K170195 · Quantitative Insights, Inc. · LLZ · May 17, 2017 · Radiology
Device Facts
Record ID
K170195
Device Name
QuantX
Applicant
Quantitative Insights, Inc.
Product Code
LLZ · Radiology
Decision Date
May 17, 2017
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.2050
Device Class
Class 2
Attributes
Software as a Medical Device
Indications for Use
QuantX is a quantitative image analysis software device used to assist radiologists in the assessment and characterization of breast abnormalities using MR image data. The software automatically registers images, and segments and analyzes user-selected regions of interest (ROI). QuantX extracts image data from the ROI to provide volumetric and surface area analysis. These imaging features are then displayed to the user in a dedicated analysis panel on the display monitor. When interpreted by a skilled physician, this device provides information that may be useful for screening and diagnosis. Patient management decisions should not be made based solely on the results of QuantX analysis. QuantX may also be used as an image viewer of multi-modality digital images, including ultrasound and mammography. The software also includes tools that allow users to measure and document images, and output in a structured report. Limitations: QuantX is not intended for primary interpretation of digital mammography images.
Device Story
QuantX is a software-based quantitative image analysis tool for breast MR imaging; assists radiologists in lesion characterization. Inputs: MR image data (T1, DCE, T2, DWI) loaded manually or via DICOM. Operation: User selects seed point or ROI; software performs automatic image registration, segmentation, and analysis. Outputs: Volumetric/surface area analysis, QI Most Enhancing Curve, and time-intensity curves displayed in a dedicated analysis panel. Also functions as a multi-modality viewer (ultrasound, mammography) with measurement and reporting tools. Used in clinical settings by radiologists to supplement diagnostic assessment; does not replace clinical judgment. Benefits: Provides quantitative metrics to aid in lesion assessment and characterization.
Clinical Evidence
No clinical data. Evidence consists of non-clinical performance testing, including verification of MR/mammography/ultrasound image loading and manipulation, accuracy of kinetic colormaps, lesion segmentation, and BIRADS reporting. Validation testing confirmed conformance to user needs and intended use.
Technological Characteristics
MRI data post-processing software; DICOM compatible. Features: 2D/3D visualization, MIPs, reformats, image registration, subtraction, kinetic analysis, parametric image maps, ADC maps, and automated/manual segmentation. Connectivity: DICOM standard interfaces. Software-based analysis.
Indications for Use
Indicated for radiologists to assist in the assessment and characterization of breast abnormalities using MR image data. May also be used as a multi-modality image viewer for ultrasound and mammography. Not intended for primary interpretation of digital mammography images.
Regulatory Classification
Identification
A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.
Special Controls
*Classification.* Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).
{1}------------------------------------------------
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
May 17, 2017
Quantitative Insights, Inc. % Mr. Robert Tomek Chief Technology Officer 1452 E 530 Street CHICAGO IL 60615
Re: K170195 Trade/Device Name: QuantX Regulation Number: 21 CFR 892.2050 Regulation Name: Picture archiving and communications system Regulatory Class: II Product Code: LLZ Dated: February 21, 2017 Received: February 21, 2017
Dear Mr. Tomek:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set
Image /page/0/Picture/10 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, with flowing lines representing hair or clothing.
{2}------------------------------------------------
forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely.
Michael D. O'Hara
For
Robert Ochs, Ph.D. Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
# Indications for Use
510(k) Number (if known) K170195
Device Name QuantX
### Indications for Use (Describe)
QuantX is a quantitative image analysis software device used to assist radiologists in the assessment and characterization of breast abnormalities using MR image data. The software automatically registers images, and segments and analyzes user-selected regions of interest (ROI). QuantX extracts image data from the ROI to provide volumetric and surface area analysis. These imaging features are then displayed to the user in a dedicated analysis panel on the display monitor.
When interpreted by a skilled physician, this device provides information that may be useful for screening and diagnosis. Patient management decisions should not be made based solely on the results of QuantX analysis.
QuantX may also be used as an image viewer of multi-modality digital images, including ultrasound and mammography. The software also includes tools that allow users to measure and document images, and output in a structured report.
Limitations: QuantX is not intended for primary interpretation of digital mammography images.
| Type of Use (Select one or both, as applicable) | |
|------------------------------------------------------------|-----------------------------------------------------------|
| <span></span> Prescription Use (Part 21 CFR 801 Subpart D) | <span></span> Over-The-Counter Use (21 CFR 801 Subpart C) |
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
# PREMARKET NOTIFICATION 510(k) Summary As required by 21 CFR §807.92(c)
# Submitter
| Manufacturer: | Quantitative Insights, Inc. |
|---------------|------------------------------------------|
| Address: | 1452 E. 53rd Street<br>Chicago, IL 60615 |
| Contact Person: |
|-----------------------|
| Contact Person Title: |
| Telephone: |
| Email: |
Robert Tomek Chief Technology Officer 872.529.2239 rtomek@guantinsights.com
Date Prepared:
May 10, 2017
## Device Information
Trade Name: QuantX Common Name: Image Processing Software Classfication Name: Picture archiving and communications system Product Code: LLZ
#### LEGALLY MARKETED PREDICATE DEVICE A.
Legally marketed predicate device is: CADStream Version 5 - K081556, K092954
#### B. DEVICE DESCRIPTION
QuantX is a software program that analyzes patient breast images, and is designed to aid radiologists in the characterization of lesions. After MR images are acquired from a third-party acquisition device, they can be loaded into the QuantX database manually, or automatically if connected to a DICOM-compatible device. Users then select and load the patient case to use the QuantX software tools in the examination of the images. Different types of MR sequences (T1, DCE, T2, DWI, etc.) can be viewed at the same time as mammography or ultrasound images from the same patient.
A variety of viewing tools are available to users. The MR images can be examined under different image planes (axial, sagittal, and coronal) as well as different image time points and slices. Users can use keyboard shortcuts or a scrolling mechanism to navigate through MR image slices. Colored axes serve as slice location markers for ease of pinpointing regions of interest (ROI). Images can be panned, changed in contrast, zoomed in or out. and measured. The Colormap feature visualizes contrast uptake (enhancement) studies, and a time intensity curve can be viewed for any location on the MR image.
{5}------------------------------------------------
QuantX includes image registration, and automated segmentation and analysis functions, based on a seed point indicated by the user. Users can select a ROI manually from the MR image, or use the automatic segmentation tool to obtain and accept a ROI, for input to the QuantX analytics.
The QuantX analytics display the QI Most Enhancing Curve, volume and surface area of the specified region. A user experienced with the significance of such data will be able to view and interpret this additional information during the diagnosis of breast lesions.
#### C. INDICATIONS FOR USE
OuantX is a quantitative image analysis software device used to assist radiologists in the assessment and characterization of breast abnormalities using MR image data. The software automatically registers images, and segments and analyzes user-selected regions of interest (ROI). QuantX extracts image data from the ROI to provide volumetric and surface area analysis. These imaging features are then displayed to the user in a dedicated analysis panel on the display monitor.
When interpreted by a skilled physician, this device provides information that may be useful for screening and diagnosis. Patient management decisions should not be made based solely on the results of QuantX analysis.
OuantX may also be used as an image viewer of multi-modality digital images, including ultrasound and mammography. The software also includes tools that allow users to measure and document images, and output in a structured report.
Limitations: QuantX is not intended for primary interpretation of digital mammography images.
An assessment has been made on the differences in indications between QuantX and the predicate device, and it has been determined that these changes do not raise any new questions regarding the safety and effectiveness of the device.
### TECHNOLOGICAL CHARACTERISTICS SUMMARY - as required by D. 807.92(a)(6)
QuantX has the following same technological characteristics as its predicate device. It is an MRI data post-processing software device, and is DICOM compatible. These similar features include:
- Visualization tools - standard 2D and 3D image viewing tools. MIPs. reformats
- Analysis registration, subtractions, kinetic analysis, parametric image maps, ● apparent diffusion coefficient (ADC) maps, automatic and manual segmentation tools
- Standardized reporting tools incorporating user selected finding and assessment ●
- User interface and interaction tools that enhance workflow efficiency .
{6}------------------------------------------------
- . Communication and storage functions including DICOM standard interfaces
The following technological differences exist between the subject and predicate device; these differences do not affect the safety and effectiveness of the device when used as labeled:
- . Subject does not support coil inhomogeneity correction
- . Subject does not support interventional planning
- Workflow efficiencies are implemented differently in the subject and predicate . devices
- . In addition to volume computation, the subject device includes a surface area computation which provides an additional indicator of lesion size
### E. NON-CLINICAL PERFORMANCE DATA TESTING AND REVIEWS - as required by 807.92(b)(1)
QuantX has been successfully tested for its software requirements and features.
Nonclinical tests included:
- Verification of the proper loading, display, and manipulation of MR image data. ●
- . Verification of the proper loading, display, and manipulation of Mammographic and Ultrasound image data.
- Verification that the calculation and display of kinetic colormaps were accurate ● on MR image data.
- Verification testing of lesion segmentation on MR image data.
- Verification that all measurements and BIRADS reporting were recorded correctly.
- Validation testing demonstrates the device conforms to user needs and intended ● use.
#### F. CONCLUSION
OuantX provides features to aid radiologists in the visualization and analysis of breast MR images. The potential hazards, including risk analysis and design considerations, have been examined and incorporated into the software development process.
This 510(k) Premarket Notification provides adequate information to establish substantial equivalence to the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.