Clinical serum samples; Venous whole blood samples from human subjects; Clinical laboratory and point-of-care (POC) site testing
Clinical samples were used to establish adult reference intervals and to perform method comparison studies against a predicate device to demonstrate substantial equivalence.
Clinical Method Comparison Study; Method comparison against predicate device
Clinical samples (serum and venous whole blood); Sample Size: 183 (serum) / 157 (whole blood); Number of Sites: 3
Access Fast hTSH (on the Access 2 system)
Correlation/comparability via weighted deming regression
Indications for Use
AFIAS TSH-SP, for use in conjunction with the AFIAS-6/SP Analyzer, is an immunofluorometric test system intended for in vitro diagnostic use at clinical laboratories and point-of-care (POC) sites for the quantitative measurement of thyroid stimulating hormone (TSH) levels in serum, sodium-heparinized plasma, or EDTA plasma samples. The test system is intended for use as an aid in the diagnosis of thyroid or pituitary disorders. AFIAS-6/SP Analyzer is a fluorescence-scanning instrument for in vitro diagnostic use at clinical laboratories and point-of-care sites in conjunction with various in vitro diagnostic AFIAS immunoassays for measuring the concentration of designated analytes in serum or plasma samples. AFIAS TSH-VB, for use in conjunction with AFIAS-6/VB Analyzer, is an immunofluorometric in vitro diagnostic test system intended for in vitro diagnostic use at clinical laboratories and point-of-care (POC) sites for the quantitative measurement of thyroid stimulating hormone (TSH) in sodium-heparinized or EDTA venous whole blood samples. The test system is intended for monitoring TSH levels in euthyroid and hypothyroid individuals. AFIAS-6/VB Analyzer is a fluorescence-scanning instrument for in vitro diagnostic use at clinical laboratories and point-of-care sites in conjunction with various in vitro diagnostic AFIAS immunoassays for measuring the concentration of designated analytes in venous whole blood samples.
Device Story
AFIAS TSH-SP/VB are immunofluorometric test systems using unitized cartridges; input is serum, plasma, or venous whole blood. Cartridges contain lyophilized detection buffer (fluorochrome-labeled anti-TSH antibodies) and nitrocellulose matrix with capture/control lines. Analyzer (AFIAS-6/SP or VB) uses semiconductor diode laser to excite fluorochrome-labeled complexes accumulated at test line; fluorescence intensity scanned and converted to electrical signal. On-board microprocessor computes TSH concentration using lot-specific calibration data from ID chip. Used in clinical labs/POC by nurses/technicians. Output is quantitative TSH concentration; aids clinicians in diagnosing thyroid/pituitary disorders or monitoring TSH levels. Benefits include rapid, automated quantitative testing at point-of-care.
Clinical Evidence
Performance evaluated via bench testing and clinical method comparison. Studies included LoB, LoD, LoQ (0.07 µIU/mL for SP; 0.3 µIU/mL for VB), linearity (0-100 µIU/mL), and high-dose hook effect (up to 3000 µIU/mL). Precision/reproducibility tested at 3 POC sites with 9 operators (Total CVs 4.0-5.6% for SP; 5.4-7.2% for VB). Method comparison against predicate (Access Fast hTSH) showed high correlation (R=0.9994 for SP; R=0.9999 for VB) using weighted Deming regression.
Technological Characteristics
Immunofluorometric sandwich immunoassay. Semiconductor diode laser excitation. Bench-top fluorescence-scanning analyzer. Unitized test cartridges with nitrocellulose matrix. Lot-specific calibration via electronic ID chip. Connectivity: manual entry, no wireless transmission. Software: microprocessor-based computation of concentration from fluorescence intensity.
Indications for Use
Indicated for quantitative measurement of TSH in serum, plasma, or venous whole blood. AFIAS TSH-SP aids in diagnosis of thyroid or pituitary disorders. AFIAS TSH-VB monitors TSH levels in euthyroid and hypothyroid individuals. Used in clinical labs and POC sites.
Regulatory Classification
Identification
A thyroid stimulating hormone test system is a device intended to measure thyroid stimulating hormone, also known as thyrotrophin and thyrotrophic hormone, in serum and plasma. Measurements of thyroid stimulating hormone produced by the anterior pituitary are used in the diagnosis of thyroid or pituitary disorders.
{1}
510(k) SUBSTANTIAL EQUVALENCE DETERMINATION DECISION SUMMARY ASSAY AND INSTRUMENT COMBINATION TEMPLATE
A. 510(k) Number:
k170232
B. Purpose for Submission:
New device
C. Measurand:
Thyroid stimulating hormone (TSH)
D. Type of Test:
Quantitative, immunofluorescent assay
E. Applicant:
Boditech Med Inc.
F. Proprietary and Established Names:
AFIAS TSH-SP
AFIAS TSH-VB
AFIAS-6/SP Analyzer
AFIAS-6/VB Analyzer
G. Regulatory Information:
| Product Code | Classification | Regulation Section | Panel |
| --- | --- | --- | --- |
| JLW | Class II | 21 CFR 862.1690 Thyroid stimulating hormone test system | Clinical Chemistry (75) |
| KHO | Class I | 21 CFR 862.2560 Fluorometer for Clinical Use | |
1
{2}
H. Intended Use:
1. Intended use(s):
See Indication(s) for use below.
2. Indication(s) for use:
AFIAS TSH-SP, for use in conjunction with the AFIAS-6/SP Analyzer, is an immunofluorometric test system intended for in vitro diagnostic use at clinical laboratories and point-of-care (POC) sites for the quantitative measurement of thyroid stimulating hormone (TSH) levels in serum, sodium-heparinized plasma, or EDTA plasma samples. The test system is intended for use as an aid in the diagnosis of thyroid or pituitary disorders.
AFIAS-6/SP Analyzer is a fluorescence-scanning instrument for in vitro diagnostic use at clinical laboratories and point-of-care sites in conjunction with various in vitro diagnostic AFIAS immunoassays for measuring the concentration of designated analytes in serum or plasma samples.
AFIAS TSH-VB, for use in conjunction with AFIAS-6/VB Analyzer, is an immunofluorometric in vitro diagnostic test system intended for in vitro diagnostic use at clinical laboratories and point-of-care (POC) sites for the quantitative measurement of thyroid stimulating hormone (TSH) in sodium-heparinized or EDTA venous whole blood samples. The test system is intended for monitoring TSH levels in euthyroid and hypothyroid individuals.
AFIAS-6/VB Analyzer is a fluorescence-scanning instrument for in vitro diagnostic use at clinical laboratories and point-of-care sites in conjunction with various in vitro diagnostic AFIAS immunoassays for measuring the concentration of designated analytes in venous whole blood samples.
3. Special conditions for use statement(s):
For prescription use only.
For use in clinical laboratories and point-of-care settings.
4. Special instrument requirements:
AFIAS-6/SP analyzer
AFIAS-6/VB analyzer
{3}
3
I. Device Description:
The AFIAS TSH-SP and the AFIAS TSH-VB test cartridges consist of two closed chambers; one containing lyophilized detection buffer and the other containing diluent/reconstitution buffer. Lyophilized detection buffer contains the fluorochrome-labeled anti-human TSH murine antibodies (0.05 µg) and fluorochrome-labeled chicken immunoglobulin Y (0.01 µg). The nitrocellulose matrix of the test strip contains a test/capture line which is tagged with anti-human TSH murine antibodies (0.929 µg); and a control line which is tagged with anti-chicken immunoglobulin Y (0.465 µg).
The AFIAS TSH-SP and AFIAS TSH-VB ID chips are electronic memory devices fitted into a plastic flat rectangular device 24 mm x 20 mm x 3 mm. The AFIAS TSH-SP ID chip is intended for use with the AFIAS-6/SP analyzer for use with serum and plasma and the AFIAS-VB ID chip is intended for use with the AFIAS-6/VB analyzer for use with whole blood samples. The ID chip contains lot specific information. Relevant information derived from the ID chip (test name/item, lot number, expiration date and sample type) will be saved in the internal memory of AFIAS-6/SP or AFIAS 6/VB analyzer. The analyzers can save ID chip information of up to 100 ID chips.
AFIAS-6/SP and AFIAS-6/VB analyzers are identical except the AFIAS-6/SP analyzer will only accept the AFIAS-6/SP ID chip and is intended for use with serum and plasma samples and the AFIAS-6/VB analyzer will only accept the AFIAS-6/VB ID chip and is intended for use with venous whole blood samples. The analyzer is a compact, bench-top, automated, fluorometric analyzer measuring 42 cm (L) x 33.6 cm (W) c 29.3 cm (H). AFIAS-6/SP and AFIAS-6/VB are fluorescence-scanning analyzers to use in conjunction with various AFIAS tests which are based on antigen-antibody reaction and fluorescence technology. The analyzers use a semiconductor diode laser as the excitation light source for illuminating the sample loaded AFIAS test cartridges inserted in its cartridge holder; thereby triggering fluorescence from the fluorochrome-labeled complexes of analytes accumulated at the test line on the cartridge membrane. The on-board microprocessor computes concentration of the analyte in the test sample base on the pre-programmed calibration derived from the AFIAS ID chip inserted in the ID chip port.
The AFIAS TSH-SP cartridge is for use with the AFIAS-6/SP analyzer; this test system is intended for use with serum and plasma samples. The AFIAS TSH-VB cartridges are for use with the AFIAS-6/VB analyzer; this test system is for use with venous whole blood samples. The test systems are identical except for the intended uses and sample types.
J. Substantial Equivalence Information:
1. Predicate device name(s):
Access Fast hTSH assay (modification to the Access HYPERsensitive hTSH)
Access 2 Immunoassay System
{4}
2. Predicate 510(k) number(s):
k042281
k922823
3. Comparison with predicate:
| Similarities and Differences | | |
| --- | --- | --- |
| Item | Candidate Device AFIAS-SP TSH Assay (k170232) | Predicate Device Access Fast hTSH (k042281) |
| Intended Use | For the quantitative measurement of thyroid stimulating hormone (TSH) | Same |
| Assay Methodology | Immunofluorometric sandwich immunoassay | Paramagnetic particle chemiluminescent immunoassay |
| Sample Matrices | Serum, K2-EDTA plasma, sodium heparin plasma | serum, heparinized plasma |
| Measuring range | 0.07-80 μIU/mL | 0.03-100 μIU/mL |
| Test time | 18 minutes | 20 minutes |
| Sample volume | 200 μL | Minimum 300 μL |
| Similarities and Differences | | |
| --- | --- | --- |
| Item | Candidate Device AFIAS-VB TSH Assay (k170232) | Predicate Device Access Fast hTSH (k042281) |
| Intended Use | For the quantitative measurement of thyroid stimulating hormone (TSH) | Same |
| Assay Methodology | Immunofluorometric sandwich immunoassay | Paramagnetic particle chemiluminescent immunoassay |
| Sample Matrices | Venous K2-EDTA whole blood and venous Na-heparin whole blood | Venous serum, venous heparinized plasma |
| Measuring range | 0.3 -80 μIU/mL | 0.03-100 μIU/mL |
| Test time | 18 minutes | 20 minutes |
| Sample volume | 150 μL | Minimum 300 μL |
{5}
| Similarities and Differences | | |
| --- | --- | --- |
| Item | Candidate Device AFIAS-6/SP Analyzer (k170232) | Predicate Device Access 2 Immunoassay System (k922823) |
| Intended Use | Analyzer for the measurement of immunoassay analytes | Same |
| Intended Use Sites | Clinical laboratories and Point-of-Care | Clinical laboratories |
| Sample type | Serum, K2-EDTA and Na-heparin plasma | Same |
| Sample capacity | Up to 6 samples at a time | Up to 60 samples at one time |
| Reagent format | Unitized test cassettes | Reagent packs |
| Analyzer Format | Bench-top | Same |
| Similarities and Differences | | |
| --- | --- | --- |
| Item | Candidate Device AFIAS-6/VB Analyzer (k170232) | Predicate Device Access 2 Immunoassay System (k922823) |
| Intended Use | Analyzer for the measurement of immunoassay analytes | Same |
| Intended Use Sites | Clinical laboratories and Point-of-Care | Clinical laboratories |
| Sample type | Venous K2-EDTA and Na-heparin whole blood | Serum and plasma |
| Sample capacity | Up to 6 samples at a time | Up to 60 samples at one time |
| Reagent format | Unitized test cassettes | Reagent packs |
| Analyzer Format | Bench-top | Same |
# K. Standard/Guidance Document Referenced (if applicable):
CLSI EP05-A3: Evaluation of Precision of Quantitative Measurement Procedures; Approved Guideline-Third Edition.
CLSI EP06-A: Evaluation of the Linearity of Quantitative Measurement Procedures: A Statistical Approach; Approved Guideline.
{6}
CLSI EP07-A2: Interference Testing in Clinical Chemistry; Approved Guideline-Second Edition.
CLSI EP09-A3: Measurement Procedure Comparison and Bias Estimation Using Patient Samples; Approved Guideline-Third Edition.
CLSI EP14-A3: Evaluation of Commutability of Processed Samples; Approved Guideline; Third Edition
CLSI EP 17-A2: Evaluation of Detection Capability for Clinical Laboratory Measurement Procedures; Approved Guideline.
CLSI EP28-A3c: Defining, Establishing, and Verifying Reference Intervals in the Clinical Laboratory; Approved Guideline; Third Edition.
## L. Test Principle:
Fluorochrome-labeled anti-human TSH murine antibodies (detector antibody) present in the reconstituted detection buffer bind with thyroid stimulating hormone present in the sample. The test mixture migrates through the nitrocellulose matrix of the test strip which contains a test capture line and a control line. The test line is tagged with anti-human TSH murine antibodies and the control line is tagged with anti-chicken immunoglobulin Y. As the test sample migrates through the test strip, anti-human TSH murine antibodies at the test line capture the detector antibody-TSH complexes. Similarly, complexes of fluorochrome-labeled chicken immunoglobulin Y-capture antibodies accumulate at the control line. At the end of the 15 minute incubation time, a laser light source illuminates the cartridge membrane thereby triggering fluorescence from the fluorochrome-labeled complexes accumulated at the test and control lines. The intensity of the fluorescence is scanned and converted into an electric signal which correlates to the concentration of human TSH in the test sample. An on-board microprocessor calculates the TSH concentration in the sample utilizing a pre-programmed, lot-specific and sample-specific master calibration curve stored in the AFIAS TSH ID chip.
## M. Performance Characteristics (if/when applicable):
### 1. Analytical performance:
#### a. Precision/Reproducibility:
In-house precision studies were performed by testing four spiked serum samples and seven spiked sodium-heparinized venous whole blood samples. The serum samples were tested once a day for 21 consecutive days. The venous whole blood samples were tested once per day for 6 consecutive days. All the studies were performed by 3 in-house laboratory technicians using 3 lots of cartridges, one lot per three
6
{7}
analyzers. The results of the in-house precision studies are summarized in the tables below:
Serum analysis for the AFIAS TSH-VB on the AFIAS-6/VB Analyzer:
| | | | Between lot | | Between day | | Total | |
| --- | --- | --- | --- | --- | --- | --- | --- | --- |
| Sample | n | Mean (μIU/mL) | SD | %CV | SD | %CV | SD | %CV |
| 1 | 63 | 0.49 | 0.005 | 1.0 | 0.012 | 2.5 | 0.020 | 4.0 |
| 2 | 63 | 4.99 | 0.058 | 1.2 | 0.151 | 3.0 | 0.275 | 5.5 |
| 3 | 63 | 14.97 | 0.244 | 1.6 | 0.451 | 3.0 | 0.826 | 5.5 |
| 4 | 63 | 53.31 | 0.295 | 0.6 | 2.844 | 5.3 | 3.044 | 5.7 |
Venous Na-heparin whole blood analysis for the AFIAS TSH-VB on the AFIAS-6/VB Analyzer:
| | | | Between lot | | Between day | | Total | |
| --- | --- | --- | --- | --- | --- | --- | --- | --- |
| Sample | n | Mean (μIU/mL) | SD | %CV | SD | %CV | SD | %CV |
| 1 | 18 | 0.49 | 0.006 | 1.3 | 0.029 | 6.0 | 0.030 | 6.1 |
| 2 | 18 | 1.18 | 0.001 | 0.1 | 0.075 | 6.4 | 0.085 | 7.2 |
| 3 | 18 | 2.66 | 0.058 | 2.2 | 0.157 | 5.9 | 0.201 | 7.6 |
| 4 | 18 | 3.95 | 0.034 | 0.9 | 0.289 | 7.3 | 0.298 | 7.5 |
| 5 | 18 | 4.89 | 0.065 | 1.3 | 0.328 | 6.7 | 0.327 | 6.7 |
| 6 | 18 | 14.39 | 0.126 | 0.9 | 1.026 | 7.1 | 0.986 | 6.8 |
| 7 | 18 | 54.53 | 0.227 | 0.4 | 3.980 | 7.3 | 3.762 | 6.9 |
External precision studies were performed at three point-of-care (POC) sites. The following sample types were tested: 2 native serum samples, 2 native Na-heparin venous whole blood samples, and 2 spiked Na-heparin venous whole blood samples. At each of the 3 sites, 3 lots of test cartridges and 3 analyzers were used (one lot per analyzer). At each of the 3 test sites, 3 operators tested each sample one time a day for 5 consecutive days for a total of 45 results per sample. The results are summarized in the tables below:
{8}
Serum analysis for the AFIAS TSH-SP on the AFIAS-6/SP Analyzer:
| Sample | Site | n | Mean (μIU/mL) | Between lot | | Between day | | Total | |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
| | | | | SD | %CV | SD | %CV | SD | %CV |
| 1 | 1 | 15 | 0.49 | 0.003 | 0.7 | 0.006 | 1.3 | 0.020 | 4.2 |
| | 2 | 15 | 0.49 | 0.003 | 0.7 | 0.009 | 1.8 | 0.018 | 3.7 |
| | 3 | 15 | 0.50 | 0.009 | 1.8 | 0.011 | 2.2 | 0.021 | 4.2 |
| 2 | 1 | 15 | 4.91 | 0.034 | 0.7 | 0.187 | 3.8 | 0.288 | 5.9 |
| | 2 | 15 | 4.96 | 0.057 | 1.1 | 0.154 | 3.1 | 0.318 | 6.4 |
| | 3 | 15 | 5.03 | 0.029 | 0.6 | 0.145 | 2.9 | 0.233 | 4.6 |
| 3 | 1 | 15 | 14.96 | 0.423 | 2.8 | 0.658 | 4.4 | 1.008 | 6.7 |
| | 2 | 15 | 14.93 | 0.240 | 1.6 | 0.280 | 1.9 | 0.749 | 5.0 |
| | 3 | 15 | 15.26 | 0.423 | 2.8 | 0.277 | 1.8 | 0.603 | 4.0 |
| 4 | 1 | 15 | 54.26 | 0.496 | 0.9 | 1.385 | 2.6 | 3.003 | 5.5 |
| | 2 | 15 | 54.35 | 2.136 | 3.9 | 1.381 | 2.5 | 3.141 | 5.8 |
| | 3 | 15 | 53.53 | 0.596 | 1.1 | 2.306 | 4.3 | 2.883 | 5.4 |
Serum (combined sites) analysis for the AFIAS TSH-SP on the AFIAS-6/SP Analyzer:
| Sample | n | Mean (μIU/mL) | Between lot | | Between day | | Total | |
| --- | --- | --- | --- | --- | --- | --- | --- | --- |
| | | | SD | %CV | SD | %CV | SD | %CV |
| 1 | 45 | 0.49 | 0.005 | 1.0 | 0.007 | 1.5 | 0.020 | 4.0 |
| 2 | 45 | 4.97 | 0.028 | 0.6 | 0.072 | 1.5 | 0.280 | 5.6 |
| 3 | 45 | 15.05 | 0.297 | 2.0 | 0.244 | 1.6 | 0.801 | 5.3 |
| 4 | 45 | 54.05 | 0.468 | 0.9 | 0.531 | 1.0 | 2.965 | 5.5 |
Venous Na-heparin whole blood analysis for the AFIAS TSH-VB on the AFIAS-6/VB Analyzer:
| Sample | Site | n | Mean (μIU/mL) | Between lot | | Between day | | Total | |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
| | | | | SD | %CV | SD | %CV | SD | %CV |
| 1 | 1 | 15 | 0.45 | 0.008 | 1.7 | 0.015 | 3.4 | 0.029 | 6.3 |
| | 2 | 15 | 0.44 | 0.005 | 1.2 | 0.013 | 2.8 | 0.031 | 7.0 |
| | 3 | 15 | 0.46 | 0.004 | 0.9 | 0.024 | 5.3 | 0.029 | 6.4 |
| 2 | 1 | 15 | 4.88 | 0.116 | 2.4 | 0.273 | 5.6 | 0.321 | 6.6 |
| | 2 | 15 | 5.00 | 0.105 | 2.1 | 0.183 | 3.7 | 0.350 | 7.0 |
| | 3 | 15 | 5.01 | 0.186 | 3.7 | 0.200 | 4.0 | 0.403 | 8.1 |
| 3 | 1 | 15 | 15.97 | 0.532 | 3.3 | 0.581 | 3.6 | 1.161 | 7.3 |
| | 2 | 15 | 15.31 | 0.425 | 2.8 | 0.563 | 3.7 | 1.156 | 7.5 |
| | 3 | 15 | 16.10 | 0.344 | 2.1 | 0.712 | 4.4 | 0.995 | 6.2 |
| 4 | 1 | 15 | 55.77 | 0.906 | 1.6 | 0.944 | 1.7 | 2.853 | 5.1 |
| | 2 | 15 | 56.57 | 0.450 | 0.8 | 1.326 | 2.3 | 2.977 | 5.3 |
| | 3 | 15 | 55.71 | 1.791 | 3.2 | 2.349 | 4.2 | 3.300 | 5.9 |
{9}
Venous Na-heparin whole blood (combined sites) analysis for the AFIAS TSH-VB on the AFIAS-6/VB Analyzer:
| Sample | n | Mean (μIU/mL) | Between lot | | Between day | | Total | |
| --- | --- | --- | --- | --- | --- | --- | --- | --- |
| | | | SD | %CV | SD | %CV | SD | %CV |
| 1 | 45 | 0.45 | 0.005 | 1.0 | 0.012 | 2.6 | 0.029 | 6.5 |
| 2 | 45 | 4.96 | 0.125 | 2.5 | 0.103 | 2.1 | 0.356 | 7.2 |
| 3 | 45 | 15.79 | 0.206 | 1.3 | 0.362 | 2.3 | 1.136 | 7.2 |
| 4 | 45 | 56.02 | 0.199 | 0.4 | 0.750 | 1.3 | 3.005 | 5.4 |
## b. Linearity/assay reportable range:
A linearity study was performed on serum and Na-heparinized whole blood samples. A serum sample with a high TSH concentration was serially diluted using a serum sample with a low concentration of TSH to produce 22 equally spaced, TSH concentrations. The serum linearity samples had the following mean TSH concentrations: 0.0, 0.03, 0.49, 0.99, 1.98, 2.99, 3.98, 4.99, 5.94, 6.99, 7.96, 9.00, 10.03, 20.16, 29.99, 40.77, 50.25, 61.63, 68.00, 79.01, 90.56, and 102.91 μIU/mL. Likewise, a Na-heparinized whole blood sample with a high TSH concentration was serially diluted with a Na- heparinized whole blood sample with a low TSH concentration to produce 22 equally spaced TSH concentrations. The Na-heparinized whole blood linearity samples had the following mean TSH concentrations: 0.00, 0.14, 0.49, 0.97, 1.92, 2.92, 3.91, 4.92, 5.94, 6.93, 7.86, 8.82, 9.81, 19.39, 30.50, 39.38, 48.77, 57.69, 70.40, 77.66, 88.69, and 98.36 μIU/mL. The samples were run in replicates of three. The following linear regression equations were obtained:
Serum analysis for the AFIAS TSH-SP on the AFIAS-6/SP Analyzer: $y = 1.0081x - 0.0785$ , $R^2 = 0.9993$
Na-heparinized whole blood analysis for the AFIAS TSH-VB on the AFIAS-6/VB Analyzer: $y = 0.9824x + 0.0216$ , $R^2 = 0.9997$
The claimed measuring range of the AFIAS TSH-SP on the AFIAS TSH/SP Analyzer for serum TSH is 0.07 to $80~\mu \mathrm{IU / mL}$ .
The claimed measuring range of the AFIAS TSH-VB on the AFIAS TSH/VB Analyzer for venous whole blood TSH is 0.3 to $80~\mu \mathrm{IU / mL}$ .
{10}
c. Traceability, Stability, Expected values (controls, calibrators, or methods):
Traceability: AFIAS TSH Controls are prepared gravimetrically and are traceable to the WHO International Standard Human Thyroid Stimulating Hormone for Immunoassay (3rd IRP, NIBSC Code 81/565).
d. Detection limit:
Detection limit studies were performed on both serum and Na-heparinized whole blood samples to determine the limit of blank, limit of detection, and limit of quantitation for both matrix types.
Limit of Blank (LoB): Five TSH-depleted human serum and whole blood samples were each tested in replicates of 5 using 3 cartridge lots and 3 analyzers for 3 days for a total of 75 test results per lot. The LoB was calculated by non-parametric analysis of the study data. The sponsor used the highest LoB from the three lots as the claimed LoB. The LoB for the serum matrix was determined to be 0.03 μIU/mL for use with the AFIAS TSH-SP on the AFIAS-6/SP Analyzer. The LoB for the whole blood matrix was determined to be 0.13 μIU/mL for use with the AFIAS TSH-VB on the AFIAS-6/VB Analyzer.
Limit of Detection (LoD): Five low level TSH spiked human serum samples were prepared to have approximate TSH concentrations of 0.04, 0.06, 0.08, 0.10, and 0.12 μIU/mL. Five low level TSH spiked venous whole blood samples were prepared to have approximate TSH concentrations of 0.2, 0.3, 0.4, 0.5, and 0.6 μIU/mL. The samples were tested in replicates of 5 using 3 cartridge lots and 3 analyzers for 3 days for a total of 75 tests per lot. The following equation was used to calculate the LoD: $$\mathrm{LoD} = \mathrm{LoB} + \mathrm{Cp} \times \mathrm{SD}_{\mathrm{L}}$$. The sponsor used the highest LoD from the three lots as the claimed LoD. The LoD for the serum matrix was determined to be 0.05 μIU/mL for use with the AFIAS TSH-SP on the AFIAS-6/SP Analyzer. The LoD for the whole blood matrix was determined to be 0.2 μIU/mL for use with the AFIAS TSH-VB on the AFIAS-6/VB Analyzer.
Limit of Quantitation (LoQ): Five low level TSH spiked human serum samples were prepared to have approximate mean TSH concentrations of 0.05, 0.06, 0.07, 0.08, and 0.09 μIU/mL. Four low level TSH spiked Na-heparin whole blood samples were prepared to have approximate mean TSH concentrations of 0.2, 0.3, 0.4, and 0.5 μIU/mL. The samples were tested and the SD and %CVs were calculated from the data. The sponsor defines LoQ as the lowest TSH concentration with an imprecision of less than or equal to 20%. The sponsor used the highest LoQ from the three lots as the claimed LoQ. The LoQ for the serum matrix was determined to be 0.07 μIU/mL for use with the AFIAS TSH-SP on the AFIAS-6/SP Analyzer. The LoQ for the whole blood matrix was determined to be 0.3 μIU/mL for use with the AFIAS TSH-VB on the AFIAS-6/VB Analyzer.
10
{11}
| | Serum (μIU/mL) for use with the AFIAS TSH-SP on the AFIAS-6/SP Analyzer | Whole Blood (μIU/mL) for use with the AFIAS TSH-VB on the AFIAS-6/VB Analyzer |
| --- | --- | --- |
| Limit of Blank | 0.03 | 0.13 |
| Limit of Detection | 0.05 | 0.20 |
| Limit of Quantitation | 0.07 | 0.30 |
The claimed measuring range of the AFIAS TSH-SP on the AFIAS TSH/SP Analyzer for serum TSH is 0.07 to $80~\mu \mathrm{IU / mL}$
The claimed measuring range of the AFIAS TSH-VB on the AFIAS TSH/VB Analyzer for venous whole blood TSH is 0.3 to $80~\mu \mathrm{IU / mL}$ .
# e. Analytical specificity:
Interference study:
The sponsor performed an interference study to evaluate the susceptibility of the AFIAS TSH test systems to interference from several endogenous substances and common thyroid medications. Two serum samples and two Na-heparin venous whole blood samples with TSH concentrations of 0.5 and $5.0~\mu \mathrm{IU / mL}$ were spiked with endogenous substances and medications. The samples were run in triplicate and the results were compared to control sample results which were not spiked with potential interfering substances. The sponsor defines significant interference as $\pm 10\%$ bias as compared to the control results. The results of the interference study are summarized in the table below and are applicable to the AFIAS TSH-SP and AFIAS TSH-VB assays:
| Substance | Highest concentration of substance tested that did not demonstrate significant interference |
| --- | --- |
| Albumin | 6 g/dL |
| Conjugated bilirubin | 20 mg/dL |
| Unconjugated bilirubin | 20 mg/dL |
| Biotin | 56 ng/mL |
| HAMA | 138 ng/mL |
| Hemoglobin | 500 mg/dL |
| Rheumatoid factor | 1500 IU/mL |
| Triglycerides | 3000 mg/dL |
| Levothyroxine | 75 μg/L |
| Methimazole | 36 mg/L |
| Metoprolol | 4.99 mg/L |
{12}
# Cross-Reactivity Study
Cross-reactivity studies were performed to evaluate the susceptibility of the AFIAS TSH test systems to cross-reactivity with the following endogenous structural analogs of human TSH: human follicle stimulating hormone (hFSH), human luteinizing hormone (hLH), and human chorionic gonadotropin (hCG). Two serum samples and two Na-heparin venous whole blood samples with TSH levels of $0.5\mu \mathrm{IU} / \mathrm{mL}$ and $5.0\mu \mathrm{IU} / \mathrm{mL}$ were spiked with $1,200\mathrm{mIU} / \mathrm{mL}$ of each of the potential cross-reactants and measured in triplicate. The sponsor defines significant cross reactivity as less than a $10\%$ bias in expected results.
The following equation was used to calculate the $\%$ cross reactivity: Mean concentration of spiked sample - mean concentration of unspiked sample x 100 Spiked concentration of the cross-reactant
The cross-reactivity study results are summarized in the table below and are applicable to the AFIAS TSH-SP and AFIAS TSH-VB assays:
| Substance | Highest concentration of substance tested that did not demonstrate significant cross-reactivity |
| --- | --- |
| hFSH | 1,200 mIU/mL |
| hLH | 1,200 mIU/mL |
| hCG | 1,200 mIU/mL |
# High Dose Hook Effect study
A hook effect study was performed by spiking a serum sample with the following TSH concentrations: 25, 50, 75, 100, 150, 200, 500, 1000, 1500, 2000, 2500, and $3,000\mu \mathrm{IU} / \mathrm{mL}$ . The samples were run in triplicate using the AFIAS TSH test system. The study showed that the samples with a TSH concentration of greater than 80 $\mu \mathrm{IU} / \mathrm{mL}$ of TSH displayed results of “ $>80~\mu \mathrm{IU} / \mathrm{mL}$ ” on the AFIAS-6 analyzer. The sponsor concludes that the AFIAS TSH test system did not show hook effect for TSH concentrations up to $3000~\mu \mathrm{IU} / \mathrm{mL}$ .
f. Assay cut-off:
Not applicable.
# 2. Comparison studies:
a. Method comparison with predicate device:
A method comparison study was performed at 3 point-of-care (POC) sites to evaluate the accuracy of the AFIAS TSH test system against the comparator method, the Access Fast hTSH test. Matching sets of serum and Na-heparinized venous whole blood were tested by the candidate devices. Serum was measured by the comparator
{13}
method at an in-house laboratory. The method comparison study included 183 serum samples and 157 Na- heparinized venous whole blood samples. Less than 15% of the samples were contrived. The method comparison study was performed by 3 POC operators at each of the 3 POC sites for a total of 9 POC operators (nurses, nursing assistants, medical assistants, office assistants, etc.). One lot of AFIAS test cartridges and 3 analyzers were used in the study. The results of the method comparison study are summarized below for each sample type:
Serum analysis for the AFIAS TSH-SP on the AFIAS-6/SP Analyzer:
y= 0.976x + 0.003, R²=0.9994 (n=183), test range: 0.07-79.9 μIU/mL
Na-heparinized venous whole blood analysis for the AFIAS TSH-VB on the AFIAS-6/VB Analyzer:
Serum results on the predicate were compared to Na-heparinized venous whole blood on the candidate device.
y=0.909 x + 0.012, R²=0.999 (n=157), test range: 0.31-76.26 μIU/mL
Analysis of individual sample results was performed to assess the classification as hypothyroid or normal using the reference range for each. The classification into hypothyroid or normal by the candidate was the same as the predicate for all samples. The difference in results are due matrix differences between serum and whole blood. Serum and whole blood results are not interchangeable and must be interpreted with the corresponding reference range only. See M.5. below.
b. Matrix comparison:
A matrix comparison study was performed by testing matched sample sets of the claimed sample types: serum, sodium heparin venous whole blood and plasma, EDTA venous whole blood and plasma. The results are summarized in the following table:
| Compared matrices | n | Linear Regression | R² | Range tested (μIU/mL) |
| --- | --- | --- | --- | --- |
| Serum vs Na-heparin plasma | 81 | y=0.956x +0.04 | 0.9998 | 0.09-77.52 |
| Serum vs K2-EDTA plasma | 79 | y=0.9654x + 0.0131 | 0.9997 | 0.09-77.52 |
| Venous Na-heparin whole blood vs K2-EDTA whole blood | 63 | y=0.993x - 0.0325 | 0.9998 | 0.40-71.74 |
The results of the matrix comparison study support the manufacturer's claim that serum, K2-EDTA plasma and Na-heparin plasma are acceptable sample types to use with the AFIAS TSH-SP test system and that venous K2-EDTA and venous Na-heparin whole blood and are acceptable sample types to use with the AFIAS TSH-VB test systems.
{14}
14
3. Clinical studies:
a. Clinical Sensitivity:
Not applicable.
b. Clinical specificity:
Not applicable.
c. Other clinical supportive data (when a. and b. are not applicable):
Not applicable.
4. Clinical cut-off:
Not applicable.
5. Expected values/Reference range:
A reference range study was performed to establish TSH reference intervals for serum, and venous Na-heparin whole blood for the candidate assay. The study included analyses of 128 serum samples (65 males and 63 females), and 133 venous sodium-heparinized whole blood samples (69 males and 64 females). All subjects were between the ages of 21 and 70 years of age. The central 95% distribution of the results are shown below:
Serum reference range for the AFIAS TSH-SP on the AFIAS-6/SP Analyzer: 0.45 to 4.49 µIU/mL
Venous heparinized whole blood reference range for the AFIAS TSH-VB on the AFIAS-6/VB Analyzer: 0.41 to 4.06 µIU/mL
The manufacturer claims 0.45 to 4.50 µIU/mL as the adult reference for serum/plasma samples analyzed on the AFIAS-6/SP test system and 0.40 to 4.0 µIU/mL for venous whole blood samples analyzed on the AFIAS-6/VB test system.
N. Instrument Name:
AFIAS-6/SP Analyzer
AFIAS-6/VB Analyzer
{15}
O. System Descriptions:
1. Modes of Operation:
Does the applicant’s device contain the ability to transmit data to a computer, webserver, or mobile device?
Yes ☐ X or No ☐
Does the applicant’s device transmit data to a computer, webserver, or mobile device using wireless transmission?
Yes ☐ or No ☐ X
2. Software:
FDA has reviewed applicant’s Hazard Analysis and software development processes for this line of product types:
Yes ☐ X or No ☐
3. Specimen Identification:
On-screen keypad or USB keyboard patient ID entry.
4. Specimen Sampling and Handling:
The test sample is manually pipetted into test cartridge prior to placing the cartridge on the analyzer.
5. Calibration:
The AFIAS TSH test system uses a lot-specific calibration factor, stored on the AFIAS ID-Chip which is inserted into the AFIAS-6 analyzer prior to testing with a new lot of AFIAS TSH test cartridges.
Periodic user-level calibration using the two-level AFIAS TSH Calibrators adjusts the two sample type specific master calibration curves (derived from each lot-specific AFIAS TSH ID chip) stored in the internal memory of the AFIAS-6.
6. Quality Control:
AFIAS TSH Controls, Level 1 and Level 2, are sold by the manufacturer and are the recommended quality control material.
15
{16}
P. Other Supportive Instrument Performance Characteristics Data Not Covered In The "Performance Characteristics" Section above:
Not applicable.
Q. Proposed Labeling:
The labeling is sufficient and it satisfies the requirements of 21 CFR Part 809.10.
R. Conclusion:
The submitted information in this premarket notification is complete and supports a substantial equivalence decision.
16
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.