K170065 · Siemens Healthcare Diagnostics, Inc. · JFM · Mar 9, 2017 · Clinical Chemistry
Device Facts
Record ID
K170065
Device Name
ADVIA® Chemistry Total Bilirubin_2 (TBIL_2)
Applicant
Siemens Healthcare Diagnostics, Inc.
Product Code
JFM · Clinical Chemistry
Decision Date
Mar 9, 2017
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 862.1110
Device Class
Class 2
Attributes
Real-World Evidence, Pediatric
Real-World Evidence
Submission
Device
Sponsor
RWD Sources
RWE Use Summary
Key Tags
K170065 · Mar 9, 2017
ADVIA® Chemistry Total Bilirubin_2 (TBIL_2)
Siemens Healthcare Diagnostics, Inc.
Patient samples (clinical laboratory samples)
Patient samples were used in a method comparison study to demonstrate the accuracy of the ADVIA® Chemistry Total Bilirubin 2 assay by comparing it to a legally marketed comparator method.
For in vitro diagnostic use in the quantitative determination of total bilirubin in human serum and plasma of adults and neonates on the ADVIA® Chemistry systems. Measurement of total bilirubin, an organic compound formed during the normal and abnormal destruction of red blood cells, is used in the diagnosis and treatment of liver, hemolytic hematological, and metabolic disorders, including hepatitis and gall bladder block. A total bilirubin measurement in newborn infants is intended to aid in indicating the risk of bilirubin encephalopathy (kernicterus).
Device Story
ADVIA® Chemistry Total Bilirubin_2 (TBIL_2) is a quantitative colorimetric assay for use on ADVIA® Chemistry systems; utilizes vanadate oxidation method to measure total bilirubin in human serum and plasma. Reagents (citrate buffer/detergent and phosphate buffer/sodium metavanadate) oxidize bilirubin to biliverdin at pH 2.9; reaction causes decrease in optical density at 451/545 nm proportional to bilirubin concentration. Used in clinical laboratory settings by trained personnel. Output provides total bilirubin concentration (mg/dL) to assist clinicians in diagnosing liver/metabolic disorders and assessing kernicterus risk in neonates.
Clinical Evidence
Method comparison study performed per CLSI EP09-A3 using 119 patient samples. Comparison against a legally marketed method showed a correlation coefficient (r) of 0.990, slope of 1.06, and y-intercept of -0.24. Analytical performance included linearity (0.15-35.0 mg/dL), LoB (0.02 mg/dL), LoD (0.06 mg/dL), and LoQ (0.08 mg/dL). Interference testing confirmed acceptable performance (≤10% bias) for Indican, Cyanokit, HbF, and HbA. No new clinical prospective studies were required as adult population data was leveraged from previous clearance (K063845).
Indicated for quantitative determination of total bilirubin in serum and plasma of adults and neonates to aid in diagnosis/treatment of liver, hemolytic, hematological, and metabolic disorders (e.g., hepatitis, gall bladder block) and to assess risk of bilirubin encephalopathy (kernicterus) in newborns.
Regulatory Classification
Identification
A bilirubin (total or direct) test system is a device intended to measure the levels of bilirubin (total or direct) in plasma or serum. Measurements of the levels of bilirubin, an organic compound formed during the normal and abnormal distruction of red blood cells, if used in the diagnosis and treatment of liver, hemolytic hematological, and metabolic disorders, including hepatitis and gall bladder block.
Predicate Devices
Abbott Laboratories Total Bilirubin (k150510)
Submission Summary (Full Text)
{1}
1
# 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY ASSAY ONLY TEMPLATE
A. 510(k) Number:
k170065
B. Purpose for Submission:
Addition of claim, for use in neonate serum and plasma
C. Measurand:
Total Bilirubin
D. Type of Test:
Quantitative colorimetric, vanadate oxidation
E. Applicant:
Siemens Healthcare Diagnostics Inc.
F. Proprietary and Established Names:
ADVIA® Chemistry Total Bilirubin_2 (TBIL_2)
G. Regulatory Information:
| Product Code | Classification | Regulation | Panel |
| --- | --- | --- | --- |
| JFM | II | 21 CFR §862.1110 | Clinical Chemistry (75) |
| MQM | I, reserved | 21 CFR §862.1113 | |
H. Intended Use:
1. Intended use(s):
See indications for use statement below.
2. Indication(s) for use:
For in vitro diagnostic use in the quantitative determination of total bilirubin in human serum and plasma of adults and neonates on the ADVIA® Chemistry systems.
{2}
Measurement of total bilirubin, an organic compound formed during the normal and abnormal destruction of red blood cells, is used in the diagnosis and treatment of liver, hemolytic hematological, and metabolic disorders, including hepatitis and gall bladder block. A total bilirubin measurement in newborn infants is intended to aid in indicating the risk of bilirubin encephalopathy (kernicterus).
3. Special conditions for use statement(s):
This device is for prescription use only.
4. Special instrument requirements:
Assay performance was demonstrated on the ADVIA® 1800 Chemistry Analyzer.
I. Device Description:
The ADVIA® Chemistry Total Bilirubin_2 (TBIL_2) consists of two reagents that are liquid and ready to use. The reagents are packaged in wedges to be used directly on the ADVIA® Chemistry Systems.
Reagent 1 contains citrate buffer, pH 2.9 (0.1 M) and detergent. Reagent 2 contains phosphate buffer, pH 7.0 (10 mM) and 4 mM sodium metavanadate.
J. Substantial Equivalence Information:
1. Predicate device name(s):
Abbott Laboratories Total Bilirubin
2. Predicate 510(k) number(s):
k150510
3. Comparison with predicate:
| Similarities | | |
| --- | --- | --- |
| Item | Candidate Device: ADVIA® Chemistry Total Bilirubin_2 (TBIL_2) (k170065) | Predicate Device: Abbott Laboratories Total Bilirubin (k150510) |
| Intended use | For in vitro diagnostic use in the quantitative determination of total bilirubin in human serum and plasma of adults and neonates. | Same |
| Measurand | Total Bilirubin | Same |
| Sample Types | Serum and plasma from adults and neonates | Same |
{3}
| Similarities | | |
| --- | --- | --- |
| Item | Candidate Device: ADVIA® Chemistry Total Bilirubin_2 (TBIL_2) (k170065) | Predicate Device: Abbott Laboratories Total Bilirubin (k150510) |
| Detection of Analyte | End-point colorimetric | Same |
| Reagent Form | Liquid, ready to use | Same |
| Differences | | |
| --- | --- | --- |
| Item | Candidate Device: ADVIA® Chemistry Total Bilirubin_2 (TBIL_2) (k170065) | Predicate Device: Abbott Laboratories Total Bilirubin (k150510) |
| Assay Principle | The bilirubin is oxidized by vanadate at about pH 2.9 to produce biliverdin. In the presence of the detergent and the vanadate, both conjugated (direct) and unconjugated bilirubin are oxidized. This oxidation reaction causes the decrease in the optical density of the yellow color, which is specific to bilirubin. The decrease in optical density at 451/545 nm is proportional to the total bilirubin concentration in the sample. | Total (conjugated and unconjugated) bilirubin couples with a diazo reagent in the presence of a surfactant to form azobilirubin. The diazo reaction is accelerated by the addition of surfactant as a solubilizing agent. The increase in absorbance at 548 nm due to azobilirubin is directly proportional to the total bilirubin concentration. |
| Measurement Protocol | Vanadate oxidation method | Diazo colorimetric method |
| Reagent Composition | R1: Citrate buffer, pH 2.9 (0.1 mol/L); Detergent R2: Phosphate buffer, pH 7.0 (10 mmol/L); Sodium metavanadate (4 mmol/L) | R1: Surfactants (10.57%), HCl (6.563 g/L) R2: 2, 4-dichloroaniline (0.81 g/L), HCl (5.563 g/L), sodium nitrate (0.345 g/L), Surfactant (1.96%) |
| Measuring Range | 0.15 - 35 mg/dL | 0.3 - 25 mg/dL |
| Expected values | Age: 0-1 day: <8.0 mg/dL 1-2 days: <12.0 mg/dL 3-5 days: <16.0 mg/dL >5 days - 60 years: 0.3-1.2 mg/dL* 60 - 90 years: 0.2-1.1 mg/dL >90 years: 0.2-0.9 mg/dL *Ages >5 days to <29 days are neonates and >29 days to 60 years are children and adults. | Premature Newborn <24 hours: <8.0 mg/dL <48 hours: <12.0 mg/dL 3 to 5 days: <15.0 mg/dL 7 days: <15.0 mg/dL Full Term Newborn <24 hours: <6.0 mg/dL <48 hours: <10.0 mg/dL 3 to 5 days: <12.0 mg/dL 7 days: <10.0 mg/dL Adults: 0.3 - 1.2 mg/dL |
{4}
| Differences | | |
| --- | --- | --- |
| Item | Candidate Device: ADVIA® Chemistry Total Bilirubin_2 (TBIL_2) (k170065) | Predicate Device: Abbott Laboratories Total Bilirubin (k150510) |
| Instrument | ADVIA® Chemistry 1800 System | ARCHITECT c8000 System |
K. Standard/Guidance Document Referenced (if applicable):
CLSI EP05-A2. Evaluation of Precision Performance of Clinical Chemistry Devices; Approved Guideline; Second Edition.
CLSI EP06-A. Evaluation of the Linearity of Quantitative Analytical Measurement Procedure: A Statistical Approach, Approved Guideline.
CLSI EP07-A2. Interference Testing in Clinical Chemistry, Approved Guideline, Second Edition.
CLSI EP09-A3. Measurement Procedure Comparison and Bias Estimation Using Patient Samples, Approved Guideline, Third Edition.
CLSI EP 17-A2. Evaluation of Detection Capability for Clinical Laboratory Measurement Procedures; Approved Guideline; Second Edition.
CLSI EP28-A3c. Defining, Establishing, and Verifying Reference Intervals in the Clinical Laboratory, Approved Guideline, Third Edition.
L. Test Principle:
The ADVIA® Chemistry Total Bilirubin_2 (TBIL_2) method is based on a chemical oxidation method using vanadate as an oxidizing agent. The bilirubin is oxidized by vanadate at about pH 2.9 to produce biliverdin. In the presence of the detergent and the vanadate, both conjugated (direct) and unconjugated bilirubin are oxidized. This oxidation reaction causes the decrease in the optical density of the yellow color, which is specific to bilirubin. The decrease in optical density at 451/545 nm is proportional to the total bilirubin concentration in the sample. The concentration is measured as an endpoint reaction.
M. Performance Characteristics (if/when applicable):
1. Analytical performance:
a. Precision/Reproducibility:
Precision for the ADVIA® Chemistry Total Bilirubin_2 (TBIL_2) was previously established in k063845.
{5}
# b. Linearity/assay reportable range:
Linearity was evaluated according to CLSI EP06-A. A high serum pool was prepared by spiking human serum with unconjugated bilirubin at a known concentration. A low serum pool was prepared by dilution of human serum with saline. Dilutions were prepared by mixing different proportions of the high and low serum pools to create nine (9) samples (0, 5, 9.9, 14.8, 19.8, 24.7, 29.6, 34.5, and $39.2\mathrm{mg / dL}$ ), which spanned the assay range. Samples were assayed in triplicate and the mean result was compared to the expected values determined by the dilution scheme. Linear regression analysis of the mean observed concentration and expected values is summarized in the table below.
| Slope | y- intercept | r |
| --- | --- | --- |
| 0.999 | 0.0510 | 0.999 |
# c. Traceability, Stability, Expected values (controls, calibrators, or methods):
The ADVIA® Chemistry Calibrator, traceable to the AACC reference method (Doumas), was previously cleared in k030169.
# d. Detection limit:
Limit of blank (LoB), limit of detection (LoD), and limit of quantitation (LoQ) were determined in accordance with CLSI EP17-A2. The LoB, LoD, and LoQ for bilirubin were determined by running replicate measurements on blank and low level analyte serum samples using three reagent lots, one calibrator lot, one system, and one operator across multiple days.
LoB was determined by running 4 blank samples in 5 replicates over 3 days (for a total of 60 replicates per lot). LoD and LoQ were determined by testing 4 low analyte samples in 20 replicates over 3 days. LoB and LoD were calculated nonparametrically. The limit of quantitation (LoQ) was calculated for each reagent lot with the maximum LoQ across all reagent lots taken as the LoQ for the assay. The results are summarized in the table below.
| LoB | LoD | LoQ |
| --- | --- | --- |
| 0.02 mg/dL | 0.06 mg/dL | 0.08 mg/dL |
The measuring range of the assay is 0.15 - 35.1 mg/dL and is supported by the linearity and detection limit studies.
# e. Analytical specificity:
Interference testing of ascorbic acid and triglycerides for ADVIA® Chemistry Total Bilirubin 2 (TBIL_2) was established in k063845. Interferences typically seen in
{6}
neonates were evaluated according to CLSI EP7-A2 using one reagent lot and one calibrator lot.
Interference from cyanokit and 3 endogenous compounds was evaluated using native serum samples with low ( $\sim 1\mathrm{mg/dL}$ ) and high ( $\sim 10\mathrm{mg/dL}$ ) bilirubin levels. Bilirubin samples were prepared by spiking a native serum pool with approximately $1\mathrm{mg/dL}$ or $10\mathrm{mg/dL}$ bilirubin. Test samples were prepared by spiking each drug or endogenous substance (at least five concentrations per interferent) into the low and high bilirubin samples. Test samples were assayed in triplicate and the mean concentration of bilirubin for each interferent concentration was compared to control samples containing no interferent. The sponsor defined significant interference as greater than or equal to $10\%$ difference between the test and control samples.
| Potentially Interfering Substance | Highest concentration of interferent tested that did not show significant interference |
| --- | --- |
| Indican | 10 mg/dL |
| Cyanokit | 40 μg/mL |
| Fetal Hemoglobin | 1000 mg/dL |
| Hemoglobin A | 1000 mg/dL |
f. Assay cut-off:
Not applicable.
2. Comparison studies:
a. Method comparison with predicate device:
A method comparison study was performed in accordance with CLSI EP09-A3. The levels of total bilirubin were measured using the ADVIA® Chemistry Total Bilirubin_2 (TBIL_2) and a legally marketed comparator method with an assay measuring range of 0.146 - 35.1 mg/dL.
One hundred and nineteen (119) serum samples from neonates with ages of 5 hours to 8 days were tested. The range of samples tested was $0.8 - 26.6\mathrm{mg / dL}$ and included five spiked samples. Data were analyzed using a weighted Deming linear regression.
| N | 119 |
| --- | --- |
| Slope | 1.06 |
| y-intercept | -0.24 |
| Correlation coefficient (r) | 0.990 |
Bias at the medical decision levels (MDLs) were calculated from the linear regression equation. MDL concentrations (bias) are as follows: $1.0\mathrm{mg / dL}$ $(-0.2\mathrm{mg / dL})$ , 8.0 mg/dL $(0.2\mathrm{mg / dL})$ , $13.0\mathrm{mg / dL}$ $(0.5\mathrm{mg / dL})$ , and $17.0\mathrm{mg / dL}$ $(0.8\mathrm{mg / dL})$ .
{7}
b. Matrix comparison:
Matrix equivalency for neonate patient samples was not evaluated. In k063845, commutability between serum and lithium heparin plasma was evaluated using adult samples.
# 3. Clinical studies:
a. Clinical Sensitivity:
Not applicable.
b. Clinical specificity:
Not applicable.
c. Other clinical supportive data (when a. and b. are not applicable):
Not applicable.
# 4. Clinical cut-off:
Not applicable.
# 5. Expected values/Reference range:
The reference interval for the ADVIA® Chemistry Total Bilirubin_2 (TBIL_2) is outlined in the literature reference: Wu AHB. Tietz Clinical Guide to Laboratory Tests, $4^{th}$ edition, Saunders Elsevier, St. Louis, MO: 2006:172. Verification of the described reference range was completed in accordance with CLSI EP28-A3c with a minimum of 20 samples for each age category listed. Each age group was analyzed separately with data summarized in the table below. The reference range verification study supports the use of the literature reference interval.
| Age Range | Acceptable Range |
| --- | --- |
| 0-1 day | <8.0 mg/dL |
| 1-2 days | <12 mg/dL |
| 3-5 days | <16.0 mg/dL |
| >5 days – 60 years | 0.3-1.2 mg/dL |
| 60-90 years | 0.2-1.1 mg/dL |
| >90 years | 0.2-0.9 mg/dL |
{8}
In addition, the clinical interpretation of the risk for hyperbilirubinemia in neonates is supported by the following literature reference: Subcommittee on Hyperbilirubinemia. Subcommittee on Hyperbilirubinemia. Management of Hyperbilirubinemia in the Newborn Infant 35 or More Weeks of Gestation. Pediatrics 2004; 114: 297-3 16. This reference indicated high risk for developing clinically significant hyperbilirubinemia in term and near-term newborns as summarized in the table below.
| Age | Total Bilirubin |
| --- | --- |
| 24 hours | ≥ 8.0 mg/dL |
| 48 hours | ≥13.0 mg/dL |
| 84 hours | ≥17.0 mg/dL |
N. Proposed Labeling:
The labeling is sufficient and it satisfies the requirements of 21 CFR Part 809.10.
O. Conclusion:
The submitted information in this premarket notification is complete and supports a substantial equivalence decision.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.