The ARCHITECT Free T3 Calibrators are for the calibration of the ARCHITECT i System when used for the quantitative determination of free triiodothyronine (Free T3) in human serum and plasma.
Device Story
ARCHITECT Free T3 Calibrators are liquid, ready-to-use human serum-based standards containing L-Thyroxine and L-Triiodothyronine; used to establish reference points for the ARCHITECT i System (i2000sr, i2000, i1000sr) during quantitative Free T3 assays. The kit consists of six levels (A-F) with concentrations ranging from 0.0 to 30.0 pg/mL. Calibrators are used by laboratory personnel in clinical settings to calibrate the assay system. By providing known concentrations, the device enables the system to generate a calibration curve, allowing for the accurate measurement of patient Free T3 levels. This measurement assists clinicians in assessing thyroid status. The device is stored at 2-8°C and is stable for 11-12 months.
Clinical Evidence
Bench testing only. Stability studies were performed per CLSI EP25-A guidelines. Real-time stability (closed vial) and in-use stability (open vial) were evaluated over 11-12 months at 2-8°C. Results support the claimed shelf life and in-use stability.
Technological Characteristics
Human serum-based calibrators; preservative: sodium azide. Six-level concentration range (0.0 to 30.0 pg/mL). Standardization via gravimetric methods using L-Triiodothyronine and L-Thyroxine (USP grade, >95% purity). Compatible with ARCHITECT i System. Storage: 2-8°C. Ready-to-use liquid form factor.
Indications for Use
Indicated for use with the ARCHITECT i System for the quantitative determination of free triiodothyronine (Free T3) in human serum and plasma to aid in the assessment of thyroid status.
Regulatory Classification
Identification
A calibrator is a device intended for medical purposes for use in a test system to establish points of reference that are used in the determination of values in the measurement of substances in human specimens. (See also § 862.2 in this part.)
Special Controls
*Classification.* Class II (special controls). The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9.
K042651 — FREE T3 CALIBRATORS ON THE ACCESS IMMUNOASSAY SYSTEMS · Beckman Coulter, Inc. · Oct 5, 2004
K061885 — DIMENSION VISTA SYSTEM TOTAL TRIIODOTHYRONINE CALIBRATOR, MODEL KC250 · Dade Behring, Inc. · Aug 14, 2006
K984319 — ARCHITECT FREE T3 MASTERCHECK, MODEL NUMBER 6C48-05 · Bio-Rad · Dec 22, 1998
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" around the perimeter. Inside the circle is an abstract symbol resembling a stylized caduceus, with three human profiles forming the staff and a stylized serpent winding around them.
February 15, 2017
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
ABBOTT LABORATORIES LINDA SOHN SR. REGULATORY SPECIALIST 100 ABBOTT PARK ROAD, DEPT 09AA, BLDG. AP08 ABBOTT PARK, IL 60064
Re: K170160
Trade/Device Name: ARCHITECT Free T3 Calibrators Regulation Number: 21 CFR 862.1150 Regulation Name: Calibrator Regulatory Class: II Product Code: JIT Dated: January 17, 2017 Received: January 18, 2017
Dear Linda Sohn:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21. Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{1}------------------------------------------------
If you desire specific advice for your device on our labeling regulations (21 CFR Parts 801 and 809), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours.
# Katherine Serrano -S
For : Courtney H. Lias, Ph.D. Director Division of Chemistry and Toxicology Devices Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) k170160
Device Name ARCHITECT Free T3 Calibrators
Indications for Use (Describe)
The ARCHITECT Free T3 Calibrators are for the ARCHITECT i System when used for the quantitative determination of free triiodothyronine (Free T3) in human serum and plasma.
| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------|
|-------------------------------------------------|
X Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
### 510(k) Summary (Summary of Safety and Effectiveness)
This summary of the 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92.
#### I. Applicant Name
Abbott Laboratories Dept. 9AA, AP8-1 100 Abbott Park Road Abbott Park, IL 60064
Primary contact person for all communications:
Linda Sohn, ADD, Sr. Regulatory Affairs Specialist Phone: (224) 667-4846 Fax: (224) 667-4836 E-Mail: linda.sohn@abbott.com
Secondary contact person for all communications:
Mary Caren Murawski, Associate Director of Regulatory Affairs Phone: (224) 667-2817 Fax: (224) 667-4836 E-Mail: mary.murawski@abbott.com
Date Summary prepared: February 13, 2017
### II. Device Name
Trade Name: ARCHITECT Free T3 Calibrators Device Classification: Class II Classification Name: Calibrator, Secondary Governing Regulation: 862.1150 Code: JIT
Classification Panel: Clinical Chemistry
510(k) Number: k170160
### III. Predicate Device
ARCHITECT Free T3 Calibrators (K983439)
{4}------------------------------------------------
## IV. Description of the Device
The calibrators are devices intended for medical purposes for use in the ARCHITECT Free T 3 assay test system to establish points of reference that are used in the quantitative determination of values in the measurement of substances in human specimens. Free T3 measurements are used as an aid in the assessment of thyroid status.
The calibrators are designed to be used on the ARCHITECT i System (i 2000gg, i 2000, and i 1000sR) with the ARCHITECT Free T3 Reagents.
| Component | Number of Bottles × Volume |
|----------------------|----------------------------|
| Calibrator A (Cal A) | 1 × 4 mL |
| Calibrator B (Cal B) | 1 × 4 mL |
| Calibrator C (Cal C) | 1 × 4 mL |
| Calibrator D (Cal D) | 1 × 4 mL |
| Calibrator E (Cal E) | 1 × 4 mL |
| Calibrator F (Cal F) | 1 × 4 mL |
The ARCHITECT Free T3 Calibrator kit contains:
- . ARCHITECT Free T3 Calibrator A contains human serum. Preservative: Sodium Azide
- ARCHITECT Free T3 Calibrators B through F contain different concentrations . of L-Thyroxine and L-Triiodothyronine prepared in human serum. Preservative: Sodium Azide.
{5}------------------------------------------------
| Component | Concentration<br>(pg/mL) | Concentration<br>(pmol/L) |
|----------------------|--------------------------|---------------------------|
| Calibrator A (Cal A) | 0.0 | 0.0 |
| Calibrator B (Cal B) | 1.4 | 2.15 |
| Calibrator C (Cal C) | 3.5 | 5.38 |
| Calibrator D (Cal D) | 7.0 | 10.75 |
| Calibrator E (Cal E) | 17.2 | 26.42 |
| Calibrator F (Cal F) | 30.0 | 46.08 |
The ARCHITECT Free T3 Calibrators are prepared to target the following concentrations:
#### Value Assignment
The ARCHITECT Free T3 standardization follows a two-step assay standardization method, where market Free T3 Calibrator lots are matched to Primary Calibrators using sample/reference (S/C ratio) specifications. The Primary Calibrators are matched to Working Reference Calibrators (Abbott internal reference standards) using sample/reference (S/C ratio) specifications. The Working Reference Calibrators are manufactured by gravimetric methods using L-Triiodothyronine (Liothyronine USP) and L-Thyroxine (Levothyroxine USP).
## V. Intended Use of the Device
The ARCHITECT Free T3 Calibrators are for the calibration of the ARCHITECT i System when used for the quantitative determination of free triiodothyronine (Free T2) in human serum and plasma.
#### VI. Summary of Technological Characteristics
The following table provides the similarities and differences between the candidate calibrators ARCHITECT Free T3 and the predicate calibrators (ARCHITECT Free T3, K983439).
{6}------------------------------------------------
# Comparison of ARCHITECT Free T3 to
Predicate ARCHITECT Free T3 Assay
| Attribute | Predicate Device<br>ARCHITECT Free T3 Calibrators<br>(2-Point Calibration), K983440 | Candidate Device<br>ARCHITECT Free T3 Calibrators<br>(6-Point Calibration) |
|---------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Intended Use | The ARCHITECT Free T3 Calibrators<br>are for the calibration of the<br>ARCHITECT i System when used for<br>the quantitative determination of free<br>triiodothyronine (Free T3) in human<br>serum and plasma. | Same |
| Instrumentation | ARCHITECT i System (i 2000gr,<br>i 2000, and i 1000sR) | Same |
| Free T3<br>Calibrators | 2 levels<br>●<br>1.4 and 30 pg/mL L-Thyroxine and<br>L-Triiodothyronine in human<br>serum | 6 levels<br>●<br>0.0, 1.4, 3.5, 7.0, 17.2, 30.0<br>pg/mL L-Thyroxine and L-<br>Triiodothyronine in human<br>serum |
| Calibrator<br>Composition | Calibrator 1:<br>Analyte: None<br>Diluent: Human Serum<br>Preservative: Sodium Azide<br>Calibrator 2:<br>Analyte: L-Thyroxine and L-<br>Triiodothyronine<br>Diluent: Human Serum<br>Preservative: Sodium Azide | Calibrator A:<br>Analyte: None<br>Diluent: Human Serum<br>Preservative: Sodium Azide<br>Calibrators B-F:<br>Analyte: L-Thyroxine and L-<br>Triiodothyronine<br>Diluent: Human Serum<br>Preservative: Sodium Azide |
{7}------------------------------------------------
| Attribute | Predicate Device<br>ARCHITECT Free T3 Calibrators<br>(2-Point Calibration), K983440 | Candidate Device<br>ARCHITECT Free T3 Calibrators<br>(6-Point Calibration) |
|--------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------|
| Standardization | The calibrators are matched to an<br>Abbott internal reference standard.<br>This internal reference standard is<br>manufactured by gravimetric methods<br>based on the Free Triiodothyronine<br>calculation (FT3c) using<br>L-Triiodothyronine (sodium salt, not<br>less than 95% pure by HPLC) and<br>L-Thyroxine (sodium pentahydrate,<br>not less than 95% pure by HPLC) at<br>each concentration level. The FT3c is<br>a calculation of the Free<br>Triiodothyronine hormone<br>concentration, which depends on the<br>amount of Total T3 and Total T4 found<br>in the serum as well as the serum 's<br>thyroid hormone binding capacity. | Same |
| Storage and<br>Stability | The calibrators are stable until the<br>expiration date when stored and<br>handled as directed at 2-8°C | Same |
| Preparation of<br>Use | Ready to use | Same |
## Stability Results
Stability studies for the ARCHITECT Free T3 Calibrators were performed based on guidance from the Clinical Laboratories and Standards Institute (CLSI) document EP25-A.
## In-Use Condition (Open Vial)
The stability of the ARCHITECT Free T3 Calibrators was evaluated. At the baseline time point (0) and each subsequent time point (1, 2, 3, 4, 6, 8, and 11 months), the ontest calibrator vials were opened and squeezed to ensure good air-exchange, then inverted and squeezed to release a minimum of 8 drops of material from each vial. The vials were then closed and stored at 2 to 8°C until the next time point. In-use testing was performed using a minimum of 10 replicates each of the on-test calibrators, reference controls, and reference panel with the designated reference reagents.
{8}------------------------------------------------
The time point results were evaluated against the stability limit evaluation criteria. The results support a stability claim for the ARCHITECT Free T3 Calibrators of 11 months at 2 to 8°C. The in-use stability is an on-going study and is scheduled to continue for up to 18 months.
#### Intended Storage Condition (Closed Vial)
The stability of the ARCHITECT Free T3 Calibrators was evaluated using a real-time stability study. For the real time stability study, three lots of test material were stored at 2 to 8°C. Samples at time points 0, 1, 2, 3, 4, 5, 6, 7, 8, 9, 10, 11, and 12 months were tested and evaluated against the stability limit evaluation criteria. The results for the Intended Use stability support a stability claim of 12 months at 2 to 8°C.
The results for In-Use and Intended Storage stability studies support a stability claim for the ARCHITECT Free T3 Calibrators of 11 months at 2 to 8°C. The in-use and intended storage stability is an on-going study and is scheduled to continue for up to 18 months.
#### Standardization
The calibrators are matched to an Abbott internal reference standard. This internal reference standard is manufactured by gravimetric methods based on the Free Triiodothyronine calculation (FT3c) using L-Triiodothyronine (sodium salt, not less than 95% pure by HPLC) and L-Thyroxine (sodium pentahydrate, not less than 95% pure by HPLC) at each concentration level. The FT3c is a calculation of the Free Triiodothyronine hormone concentration, which depends on the amount of Total T3 and Total T4 found in the serum as well as the serum 's thyroid hormone binding capacity.
#### VII. Conclusion
Substantial equivalence for the candidate device, ARCHITECT Free T3 Calibrators, is claimed to the predicate device cleared in K983439, ARCHITECT Free T3 Calibrators. The modifications to the calibrators have not changed the intended use, as described in its labeling, nor have the modifications altered the fundamental scientific technology of this device.
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.