← Product Code [NHA](/productcode/NHA) · K170013

# Eztetic BellaTek Encode Healing Abutments (K170013)

_Biomet 3i · NHA · Sep 21, 2017 · Dental · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K170013

## Device Facts

- **Applicant:** Biomet 3i
- **Product Code:** [NHA](/productcode/NHA.md)
- **Decision Date:** Sep 21, 2017
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 872.3630
- **Device Class:** Class 2
- **Review Panel:** Dental

## Indications for Use

The Eztetic™ BellaTek® Encode® Healing Abutments are intended for use as an accessory to endosseous dental implants during endosseous and gingival healing to prepare gingival tissue for acceptance of a final abutment and restoration.

## Device Story

Healing abutment accessory for endosseous dental implants; facilitates gingival tissue healing; consists of abutment and retaining screw; machined from Ti-6Al-4V ELI titanium alloy; features Encode coding scheme (machined markings) for implant position/orientation; used in dental clinics by clinicians; fixated on implant via retaining screw in one-stage or two-stage surgical protocols; soft tissue sutured around device; remains in mouth until soft tissue development; provides passive/clearance fit within implant internal connection; supports gingival tissue preparation for final restoration.

## Clinical Evidence

Bench testing only. No clinical data. Testing included torque to failure, physical fit evaluation, tolerance analysis, and print verification. Biocompatibility and MR compatibility leveraged from identical materials in previously tested Biomet 3i devices. Packaging integrity validated per ISO 11607-1 and ASTM D4169.

## Technological Characteristics

Material: Titanium Alloy (Ti-6Al-4V ELI) per ASTM F136. Form: 2.9mm restorative platform, 3-8mm heights, 3.8-5mm emergence profiles. Connection: Flat-on-flat engagement with internal implant connection. Sterilization: Gamma radiation (25-38 kGy) per ISO 11137-1/2. Shelf life: 5 years. Coding: Machined markings for implant orientation.

## Regulatory Identification

An endosseous dental implant abutment is a premanufactured prosthetic component directly connected to the endosseous dental implant and is intended for use as an aid in prosthetic rehabilitation.

## Special Controls

*Classification.* Class II (special controls). The guidance document entitled “Class II Special Controls Guidance Document: Root-Form Endosseous Dental Implants and Endosseous Dental Implant Abutments” will serve as the special control. (See § 872.1(e) for the availability of this guidance document.)

## Predicate Devices

- BellaTek® Encode® Healing Abutments ([K072642](/device/K072642.md))

## Reference Devices

- Healing Collars ([K142082](/device/K142082.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image contains the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized graphic of three human profiles facing to the right, stacked on top of each other.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

September 21, 2017

BIOMET 3i Krupal Patel Senior Regulatory Specialist 4555 Riverside Drive Palm Beach Gardens, Florida 33410

Re: K170013

Trade/Device Name: Eztetic™ BellaTek® Encode® Healing Abutments Regulation Number: 21 CFR 872.3630 Regulation Name: Endosseous Dental Implant Abutment Regulatory Class: Class II Product Code: NHA Dated: September 1, 2017 Received: September 5, 2017

Dear Krupal Patel:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices. good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{1}------------------------------------------------

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely,

Mary S. Runner -S

for Tina Kiang, Ph.D. Acting Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

# INDICATIONS FOR USE

## 510(k) Number (if known): K170013

Device Name: Eztetic™ BellaTek® Encode® Healing Abutments

### Indications for Use:

The Eztetic™ BellaTek® Encode® Healing Abutments are intended for use as an accessory to endosseous dental implants during endosseous and gingival healing to prepare gingival tissue for acceptance of a final abutment and restoration.

| Prescription Use            | X |
|-----------------------------|---|
| (Part 21 CFR 801 Subpart D) |   |
| Over-The-Counter Use        |   |
| (21 CFR 801 Subpart C)      |   |

AND/OR

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

{3}------------------------------------------------

# Eztetic™ BellaTek® Encode® Healing Abutments 510(K) Summary 09/21/2017

#### I. Submitter Information:

|      | Name:                       | Biomet 3i                                            |  |
|------|-----------------------------|------------------------------------------------------|--|
|      | Address:                    | 4555 Riverside Drive                                 |  |
|      |                             | Palm Beach Gardens, Florida 33410                    |  |
|      | Phone:                      | (561) 776-6700                                       |  |
|      | Contact Person:             | Krupal Patel                                         |  |
|      | Job Title:                  | Senior Regulatory Specialist                         |  |
|      | Email:                      | krupal.patel@zimmerbiomet.com                        |  |
|      | Telephone:                  | (561) 776-6923                                       |  |
|      | Fax:                        | (561) 514-6316                                       |  |
| II.  | Proprietary Trade Name:     | Eztetic™ BellaTek® Encode® Healing Abutments         |  |
| III. | Device Classification Name: | Endosseous Dental Implant Abutment (21 CFR 872.3630) |  |
| IV.  | Regulatory Class:           | Class II                                             |  |
| V.   | Product Code:               | NHA                                                  |  |

### VI. Predicate Devices:

Primary predicate device:

- BellaTek® Encode® Healing Abutments referenced in Biomet 3i Dental Abutments and Restorative Components (K072642 / SE 12/20/2007)
Reference device:

- . Healing Collars referenced in Zimmer 3.1mm Dental Implant System (K142082 / SE 07/28/2014)
### VII. Product Description:

The Eztetic™ BellaTek® Encode® Healing Abutment is a healing abutment designed to facilitate gingival tissue healing. It consists of an abutment and a retaining screw that are assembled together. Both components are machined from Titanium Alloy (Ti-6AL-4V ELI). The shelf life of the Eztetic™ BellaTek® Encode® Healing Abutment is 5 years and they are intended for single use only. The device is packaged in a sealed nylon bag and sold sterile. The device is sterilized using gamma irradiation method.

{4}------------------------------------------------

#### VIII. Indications for Use:

The Eztetic™ BellaTek® Encode® Healing Abutments are intended for use as an accessory to endosseous dental implants during endosseous and gingival healing to prepare gingival tissue for acceptance of a final abutment and restoration.

#### IX. Summary of the Technological Characteristics:

The Eztetic™ BellaTek® Encode® Healing Abutments is available in 2.9mm restorative platform diameter, four heights: 3, 4, 6 and 8mm and two emergence profiles: 3.8 and 5mm. The height, emergence profile and the screw method of attachment with mating implant are same as primary predicate devices. The major change from the primary predicate device is the restorative platform diameter. However the restorative platform diameter is same as the reference device. The connection geometry of subject device is similar to reference device and they both engage Eztetic™ 3.1mm Dental Implant. The abutment form a passive or clearance fit within the implant internal connection. The retaining screw is used to secure the abutment to the implant. The abutment makes a flaton-flat engagement with top surface of the implant platform. A general device comparison of subject and predicate device is provided in table below.

| Feature                                 | Subject Device<br>Eztetic™ BellaTek®<br>Encode® Healing<br>Abutment                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   | Primary Predicate<br>Device<br>BellaTek® Encode®<br>Healing Abutment<br>(K072642)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  | Reference Device<br>Healing Collars<br>(K142082)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               |
|-----------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Intended Use/<br>Indications for<br>Use | The Eztetic™<br>BellaTek® Encode®<br>Healing Abutments<br>are intended for use<br>as an accessory to<br>endosseous dental<br>implants during<br>endosseous and<br>gingival healing to<br>prepare gingival<br>tissue for acceptance<br>of a final abutment<br>and restoration.                                                                                                                                                                                                                                                                                         | Biomet 3i Dental<br>Abutments are intended<br>for use as accessories to<br>endosseous dental implant<br>to support a prosthetic<br>device in a partially or<br>completely edentulous<br>patient. A dental abutment<br>is intended for use to<br>support single and<br>multiple tooth prosthesis,<br>in the mandible or<br>maxilla. The prosthesis<br>can be screw retained or<br>cement retained to the<br>abutment.<br>Restorative Components:<br>● Temporary Healing                                                                                                                                                                                                                                                             | Zimmer 3.1mmD Dental<br>Implants are designed for use<br>in the anterior maxilla or<br>mandible for immediate<br>loading or for loading after a<br>conventional healing period.<br>Implants may be used to<br>replace one or more missing<br>teeth. Immediate loading is<br>indicated when there is good<br>primary stability and an<br>appropriate occlusal load.<br>Zimmer 3.1mmD Dental<br>Implants may be placed<br>immediately following an<br>extraction or loss of natural<br>teeth provided there is<br>sufficient volume of alveolar<br>bone to minimally support the                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 |
|                                         | Subject Device                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        | Primary Predicate<br>Device                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        | Reference Device                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               |
| Feature                                 | Eztetic™ BellaTek®<br>Encode® Healing<br>Abutment                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     | BellaTek® Encode®<br>Healing Abutment<br>(K072642)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 | Healing Collars<br>(K142082)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   |
|                                         |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       | Abutments are<br>intended for use to<br>shape and maintain<br>the soft tissue opening<br>during healing.<br>Castable restorative<br>components are<br>intended for use as<br>accessories to<br>endosseous dental<br>implants to aid in the<br>fabrication of dental<br>prosthetics. Screw components are<br>intended for use as<br>accessories to<br>endosseous dental<br>implants for retention<br>of screw retained<br>abutments to the<br>dental implant.                                                                                                                                                                                                                                                                       | implant (minimum 1mm<br>circumferential and 2mm<br>apical). The Zimmer 3.1mmD<br>Dental Implants should be<br>splinted to additional implants<br>when used in the pre-molar<br>region and should not be used<br>in the molar region.<br><br>The 2.9mm Angled Abutment<br>and the 2.9mm Angled<br>Abutment, Straight Hex are<br>used for attachment of<br>restorations requiring off-axis<br>correction. The 2.9mm Angled<br>Abutment and the 2.9mm<br>Angled Abutment, Straight<br>Hex are designed to be used in<br>edentulous or partially<br>edentulous mandibles or<br>maxillae for attachment of<br>complete denture prostheses,<br>or as a terminal or<br>intermediary abutment for<br>fixed or removable<br>bridgework, or as a<br>freestanding single tooth<br>replacement.<br><br>The 2.9mm Contour Abutment<br>and 2.9mm Contour Abutment<br>Straight Hex are used as a<br>terminal or intermediate<br>abutment for a cemented<br>prosthesis. Abutment can be<br>used for a single- or multiple-<br>unit restoration. The 2.9mm<br>Angled Contour Abutment and<br>2.9mm Angled Contour<br>Abutment, Straight Hex are |
|                                         |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    | designed to be used as a                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       |
| Feature                                 | Subject Device                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        | Primary Predicate<br>Device                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        | Reference Device                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               |
|                                         | EzteticTM BellaTek®<br>Encode® Healing<br>Abutment                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    | BellaTek® Encode®<br>Healing Abutment<br>(K072642)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 | Healing Collars<br>(K142082)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   |
|                                         |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    | terminal or intermediate<br>abutment for a cemented<br>prosthesis where the angle<br>needs to be offset by 17º.<br>Abutment can be used for a<br>single-or multiple-unit<br>restoration.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       |
|                                         |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    | The 2.9mm Temporary<br>Abutment is intended to be<br>used to fabricate and support<br>provisional restorations that<br>aid in creating an esthetic<br>emergence through the gingiva<br>during the healing period and<br>prior to final restoration. The<br>2.9mm Temporary Abutment<br>can be used for cement-<br>retained or screw-retained<br>provisional restorations. The<br>abutments can be used for<br>single-unit and multiple-unit<br>restorations.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   |
|                                         |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    | The Ball Abutment is used for<br>retaining overdentures or<br>partial dentures when<br>resilience and facilitated oral<br>hygiene are desired.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 |
|                                         |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    | The Healing Collar is used to<br>assist in the forming of the soft<br>tissue during healing before a<br>final restoration is placed.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           |
|                                         |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    | The Healing Screw is used to<br>seal the implant internal<br>connection and separate it<br>from the soft tissue which is<br>sutured over the implant                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           |
| Feature                                 | Subject Device<br>Eztetic™ BellaTek®<br>Encode® Healing<br>Abutment                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   | Primary Predicate<br>Device<br>BellaTek® Encode®<br>Healing Abutment<br>(K072642)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  | Reference Device<br>Healing Collars<br>(K142082)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               |
|                                         |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    | The Retaining Screws are<br>intended to be used for<br>securing the temporary<br>abutments, final abutments and<br>impression transfers to the<br>implant or implant analog. The<br>long Retaining Screw is<br>intended to be used with<br>Temporary Abutments for<br>fabrication of screw-retained<br>provisional restorations and<br>with Impression Transfers for<br>direct impressions.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    |
| Operating<br>Principle                  | The Eztetic™<br>BellaTek® Encode®<br>Healing Abutment<br>aids in prosthetic<br>rehabilitation by<br>supporting the<br>surrounding gingival<br>tissue during the<br>healing period. The<br>healing abutment is<br>fixated on the<br>implant using a<br>retaining screw<br>immediately after<br>implant placement<br>in a one-stage<br>surgical protocol. In<br>a two-stage surgical<br>protocol, the healing<br>abutment is fixated<br>on the implant<br>following the bone<br>healing period. The<br>soft tissue is sutured<br>around the healing<br>abutment and the | The BellaTek® Encode®<br>Healing Abutment aids in<br>prosthetic rehabilitation<br>by supporting the<br>surrounding gingival<br>tissue during the healing<br>period. The healing<br>abutment is fixated on the<br>implant using a retaining<br>screw immediately after<br>implant placement in a<br>one-stage surgical<br>protocol. In a two-stage<br>surgical protocol, the<br>healing abutment is<br>fixated on the implant<br>following the bone<br>healing period. The soft<br>tissue is sutured around<br>the healing abutment and<br>the device remains in the<br>mouth until the soft tissue<br>fully develops. The<br>occlusal surface of the<br>device includes machined<br>markings that provide<br>information about the | The Eztetic™ Healing Collar<br>is designed to facilitate<br>gingival tissue healing. The<br>Eztetic™ Healing Collar aids<br>in prosthetic rehabilitation by<br>supporting the surrounding<br>gingival tissue during the<br>healing period. The soft tissue<br>is sutured around the healing<br>collar and the device remains<br>in the mouth until the soft<br>tissue fully develops. The<br>healing collar is threaded onto<br>the implant immediately after<br>implant placement in a one-<br>stage surgical protocol. In a<br>two-stage surgical protocol,<br>the healing collar is threaded<br>on the implant following the<br>bone healing period.                                                                                                                                                                                                                                                                                                                                                                                                                                                                        |
|                                         | Subject Device                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        | Primary Predicate<br>Device                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        | Reference Device                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               |
| Feature                                 | Eztetic™ BellaTek®<br>Encode® Healing<br>Abutment                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     | BellaTek® Encode®<br>Healing Abutment<br>(K072642)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 | Healing Collars<br>(K142082)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   |
|                                         | device remains in<br>the mouth until the<br>soft tissue fully<br>develops. The<br>occlusal surface of<br>the device includes<br>machined markings<br>that provide<br>information about<br>the mating implant's<br>position and<br>orientation.                                                                                                                                                                                                                                                                                                                        | mating implant's position<br>and orientation.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                |
| Technological<br>Characteristics        | Encode® Coding<br>Scheme (Machined<br>Markings)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       | Encode® Coding Scheme<br>(Machined Markings)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       | No Machined Markings                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           |
| Method of<br>Attachment<br>(Mating)     | Attached to the<br>implant by a<br>retaining screw                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    | Attached to the implant<br>by a retaining screw                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    | Threads into implant                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           |
| Emergence<br>Profile                    | 3.8mm and 5.0mm                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       | 3.8mm - 7.5mm                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      | 3.7mm and 4.5mm                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                |
| Height                                  | 3, 4, 6 and 8mm                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       | 3, 4, 6 and 8mm                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    | 1.5, 3.0 and 4.5mm                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             |
| Restorative<br>Platform<br>Diameter     | 2.9mm                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 | 3.4, 4.1, 5.0 and 6.0mm                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            | 2.9mm                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          |
| Connection<br>Geometry                  | Flat-to-flat mating<br>with a passive fit<br>within the implant<br>internal connection<br>to engage Eztetic™<br>3.1mm Dental<br>Implant                                                                                                                                                                                                                                                                                                                                                                                                                               | Flat-to-flat mating with a<br>passive fit within the<br>implant internal<br>connection to engage<br>Biomet 3i External and<br>Internal Hexagon<br>Implants                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         | Flat-to-flat mating with a<br>passive fit within the implant<br>internal connection to engage<br>Eztetic™ 3.1mm Dental<br>Implant                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              |
| Mating Implant                          | Zimmer Dental<br>3.1mm Eztetic™<br>Implant System                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     | Biomet 3i Internal and<br>External Hexagon Implant<br>System                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       | Zimmer Dental 3.1mm<br>Eztetic™ Implant System                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 |
| Packaging                               | Nylon Bag                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             | Nylon Bag                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          | Sealed Tray in an outer<br>chipboard box                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       |
| Materials                               | Titanium Alloy<br>(Ti-6Al-4V ELI)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     | Titanium Alloy<br>(Ti-6Al-4V ELI)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  | Titanium Alloy<br>(Ti-6Al-4V ELI)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              |
| Feature                                 | Subject Device<br>EzteticTM BellaTek®<br>Encode® Healing<br>Abutment                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  | Primary Predicate<br>Device<br>BellaTek® Encode®<br>Healing Abutment<br>(K072642)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  | Reference Device<br>Healing Collars<br>(K142082)…

---

**Source:** [https://fda.innolitics.com/device/K170013](https://fda.innolitics.com/device/K170013)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
