← Product Code [DQY](/productcode/DQY) · K170041

# XableCath Support Catheter (K170041)

_Xablecath, Inc. · DQY · Sep 7, 2017 · Cardiovascular · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K170041

## Device Facts

- **Applicant:** Xablecath, Inc.
- **Product Code:** [DQY](/productcode/DQY.md)
- **Decision Date:** Sep 7, 2017
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 870.1250
- **Device Class:** Class 2
- **Review Panel:** Cardiovascular

## Indications for Use

The XableCath catheter is intended to be used to facilitate access to discrete regions of the peripheral vasculature in conjunction with steerable guidewires. This device may be used to facilitate placement and exchange of guidewires and other interventional devices.

## Device Story

XableCath Support Catheter is an over-the-wire, single-lumen catheter designed for peripheral vascular access. It features a blunt tip, cortically reinforced co-polymeric shaft, and radiopaque cobalt-chromium-tungsten-nickel distal tip. The device is used in hospital surgical suites or labs by physicians. The operator advances the catheter manually over a steerable guidewire under fluoroscopic imaging to reach target sites in the peripheral vasculature. It serves as a conduit for guidewire placement and exchange of interventional devices. The device benefits patients by facilitating navigation through stenotic regions of the vasculature. It is provided sterile for single-use.

## Clinical Evidence

No clinical tests were performed. Substantial equivalence is supported by bench testing, including visual/technical inspections, luer/sheath compatibility, guidewire retraction/reinsertion, radiopacity, torque transmission, simulated use (iliac model), dimensional assessment, leak testing, kink-resistance, tensile testing, and corrosion testing. Biocompatibility was confirmed via a full panel of GLP-compliant tests (cytotoxicity, sensitization, irritation, systemic toxicity, hemolysis, thromboresistance, etc.).

## Technological Characteristics

Single-lumen catheter; cortically reinforced co-polymeric shaft; radiopaque cobalt-chromium-tungsten-nickel distal tip; olefin strain relief; polycarbonate hub with female luer. Dimensions vary by model (5Fr-7Fr). Manual advancement; fluoroscopic guidance. Sterilized via Gamma radiation (SAL 10^-6). Complies with ISO 10555-1.

## Regulatory Identification

A percutaneous catheter is a device that is introduced into a vein or artery through the skin using a dilator and a sheath (introducer) or guide wire.

## Predicate Devices

- Merit Medical SureCross® Catheter ([K123311](/device/K123311.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the Department of Health & Human Services - USA. The logo is a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized image of three human profiles facing right, with flowing lines suggesting movement or connection. The logo is black and white.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

September 7, 2017

XableCath, Inc. Mr. Rick Gaykowski Chief Regulatory Officer 417 S. Wakara Way, Suite 3510 Salt Lake City, UT 84108

Re: K170041

Trade/Device Name: XableCath Support Catheter Regulation Number: 21 CFR 870.1250 Regulation Name: Percutaneous Catheter Regulatory Class: Class II Product Code: DOY Dated: July 19, 2017 Received: July 24, 2017

Dear Mr. Gaykowski:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{1}------------------------------------------------

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely,

Kenneth J. Cavanaugh -S

for

Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

### Indications for Use

510(k) Number (if known) K170041

Device Name XableCath Support Catheter

Indications for Use (Describe)

The XableCath catheter is intended to be used to facilitate access to discrete regions of the peripheral vasculature in conjunction with steerable guidewires. This device may be used to facilitate placement and exchange of guidewires and other interventional devices.

Type of Use (Select one or both, as applicable)

| <span style="font-size: 10pt;">☑ Prescription Use (Part 21 CFR 801 Subpart D)</span> | <span style="font-size: 10pt;">☐ Over-The-Counter Use (21 CFR 801 Subpart)</span> |
|--------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------|
|--------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------|

__ Over-The-Counter Use (21 CFR 801 Subpart C)

### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

# 510(k) Section 6 510(k) SUMMARY

## XableCath Support Catheter

| Date prepared:                                                                                                                                                                                                            | August 28, 2017                                              |                     |                                   |
|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------|---------------------|-----------------------------------|
| Submitter information [21 CFR 807.929(a)(1)]                                                                                                                                                                              |                                                              |                     |                                   |
| Name:                                                                                                                                                                                                                     | XableCath, Inc.                                              |                     |                                   |
| Address:                                                                                                                                                                                                                  | 417 S. Wakara Way, Ste 3510, Salt Lake City, Utah 84108-1457 |                     |                                   |
| Phone Number:                                                                                                                                                                                                             | (617) 447-4000 Mobile                                        |                     |                                   |
| Fax number:                                                                                                                                                                                                               | N/A                                                          |                     |                                   |
| Establishment Registration Number:                                                                                                                                                                                        | Yet to be secured                                            |                     |                                   |
| Name of contact person:                                                                                                                                                                                                   | Rick Gaykowski, Chief Regulatory Officer                     |                     |                                   |
| Name of the device [21 CFR 807.92(a)(2)]                                                                                                                                                                                  |                                                              |                     |                                   |
| Trade or proprietary name:                                                                                                                                                                                                | XableCath Support Catheter                                   |                     |                                   |
| Common or usual name:                                                                                                                                                                                                     | Peripheral Vascular Support Catheter                         |                     |                                   |
| Classification name:                                                                                                                                                                                                      | Percutaneous Catheter                                        |                     |                                   |
| Classification number:                                                                                                                                                                                                    | 21 CFRR 870.1250, Procode DQY                                |                     |                                   |
| Legally marketed device(s) to which equivalence is claimed [21 CFR 807.92(a)(3)]                                                                                                                                          |                                                              |                     |                                   |
| Merit Medical SureCross® Catheter (K123311)                                                                                                                                                                               |                                                              |                     |                                   |
| Device description [21 CFR 807.92(a)(4)]                                                                                                                                                                                  |                                                              |                     |                                   |
| The XableCath Support Catheter device is an over-the-wire (OTW) single-lumen catheter, with blunt tip design. The catheter is offered in a product family configuration, with size models as outlined in the table below: |                                                              |                     |                                   |
| Catalog Number                                                                                                                                                                                                            |                                                              | Dimensions          |                                   |
| Blunt Tip (B)                                                                                                                                                                                                             | French Size                                                  | Working Length (cm) | Recommended Guidewire Inches (mm) |
| XC-014-065B                                                                                                                                                                                                               | 5Fr                                                          | 65                  | 0.014" (0.36 mm)                  |
| XC-014-090B                                                                                                                                                                                                               | 5Fr                                                          | 90                  | 0.014" (0.36 mm)                  |
| XC-014-145B                                                                                                                                                                                                               | 5Fr                                                          | 145                 | 0.014" (0.36 mm)                  |
| XC-018-065B                                                                                                                                                                                                               | 6Fr                                                          | 65                  | 0.018" (0.46 mm)                  |
| XC-018-090B                                                                                                                                                                                                               | 6Fr                                                          | 90                  | 0.018" (0.46 mm)                  |
| XC-018-145B                                                                                                                                                                                                               | 6Fr                                                          | 145                 | 0.018" (0.46 mm)                  |
| XC-035-065B                                                                                                                                                                                                               | 7Fr                                                          | 65                  | 0.035" (0.89 mm)                  |
| XC-035-090B                                                                                                                                                                                                               | 7Fr                                                          | 90                  | 0.035" (0.89 mm)                  |
| XC-035-145B                                                                                                                                                                                                               | 7Fr                                                          | 145                 | 0.035" (0.89 mm)                  |

{4}------------------------------------------------

The catheter shaft consists of a cortically reinforced co-polymeric material blend, with radiopaque cobalt chromiumtungsten-nickel distal tip. The proximal portion is comprised of an olefin strain relief and polycarbonate hub, which includes a female luer connection with the catheter lumen. The catheter pass-through lumen is used to advance the catheter over an appropriately sized guidewire for intravascular advancement to the target site. Component connections are ensured through inter-mechanical locks and utilization of medical grade adhesives. Catheter quidewire compatibility size/length are declared on the XableCath Support Catheter strain relief. All provided family member models are compatible with ≥4Fr introducer sheaths. All configurations are provided sterile, single-use only.

### Indications for Use [21 CFR 807.92(a)(5)]

The XableCath catheter is intended to facilitate access to discrete regions of the peripheral vasculature in conjunction with steerable quidewires. This device may be used to facilitate placement and exchange of quidewires and other interventional devices.

### Comparison of the technological characteristics with the predicate device [21 CFR 807.92(a)(6)]

### The subject devices and predicate devices are based on the following similar technological elements:

The XableCath Support Catheter indication for use is comparable to the predicate device, with direct overlap for peripheral application. Both device systems provide support access to the vasculature, utilize similar insertion and deployment principles, and allow for quidewire and other interventional device exchanges. Both products are for use in endovascular catheter utilization and deployment. Both products are configured with uniquely design distal tip regions for intravascular passage across stenotic regions. Furthermore, both catheters possess similarly offered working lengths, and identical dimensional quidewire compatibility (i.e., 0.014", and 0.035").

Both products have similar intended therapeutic applications, used in similar patient populations, and employ equivalent technological characteristics in achieving desired technical methods-of-action through manual advancement under fluoroscopic imaging. Similar categories of product materials comprise the subject and predicate products, with both containing hub/luer, shaft body, and distal tip design regions.

The following table illustrates comparable features of the subject and predicate products:

#### Substantial Equivalence Comparison

| Substantial Equivalence Comparison |                                                                                                                                                                                                                                                                                                     |                                                                                                                                                                                                                                                                                                                                                              |
|------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Product Information                | Subject<br>XableCath Support Catheter                                                                                                                                                                                                                                                               | Predicate<br>Merit Medical SureCross®<br>Catheter                                                                                                                                                                                                                                                                                                            |
| 510(k) #                           | TBD<br>This Submission                                                                                                                                                                                                                                                                              | K123311                                                                                                                                                                                                                                                                                                                                                      |
| FDA Classification                 | Class II                                                                                                                                                                                                                                                                                            | Class II                                                                                                                                                                                                                                                                                                                                                     |
| Regulation Number                  | 870.1250                                                                                                                                                                                                                                                                                            | 870.1250                                                                                                                                                                                                                                                                                                                                                     |
| Product Code                       | DQY                                                                                                                                                                                                                                                                                                 | DQY                                                                                                                                                                                                                                                                                                                                                          |
| Indications for Use                | The XableCath catheter is<br>intended to be used to facilitate<br>access to discrete regions of the<br>peripheral vasculature in<br>conjunction with steerable<br>guidewires. This device may be<br>used to facilitate placement and<br>exchange of guidewires and other<br>interventional devices. | The Access & Support Catheter is<br>intended for use during coronary and<br>peripheral interventional procedures<br>to guide and support guidewires,<br>traverse discrete portions of the<br>vasculature, allow for guidewire<br>exchange and provide a conduit for<br>infusion of saline solution, diagnostic<br>contrast agents and therapeutic<br>agents. |
| Proximal Luer/Hub                  | HDPE                                                                                                                                                                                                                                                                                                | HDPE                                                                                                                                                                                                                                                                                                                                                         |
| Strain Relief                      | Olefin                                                                                                                                                                                                                                                                                              | PEBA                                                                                                                                                                                                                                                                                                                                                         |
| Catheter Shaft                     | Nylon, Single lumen                                                                                                                                                                                                                                                                                 | HDPE, Single lumen                                                                                                                                                                                                                                                                                                                                           |
| Markers                            | Cobalt-chromium/SS                                                                                                                                                                                                                                                                                  | Pt/Ir                                                                                                                                                                                                                                                                                                                                                        |
| Coating                            | N/A                                                                                                                                                                                                                                                                                                 | Hydrophilic                                                                                                                                                                                                                                                                                                                                                  |
| Infusion Pressure/Burst Strength   | ≥300 psi                                                                                                                                                                                                                                                                                            | ≥300 psi                                                                                                                                                                                                                                                                                                                                                     |
| Outer Diameter                     | Varies by Model<br>Proximal: 0.043", 0.053", 0.062"<br>Distal: 0.060", 0.070", 0.080"<br>Tip: 0.052", 0.063", 0.071"                                                                                                                                                                                | Varies by Model<br>Proximal: 0.039", 0.044", 0.063"<br>Distal: 0.0265", 0.0305", 0.0505"<br>Tip: 0.0195", 0.0225", 0.0405"                                                                                                                                                                                                                                   |
| Guidewire Compatibility            | 0.014", 0.018", 0.035"                                                                                                                                                                                                                                                                              | 0.014", 0.018", 0.035"                                                                                                                                                                                                                                                                                                                                       |
| Guide Catheter (max)               | ≥4Fr                                                                                                                                                                                                                                                                                                | 5Fr                                                                                                                                                                                                                                                                                                                                                          |
| Access Sheath (max)                | ≥4Fr                                                                                                                                                                                                                                                                                                | 5Fr                                                                                                                                                                                                                                                                                                                                                          |
| Effective Working Length           | 65, 90, 145cm                                                                                                                                                                                                                                                                                       | 65, 90, 135, 150cm                                                                                                                                                                                                                                                                                                                                           |
| Deployment                         | OTW - Manual                                                                                                                                                                                                                                                                                        | OTW - Manual                                                                                                                                                                                                                                                                                                                                                 |
| Sterilization Method               | Gamma<br>(SAL – 10-6)                                                                                                                                                                                                                                                                               | EO<br>(SAL – 10-6)                                                                                                                                                                                                                                                                                                                                           |
| Single Use, Sterile                | Yes                                                                                                                                                                                                                                                                                                 | Yes                                                                                                                                                                                                                                                                                                                                                          |
| Labeling                           | Individual IFU: Warning, Cautions,<br>Contraindications, tables, images,<br>organized outline.                                                                                                                                                                                                      | Individual IFU: Warning, Cautions,<br>Contraindications, tables, images,<br>organized outline.                                                                                                                                                                                                                                                               |
| Packaging                          | Sterile thermal sealed Tyvek/PET<br>Pouch, SBS Carton                                                                                                                                                                                                                                               | Sterile thermal-sealed Pouch, SBS<br>Carton                                                                                                                                                                                                                                                                                                                  |
| Use Environment                    | Rx Only – By or on the order of a<br>physician. Hospital, Lab/Surgical<br>Suite                                                                                                                                                                                                                     | Rx Only – By or on the order of a<br>physician. Hospital, Lab/Surgical<br>Suite                                                                                                                                                                                                                                                                              |
| Performance Testing                | ISO 10555-1 Second Edition 2013-<br>07-01. Intravascular catheters --<br>Sterile, single-use intravascular<br>catheters -- Part 1: General                                                                                                                                                          | ISO 10555-1 Second Edition 2013-<br>07-01. Intravascular catheters --<br>Sterile, single-use intravascular<br>catheters -- Part 1: General                                                                                                                                                                                                                   |

{5}------------------------------------------------

The following technological differences exist between the subject and predicate devices:

The subject and predicate devices have slightly differing distal tip configurations, unique to their particular design, .

{6}------------------------------------------------

| and overall product composition. Nevertheless, both product distal tip designs are intended to achieve the same outcomes and have been shown to be commensurate in performance through comparative assessment.                                                                                                                                                                                                                                                                                                                                                                                                                          |                                                                                                                                                                                                                                                                                        |                                                                                                                                                                                                                                                                         |
|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| •                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       | The predicate product contains an exterior hydrophilic coating over the distal 40cm of length, whilst the subject product does not. Nevertheless, ease of insertion and tracking within applied models yields equivalent product delivery performance through comparative assessment.  |                                                                                                                                                                                                                                                                         |
| •                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       | The predicate product is indicated for coronary application, and for delivery of diagnostic/therapeutic solutions, whilst the subject product is not.                                                                                                                                  |                                                                                                                                                                                                                                                                         |
| Despite these noted differences between the subject and predicate products, these attributes are determined to be cumulatively insignificant as demonstrated through direct product comparative assessment.                                                                                                                                                                                                                                                                                                                                                                                                                             |                                                                                                                                                                                                                                                                                        |                                                                                                                                                                                                                                                                         |
| <b>Performance Data [21 CFR 807.92(b)]</b>                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              |                                                                                                                                                                                                                                                                                        |                                                                                                                                                                                                                                                                         |
| <b>Summary of non-clinical tests conducted for determination of substantial equivalence [21 CFR 807.92(b)(1)]</b>                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       |                                                                                                                                                                                                                                                                                        |                                                                                                                                                                                                                                                                         |
| •                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       | Direct product bench <i>in-vitro</i> comparison testing has shown the subject & predicate products to be substantially equivalent, via assessment within the following areas:                                                                                                          |                                                                                                                                                                                                                                                                         |
|                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         | <ul><li>Visual &amp; technical inspections</li><li>Luer syringe compatibility</li><li>Sheath compatibility</li><li>Guidewire retraction/reinsertion</li><li>Relative radiopacity comparison</li><li>Torque transmission &amp; capability</li><li>Simulated use (iliac model)</li></ul> | <ul><li>Dimensional assessment &amp; comparisons</li><li>Guidewire compatibility</li><li>Leak Testing</li><li>Catheter kink-resistance</li><li>Tensile testing (proximal &amp; distal)</li><li>Corrosion testing</li><li>General packaging, shelf-life/expiry</li></ul> |
| •                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       | Full panel biocompatibility was successfully performed in accord with product classification, under GLP rigors, demonstrating that all utilized materials and methods of construction/processing passed biocompatibility rigors. Conducted test included:                              |                                                                                                                                                                                                                                                                         |
|                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         | <ul><li>Cytotoxicity</li><li>Irritation/Intracutaneous Reactivity</li><li>Hemolysis</li><li>Complement Activation</li><li>Platelet and Leukocyte Count</li></ul>                                                                                                                       | <ul><li>Sensitization</li><li>Systemic Toxicity (Acute)</li><li>Thromboresistance</li><li>Partial Thromboplastin Time</li><li>Materials Mediated Pyrogenicity</li></ul>                                                                                                 |
| •                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       | Packaging integrity, transport challenge testing, and shelf-life testing were applied and successfully completed in accordance with established acceptance criteria demonstrating configurational adequacy.                                                                            |                                                                                                                                                                                                                                                                         |
| <b>Summary of clinical tests conducted for determination of substantial equivalence or of clinical information [21 CFR 807.92(b)(2)]</b>                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                |                                                                                                                                                                                                                                                                                        |                                                                                                                                                                                                                                                                         |
| No clinical tests were necessary. However human factors engineering and usability assessments were performed.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           |                                                                                                                                                                                                                                                                                        |                                                                                                                                                                                                                                                                         |
| <b>Conclusions drawn [21 CFR 807.92(b)(3)]</b>                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          |                                                                                                                                                                                                                                                                                        |                                                                                                                                                                                                                                                                         |
| Based upon the overlapping intended use, basal product designs, methods of deployment, target population and anatomical site overlap, and direct bench comparative assessment the preponderance of gathered evidence within this pre-market notification application demonstrates the XableCath Support Catheter has been shown to be substantially equivalent to the predicate device. Furthermore, the XableCath Support Catheter raises no new questions of safety or effectiveness when compared directly to the predicate device, and is therefore justifiably concluded to be substantially equivalent for declared intended use. |                                                                                                                                                                                                                                                                                        |                                                                                                                                                                                                                                                                         |

---

**Source:** [https://fda.innolitics.com/device/K170041](https://fda.innolitics.com/device/K170041)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
