← Product Code [NKB](/productcode/NKB) · K170126

# NuVasive® Reline® 4.5-5.0 System (K170126)

_Nu Vasive, Incorporated · NKB · Mar 9, 2017 · Orthopedic · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K170126

## Device Facts

- **Applicant:** Nu Vasive, Incorporated
- **Product Code:** [NKB](/productcode/NKB.md)
- **Decision Date:** Mar 9, 2017
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 888.3070
- **Device Class:** Class 2
- **Review Panel:** Orthopedic
- **Attributes:** Therapeutic, Pediatric

## Indications for Use

When used as a pedicle screw fixation system, the NuVasive® 4.5-5.0 System is intended to provide immobilization and stabilization of spinal segments in skeletally mature patients as an adjunct to fusion of the following acute and chronic instabilities or deformities of the posterior thoracic, lumbar, and sacral spine: 1. Degenerative disc disease (as defined by back pain of discogenic origin with degeneration of the disc confirmed by patient history and radiographic studies) 2. Degenerative spondylolisthesis with objective evidence of neurologic impairment 3. Fracture 4. Dislocation 5. Scoliosis 6. Kyphosis 7. Spinal tumor and/or 8. Failed previous fusion (pseudoarthrosis) The NuVasive Reline 4.5-5.0 System is also indicated for the treatment of severe spondylolisthesis (Grades 3 and 4) of the L5-S1 vertebral joint in skeletally mature patients receiving fusion by autogenous bone graft, having the device fixed or attached to the lumbar and sacral spine (L3 to sacrum), with removal of the implants after attainment of a solid fusion. When used as an anterolateral non-pedicle screw system in the thoracic and lumbar spine, the NuVasive Reline 4.5-5.0 System is also intended for the following indications: 1. Degenerative disc disease (as defined by back pain of discogenic origin with degeneration of the disc confirmed by patient history and radiographic studies) 2. Spinal stenosis 3. Spondylolisthesis 4. Spinal deformities 5. Fracture 6. Pseudoarthosis 7. Tumor resection and/or 8. Failed previous fusion When used for posterior non-cervical screw fixation in pediatric patients, NuVasive Reline 4.5-5.0 System is indicated as an adjunct to fusion to treat adolescent idiopathic scoliosis. Additionally the NuVasive Reline 4.5-5.0 System is intended to treat pediatic patients diagnosed with the following conditions: spondylolysis, and fracture caused by turnor and/or trauma. Pediatric pedicle screw fixation is limited to a posterior approach and is intended to be used with autograft and/or allograft. In order to achieve additional levels of fixation, the NuVasive Reline 4.5-5.0 System rods may be connected to the Reline System.

## Device Story

Pedicle screw system; consists of screws, hooks, rods, lock screws, transverse connectors, rod-to-rod connectors, and iliac connectors. Used by surgeons in clinical settings to provide spinal immobilization and stabilization. Components are assembled and rigidly locked in configurations tailored to patient anatomy and pathology. Provides mechanical support to spinal segments as an adjunct to fusion; facilitates bone healing and stabilization of deformities or instabilities. Rods can connect to existing Reline System for additional fixation levels.

## Clinical Evidence

Bench testing only. No clinical data presented. Testing included static compression bending, static torsion, dynamic compression bending, and static tulip pull-off per ASTM F1717 and ASTM F1798.

## Technological Characteristics

Thoracolumbosacral pedicle screw system. Components include screws, hooks, rods, and connectors. Mechanical testing performed per ASTM F1717 (static/dynamic compression bending, static torsion) and ASTM F1798 (static tulip pull-off).

## Regulatory Identification

(1) Rigid pedicle screw systems are comprised of multiple components, made from a variety of materials that allow the surgeon to build an implant system to fit the patient's anatomical and physiological requirements. Such a spinal implant assembly consists of a combination of screws, longitudinal members (e.g., plates, rods including dual diameter rods, plate/rod combinations), transverse or cross connectors, and interconnection mechanisms (e.g., rod-to-rod connectors, offset connectors).(2) Semi-rigid systems are defined as systems that contain one or more of the following features (including but not limited to): Non-uniform longitudinal elements, or features that allow more motion or flexibility compared to rigid systems.

## Special Controls

*Classification.* (1) Class II (special controls), when intended to provide immobilization and stabilization of spinal segments in skeletally mature patients as an adjunct to fusion in the treatment of the following acute and chronic instabilities or deformities of the thoracic, lumbar, and sacral spine: severe spondylolisthesis (grades 3 and 4) of the L5-S1 vertebra; degenerative spondylolisthesis with objective evidence of neurologic impairment; fracture; dislocation; scoliosis; kyphosis; spinal tumor; and failed previous fusion (pseudarthrosis). These pedicle screw spinal systems must comply with the following special controls:(i) Compliance with material standards;
(ii) Compliance with mechanical testing standards;
(iii) Compliance with biocompatibility standards; and
(iv) Labeling that contains these two statements in addition to other appropriate labeling information:
 “Warning: The safety and effectiveness of pedicle screw spinal systems have been established only for spinal conditions with significant mechanical instability or deformity requiring fusion with instrumentation. These conditions are significant mechanical instability or deformity of the thoracic, lumbar, and sacral spine secondary to severe spondylolisthesis (grades 3 and 4) of the L5-S1 vertebra, degenerative spondylolisthesis with objective evidence of neurologic impairment, fracture, dislocation, scoliosis, kyphosis, spinal tumor, and failed previous fusion (pseudarthrosis). The safety and effectiveness of these devices for any other conditions are unknown.”
“Precaution: The implantation of pedicle screw spinal systems should be performed only by experienced spinal surgeons with specific training in the use of this pedicle screw spinal system because this is a technically demanding procedure presenting a risk of serious injury to the patient.”
(2) Class II (special controls), when a rigid pedicle screw system is intended to provide immobilization and stabilization of spinal segments in the thoracic, lumbar, and sacral spine as an adjunct to fusion in the treatment of degenerative disc disease and spondylolisthesis other than either severe spondylolisthesis (grades 3 and 4) at L5-S1 or degenerative spondylolisthesis with objective evidence of neurologic impairment. These pedicle screw systems must comply with the following special controls:
(i) The design characteristics of the device, including engineering schematics, must ensure that the geometry and material composition are consistent with the intended use.
(ii) Non-clinical performance testing must demonstrate the mechanical function and durability of the implant.
(iii) Device components must be demonstrated to be biocompatible.
(iv) Validation testing must demonstrate the cleanliness and sterility of, or the ability to clean and sterilize, the device components and device-specific instruments.
(v) Labeling must include the following:
(A) A clear description of the technological features of the device including identification of device materials and the principles of device operation;
(B) Intended use and indications for use, including levels of fixation;
(C) Identification of magnetic resonance (MR) compatibility status;
(D) Cleaning and sterilization instructions for devices and instruments that are provided non-sterile to the end user; and
(E) Detailed instructions of each surgical step, including device removal.
(3) Class II (special controls), when a semi-rigid system is intended to provide immobilization and stabilization of spinal segments in the thoracic, lumbar, and sacral spine as an adjunct to fusion for any indication. In addition to complying with the special controls in paragraphs (b)(2)(i) through (v) of this section, these pedicle screw systems must comply with the following special controls:
(i) Demonstration that clinical performance characteristics of the device support the intended use of the product, including assessment of fusion compared to a clinically acceptable fusion rate.
(ii) Semi-rigid systems marketed prior to the effective date of this reclassification must submit an amendment to their previously cleared premarket notification (510(k)) demonstrating compliance with the special controls in paragraphs (b)(2)(i) through (v) and paragraph (b)(3)(i) of this section.

## Predicate Devices

- NuVasive Reline System ([K160989](/device/K160989.md))
- NuVasive Reline ([K161014](/device/K161014.md))
- NuVasive GSB Global Spinal Balance System ([K132014](/device/K132014.md))
- NuVasive SpheRx II System ([K062317](/device/K062317.md))
- DePuy Spine Expedium Spine System ([K090230](/device/K090230.md))
- Medtronic CD HORIZON Spinal System ([K082236](/device/K082236.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

March 9, 2017

NuVasive, Incorporated Ms. Olga Lewis Regulatory Affairs Manager 7475 Lusk Boulevard San Diego, California 92121

Re: K170126

Trade/Device Name: NuVasive® Reline® 4.5-5.0 System Regulation Number: 21 CFR 888.3070 Regulation Name: Thoracolumbosacral pedicle screw system Regulatory Class: Class II Product Code: NKB, KWP, KWQ Dated: January 12, 2017 Received: January 13, 2017

Dear Ms. Lewis:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-

Image /page/0/Picture/11 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, stacked on top of each other.

{1}------------------------------------------------

related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely,

Mark N. Melkerson -S

Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

# Indications for Use

510(k) Number (if known) K170126

### Device Name

NuVasive® Reline® 4.5-5.0 System

## Indications for Use (Describe)

When used as a pedicle screw fixation system, the NuVasive® 4.5-5.0 System is intended to provide immobilization and stabilization of spinal segments in skeletally mature patients as an adjunct to fusion of the following acute and chronic instabilities or deformities of the posterior thoracic, lumbar, and sacral spine: 1. Degenerative disc disease (as defined by back pain of discogenic origin with degeneration of the disc confirmed by

patient history and radiographic studies)

- 2. Degenerative spondylolisthesis with objective evidence of neurologic impairment
- 3. Fracture
- 4. Dislocation
- 5. Scoliosis
- 6. Kyphosis
- 7. Spinal tumor and/or
- 8. Failed previous fusion (pseudoarthrosis)

The NuVasive Reline 4.5-5.0 System is also indicated for the treatment of severe spondylolisthesis (Grades 3 and 4) of the L5-S1 vertebral joint in skeletally mature patients receiving fusion by autogenous bone graft, having the device fixed or attached to the lumbar and sacral spine (L3 to sacrum), with removal of the implants after attainment of a solid fusion. When used as an anterolateral non-pedicle screw system in the thoracic and lumbar spine, the NuVasive Reline 4.5-5.0 System is also intended for the following indications:

1. Degenerative disc disease (as defined by back pain of discogenic origin with degeneration of the disc confirmed by patient history and radiographic studies)

- 2. Spinal stenosis
- 3. Spondylolisthesis
- 4. Spinal deformities
- 5. Fracture
- 6. Pseudoarthosis
- 7. Tumor resection and/or
- 8. Failed previous fusion

When used for posterior non-cervical screw fixation in pediatric patients, NuVasive Reline 4.5-5.0 System is indicated as an adjunct to fusion to treat adolescent idiopathic scoliosis. Additionally the NuVasive Reline 4.5-5.0 System is intended to treat pediatic patients diagnosed with the following conditions: spondylolysis, and fracture caused by turnor and/or trauma. Pediatric pedicle screw fixation is limited to a posterior approach and is intended to be used with autograft and/or allograft.

In order to achieve additional levels of fixation, the NuVasive Reline 4.5-5.0 System rods may be connected to the Reline System.

Type of Use (Select one or both, as applicable)

2 Prescription Use (Part 21 CFR 801 Subpart D)

_ Over-The-Counter Use (21 CFR 801 Subpart C)

{3}------------------------------------------------

PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.

## FOR FDA USE ONLY

Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)

This section applies only to requirements of the Paperwork Reduction Act of 1995.

## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW *

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.qov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{4}------------------------------------------------

Image /page/4/Picture/0 description: The image shows the Nuvasive logo. The logo consists of a purple and gray leaf-like shape on the left, followed by the word "NUVASIVE" in gray, sans-serif font. The logo is simple and modern, and the colors are muted and professional.

## 510(k) Summary

In accordance with Title 21 of the Code of Federal Regulations, Part 807, and in particular 21 CFR §807.92, the following summary of information is provided:

#### Submitted by: A.

Olga Lewis Regulatory Affairs Manager NuVasive, Incorporated 7475 Lusk Blvd. San Diego, California 92121 Telephone: (858) 909-1800

Date Prepared: January 30, 2017

#### B. Device Name

| Trade or Proprietary Name: | NuVasive® Reline® 4.5-5.0 System                                                                                                   |
|----------------------------|------------------------------------------------------------------------------------------------------------------------------------|
| Common or Usual Name:      | Pedicle Screw System                                                                                                               |
| Classification Name:       | Thoracolumbosacral Pedicle Screw System, Spinal Interlaminal<br>Fixation Orthosis, Spinal Intervertebral Body Fixation<br>Orthosis |
| Device Class:              | Class II                                                                                                                           |
| Classification:            | 21 CFR § 888.3050, 888.3060, 888.3070                                                                                              |
| Product Code:              | NKB, KWP, KWQ                                                                                                                      |

#### C. Predicate Devices

The subject device is substantially equivalent to the primary predicate device NuVasive Reline System (K160989) and additional predicate devices. NuVasive Reline (K161014), NuVasive GSB Global Spinal Balance System (K132014), NuVasive SpheRx II System (K062317), DePuy Spine Expedium Spine System (K090230), and Medtronic CD HORIZON Spinal System (K082236).

#### D. Device Description

The NuVasive Reline 4.5-5.0 System is a pedicle screw system that consists of a varietv screws, hooks, rods, lock screws, transverse connectors, rod-to-rod connectors, iliac connectors and associated general instruments. Implant components are available in a variety sizes and can be rigidly locked into a variety of different configurations to suit the individual pathology and anatomical conditions of the patient.

#### E. Indications for Use

When used as a pedicle screw fixation system, the NuVasive Reline 4.5-5.0 System is intended to provide immobilization and stabilization of spinal segments in skeletally mature patients as an adjunct to fusion in the treatment of the following acute and chronic instabilities or deformities of the posterior thoracic, lumbar, and sacral spine:

- 1.Degenerative disc disease (as defined by back pain of discogenic origin with degeneration of the disc confirmed by patient history and radiographic studies)
- 2. Degenerative spondylolisthesis with objective evidence of neurologic impairment
- 3.Fracture

{5}------------------------------------------------

Image /page/5/Picture/1 description: The image shows the Nuvasive logo. The logo consists of a stylized leaf shape in purple and gray on the left, followed by the word "NUVASIVE" in gray. The leaf shape is divided into two parts, with the top part in purple and the bottom part in gray. The word "NUVASIVE" is in a sans-serif font.

- 4. Dislocation
- 5. Scoliosis
- 6. Kyphosis
- 7. Spinal tumor and/or
- 8. Failed previous fusion (pseudoarthrosis)

The NuVasive Reline 4.5-5.0 System is also indicated for the treatment of severe spondylolisthesis (Grades 3 and 4) of the L5-S1 vertebral joint in skeletally mature patients receiving fusion by autogenous bone graft, having the device fixed or attached to the lumbar and sacral spine (L3 to sacrum), with removal of the implants after attainment of a solid fusion.

When used as an anterolateral non-pedicle screw system in the thoracic and lumbar spine, the NuVasive Reline 4.5-5.0 System is also intended for the following indications:

- 1. Degenerative disc disease (as defined by back pain of discogenic origin with degeneration of the disc confirmed by patient history and radiographic studies)
- 2. Spinal stenosis
- 3. Spondylolisthesis
- 4. Spinal deformities
- 5.Fracture
- 6. Pseudoarthosis
- 7. Tumor resection and/or
- 8.Failed previous fusion

When used for posterior non-cervical screw fixation in pediatric patients, NuVasive Reline 4.5-5.0 System is indicated as an adjunct to fusion to treat adolescent idiopathic scollosis. Additionally the NuVasive Reline 4.5-5.0 System is intended to treat pediatric patients diagnosed with the following conditions: spondylolisthesis, spondylolysis, and fracture caused by tumor and/or trauma. Pediatric pedicle screw fixation is limited to a posterior approach and is intended to be used with autograft and/or allograft.

In order to achieve additional levels of fixation, the NuVasive Reline 4.5-5.0 System rods may be connected to the Reline System.

#### F. Technological Characteristics

As was established in this submission, the subject NuVasive Reline 4.5-5.0 System is substantially equivalent to other predicate devices cleared by the FDA for commercial distribution in the United States. The subject device was shown to be substantially equivalent and have equivalent technological characteristics to its predicate device through comparison in areas including design, labeling/intended use, material composition, and function.

#### G. Performance Data

Nonclinical testing was performed to demonstrate that the subject NuVasive Reline 4.5-5.0 System is substantially equivalent to the predicate device. The following testing was performed:

- Static Compression Bending per ASTM F1717
- Static Torsion per ASTM F1717 ●

{6}------------------------------------------------

Image /page/6/Picture/0 description: The image shows the logo for Nuvasive. The logo consists of a stylized leaf shape in purple and gray, followed by the word "NUVASIVE" in gray sans-serif font. The leaf shape is on the left side of the logo, and the word "NUVASIVE" is on the right side.

- Dynamic Compression Bending per ASTM F1717 .
- Static Tulip Pull-off per ASTM F1798 ●

The results demonstrate that the subject NuVasive Reline 4.5-5.0 System is substantially equivalent to the predicate.

#### Conclusions H.

The subject NuVasive Reline 4.5-5.0 System has been shown to be substantially equivalent to legally marketed predicate devices for its intended use.

---

**Source:** [https://fda.innolitics.com/device/K170126](https://fda.innolitics.com/device/K170126)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
