# Fortress™ Pedicular Fixation System (K170268)

_Spineology, Inc. · NKB · Apr 14, 2017 · Orthopedic · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K170268

## Device Facts

- **Applicant:** Spineology, Inc.
- **Product Code:** [NKB](/productcode/NKB.md)
- **Decision Date:** Apr 14, 2017
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 888.3070
- **Device Class:** Class 2
- **Review Panel:** Orthopedic
- **Attributes:** Therapeutic

## Indications for Use

The Spineology Fortress™ Pedicular Fixation System is intended for posterior, non-cervical fixation as an adjunct to fusion in skeletally mature patients for the following indications: degenerative disc disease (defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies) spondylolisthesis; trauma (i.e., fracture or dislocation); spinal stenosis; curvatures (i.e., scoliosis, kyphosis and/or lordosis); tumor; pseudoarthrosis; and/ or failed previous fusion.

## Device Story

Fortress™ Pedicular Fixation System is a spinal implant assembly used by surgeons in clinical settings to provide immobilization and stabilization of thoracic, lumbar, and sacral spinal segments. The system comprises titanium alloy pedicle screws (optionally hydroxyapatite-coated), cobalt chrome rods, and titanium alloy ConneX™ cross and rod connectors. Surgeons select and assemble these components to fit patient-specific anatomical requirements, placing screws via various trajectories (transpedicular, caudocephalad, or lateral). The device acts as an internal fixation construct to support spinal fusion. By stabilizing the affected spinal segments, the system facilitates fusion, potentially reducing pain and correcting deformities associated with degenerative, traumatic, or pathological spinal conditions.

## Clinical Evidence

No clinical data provided. Substantial equivalence is based on non-clinical bench testing, including static and dynamic compression bending, flexural grip, and static cantilever bending per ASTM F1717, F1798, and F2193. Bacterial endotoxin testing was performed per ANSI/AAMI ST-72:2011.

## Technological Characteristics

System components include titanium alloy screws and connectors, and cobalt chrome rods. Mechanical testing performed per ASTM F1717, ASTM F1798, ASTM F2193. Bacterial endotoxin testing per ANSI/AAMI ST-72:2011. System is a modular pedicle screw construct for posterior spinal fixation.

## Regulatory Identification

(1) Rigid pedicle screw systems are comprised of multiple components, made from a variety of materials that allow the surgeon to build an implant system to fit the patient's anatomical and physiological requirements. Such a spinal implant assembly consists of a combination of screws, longitudinal members (e.g., plates, rods including dual diameter rods, plate/rod combinations), transverse or cross connectors, and interconnection mechanisms (e.g., rod-to-rod connectors, offset connectors).(2) Semi-rigid systems are defined as systems that contain one or more of the following features (including but not limited to): Non-uniform longitudinal elements, or features that allow more motion or flexibility compared to rigid systems.

## Special Controls

*Classification.* (1) Class II (special controls), when intended to provide immobilization and stabilization of spinal segments in skeletally mature patients as an adjunct to fusion in the treatment of the following acute and chronic instabilities or deformities of the thoracic, lumbar, and sacral spine: severe spondylolisthesis (grades 3 and 4) of the L5-S1 vertebra; degenerative spondylolisthesis with objective evidence of neurologic impairment; fracture; dislocation; scoliosis; kyphosis; spinal tumor; and failed previous fusion (pseudarthrosis). These pedicle screw spinal systems must comply with the following special controls:(i) Compliance with material standards;
(ii) Compliance with mechanical testing standards;
(iii) Compliance with biocompatibility standards; and
(iv) Labeling that contains these two statements in addition to other appropriate labeling information:
 “Warning: The safety and effectiveness of pedicle screw spinal systems have been established only for spinal conditions with significant mechanical instability or deformity requiring fusion with instrumentation. These conditions are significant mechanical instability or deformity of the thoracic, lumbar, and sacral spine secondary to severe spondylolisthesis (grades 3 and 4) of the L5-S1 vertebra, degenerative spondylolisthesis with objective evidence of neurologic impairment, fracture, dislocation, scoliosis, kyphosis, spinal tumor, and failed previous fusion (pseudarthrosis). The safety and effectiveness of these devices for any other conditions are unknown.”
“Precaution: The implantation of pedicle screw spinal systems should be performed only by experienced spinal surgeons with specific training in the use of this pedicle screw spinal system because this is a technically demanding procedure presenting a risk of serious injury to the patient.”
(2) Class II (special controls), when a rigid pedicle screw system is intended to provide immobilization and stabilization of spinal segments in the thoracic, lumbar, and sacral spine as an adjunct to fusion in the treatment of degenerative disc disease and spondylolisthesis other than either severe spondylolisthesis (grades 3 and 4) at L5-S1 or degenerative spondylolisthesis with objective evidence of neurologic impairment. These pedicle screw systems must comply with the following special controls:
(i) The design characteristics of the device, including engineering schematics, must ensure that the geometry and material composition are consistent with the intended use.
(ii) Non-clinical performance testing must demonstrate the mechanical function and durability of the implant.
(iii) Device components must be demonstrated to be biocompatible.
(iv) Validation testing must demonstrate the cleanliness and sterility of, or the ability to clean and sterilize, the device components and device-specific instruments.
(v) Labeling must include the following:
(A) A clear description of the technological features of the device including identification of device materials and the principles of device operation;
(B) Intended use and indications for use, including levels of fixation;
(C) Identification of magnetic resonance (MR) compatibility status;
(D) Cleaning and sterilization instructions for devices and instruments that are provided non-sterile to the end user; and
(E) Detailed instructions of each surgical step, including device removal.
(3) Class II (special controls), when a semi-rigid system is intended to provide immobilization and stabilization of spinal segments in the thoracic, lumbar, and sacral spine as an adjunct to fusion for any indication. In addition to complying with the special controls in paragraphs (b)(2)(i) through (v) of this section, these pedicle screw systems must comply with the following special controls:
(i) Demonstration that clinical performance characteristics of the device support the intended use of the product, including assessment of fusion compared to a clinically acceptable fusion rate.
(ii) Semi-rigid systems marketed prior to the effective date of this reclassification must submit an amendment to their previously cleared premarket notification (510(k)) demonstrating compliance with the special controls in paragraphs (b)(2)(i) through (v) and paragraph (b)(3)(i) of this section.

## Predicate Devices

- Fortress™ Pedicular Fixation System ([K152148](/device/K152148.md))
- Fortress™ Pedicular Fixation System ([K140010](/device/K140010.md))
- S4™ Spinal System ([K123352](/device/K123352.md))
- CD Horizon™ Spinal System ([K162494](/device/K162494.md))
- Arsenal™ Spinal Fixation System ([K152968](/device/K152968.md))
- Synthes Matrix™ System ([K120838](/device/K120838.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

Spineology Inc. Ms. Jacqueline A. Hauge Regulatory Affairs Manager 7800 3rd Street North. Suite 600 Saint Paul, Minnesota 55128

April 14, 2017

Re: K170268

Trade/Device Name: Fortress™ Pedicular Fixation System Regulation Number: 21 CFR 888.3070 Regulation Name: Thoracolumbosacral pedicle screw system Regulatory Class: Class II Product Code: NKB Dated: January 26, 2017 Received: January 27, 2017

Dear Ms. Hauge:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-

Image /page/0/Picture/11 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo is a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, stacked on top of each other.

{1}------------------------------------------------

related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely,

Image /page/1/Picture/9 description: The image shows the name "Mark N. Melkerson -S" in a clear, sans-serif font. The text is horizontally aligned and appears to be part of a document or label. The letters are uniformly sized and spaced, contributing to the legibility of the name.

Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

# Indications for Use

510(k) Number (if known) K170268

Device Name Fortress™ Pedicular Fixation System

### Indications for Use (Describe)

The Spineology Fortress™ Pedicular Fixation System is intended for posterior, non-cervical fixation as an adjunct to fusion in skeletally mature patients for the following indications: degenerative disc disease (defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies) spondylolisthesis; trauma (i.e., fracture or dislocation); spinal stenosis; curvatures (i.e., scoliosis, kyphosis and/or lordosis); tumor; pseudoarthrosis; and/ or failed previous fusion.

| Type of Use (Select one or both, as applicable)                                             |                                                                                           |
|---------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------|
| <div> <span>            Prescription Use (Part 21 CFR 801 Subpart D)         </span> </div> | <div> <span>           Over-The-Counter Use (21 CFR 801 Subpart C)         </span> </div> |

CONTINUE ON SEPARATE PAGE IF NEEDED

# CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/11 description: The image shows the word "Spineology" in a bold, sans-serif font. There is a horizontal line underneath the word. To the right of the word is a graphic of a cluster of dots, with one larger dot in the center and several smaller dots surrounding it. The dots are arranged in a circular pattern.

| Date Prepared:  | April 11, 2017                                                                                                                |
|-----------------|-------------------------------------------------------------------------------------------------------------------------------|
| Submitter:      | Spineology Inc.<br>7800 3rd Street North<br>Suite 600<br>Saint Paul, MN 55128<br>Establishment Registration Number:  2135156  |
| Contact Person: | Jacqueline A. Hauge<br>Regulatory Affairs Manager<br>Phone: 651.256.8534<br>Fax: 651.256.8505<br>Email: jhauge@spineology.com |

# Device Name and Classification

| Trade Name:          | Fortress™ Pedicular Fixation System     |
|----------------------|-----------------------------------------|
| Common Name:         | Thoracolumbosacral Pedicle Screw System |
| Classification Name: | Orthosis, Spinal Pedicle Fixation       |
| Product Codes:       | NKB                                     |
| Regulatory Class:    | Class II                                |
| Regulation Number:   | 21 CFR 888.3070                         |
| Panel:               | Orthopedic                              |

## Predicate Devices

| Primary:    | K152148 | Fortress™ Pedicular Fixation System (Spineology Inc.)  |
|-------------|---------|--------------------------------------------------------|
| Additional: | K140010 | Fortress™ Pedicular Fixation System (Spineology Inc.)  |
|             | K123352 | S4™ Spinal System (Aesculap Implant Systems, Inc.)     |
|             | K162494 | CD Horizon™ Spinal System (Medtronic Sofamor Danek)    |
|             | K152968 | Arsenal™ Spinal Fixation System (Alphatec Spine, Inc.) |
|             | K120838 | Synthes Matrix™ System (Synthes Spine)                 |

#### I. Purpose

The purpose of this submission is to obtain FDA clearance for the addition of Spineology's ConneX™ Cross Connector, ConneX™ Rod Connector, and 4.75mm cobalt chrome Straight Rod devices to its Fortress Pedicular Fixation System.

#### II. FDA Clearance History

| Device Name                         | 510(k) # | Purpose of 510(k)                                     |
|-------------------------------------|----------|-------------------------------------------------------|
| Fortress™ Pedicular Fixation System | K140010  | Initial FDA Clearance                                 |
|                                     | K152148  | Addition of Hydroxyapatite (HA) coated pedicle screws |

{4}------------------------------------------------

#### Device Description III.

#### A. Fortress™ System

The Spineology Fortress Pedicular Fixation System consists of screws (titanium alloy), curved and straight rods (cobalt chrome), adjustable ConneX Cross Connector (titanium alloy), and fixed ConneX Rod Connector (titanium alloy) devices to allow the surgeon to build an implant system to fit the patient's anatomical and physiological requirements. The Fortress System screws are available with or without a hydroxyapatite coating. The system is intended to provide immobilization and stabilization of spinal segments in skeletally mature patients as an adjunct to fusion in the treatment of acute and chronic instabilities or deformities of the thoracic, lumbar, and sacral spine. The screws can be placed in the pedicles in a variety of trajectories ranging from the standard anatomic transpedicular path projected medially toward the ventral vertebral body, to a caudocephalad path sagittally and a laterally directed path in the transverse plane.

#### B. ConneX™ Cross Connectors

Spineology Connex Cross Connector devices are transversely-placed implants that are intended to connect the rod on one side of the spinal construct to the rod on the other side. These devices are manufactured from titanium alloy and are adjustable to allow the surgeon to build an implant system to fit the patient's anatomical and physiological requirements. ConneX Cross Connector devices are intended to be used with Spineology's Fortress™, Threshold V2™, and Palisade™ Pedicular Fixation Systems which are intended to provide immobilization and stabilization of spinal segments in skeletally mature patients as an adjunct to fusion in the treatment of acute and chronic instabilities or deformities of the thoracic, lumbar, and sacral spine.

#### C. ConneX™ Rod Connectors

Spineology Connex Rod Connector devices are implants that are intended to connect two rods in a spinal construct. These devices are manufactured from titanium alloy and the components allow the surgeon to build an implant system to fit the patient's anatomical and physiological requirements. Connex Rod Connector devices are intended to be used with Spineology's Fortress™, Threshold™, Threshold V2™, and Palisade™ Pedicular Fixation Systems which are intended to provide immobilization and stabilization of spinal segments in skeletally mature patients as an adjunct to fusion in the treatment of acute and chronic instabilities or deformities of the thoracic, lumbar, and sacral spine.

#### IV. Indications for Use

The Spineology Fortress™ Pedicular Fixation System is intended for posterior, non-cervical fixation as an adjunct to fusion in skeletally mature patients for the following indications: degenerative disc disease (defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies) spondylolisthesis; trauma (i.e., fracture or dislocation); spinal stenosis; curvatures (i.e., scoliosis and/or lordosis); tumor; pseudoarthrosis; and/or failed previous fusion.

#### V. Comparison to Predicate

When compared to the predicate devices, the Connex Cross Connector, ConneX Rod Connector, and Straight Rod devices have the same:

- Intended Use ●
- Indications for Use
- Fundamental Scientific Technology ●
- Principle of Operation
- Materials of Construction

{5}------------------------------------------------

#### VI. Non-Clinical Testing

#### A. ConneX™ Cross Connectors and ConneX™ Rod Connectors

The following mechanical testing was conducted on representative worst case ConneX Cross Connector and Connex Rod Connector devices in accordance with FDA Guidance: Spinal System 510(k)s (2004) and applicable American Society for Testing and Materials (ASTM) standards:

### ASTM F1717

ASTM F1798 • Flexural Grip • ASTM F2193 • Static Cantilever Bending

- Static Compression Bending • Dynamic Compression Bending
Bacterial endotoxin testing (BET) was conducted in accordance with ANSI/AAMI ST-72:2011.

#### B. Straight Rods

The following mechanical testing was conducted on representative Straight Rod devices in accordance with FDA Guidance: Spinal System 510(k)s (2004) and applicable American Society for Testing and Materials (ASTM) standards:

## ASTM F1717-15

- · Static Compression Bending
Bacterial endotoxin testing (BET) was conducted in accordance with ANSI/AAMI ST-72:2011.

#### VII. Conclusion

Based on the indications for use, technological characteristics, and comparison to the predicate devices, the ConneX Cross Connector, Connex Rod Connector, and Straight Rod devices have been shown to be substantially equivalent to the legally marketed predicate devices.

---

**Source:** [https://fda.innolitics.com/device/K170268](https://fda.innolitics.com/device/K170268)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).