← Product Code [NPM](/productcode/NPM) · K170245

# DSM Biomedical Dental Bone Graft (K170245)

_Kensey Nash Corporation Dba Dsm Biomedical · NPM · May 16, 2017 · Dental · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K170245

## Device Facts

- **Applicant:** Kensey Nash Corporation Dba Dsm Biomedical
- **Product Code:** [NPM](/productcode/NPM.md)
- **Decision Date:** May 16, 2017
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 872.3930
- **Device Class:** Class 2
- **Review Panel:** Dental
- **Attributes:** Therapeutic

## Indications for Use

DSM Biomedical Dental Bone Graft is indicated for: - Augmentation or reconstructive treatment of the alveolar ridge - Filling of infrabony periodontal defects - Filling of defects after root resection, apicoectomy, and cystectomy - Filling of extraction sockets to enhance preservation of the alveolar ridge - Elevation of the maxillary sinus floor - Filling of periodontal defects in conjuncts intended for Guided Tissue Regeneration (GTR) and Guided Bone Regeneration (GBR) - Filling of peri-implant defects in conjunction with products intended for Guided Bone Regeneration (GBR)

## Device Story

DSM Biomedical Dental Bone Graft is a non-pyrogenic, porous, osteoconductive bone mineral matrix derived from porcine bone. It serves as a scaffold for bone formation in periodontal, oral, and maxillofacial surgical procedures. The device consists of cancellous granules (0.25–1.0mm) with trabecular, interconnecting macro and micro pores. It is supplied sterile via gamma irradiation for single-use. During surgery, the clinician packs the granules into the defect site, often in conjunction with GTR or GBR products. The material acts as a space-maintaining scaffold to support natural bone regeneration. It is biocompatible and designed to be replaced or integrated by host bone over time.

## Clinical Evidence

Evidence includes material characterization (ASTM F 1581), bench testing for hydration/mixing, and a canine intrabony defect animal study comparing the subject device to the predicate. Biocompatibility testing followed ISO 10993-1:2009 for permanent implants, including cytotoxicity, sensitization, irritation, systemic toxicity, genotoxicity, hemocompatibility, and implantation. Viral inactivation and residual chemical assessments were also performed. Results confirm an acceptable biocompatibility and performance profile.

## Technological Characteristics

Material: Porcine-derived osteoconductive hydroxyapatite bone mineral; meets ASTM F 1581. Form: Cancellous granules (0.25–1.0mm). Morphology: Trabecular with interconnecting macro/micro pores. Sterilization: Gamma irradiation. Single-use. No electronic components or software.

## Regulatory Identification

Bone grafting material is a material such as hydroxyapatite, tricalcium phosphate, polylactic and polyglycolic acids, or collagen, that is intended to fill, augment, or reconstruct periodontal or bony defects of the oral and maxillofacial region.

## Special Controls

*Classification.* (1) Class II (special controls) for bone grafting materials that do not contain a drug that is a therapeutic biologic. The special control is FDA's “Class II Special Controls Guidance Document: Dental Bone Grafting Material Devices.” (See § 872.1(e) for the availability of this guidance document.)(2) Class III (premarket approval) for bone grafting materials that contain a drug that is a therapeutic biologic. Bone grafting materials that contain a drug that is a therapeutic biologic, such as biological response modifiers, require premarket approval.
(c)
*Date premarket approval application (PMA) or notice of product development protocol (PDP) is required.* Devices described in paragraph (b)(2) of this section shall have an approved PMA or a declared completed PDP in effect before being placed in commercial distribution.

## Predicate Devices

- Bio-OSS® ([K033815](/device/K033815.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/2 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, stacked on top of each other.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

May 16, 2017

Kensey Nash Corporation Dba DSM Biomedical Susan Pileggi Senior Regulatory Specialist 735 Pennsylvania Drive Exton, Pennsylvania 19341

Re: K170245

Trade/Device Name: DSM Biomedical Dental Bone Graft Regulation Number: 21 CFR 872.3930 Regulation Name: Bone Grafting Material Regulatory Class: Class II Product Code: NPM Dated: April 12, 2017 Received: April 13, 2017

Dear Susan Pileggi:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device

{1}------------------------------------------------

related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely,

# Michael J. Ryan -S

for Tina Kiang, Ph.D. Acting Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

# Indications for Use

510(k) Number (if known)

K170245

Device Name DSM Biomedical Dental Bone Graft

Indications for Use (Describe)

DSM Biomedical Dental Bone Graft is indicated for:

- Augmentation or reconstructive treatment of the alveolar ridge
- Filling of infrabony periodontal defects
- Filling of defects after root resection, apicoectomy, and cystectomy
- Filling of extraction sockets to enhance preservation of the alveolar ridge
- Elevation of the maxillary sinus floor

- Filling of periodontal defects in conjuncts intended for Guided Tissue Regeneration (GTR) and Guided Bone Regeneration (GBR)

- Filling of peri-implant defects in conjunction with products intended for Guided Bone Regeneration (GBR)

| Type of Use (Select one or both, as applicable) |  |
|-------------------------------------------------|--|
|                                                 |  |

X Prescription Use (Part 21 CFR 801 Subpart D)

| | Over-The-Counter Use (21 CFR 801 Subpart C)

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows the logo for DSM, a global science-based company active in health, nutrition, and materials. The logo consists of a colorful abstract graphic on the left and the letters "DSM" in bold blue font on the right. The graphic is composed of overlapping curved shapes in various colors, including blue, green, orange, and purple, forming a circular design with a hexagonal shape in the center.

#### 510(k) Summary

| Submitted By:   | DSM Biomedical<br>735 Pennsylvania Drive<br>Exton, PA 19341                    |
|-----------------|--------------------------------------------------------------------------------|
| Contact Person: | Susan Pileggi<br>susan.pileggi@dsm.com<br>(P) 484-713-2100<br>(F) 484-713-2903 |
| Date Prepared:  | April 11, 2017                                                                 |

Date Prepared:

Device:

| Trade Name:                | DSM Biomedical Dental Bone Graft |
|----------------------------|----------------------------------|
| Common/Usual Name:         | Bone Grafting Material           |
| Classification Name:       | Bone Grafting Material           |
| Classification Regulation: | 21 CFR 872.3930                  |
| Device Class:              | II                               |
| Device Code:               | NPM                              |
| Advisory Panel:            | Dental                           |

Predicate Device:

Bio-OSS® (K033815) [Geistlich-Pharma]

## Device Description

The DSM Biomedical Dental Bone Graft is a non-pyrogenic porous bone mineral matrix for use in periodontal, oral, and maxillofacial surgery. It is produced by removing organic components from porcine bone. The composition of DSM Biomedical Dental Bone Graft meets the requirements of ASTM F 1581 Standard Specification for Composition of Anorganic Bone for Surgical Implants. The device is provided as cancellous granules approximately 0.25 - 1.0mm in size. It is supplied sterile by gamma irradiation and is for single use only.

{4}------------------------------------------------

Image /page/4/Picture/0 description: The image shows the logo for DSM, a global science-based company. The logo consists of a colorful, abstract swirl design on the left and the letters "DSM" in bold, blue font on the right. The swirl design is made up of overlapping shapes in various colors, including blue, green, yellow, and purple. The overall design is modern and eye-catching.

#### Indications For Use:

DSM Biomedical Dental Bone Graft is indicated for:

- Augmentation or reconstructive treatment of the alveolar ridge
- Filling of infrabony periodontal defects
- Filling of defects after root resection, apicoectomy, and cystectomy
- Filling of extraction sockets to enhance preservation of the alveolar ridge
- Elevation of the maxillary sinus floor
- Filling of periodontal defects in conjunction with products intended for Guided Tissue Regeneration (GTR) and Guided Bone Regeneration (GBR)
- Filling of peri-implant defects in conjunction with products intended for Guided Bone Regeneration (GBR)

#### Basis for Substantial Equivalence

The DSM Biomedical Dental Bone Graft is substantially equivalent in terms of indications for use, material composition, technological characteristics and performance characteristics to the predicate device, Bio-OSS® [Geistlich-Pharma] K033815.

DSM Biomedical Dental Bone Graft and the predicate device have identical indications for use. Both are for use in the same target population. They are similar in sizes offered with the subject device sizes falling within the range of the predicate sizes. Both are sterilized by gamma irradation and supplied in a single use vial. The materials are shown to be an osteoconductive hydroxyapatite bone mineral to aid in bone formation. The devices are both trabecular and contain interconnecting macro and micro pores. The devices are both biocompatible. The subject device is derived from porcine bone while the predicate is derived from bovine bone. This does not raise questions of substantial equivalence as both types of bone are subjected to manufacturing processes which remove organic content from the source bone. Chemical and physical characteristics have been shown to be substantially equivalent in benchtop and animal testing.

{5}------------------------------------------------

Image /page/5/Picture/0 description: The image shows the logo for DSM, a global science-based company active in health, nutrition, and materials. The logo consists of a colorful abstract graphic to the left of the company name, "DSM," which is written in a bold, blue sans-serif font. The graphic is composed of overlapping, curved shapes in various colors, including blue, green, purple, and orange, creating a sense of dynamism and innovation. The overall design is modern and corporate, reflecting the company's focus on scientific solutions.

| Characteristic         | DSM Biomedical Dental Bone Graft<br>( <i>Subject Device</i> )                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               | Bio-OSS®<br>( <i>K033815: Predicate Device</i> )                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 |
|------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Indications for Use    | DSM Biomedical Dental Bone Graft is<br>indicated for augmentation or<br>reconstructive treatment of the alveolar<br>ridge; filling of infrabony periodontal<br>defects; filling of defects after root<br>resection, apicoectomy, and cystectomy;<br>filling of extraction sockets to enhance<br>preservation of the alveolar ridge;<br>elevation of the maxillary sinus floor;<br>filling of periodontal defects in<br>conjunction with products intended for<br>Guided Tissue Regeneration (GTR) and<br>Guided Bone Regeneration (GBR); filling<br>of peri-implant defects in conjunction<br>with products intended for Guided Bone<br>Regeneration (GBR). | Bio-OSS® is indicated for augmentation<br>or reconstructive treatment of the<br>alveolar ridge; filling of infrabony<br>periodontal defects; filling of defects<br>after root resection, apicoectomy, and<br>cystectomy; filling of extraction sockets<br>to enhance preservation of the alveolar<br>ridge; elevation of the maxillary sinus<br>floor; filling of periodontal defects in<br>conjunction with products intended for<br>Guided Tissue Regeneration (GTR) and<br>Guided Bone Regeneration (GBR); filling<br>of peri-implant defects in conjunction<br>with products intended for Guided Bone<br>Regeneration (GBR). |
| Target Population      | Human, oral, periodontal                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    | Human, oral, periodontal                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         |
| Source Bone            | Porcine bone                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                | Bovine bone                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      |
| Form                   | 0.25 - 1.0mm granules                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       | 0.25mm - 1.0mm or 1.00mm - 2.0mm<br>granules                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     |
| Supplied As            | Granules supplied in a single use<br>sterilized vial                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        | Granules supplied in a single use<br>sterilized vial                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             |
| Material               | Porcine-derived osteoconductive<br>hydroxyapatite bone mineral                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              | Bovine-derived osteoconductive<br>hydroxyapatite bone mineral                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    |
| Physical<br>Morphology | Trabecular, interconnecting macro and<br>micro pores                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        | Trabecular, interconnecting macro and<br>micro pores                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             |
| Performance            | Bone formation                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              | Bone formation                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   |
| Biocompatible          | Yes                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         | Yes                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              |
| Characteristic         | DSM Biomedical Dental Bone Graft<br>( <i>Subject Device</i> )                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               | Bio-OSS®<br>( <i>K033815: Predicate Device</i> )                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 |
| Reusable               | Single Use Device                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           | Single Use Device                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                |
| Sterilization Method   | Gamma Irradiation                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           | Gamma Irradiation                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                |

{6}------------------------------------------------

Image /page/6/Picture/0 description: The image shows the logo for DSM, a global science-based company. The logo consists of a colorful abstract graphic on the left and the letters "DSM" in bold blue font on the right. The graphic is composed of swirling shapes in shades of blue, green, yellow, and pink, creating a sense of movement and dynamism. The overall design is modern and corporate, reflecting the company's focus on innovation and technology.

## Biocompatibility and Performance Data:

Material characterization, bench testing, biocompatibility testing, viral inactivation, and an animal study have been conducted to evaluate the performance characteristics and biological safety of DSM Biomedical Dental Bone Graft.

Material characterization of the DSM Biomedical Dental Bone Graft was completed in accordance with the FDA Guidance Document, Class II Special Controls Guidance Document: Dental Bone Grafting Material Devices and ASTM F 1581, Standard Specifications for Composition or Anorganic Bone for Surgical Implants. Bench testing was completed to verify that the device can be hydrated and mixed in the provided packaging.

A canine intrabony defect animal study was performed to evaluate bone healing following treatment with the DSM Biomedical Dental Bone Graft compared to the predicate device.

Biocompatibility testing was completed in accordance with the requirements of ISO 10993-1: 2009, Biological Evaluation of Medical Devices—Part 1: Evaluation and Testing within a Risk Management Process for a permanent implant device with tissue/bone contact. Testing included Cytotoxicity, Sensitization, Irritation, Acute Systemic Toxicity, Genotoxicity, Hemocompatibility, Subacute Systemic Toxicity, and Chronic Systemic Toxicity, and Implantation. Other safety testing included a viral inactivation study and residual chemical assessment. Results indicate that the device's biocompatibility profile is acceptable.

#### Conclusion:

Pursuant to section 510(k), DSM Biomedical Dental Bone Graft is substantially equivalent to the predicate device Bio-OSS® with regard to indication for use, material, technological characteristics, including principles of operation, and performance characteristics as shown in an anatomically relevant animal model.

---

**Source:** [https://fda.innolitics.com/device/K170245](https://fda.innolitics.com/device/K170245)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
