | 32 | 0 | 0 | 100% (89.1%, 100%) | 100% (89.1%, 100%) | | B | COSM763 | PIK3C A E545K | 32 | 32 | 0 | 0 | 100% (89.1%, 100%) | 100% (89.1%, 100%) | | C | ROS1 | N/A | 32 | 30 | 2 | 0 | 93.8% (79.2%, 99.2%) | 93.8% (79.2%, 99.2%) | | D | COSM6225 | EGFR Exon19del | 32 | 32 | 0 | 0 | 100% (89.1%, 100%) | 100% (89.1%, 100%) | | E | COSM476 | BRAF V600E | 32 | 32 | 0 | 0 | 100% (89.1%, 100%) | 100% (89.1%, 100%) | | F | COSM521 | KRAS G12D | 32 | 30 | 0 | 2 | 93.8% (79.2%, 99.2%) | 100% (88.4%, 100%) | | F | COSM29313 | PIK3C A M1043I | 32 | 30 | 0 | 2 | 93.8% (79.2%, 99.2%) | 100% (88.4%, 100%) | | G | COSM6224 | EGFR L858R | 32 | 32 | 0 | 0 | 100% (89.1%, 100%) | 100% (89.1%, 100%) | | J | COSM87298 | KRAS Q61K | 32 | 32 | 0 | 0 | 100% (89.1%, | 100% (89.1%, | | | | | | | | | | 100%) | | | | | | | | | | 100%) | {9} | Sample | Variant identification | Variant location | # of valid sample results (N) | # of positive calls (A) | # of negative calls (B) | # of No Calls (C) | Positive call rate + 95% CI | | | --- | --- | --- | --- | --- | --- | --- | --- | --- | | | | | | | | | Including no calls (A/N) | Excluding no calls (A/(A+B)) | | | | | | | | | 100%) | 100%) | | J | COSM172423 | ERBB3 V104M | 32 | 32 | 0 | 0 | 100% (89.1%, 100%) | 100% (89.1%, 100%) | | K | COSM775 | PIK3 H1047R | 30 | 29 | 0 | 1 | 96.7% (82.8%, 99.9%) | 100% (88.1%, 100%) | | M | COSM715 | FGR3 S249C | 32 | 32 | 0 | 0 | 100% (89.1%, 100%) | 100% (89.1%, 100%) | {10} | Sample | Variant identification | Variant location | # of valid sample results (N) | # of positive calls (A) | # of negative calls (B) | # of No Calls (C) | Negative call rate + 95% CI | | Within-run repeatability + 95% CI | | | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | | | | | | | | | Including no calls (B/N) | Excluding no calls (B/(A+B)) | Including no calls | Excluding no calls | | B | COSM622 3 | EGRF Exon19 del | 32 | 32 | 0 | 0 | 0% (0%, 10.9%) | 0% (0%, 10.9%) | 100% (79.4%, 100%) | 100% (79.4%, 100%) | | B | COSM763 | PIK3CA E545K | 32 | 32 | 0 | 0 | 0% (0%, 10.9%) | 0% (0%, 10.9%) | 100% (79.4%, 100%) | 100% (79.4%, 100%) | | C | ROS1 | N/A | 32 | 30 | 2 | 0 | 6.3% (0.8%, 20.8%) | 6.3% (0.8%, 20.8%) | 87.5% (61.7%, 98.4%) | 87.5% (61.7%, 98.4%) | | D | COSM622 5 | EGFR Exon19 del | 32 | 32 | 0 | 0 | 0% (0%, 10.9%) | 0% (0%, 10.9%) | 100% (79.4%, 100%) | 100% (79.4%, 100%) | | E | COSM476 | BRAF V600E | 32 | 32 | 0 | 0 | 0% (0%, 10.9%) | 0% (0%, 10.9%) | 100% (79.4%, 100%) | 100% (79.4%, 100%) | | F | COSM521 | KRAS G12D | 32 | 30 | 0 | 2 | 0% (0%, 10.9%) | 0% (0%, 11.6%) | 87.5% (61.7%, 98.4%) | 100% (76.8%, 100%) | | F | COSM293 13 | PIK3CA M1043I | 32 | 30 | 0 | 2 | 0% (0%, 10.9%) | 0% (0%, 11.6%) | 87.5% (61.7%, 98.4%) | 100% (76.8%, 100%) | | G | COSM622 4 | EGFR L858R | 32 | 32 | 0 | 0 | 0% (0%, | 0% (0%, | 100% (79.4%, | 100% (79.4%, | | | | | | | | | 10.9%) | | | | | H | COSM622 6 | EGFR L858R | 32 | 32 | 0 | 0 | 0% (0%, 10.9%) | 0% (0%, 11.6%) | 100% (79.4%, 100%) | 100% (79.4%, 100%) | | I | COSM293 14 | PIK3CA M1043I | 32 | 32 | 0 | 0 | 0% (0%, 10.9%) | 0% (0%, 11.6%) | 100% (79.4%, 100%) | 100% (79.4%, 100%) | {11} | Sample | Variant identification | Variant location | # of valid sample results (N) | # of positive calls (A) | # of negative calls (B) | # of No Calls (C) | Negative call rate + 95% CI | | Within-run repeatability + 95% CI | | | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | | | | | | | | | Including no calls (B/N) | Excluding no calls (B/(A+B)) | Including no calls | Excluding no calls | | | | | | | | | 10.9%) | 10.9%) | 100%) | 100%) | | J | COSM872 98 | KRAS Q61K | 32 | 32 | 0 | 0 | 0% (0%, 10.9%) | 0% (0%, 10.9%) | 100% (79.4%, 100%) | 100% (79.4%, 100%) | | J | COSM172 423 | ERBB3 V104M | 32 | 32 | 0 | 0 | 0% (0%, 10.9%) | 0% (0%, 10.9%) | 100% (79.4%, 100%) | 100% (79.4%, 100%) | | K | COSM775 | PIK3 H1047 R | 30 | 29 | 0 | 1 | 0% (0%, 11.6%) | 0% (0%, 11.9%) | 93.3% (68.1%, 99.8%) | 100% (76.8%, 100%) | | M | COSM715 | FGR3 S249C | 32 | 32 | 0 | 0 | 0% (0%, 10.9%) | 0% (0%, 10.9%) | 100% (79.4%, 100%) | 100% (79.4%, 100%) | # Assay Reproducibility The reproducibility and repeatability of a representative assay on the Ion PGM Dx System was evaluated for 30 representative variants from 18 DNA and 9 RNA samples. The study was designed to evaluate within-run precision performance (repeatability) and variability across sites, operators, and instrument platforms (reproducibility). Six of the 18 DNA samples were mixtures of plasmid and clinical DNA. Seven of the 12 deletion variants were represented by these plasmid blends. All other variant types were represented by clinical sample DNA. Due to the large number of variants detected by the test and the rarity of some of the variants, a representative variant approach was used. Variants were selected in the following categories: {12} | Variant Category | No. of Plasmid Blends Used | No. of Clinical Samples Used | | --- | --- | --- | | 6-bp deletion | 6 | 0 | | 9-bp deletion | 4 | 2 | | 15-bp deletion | 2 | 4 | | 18-bp deletion | 2 | 4 | | Simple SNV | 0 | 8 | | Complex SNVs and MNPs* | 0 | 6 | | Fusion | 0 | 12 | *Including SNVs in di- or tri-nucleotide repeat regions and SNVs in high-GC (>60%) or low-GC (<40%) content regions Two of the 18 DNA samples were WT at all locations, and the remaining 16 contained DNA from one or more DNA variants. One of the 9 RNA samples contained no fusion molecules, and the remaining 8 samples each contained RNA from an RNA variant. Each pre-extracted DNA or RNA sample was sequenced at 4 sites by 2 operators on 2 systems at each site. At each site, operators were grouped into 2 pairs, with each pair being assigned to 2 instrument systems, and each pair being responsible for testing of 9 of the DNA samples and all 9 of the RNA samples. Samples were run in duplicate using 2 different reagent lots at 3 of the study sites and on all 3 reagent lots at 1 study site. The design resulted in a total of 72 test determinations per DNA sample (3 sites $\times$ 2 lots $\times$ 2 operators $\times$ 2 systems $\times$ 2 replicates + 1 site $\times$ 3 lots $\times$ 2 operators $\times$ 2 systems $\times$ 2 replicates). Because there were half as many RNA samples as DNA samples, each RNA sample was tested twice as many times (n=144). In total, at least 1,296 sequencing reactions were performed, and all variant locations were assessed for each sample. {13} The reproducibility results are summarized in the following table: | Description | No. of variant samples | Call rate excluding no calls | | Call rate including no calls | | Call rate including no calls and insulids | | | --- | --- | --- | --- | --- | --- | --- | --- | | | | Mean | Median | Mean | Median | Mean | Median | | DNApositive variants (positive calls) | 46 | 96.60% | 97.10% | 94.50% | 95.80% | 94.50% | 95.80% | | RNApositive variants (positive calls) | 6 | 94.80% | 95.50% | 94.80% | 95.50% | 94.80% | 95.50% | | WTDNA variant locations | 872 | 96.10% | 95.00% | 96.10% | 95.00% | 90.70% | 93.10% | | WTRNA variant locations | 170 | 99.30% | 99.30% | 99.30% | 99.30% | 99.30% | 99.30% | Excluding no calls, the estimate of repeatability at each DNA variant location across all the samples was $\geq 98.8\%$ (95% CI lower limit of $\geq 97.5\%$ ). The coefficient of variation (CV) across all DNA clinical variants ranged from $9.8\%$ to $39\%$ . The highest CVs (24.9–39.2%) were observed for one specific variant (BRAF V600E variant); the root cause for the higher percent CV for this sample was not definitively determined. The CVs for other specific variants (e.g., EGFR L858R variant and EGFR deletion variants) ranged from $9.8\%$ to $11.3\%$ , and $11.2\%$ to $25.5\%$ respectively. Excluding no calls, the estimate of repeatability at each RNA clinical variant location was $94.4\%$ . The CV across all RNA locations ranged from $72\%$ to $78\%$ . # c. Linearity: Not applicable. # d. Carryover: A study was performed to evaluate the potential for inter-run and intra-run sample carryover (cross-contamination). Eight FFPE cell line samples were evaluated to determine the percentage of false positive results caused by contamination from one sample to another within the same sequencing run and by contamination from a previous run on the same instrument system. Samples that were WT and mutant were tested in consecutive sequencing runs on the same instruments, and 5 DNA variant locations and 2 RNA variant locations that were expected to be WT for a sample were evaluated for contamination. Out of 100 DNA and 80 RNA data points analyzed, no false positive results were reported in the DNA variants, and 1 false positive result was reported in an RNA variant. Therefore, the false-positive rate at the DNA variant {14} locations was 0% (0/100) and the false-positive rate at the RNA variant locations was 1.25% (1/80). e. Interfering Substances: Endogenous and Exogenous Substances To assess the impact of endogenous and exogenous interfering substances on the instrument, a representative assay was evaluated in the presence and absence of potential interferents. Six potentially interfering substances were evaluated, including five potentially interfering substances used to extract DNA and RNA from FFPE tissue samples, in addition to hemoglobin, a potentially interfering endogenous substance. A total of 8 FFPE samples (1 WT and 7 mutants) with 6 replicates each were processed through the entire assay workflow, and the samples were spiked with additional concentrations or amounts of the listed substances at the relevant processing step, as shown in the table. Replicates of a control sample with no spike-ins were also analyzed. The concordance between variant calls in samples with and without interfering substances was computed for each substance under investigation. Amounts of the tested substances and the steps at which they were added during the assay are summarized in the table below: | Potential interfering substance | Step | Amount of substance | Call Rate | | --- | --- | --- | --- | | Paraffin | At the deparaffinization step, extra paraffin was added to the xylene bath that contained 250mL of xylene. | 4X of normally expected levels | | | Xylene | Extra xylene was added into the ethanol bath that contained 250mL of ethanol. | 6X of normally expected residual volume | | | Ethanol | Extra ethanol was added into the Protease digestion step before digestion | >4X of normally expected residual volume | | | Hemoglobin | After deparaffinization, hemoglobin was added to the Digestion Buffer used to pre-wet the tissue section | 4 mg/mL | | | Protease | Extra Protease was added into the reaction after the digestion step and before column purification | >10X of expected residual Protease after the heat-kill step | | {15} | Wash buffer | Wash buffer used to isolate DNA and RNA from deparaffinized and digested samples was added into an aliquot of Dilution Solution, which was subsequently used to dilute the RNA and DNA to the appropriate concentration before library preparation. | 1% wash buffer (equivalent to ~10% wash buffer carried over into eluate) | | | --- | --- | --- | --- | | Control | Tissue sections were processed using the standard protocol, without the addition of any potentially interfering substances. | NA | | No false positive or false negative results were observed with any of the tested exogenous interferents. With no calls excluded, for each potential interferent, the positive and negative concordance with the control condition across all samples was 100%, and the overall concordance with the control condition across all samples was 100%. With no calls excluded, the results of testing with hemoglobin showed positive concordance with the control condition of 100% (only samples with a positive control condition were analyzed), negative concordance of 99.99%, and overall concordance of 99.99%. For one sample, 1 of 6 replicates of a control condition was a false negative. For another sample, 1 of 6 replicates of a hemoglobin condition was a false positive. ## Necrosis It is currently unknown whether necrotic tissue in the region of interest in FFPE tissue samples interferes with sequencing of FFPE samples. Therefore, users should macrodissect highly necrotic areas or select alternate samples if possible. ## 2. Other Supportive Instrument Performance Data Not Covered Above: ## Tissue input study: Sixty slide-mounted FFPE samples were analyzed to determine if samples extracted using the Ion Torrent Dx Total Nucleic Acid Isolation Kit yield DNA and RNA at the concentrations required by a representative assay when tissue input requirements are met. The test requires DNA at a concentration of ≥0.83 ng/μL and RNA at a concentration of ≥1.43 ng/μL. Thirty resection samples with ≥20% tumor content were prepared without macrodissection, 15 resection samples with <20% to ≥10% tumor cell content were macrodissected, and 15 samples were collected by core needle biopsy (CNB). For the resection samples, 2 × 5 μm sections were used per extraction. For CNBs, 9 × 5 μm sections were used per extraction. DNA and RNA concentrations were determined using 16 {16} the Ion Torrent Dx DNA and RNA Quantification Kits, respectively. No sequencing was performed on the extracted samples. Of the 60 samples tested, 98.3% (59/60) had a DNA concentration of ≥0.83 ng/μL and an RNA concentration of ≥1.43 ng/μL. One CNB sample failed the minimum DNA and RNA concentration specifications, with values of 0.52 ng/μL and 1.23 ng/μL respectively. ## DNA input study: To evaluate the amount of input FFPE-excted DNA and RNA required to reproducibly make accurate variant calls, eight cell-line samples were prepared as FFPE sections, and DNA and RNA were extracted and quantified from multiple sections from each cell line for blending and testing. Sample blends were prepared with known variants at various DNA and RNA input-level combinations within the range of 5–15 ng. The DNA and RNA blends had a target allele frequency of 15% for SNVs and deletions and target fusion reads of 300–600 for a specific variant (ROS1). A total of 540 individual DNA and RNA libraries were tested, including positive controls and NTC controls, with 6 replicate libraries each for DNA and RNA per test condition. The study demonstrated a 100% positive variant call rate within the input range tested, supporting an input amount of 10 ng each for DNA and RNA for a specific representative assay. The negative variant call rate was >95% for all except 4 sample and DNA/RNA input-level combinations. All cases with a negative variant call rate of <95% were due to no calls, 3 of which occurred with a DNA or RNA input amount of 5 ng and 1 of which occurred in a single sample with DNA and RNA inputs of 10 ng each. There were no false-positive calls. Additionally, 4 clinical samples prepared as FFPE sections were tested: two samples containing DNA variants and two containing the specific CD74-ROS1 fusion variant. The DNA variant samples were paired with wild-type RNA from the same sample at various input combinations within the range of 5–15 ng, and the RNA variant samples were paired with wild-type DNA at input combinations within the same range. The study demonstrated positive and negative call rates of >95% for the DNA variants at all input combinations, and 100% for one of the CD74-ROS1 fusion variants at all input combinations. The second CD74-ROS1 clinical sample showed 100% negative call rates for all test conditions, and 100% positive call rates except for Test Condition 4 (8.5 ng RNA/15 ng DNA), where the call rate was 83%, and Test Condition 6 (15 ng RNA/15 ng DNA), where the call rate was 50%. ## Sample indexing study: Sample barcode adapters are used in all runs to assign a unique nucleic acid barcode to each sample DNA, allowing the ability to pool up to sixteen (16) barcoded sample 17 {17} 18 libraries into a single Ion PGM Dx sequencing run. All 16 barcode adapters were tested with the SVA panel and genomic DNA samples (NA12878 and NA19240) to determine variant call (SNV and Indel) reproducibility between adapter sequences. Sample results were compared to reference database sequences for the NA12878 and NA19240 genomes at all variant and non-variant positions. Comparison of data from 60 runs consisting of 4 sequencing reactions (2 DNA samples × 2 barcodes each) and 24 runs consisting of 16 sequencing reactions (2 DNA samples × 8 barcodes each) indicate that the barcode adapter used and the number of samples pooled together had no effect on variant call reproducibility. Indexing chemistry and software is identical for all libraries, independent of the type and source of nucleic acid. The previously conducted indexing study for indexing genomic DNA extracted from whole blood was therefore used to support indexing of FFPE RNA and DNA samples. ## K. Proposed Labeling: Labeling satisfies the requirements of 21 CFR 809.10, 21 CFR 801.109, including an appropriate prescription statement as required by 21 CFR 801.109(b), and the special controls for this type of device. ## L. Conclusion: The submitted information in this premarket notification is complete and supports a substantial equivalence decision.
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...