The MyoCycle is intended for general rehabilitation for: 1. Relaxation of muscle spasms 2. Prevention or retardation of disuse atrophy 3. Increasing local blood circulation 4. Maintaining or increasing range of motion
Device Story
MyoCycle is a stationary functional electrical stimulation (FES) cycle ergometer; consists of motorized ergometer, FES controller, stimulation cables, and cutaneous electrodes (up to 12). Used in home or clinical settings by patients with impaired lower extremity movement; operated by patients or clinicians. System provides active cycling via FES-evoked muscle contractions and passive cycling via motor power. Motor maintains constant 35 r/min (isokinetic cycling); automatically assists or resists based on patient torque. Software manages stimulation parameters and user data (Pro mode for multiple users; Home mode for single user). Output allows patients to perform cycling exercises despite motor impairment; benefits include muscle spasm reduction, atrophy prevention, improved circulation, and range of motion maintenance.
Clinical Evidence
No clinical trials; evidence based on bench testing and validation studies. System testing verified performance against specifications and consensus standards (IEC 60601-2-10, IEC 60601-1, IEC 60601-1-2, IEC 60601-1-11). Validation included usability testing with five lay operators (simulated patient), seven clinician operators (six spinal cord injured patients), and five spinal cord injured patients (unassisted).
Technological Characteristics
Stationary FES cycle ergometer; mains powered. Uses embedded microcontrollers with custom software. Stimulation: 0-126 mA charge-balanced stimulator. Connectivity: supports unique user ID codes. Standards: IEC 60601-2-10, AAMI ANSI ES60601-1, IEC 60601-1-2, IEC 60601-1-11. Form factor: motorized ergometer allowing use from existing seating (e.g., wheelchair).
Indications for Use
Indicated for general rehabilitation in patients requiring relaxation of muscle spasms, prevention/retardation of disuse atrophy, increased local blood circulation, and maintenance/increase of range of motion. Prescription use only.
Regulatory Classification
Identification
An external functional neuromuscular stimulator is an electrical stimulator that uses external electrodes for stimulating muscles in the leg and ankle of partially paralyzed patients (e.g., after stroke) to provide flexion of the foot and thus improve the patient's gait.
{0}------------------------------------------------
Image /page/0/Picture/2 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo is a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the top half of the circle. Inside the circle is a stylized image of three human profiles facing to the right, with flowing lines beneath them.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
April 25. 2017
Myolyn, LLC Matthew Bellman Chief Technology Officer 7731 W. Newberry Rd., Suite A-2 Gainesville, Florida 32606
Re: K170132
Trade/Device Name: Myocycle Home, Myocycle Pro Regulation Number: 21 CFR 882.5810 Regulation Name: External Functional Neuromuscular Stimulator Regulatory Class: Class II Product Code: GZI Dated: March 21, 2017 Received: March 23, 2017
Dear Mr. Bellman:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device
{1}------------------------------------------------
related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely,
# Michael J. Hoffmann -S
Carlos L. Peña, PhD, MS for Director
Division of Neurological and Physical Medicine Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K170132
Device Name MyoCycle Home, MyoCycle Pro
Indications for Use (Describe)
The MyoCycle is intended for general rehabilitation for:
- 1. Relaxation of muscle spasms
- 2. Prevention or retardation of disuse atrophy
- 3. Increasing local blood circulation
- 4. Maintaining or increasing range of motion
| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------|
|-------------------------------------------------|
X Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the word "MYOLYN" in a stylized font. The letters are all capitalized and in a thin, sans-serif font. The "O" in the word is replaced by a circular graphic containing a complex, branching pattern resembling a neuron or nerve cell. The overall design is clean and modern.
# 510(k) Summary
- (1) Submitter's name, address, telephone number, a contact person, and the date the summary was prepared:
| Company: | MYOLYN, LLC<br>7731 W. Newberry. Rd., Suite A-2<br>Gainesville, FL 32606<br>Phone: (352) 354-2749 |
|----------|---------------------------------------------------------------------------------------------------|
| | |
Contact person: Matthew Bellman, PhD Chief Technology Officer Phone: (850) 832-1842 E-mail: mjbellman@myolyn.com
Date prepared: 04-12-2017
- (2) Name of the device, including the trade or proprietary name if applicable, the common or usual name, and the classification name:
| Proprietary name: | MyoCycle Home, MyoCycle Pro |
|------------------------------|---------------------------------------------------------|
| Common name: | Functional electrical stimulation (FES) cycle ergometer |
| Classification name: | External functional neuromuscular stimulator |
| Device class: | 2 |
| Classification product code: | GZI |
| Classification panel: | Neurology |
| Regulation number: | 882.5810 |
- (3) Identification of the legally marketed device to which the submitter claims equivalence:
Manufacturer: Restorative Therapies Inc. Product: RT300-S 510(k) number: K050036 Device class: ನ
#### (4) A description of the device that is subject of the premarket notification submission:
The MyoCycle is a stationary FES cycle ergometer which is composed of:
- 1. a motorized cycle ergometer
- 2. an FES controller / stimulator
- 3. a stimulation cable which connects the controller / stimulator to cutaneous electrodes (at maximum 12 cutaneous electrodes for 6 channels)
- 4. cutaneous electrodes
This system allows a person with impaired lower extremity movement to undertake cycle ergometry both actively (utilized FES evoked lower extremity muscle contractions) and passively (utilizing power developed by the ergometer's motor). The electric motor inside
{4}------------------------------------------------
Image /page/4/Picture/1 description: The image shows the word "MYOLYN" in a stylized font. The "O" in the word is replaced by a circular graphic containing a neuron-like structure. The letters are all in uppercase and have a minimalist, geometric design. The overall design is clean and modern.
the MyoCycle ensures that the pedals always rotate at 35 revolutions per minute (r/min) (isokinetic cycling), so it automatically assists or resists the patient, depending on the magnitude and direction of the torque applied to the crank by the patient. The system software operates in two modes:
- 1. Pro mode allows for multiple users by storing unique user data that can be accessed by a unique user identification code
- 2. Home mode only stores data for one user
### (5) Indications for Use Statement:
The MyoCycle's Indications for Use (IFU) Statement is substantially equivalent to that of the predicate device, the RT300-S (K050036), as described in the table below. The MyoCycle does not have the same IFU Statement as the RT300-SP (pediatric version).
| MyoCycle IFU Statement | RT300-S (K050036) IFU Statement |
|-------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------|
| The MyoCycle is intended for general<br>rehabilitation for: | The RT300-S (adult version) and RT300-SP (pediatric version) are intended for general rehabilitation for: |
| 1. Relaxation of muscle spasms | 1. Relaxation of muscle spasms |
| 2. Prevention or retardation of disuse atrophy | 2. Prevention or retardation of disuse atrophy |
| 3. Increasing local blood circulation | 3. Increasing local blood circulation |
| 4. Maintaining or increasing range of motion | 4. Maintaining or increasing range of motion |
| | The RT300-SP (pediatric version), is intended for population ages 4 to 12 years. |
The MyoCycle is for prescription use only.
#### (6) Technological Characteristics
The function of the MyoCycle is substantially equivalent to the predicate device; however, there are certain technological similarities and differences, as described below:
| Technology | MyoCycle | RT300-S (K050036) |
|-------------------------------|--------------------------------------------------------------|--------------------------------------------------|
| Power source (energy used) | Mains power | Mains power |
| Controller | Uses embedded<br>microcontrollers running<br>custom software | Based on Pocket PC<br>running custom<br>software |
| Stimulator (energy delivered) | 0-126 mA charge<br>balanced stimulator | 0-140 mA charge<br>balanced stimulator |
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the word "MYOLYN" in a stylized font. The letters are all in uppercase and are black. The "O" in the word is replaced by a circle containing a black neuron-like structure, adding a biological or medical theme to the design. The overall design is clean and modern.
| Flywheel | Uses motor to create<br>flywheel effect with<br>reduced weight and<br>space | Uses motor to create<br>flywheel effect with<br>reduced weight and<br>space |
|-----------------|--------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------|
| Seating | Allows user to remain in<br>their own seating, e.g.<br>wheelchair eliminating<br>the need for transfer | Allows user to remain in<br>their own seating, e.g.<br>wheelchair eliminating<br>the need for transfer |
| Passive cycling | Utilizes motor to provide<br>assistance during<br>passive cycling | Utilizes motor to provide<br>assistance during<br>passive cycling |
## (b) Performance data
Testing to determine equivalence has been primarily composed of the following tests:
| Test or procedure | Description |
|------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------|
| Review of user documentation for<br>predicate device | Ensure that equivalent functionality is<br>specified and implemented in the new<br>device. |
| Review of 510(k) submission for<br>predicate device | Confirm technical specifications for<br>completion of predicate details in<br>comparison tables. |
| Output characteristic measurement of<br>new device | Confirm technical specifications for<br>completion of new device details in<br>comparison tables. |
| Conduct of system testing | Conduct system testing to verify<br>performance to specification. |
| Testing with lay operators | The MyoCycle was validated with five lay<br>operators using the device with an able<br>bodied, simulated patient. |
| Testing with clinician operators | The MyoCycle was validated with seven<br>clinician operators using the device with six<br>spinal cord injured patients. |
| Testing with patient operators | The MyoCycle was validated with five<br>spinal cord injured patients using the<br>device without assistance. |
As part of the system testing described above, the MyoCycle met the following recognized consensus standards:
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image shows the word "MYOLYN" in a stylized font. The letters are all capitalized and made of thin lines. The "O" in the word is replaced by a circle containing a black neuron-like structure with branching dendrites, adding a biological or scientific element to the design.
| Recognition<br>Number | Standards<br>Organization | Standard Designation<br>Number and Date | Title of Standard |
|-----------------------|---------------------------|-------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 17-11 | IEC | 60601-2-10 Edition 2.0<br>2012-06 | Medical Electrical Equipment -- Part<br>2-10: Particular Requirements For<br>The Basic Safety And Essential<br>Performance Of Nerve And Muscle<br>Stimulators |
| 19-4 | AAMI ANSI | ES60601-<br>1:2005/(R)2012 And<br>A1:2012 | C1:2009/(R)2012 And<br>A2:2010/(R)2012 (Consolidated Text)<br>Medical Electrical Equipment - Part<br>1: General Requirements For Basic<br>Safety And Essential Performance<br>(IEC 60601-1:2005, MOD) |
| 19-8 | IEC | 60601-1-2 Edition 4.0<br>2014-02 | Medical Electrical Equipment - Part<br>1-2: General Requirements For Basic<br>Safety And Essential Performance -<br>Collateral Standard: Electromagnetic<br>Disturbances - Requirements And<br>Tests |
| 19-14 | IEC | 60601-1-11 Edition 2.0<br>2015-01 | Medical Electrical Equipment - Part<br>1-11: General Requirements For<br>Basic Safety And Essential<br>Performance - Collateral Standard:<br>Requirements For Medical Electrical<br>Equipment And Medical Electrical<br>Systems Used In The Home<br>Healthcare Environment |
| 19-16 | AAMI ANSI | HA60601-1-11:2015 | Medical Electrical Equipment -- Part<br>1-11: General Requirements For<br>Basic Safety And Essential<br>Performance -- Collateral Standard:<br>Requirements For Medical Electrical<br>Equipment And Medical Electrical<br>Equipment And Medical Electrical<br>Systems Used In The Home<br>Healthcare Environment (IEC 60601-<br>1-11:2015 MOD) |
#### MYOLYN concludes that:
The intended use of the MyoCycle is substantially equivalent to that of the predicate device.
The MyoCycle has similar output characteristics to those of the predicate device. The different technological characteristics do not raise new questions of safety and effectiveness.
MYOLYN's performance testing has demonstrated that the MyoCycle is as safe and effective for the intended use as the predicate device and is therefore substantially equivalent.
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.