← Product Code [LZO](/productcode/LZO) · K170089

# U-Motion II Acetabular System— Additional sizes (K170089)

_United Orthopedic Corporation · LZO · Oct 6, 2017 · Orthopedic · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K170089

## Device Facts

- **Applicant:** United Orthopedic Corporation
- **Product Code:** [LZO](/productcode/LZO.md)
- **Decision Date:** Oct 6, 2017
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 888.3353
- **Device Class:** Class 2
- **Review Panel:** Orthopedic
- **Attributes:** Therapeutic

## Indications for Use

The device is used for reduction or relief of pain and/or improved hip function in skeletally mature patients with the following conditions: 1. Painful, disabling joint disease of the hip resulting from: degenerative arthritis, rheumatoid arthritis, post-traumatic arthritis or late stage avascular necrosis. 2. Revision of pervious unsuccessful femoral head replacement, cup arthroplasty or other procedure. 3. Clinical management problems where arthrodesis or alternative reconstructive techniques are less likely to achieve satisfactory results. 4. Correction of function deformity. 5. Treatment of nonunion femoral neck and trochanteric fracture of the proximal femur with head involvement that is unmanageable using other techniques. The device is intended for cementless use.

## Device Story

The U-Motion II Acetabular System is a total hip prosthesis component consisting of an acetabular cup and XPE cup liner. This submission introduces additional sizes (46 mm and 50 mm O.D. cups; 32 mm and 36 mm I.D. liners) to the previously cleared system. The device is implanted by orthopedic surgeons in a clinical setting to replace diseased or damaged hip joints. It functions as a semi-constrained prosthetic to restore hip function and reduce pain. The device is intended for cementless fixation. Clinical benefit is derived from the restoration of joint mechanics and relief of pain associated with degenerative or traumatic hip conditions.

## Clinical Evidence

Bench testing only. Mechanical property evaluations were performed to compare the new sizes against the existing system, confirming that the additional sizes do not negatively impact safety or effectiveness.

## Technological Characteristics

Total hip prosthesis (acetabular cup and XPE liner). Cementless design. Materials, design, and manufacturing methods are identical to the predicate U-Motion II system. Dimensions: 46 mm and 50 mm O.D. cups; 32 mm and 36 mm I.D. liners.

## Regulatory Identification

A hip joint metal/ceramic/polymer semi-constrained cemented or nonporous uncemented prosthesis is a device intended to be implanted to replace a hip joint. This device limits translation and rotation in one or more planes via the geometry of its articulating surfaces. It has no linkage across-the-joint. The two-part femoral component consists of a femoral stem made of alloys to be fixed in the intramedullary canal of the femur by impaction with or without use of bone cement. The proximal end of the femoral stem is tapered with a surface that ensures positive locking with the spherical ceramic (aluminium oxide, A12 03 ) head of the femoral component. The acetabular component is made of ultra-high molecular weight polyethylene or ultra-high molecular weight polyethylene reinforced with nonporous metal alloys, and used with or without bone cement.

## Predicate Devices

- U-Motion II Acetabular System ([K122185](/device/K122185.md))
- U-Motion II PS+ Cup ([K132455](/device/K132455.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logos of the U.S. Department of Health & Human Services and the Food and Drug Administration (FDA). The Department of Health & Human Services logo is on the left, featuring a stylized eagle. To the right of that is the FDA logo, which includes the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.

United Orthopedic Corporation October 6, 2017 Gimpel Chien Regulatory Affairs Manager No 57, Park Ave 2, Science Park Hsinchu 300 Taiwan Re: K170089 Trade/Device Name: U-motion II Acetabular System- Additional Sizes Regulation Number: 21 CFR 888.3353 Regulation Name: Hip Joint Metal/Ceramic/Polymer Semi-Constrained Cemented Or Nonporous Uncemented Prosthesis Regulatory Class: Class II Product Code: LZO, MEH, KWY, LWJ Dated: September 1, 2017 Received: September 05, 2017

Dear Ms. Chien:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820);

{1}------------------------------------------------

and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education (DICE) at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education (DICE) at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely.

## Mark N. Melkerson -S

Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

| DEPARTMENT OF HEALTH AND HUMAN SERVICES |                                   |
|-----------------------------------------|-----------------------------------|
| Food and Drug Administration            |                                   |
|                                         | Form Approved: OMB No. 0910-0120  |
| <i>Indications for Use</i>              | Expiration Date: January 31, 2017 |
|                                         | See PRA Statement below.          |

| 510(k) Number (if known)                                                                                                                                                | K170089                                         |
|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------|
| Device Name                                                                                                                                                             | U-Motion II Acetabular System- Additional sizes |
| Indications for Use (Describe)                                                                                                                                          |                                                 |
| The device is used for reduction or relief of pain and/or improved hip function in skeletally mature patients with the following conditions:                            |                                                 |
| 1. Painful, disabling joint disease of the hip resulting from: degenerative arthritis, rheumatoid arthritis, post-traumatic arthritis or late stage avascular necrosis. |                                                 |
| 2. Revision of pervious unsuccessful femoral head replacement, cup arthroplasty or other procedure.                                                                     |                                                 |
| 3. Clinical management problems where arthrodesis or alternative reconstructive techniques are less likely to achieve satisfactory results.                             |                                                 |
| 4. Correction of function deformity.                                                                                                                                    |                                                 |
| 5. Treatment of nonunion femoral neck and trochanteric fracture of the proximal femur with head involvement that is unmanageable using other techniques.                |                                                 |
| The device is intended for cementless use.                                                                                                                              |                                                 |

| Type of Use (Select one or both, as applicable)                                                               |                                                                                               |
|---------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------|
| <div> <input checked="true" type="checkbox"/> Prescription Use (Part 21 CFR 801 Subpart D)             </div> | <div> <input type="checkbox"/> Over-The-Counter Use (21 CFR 801 Subpart C)             </div> |

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

***DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.***

The burden time for this collection of information is estimated to average 79 hours per response, including the
time to review instructions, search existing data sources, gather and maintain the data needed and complete
and review the collection of information. Send comments regarding this burden estimate or any other aspect
of this information collection, including suggestions for reducing this burden, to:

 Department of Health and Human Services  

 Food and Drug Administration  

 Office of Chief Information Officer  

 Paperwork Reduction Act (PRA) Staff  

 PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of
information unless it displays a currently valid OMB number."

| FORM FDA 3881 (8/14) | Page 1 of 1 |
|----------------------|-------------|
|----------------------|-------------|

K170089 page 1 of 1

12

UOC-FDA-047

•|

్రదేశం

{3}------------------------------------------------

# യ U-Motion II Acetabular System— Additional sizes

### 510(k) Summary of Safety and Effectiveness

### Submitter Information

| Name                          | United Orthopedic Corporation                                                                                                                                                                                                                                                                                  |
|-------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Address                       | No 57, Park Ave 2, Science Park, Hsinchu 300, Taiwan                                                                                                                                                                                                                                                           |
| Phone Number                  | +886-3-5773351 ext. 2217                                                                                                                                                                                                                                                                                       |
| Fax Number                    | +886-3-577156                                                                                                                                                                                                                                                                                                  |
| Name of Contact Person        | Gimpel Chien<br>Regulation and Document Management                                                                                                                                                                                                                                                             |
| Date prepared                 | January 05, 2017                                                                                                                                                                                                                                                                                               |
| Name of Device                |                                                                                                                                                                                                                                                                                                                |
| Trade Name                    | U-Motion II Acetabular System- Additional sizes                                                                                                                                                                                                                                                                |
| Common Name                   | Total hip prostheses                                                                                                                                                                                                                                                                                           |
| Regulation Name and<br>Number | The device classification for U-Motion II Acetabular<br>System— Additional sizes is “Hip join<br>metal/ceramic/polymer semi-constrained cemented or<br>nonporous uncemented prosthesis” and is contained in the<br>Code of Federal Regulation, under 21CFR 888.3353. This<br>falls under the Orthopedic Panel. |
| Device Class                  | Class II                                                                                                                                                                                                                                                                                                       |
| Classification Panel          | Orthopaedics                                                                                                                                                                                                                                                                                                   |
| Product Code                  | LZO, KWY, LWJ, MEH                                                                                                                                                                                                                                                                                             |
| Predicate Device              | 1. U-Motion II Acetabular System (K122185)<br>2. U-Motion II PS+ Cup (K132455)                                                                                                                                                                                                                                 |

{4}------------------------------------------------

### U-Motion II Acetabular System— Additional sizes

#### Device Description:

This device is an extension in terms of additional size to the previously cleared U-Motion II Acetabular System (K122185) and U-Motion II PS+ Cup (K132455). U-Motion II acetabular component includes U-Motion II Cup and U-Motion II XPE Cup Liner. The indications, materials, design, safety and effectiveness of U-Motion II Cup and Cup Liner are identical to the cleared U-Motion II Acetabular System (K122185) and U-Motion II PS* Cup (K132455) except the dimension. The subject U-Motion II Cups are available in size Ø 50 mm and Ø 46 mm O.D. and the subject U-Motion II Cup Liners are available in 32 mm and 36 mm I.D. The 32 mm and 36 mm inserts fit the acetabular shells with 46 mm O.D. and 50 mm O.D., respectively.

#### Indications:

The device is used for reduction or relief of pain and/or improved hip function in skeletally mature patients with the following conditions:

- 1. Painful, disabling joint disease of the hip resulting from: degenerative arthritis, rheumatoid arthritis, post-traumatic arthritis or late stage avascular necrosis.
- 2. Revision of previous unsuccessful femoral head replacement, cup arthroplasty or other procedure.
- 3. Clinical management problems where arthrodesis or alternative reconstructive techniques are less likely to achieve satisfactory results.
- 4. Correction of functional deformity.
- Treatment of nonunion femoral neck and trochanteric fracture of the proximal 5. femur with head involvement that is unmanageable using other techniques.

The device is intended for cementless use.

#### Comparison to Predicate Device:

U-Motion II Acetabular System— Additional sizes has the same basic design, intended use, materials and the same manufacturing method as device of the U-Motion II Acetabular System (K122185) and U-Motion II PS+ Cup (K132455).

{5}------------------------------------------------

# യ U-Motion II Acetabular System— Additional sizes

#### Performance Data:

The mechanical properties of this device have been evaluated, and the analysis results shown that this subjected device is not the worst case within all sizes of U-Motion II Acetabular System. These adding sizes would not affect the safety and effectiveness.

---

**Source:** [https://fda.innolitics.com/device/K170089](https://fda.innolitics.com/device/K170089)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
