The Leica FL560 is a surgical microscope accessory used in viewing fluorescence of fluorophores with an excitation peak between ~460 nm and ~500 nm (blue) and the fluorescence emission comprising the green, yellow and red spectrum in a spectral band above ~510 nm.
Device Story
Leica FL560 is a surgical microscope accessory for visualizing intra-operative blood flow in cerebral vascular surgery. It attaches to Leica surgical operating microscopes (SOM). The device uses the SOM's standard illumination light source to produce excitation light (460-500nm). An observation filter in the optics carrier enables visualization of fluorescence emission (>510nm) from a fluorophore (Fluorescein Sodium). Used in the OR by neurosurgeons and OR staff. The surgeon views the fluorescence through the microscope eyepieces to assess blood flow and vessel architecture. This real-time visualization aids in surgical decision-making, such as confirming vessel patency or identifying occlusions, potentially improving patient outcomes during neurosurgical procedures.
Clinical Evidence
No human clinical data. Evidence includes bench testing (optical/mechanical verification) and a prospective preclinical porcine study (n=18 comparative reviews). The porcine study compared the subject device (using Fluorescein Sodium) to the predicate (using ICG) for visualizing cerebrovascular blood flow. Results showed 100% confirmation of functional equivalence in visualizing blood flow and vessel architecture, with the subject device additionally enabling concurrent visualization of background anatomical structures.
Indicated for viewing intra-operative blood flow in the cerebral vascular area.
Regulatory Classification
Identification
An angiographic x-ray system is a device intended for radiologic visualization of the heart, blood vessels, or lymphatic system during or after injection of a contrast medium. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
CIS EyeScan Portable Modular Imaging System (K092374)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/2 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo is a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, stacked on top of each other.
July 5, 2017
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
Leica Microsystems (Schweiz) AG Ms. Grainne Griffin Senior RA Specialist Max Schmidheiny-Strasse 201 Heerbrugg, Sankt Gallen, Switzerland 9435 CH
Re: K170239
Trade/Device Name: Leica FL560 Regulation Number: 21 CFR 892.1600 Regulation Name: Angiographic X-Ray System Regulatory Class: Class II Product Code: IZI Dated: June 5, 2017 Received: June 5, 2017
Dear Ms. Griffin:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. However, you are responsible to determine that the medical devices you use as components in the kit have either been determined as substantially equivalent under the premarket notification process (Section 510(k) of the act), or were legally on the market prior to May 28, 1976, the enactment date of the Medical Device Amendments. Please note: If you purchase your device components in bulk (i.e., unfinished) and further process (e.g., sterilize) you must submit a new 510(k) before including these components in your kit/tray. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be
{1}------------------------------------------------
found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
In addition, we have determined that your device kit contains Sodium Fluorescein which is subject to regulation as a drug.
Our substantially equivalent determination does not apply to the drug component of your device. We recommend you first contact the Center for Drug Evaluation and Research before marketing your device with the drug component. For information on applicable Agency requirements for marketing this drug, we suggest you contact:
Director, Division of Drug Labeling Compliance Center for Drug Evaluation and Research Food and Drug Administration 5600 Fishers Lane Rockville, Maryland 20857 (301) 594-0101
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
{2}------------------------------------------------
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely,
# Michael J. Hoffmann -S
for
Carlos L. Peña, PhD, MS Director Division of Neurological and Physical Medicine Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
#### Indications for Use
510(k) Number (if known) K170239
Device Name Leica FL560
Indications for Use (Describe)
The Leica FL560 is a surgical microscope accessory used in viewing intra-operative blood flow in the cerebral vascular area.
Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
## From Eye to Insight
Image /page/4/Picture/1 description: The image shows the logo for Leica Microsystems. The logo features the word "Leica" in a stylized, red, cursive font. Below "Leica" is the word "MICROSYSTEMS" in a smaller, sans-serif, black font. The logo is simple and modern.
#### Overview
This 510(k) summary has been prepared in accordance with the requirements of 21CFR §807.92.
| Trade Name | Leica FL560 |
|------------------|--------------------------------------------------------------------------------------------------------------------------------------------------|
| Common Name | Fluorescent Angiographic System |
| Classification | Class II; Angiographic x-ray system (21 CFR 892.1600) |
| Product Code | IZI |
| Manufacturer | Leica Microsystems (Schweiz) AG., (Registration # 3003974370)<br>Max Schmidheiny-Strasse 201<br>Heerbrugg,<br>Sankt Gallen 9435,<br>Switzerland. |
| Contact Name | Grainne Griffin |
| Tel | +353 86 7710135 |
| email | grainne.griffin@leica-microsystems.com |
| Predicate Device | Primary Leica FL800 ULT – clearance K141136<br>Auxiliary Leica FL800 – clearance K061871 & K080612 |
| Preparation Date | 20th December 2016 |
#### Device Description
Similar to the predicate device (Leica FL800 ULT approved under K141136), the Leica FL560 is an accessory to the Leica Microsystems (LMS) Class I 510(k) exempt surgical operating microscope (SOM).
The Leica FL560 utilizes the illumination light source, supplied as standard with the SOM to produce excitation light which is filtered using an illumination filter (also referred to as the excitation filter) within the 460 - 500nm wavelength.
An observation filter is introduced into the observer light path within the optics carrier of the SOM to enable visualization of the resulting fluorescence emission comprising of the green, yellow and red spectrum in a spectral band above ~510nm.
{5}------------------------------------------------
Image /page/5/Picture/1 description: The image shows the logo for Leica Microsystems. The logo consists of the word "Leica" in a stylized red font, with the word "MICROSYSTEMS" in a smaller, black, sans-serif font below it. The word "Leica" is written in a cursive font, and the letters are connected.
### Device Intended Use
The Leica FL560 is a surgical microscope accessory used in viewing fluorescence of fluorophores with an excitation peak between ~460 nm and ~500 nm (blue) and the fluorescence emission comprising the green, yellow and red spectrum in a spectral band above ~510 nm.
#### Device Indication for Use
The Leica FL560 is a surgical microscope accessory used in viewing intra-operative blood flow in the cerebral vascular area.
#### Testing
Pre-clinical studies, human factors studies, electrical safety testing have been conducted to demonstrate the substantial equivalence of the Leica FL560 to the Leica FL800 ULT.
#### Summary Table
See 'Substantial Equivalence Summary Table: Comparison to Predicate Device' on following pages for a summary of all predicate and subject device comparative features and supporting substantial equivalence testing.
#### Conclusion
Based on the technological characteristics, principle of operation, intended use, environment of use, and indications for use, the Leica FL560 has been determined to be substantially equivalent to the predicate device, the Leica FL800 ULT (K141136) in terms of safety, effectiveness and performance.
{6}------------------------------------------------
# From Eye to Insight
Image /page/6/Picture/1 description: The image shows the logo for Leica Microsystems. The logo consists of the word "Leica" in a stylized red font, with a curved line underneath. Below the word "Leica" is the word "MICROSYSTEMS" in a smaller, sans-serif font. The logo is simple and modern, and the red color is eye-catching.
| | Primary Predicate | Subject Device | |
|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Device → | Leica FL800 ULT | Leica FL560 | Demonstration of |
| What↓ | (K141136) | (Proposed new accessory) | Substantial Equivalence<br>(SE) |
| | SUMMARY OF GENERAL FEATURES | | |
| Indications<br>for use | The Leica FL800 is a surgical<br>Microscope accessory used in<br>viewing intra-operative blood<br>flow in the cerebral vascular<br>area and by-pass grafts during<br>coronary artery bypass (CABG)<br>surgery, as well as blood flow<br>during plastic and reconstructive<br>surgery. | The Leica FL560 is a surgical<br>Microscope accessory used in<br>viewing intra-operative blood<br>flow in the cerebral vascular<br>area. | Identical to subset of predicate<br>indications for use |
| For use with | Standard Leica Surgical<br>Operating Microscope<br>M520 / M525 / M720 & M530<br>product range | Standard Leica M530 OH6/OHX<br>Surgical Operating Microscope<br>(Class I Exempt)<br>Note: FL560 will not be available<br>for the M520 / M525 / M720 SOM<br>as these are older models<br>scheduled for phase out | Identical to subset of predicate<br>equipment platforms |
| Device<br>Components | • Observation filter<br>• Illumination filter<br>• Filter housing ICG Filter<br>• Beam Splitter<br>• Built in Dual Video Adaptor<br>consisting of Internal NIR<br>Camera | • Observation filter<br>• Illumination filter<br>• Filter housing | Equivalent to subset of predicate<br>components (filters equivalent but<br>for different wavelength<br>specifications) |
| Software | No Software | No Software | Identical |
| Required but<br>not supplied | • Leica Surgical Microscope<br>• Recording device | • Leica Surgical Microscope | Identical to subset of predicate |
| Drug | ICG | Fluorescein Sodium<br>(fluorescein)* | Difference does not impact<br>substantial equivalence, both<br>drugs are fluorophores for<br>visualizing blood flow and have<br>longstanding FDA approvals |
| | Primary Predicate | Subject Device | |
| Device → | Leica FL800 ULT | Leica FL560 | Demonstration of |
| What↓ | (K141136) | (Proposed new accessory) | Substantial Equivalence<br>(SE) |
| | SUMMARY OF GENERAL FEATURES, CONTINUED | | |
| Illumination<br>Filter [nm] | 400 – 780nm | 460 - 500nm | Equivalent, both are band pass<br>filters with overlapping ranges |
| Observation<br>Filter [nm] | 800 – 880nm | Above ~ 510nm | Equivalent, the FL560 high pass<br>filter range contains the FL800<br>band pass filter range |
| Light Source | Xenon 300 - 400watt | Xenon 400watt | Identical to subset of predicate |
| Test Card | Pre-operative check test card | Pre-operative check test card | Equivalent |
| Procedure Kit | None supplied | FL560 procedure kit containing:<br>• AK-FLUOR (fluorescein)<br>• FL560 test card<br>• Procedure kit IFU | Difference does not impact<br>substantial equivalence |
| *Reference Device CIS EyeScan Portable Modular Imaging System, K092374, also enables visualization of fluorescein (a<br>technical feature of both the subject and reference device) for a different Indication for Use (ophthalmology imaging) but a<br>generally equivalent intended use (fluorescence vascular angiography) | | | |
| | SUMMARY OF SUPPORTING TESTING | | |
| Electrical<br>Safety | Conformance to the following<br>standards tested and confirmed:<br>• IEC 60601-1:2005: Medical<br>electrical equipment - Part 1:<br>General requirements for<br>basic safety and essential<br>performance<br>• IEC 60601-1-2:2007<br>(Modified): Medical electrical<br>equipment - Part 1-2:<br>General requirements for<br>basic safety and essential<br>performance - Collateral<br>standard: Electromagnetic<br…
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.