K170165 · Molecular Matrix, Inc. · MQV · Dec 29, 2017 · Orthopedic
Device Facts
Record ID
K170165
Device Name
Osteo-P Bone Graft Substitute
Applicant
Molecular Matrix, Inc.
Product Code
MQV · Orthopedic
Decision Date
Dec 29, 2017
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 888.3045
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
Osteo-PTM Bone Graft Substitute is indicated for use only in the treatment of bony voids or gaps that are not intrinsic to the stability of the bony structure. It is intended to be gently packed into bony voids or gaps of the skeletal system (i.e., the extremities and pelvis), and the opening to the defect site sealed with bone wax. These defects may be surgically created osseous defects or osseous defects created from traumatic or degenerative injury to the bone. Subsequent to implantation, Osteo-PTM Bone Graft Substitute is resorbed and replaced with bone during the healing process.
Device Story
Osteo-PTM is a radiolucent, porous, osteoconductive polymeric carbohydrate bone void filler; trabecular structure mimics human cancellous bone. Device provided sterile in various forms (granules, sheets, cubes, wedges, cylinders). Surgeon manually packs material into bony voids or gaps; defect site opening sealed with bone wax. Device supports/guides new bone ingrowth; resorbed during healing. Used in clinical settings for orthopedic/trauma procedures. Benefits patient by providing scaffold for bone regeneration in non-load-bearing defects.
Clinical Evidence
No human clinical data. Evidence based on bench testing and animal study. NZW Rabbit femoral condyle critical-sized defect study compared Osteo-PTM to predicate and autograft, demonstrating new bone formation consistent with FDA requirements for Class II Bone Void Fillers. Extensive in vitro and in vivo biocompatibility testing performed.
Technological Characteristics
Porous osteoconductive polymeric carbohydrate matrix; radiolucent. Porosity 75-95%; pore size ~1-700 µm. Non-pyrogenic. Sterile (SAL 10^-6) via E-Beam. Supplied in various preformed shapes (granules, sheets, cubes, wedges, cylinders). Does not impart mechanical strength.
Indications for Use
Indicated for treatment of bony voids or gaps not intrinsic to skeletal stability in extremities and pelvis. Applicable to surgically created osseous defects or those resulting from traumatic or degenerative injury.
Regulatory Classification
Identification
A resorbable calcium salt bone void filler device is a resorbable implant intended to fill bony voids or gaps of the extremities, spine, and pelvis that are caused by trauma or surgery and are not intrinsic to the stability of the bony structure.
Special Controls
*Classification.* Class II (special controls). The special control for this device is the FDA guidance document entitled “Class II Special Controls Guidance: Resorbable Calcium Salt Bone Void Filler Device; Guidance for Industry and FDA.” See § 888.1(e) of this chapter for the availability of this guidance.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
December 29, 2017
Molecular Matrix, Inc. % Mr. John Kapitan Chief Executive Officer Kapstone Medical P.O. Box 969 Leicester, North Carolina 28748
Re: K170165
Trade/Device Name: Osteo-PTM Bone Graft Substitute Regulation Number: 21 CFR 888.3045 Regulation Name: Resorbable calcium salt bone void filler device Regulatory Class: Class II Product Code: MOV Dated: November 27, 2017 Received: November 30, 2017
Dear Mr. Kapitan:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{1}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
# Mark N. Melkerson -S
Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2020 See PRA Statement below.
510(k) Number (if known) K170165
Device Name Osteo-PTM Bone Graft Substitute
#### Indications for Use (Describe)
Osteo-PIM Bone Graft Substitute is indicated for use only in the treatment of bony voids or gaps that are not intrinsic to the stability of the bony structure. It is intended to be gently packed into bony voids or gaps of the skeletal system (i.e., the extremities and pelvis), and the opening to the defect site sealed with bone wax. These defects may be surgically created osseous defects or osseous defects created from traumatic or degenerative injury to the bone. Subsequent to implantation, Osteo-PTM Bone Graft Substitute is resorbed and replaced with bone during the healing process.
| Type of Use (Select one or both, as applicable) | |
|---------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------|
| <div> <span> Prescription Use (Part 21 CFR 801 Subpart D) </span> </div> | <div> <span> Over-The-Counter Use (21 CFR 801 Subpart C) </span> </div> |
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) Summary
Image /page/3/Picture/3 description: The image shows the logo for Molecular Matrix. The logo consists of a stylized "M" formed by interconnected blue spheres, followed by the text "Molecular Matrix" in blue font. The word "Matrix" has a trademark symbol next to it.
In accordance with 21 CFR 807.87(h) and 21 CFR 807.92, the 510(k) summary for Osteo-P™ is provided below.
| Date Summary Prepared: | November 17, 2017 |
|-----------------------------------------------------------------|-------------------------------------------------------------------------|
| Submitter Name: | Molecular Matrix, Inc. |
| Submitter Address: | 3410 Industrial Blvd., Suite 103<br>West Sacramento, CA. 95691 |
| Contact Person: | John Kapitan<br>Kapstone Medical, LLC |
| Phone Number:<br>Fax Number: | 916-374-9404<br>530-231-6126 |
| Device Trade Name: | Osteo-PTM Bone Graft Substitute |
| Device Common Name: | Resorbable bone void filler |
| Classification Name:<br>Classification Number:<br>Product Code: | Bone Void Filler Device<br>21 CFR 888.3045<br>MQV |
| Primary Predicate Device:<br>Reference Device: | K063359 ISTO Technologies, Inc. InQu®<br>K043045 Synthes (USA) chronOS® |
### Device Description
Osteo-PTM Bone Graft Substitute is a radiolucent bone void filler composed of a porous, osteoconductive polymeric carbohydrate with a trabecular structure simulating the multidimensional interconnectivity of human cancellous bone. When placed in direct contact with host bone, Osteo-PTM supports and guides the ingrowth of new bone across the graft site and is resorbed as the healing process occurs. The consistency of Osteo-P™ allows the surgeon to easily manipulate and maximize the direct contact with viable host bone. It is intended for single patient use only and is provided sterile in single-use double sterile packaging. Osteo-P™ is available in various forms to include granules, sheets, cubes, wedges, and cylinders.
{4}------------------------------------------------
## Indications for Use
Osteo-PTM Bone Graft Substitute is indicated for use only in the treatment of bony voids or gaps that are not intrinsic to the stability of the bony structure. It is intended to be gently packed into bony voids or gaps of the skeletal system (i.e., the extremities and pelvis), and the opening to the defect site sealed with bone wax. These defects may be surgically created osseous defects or osseous defects created from traumatic or degenerative injury to the bone. Subsequent to implantation. Osteo-PTM Bone Graft Substitute is resorbed and replaced with bone during the healing process.
## Technological Characteristics
Testing has confirmed Osteo-P™ Bone Graft Substitute to be biocompatible as a bone void filler device. Animal testing demonstrated evidence of new bone formation consistent with FDA's recommendations for performance testing of Class II Bone Void Filler Devices. Osteo-P™ is substantially equivalent to the predicate device in terms of design, function, intended use and bone healing performance in an animal model. Results from a NZW Rabbit femoral condyle critical-sized defect study comparing Osteo-P™ to the predicate device and autograft as well as extensive biocompatibility testing demonstrated that the differences in raw material composition do not raise new issues of safety and effectiveness.
| Characteristic | Osteo-PTM Bone Graft<br>Substitute<br>(subject device) | InQu® Bone Graft<br>Substitute<br>(predicate device) | chronOS® Bone Void<br>Filler<br>(reference device) | Comparison |
|----------------------------------|-----------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------|--------------------------------------|
| 510(k) | | K063359 | K013072 | N/A |
| Intended Use | Resorbable Bone void<br>Filler | Resorbable Bone void<br>Filler | Resorbable Bone void<br>Filler | Equivalent |
| Target Population | Individuals with bony<br>defects resulting from<br>surgery or trauma; not<br>intrinsic to the stability of<br>the structure | Individuals with bony<br>defects resulting from<br>surgery or trauma; not<br>intrinsic to the stability of<br>the structure | Individuals with bony<br>defects resulting from<br>surgery or trauma; not<br>intrinsic to the stability of<br>the structure | Equivalent |
| Anatomical<br>Location | Bony voids or gaps of the<br>skeletal system, (i.e., the<br>extremities and pelvis) | Bony voids or gaps of the<br>skeletal system, (i.e., the<br>extremities and pelvis) | Bony voids or gaps of the<br>skeletal system, (i.e., the<br>extremities, spine and<br>pelvis) | Equivalent to<br>predicate<br>device |
| Technological<br>Characteristics | | | | |
| Principle of<br>Operation | Provide Osteoconductive<br>Matrix | Provide Osteoconductive<br>Matrix | Provide Osteoconductive<br>Matrix | Equivalent |
| Material of<br>Construction | Porous osteoconductive<br>polymer matrix | Porous osteoconductive<br>polymer matrix | Porous osteoconductive<br>polymer matrix | Equivalent |
| Shapes | Variety (to include):<br>Granules, Sheets, Cubes,<br>Wedges, and Cylinders | Variety: Granules and<br>Sheets | Variety (to include):<br>Granules, Blocks, Wedges,<br>and Cylinders | Equivalent to<br>reference<br>device |
| Sizes | Granules: 1-5mm<br>Preformed shapes: Variety | Granules: 2-5mm<br>Sheets: Variety | Granules: 1.4mm-2.8mm &<br>2.8mm-5.6mm.<br>Preformed shapes: Variety | Equivalent |
## Substantial Equivalence
{5}------------------------------------------------
| Molecular Matrix, Inc. | | | | Traditional 510(k) |
|---------------------------------------------|-----------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------|--------------------------------------|
| Porosity | 75-95% | 75-90% | 60-70% | Equivalent to<br>predicate<br>device |
| Pore Size | ~1-700 µm<br>main distribution: 50-500<br>µm | ~1-300µm | Macropores: 100-500µm<br>(main distribution)<br>Micropores: <10 µm | Equivalent |
| Radiolucency | Radiolucent | Radiolucent | Radiopaque | Equivalent to<br>predicate<br>device |
| Performance | | | | |
| Surgical<br>Application<br>Restrictions/Use | Gently packed into defect.<br>Requires appropriate<br>fixation/stabilization.<br>Requires hydration | Gently packed into defect.<br>Requires appropriate<br>fixation/stabilization.<br>Requires hydration. | Gently packed into defect.<br>Requires appropriate<br>fixation/stabilization.<br>Requires hydration | Equivalent |
| Osteoconductivity | Osteoconductive | Osteoconductive | Osteoconductive | Equivalent |
| Resorbable | Yes | Yes | Yes | Equivalent |
| Mechanical<br>Strength | Does not impart mechanical<br>strength to surgical site. | Does not impart mechanical<br>strength to surgical site. | Compressive strength<br>~5MPa | Equivalent to<br>predicate<br>device |
| Packaging | Double sterile pack. | Double sterile pack | Double sterile pack. | Equivalent |
| Sterility | Sterile, SAL 10-6 | Sterile, SAL 10-6 | Sterile, SAL 10-6 | Equivalent |
| Sterilization<br>Method | E-Beam | E-Beam | E-Beam | Equivalent |
| Pyrogenicity | Non-pyrogenic<br>Limulus Amebocyte Lysate<br>(LAL) Chromogenic<br>Endotoxin Quantitation<br>Assay | Non-pyrogenic | Non-pyrogenic | Equivalent |
| Biocompatibility | Biocompatible | Biocompatible | Biocompatible | Equivalent |
The following technological differences exist between the subject and predicate or reference devices:
| Characteristic | Osteo-PTM | InQu® | chronOS® |
|----------------------|-----------------------------------------------------|-------------------------|----------------------------------------|
| | (subject device) | (predicate device) | (reference device) |
| Material Composition | Porous Hyper Cross-linked<br>Polymeric Carbohydrate | Porous PGLA / HA matrix | Porous β-Tricalcium<br>Phosphate (TCP) |
Pre-Clinical Testing confirmed that despite differences in material composition, Osteo-P™ is equivalent to the predicate device in function, indication for use, device classification product code, environment of use, and principles of operation to the predicate device. The biocompatibility of Osteo-P™ has also been confirmed through extensive in vitro and in vivo testing.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.