Esophageal Cooling Device

K170009 · Advanced Cooling Therapy, LLC · PLA · Mar 30, 2017 · Cardiovascular

Device Facts

Record IDK170009
Device NameEsophageal Cooling Device
ApplicantAdvanced Cooling Therapy, LLC
Product CodePLA · Cardiovascular
Decision DateMar 30, 2017
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 870.5910
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The Esophageal Cooling Device is a thermal regulating device, intended to: - connect to a Gaymar Medi-Therm III Conductive Hyper/Hypothermia System or Stryker Altrix Precision Temperature Management System to control patient temperature, and - provide gastric decompression and suctioning.

Device Story

Multi-lumen silicone tube inserted into esophagus like standard orogastric tube; connects to external heat exchanger (Gaymar Medi-Therm III or Stryker Altrix) to modulate patient temperature; third central lumen connects to wall suction for gastric decompression/drainage. Single-use, disposable, non-implantable; duration ≤36 hours. Used in clinical settings by healthcare providers. Device enables simultaneous temperature control and gastric access; benefits patient by providing thermal regulation while maintaining standard orogastric functionality.

Clinical Evidence

Bench testing only. Flow rate testing was performed to demonstrate compatibility and thermal transfer capabilities with the Stryker Altrix Precision Temperature Management System.

Technological Characteristics

Multi-lumen silicone tube; single-use; non-implantable. Functions as a heat exchanger interface. Connects to external hyper/hypothermia systems. No electronic components; purely mechanical fluid-based thermal transfer.

Indications for Use

Indicated for patients requiring thermal regulation and gastric decompression/suctioning via an orogastric tube. Used in conjunction with compatible external heat exchangers (Gaymar Medi-Therm III or Stryker Altrix).

Regulatory Classification

Identification

An esophageal thermal regulation device is a prescription device used to apply a specified temperature to the endoluminal surface of the esophagus via an external controller. This device may incorporate a mechanism for gastric decompression and suctioning. The device is used to regulate patient temperature.

Special Controls

In combination with the general controls of the FD&C Act, the Esophageal Cooling Device is subject to the following special controls:

*Classification.* Class II (special controls). The special controls for this device are:(1) The patient contacting materials must be demonstrated to be biocompatible. (2) Non-clinical performance evaluation must demonstrate that the device performs as intended under anticipated conditions of use. The following performance characteristics must be tested: (i) Mechanical integrity testing. (ii) Testing to determine temperature change rate(s). (iii) Testing to demonstrate compatibility with the indicated external controller. (iv) Shelf life testing. (3) Animal testing must demonstrate that the device does not cause esophageal injury and that body temperature remains within appropriate boundaries under anticipated conditions of use. (4) Labeling must include the following: (i) Detailed insertion instructions. (ii) Warning against attaching the device to unintended connections, such as external controllers for which the device is not indicated, or pressurized air outlets instead of vacuum outlets for those devices, including gastric suction. (iii) The operating parameters, name, and model number of the indicated external controller. (iv) The intended duration of use.

Predicate Devices

Reference Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized caduceus symbol, which is a staff with two snakes coiled around it. The caduceus is surrounded by the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" in a circular arrangement. Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002 March 30, 2017 Advanced Cooling Therapy, LLC Erik Kulstad, M.D., M.S. President/CEO 3440 S. Dearborn St. #215-south Chicago, Illinois 60616 Re: K170009 Trade/Device Name: Esophageal Cooling Device Regulation Number: 21 CFR 870.5910 Regulation Name: Esophageal Thermal Regulation Device Regulatory Class: Class II Product Code: PLA Dated: December 30, 2016 Received: January 6, 2017 Dear Dr. Kulstad: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. {1}------------------------------------------------ If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Sincerely, M.A. Hilleman for Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ ## Indications for Use 510(k) Number (if known) K170009 Device Name Esophageal Cooling Device Indications for Use (Describe) The Esophageal Cooling Device is a thermal regulating device, intended to: - · connect to a Gaymar Medi-Therm III Conductive Hyper/Hypothermia System or Stryker Altrix Precision Temperature Management System to control patient temperature, and · provide gastric decompression and suctioning. | Type of Use (Select one or both, as applicable) | | |----------------------------------------------------------------------------------|---------------------------------------------------------------------------------| | <div> <span> Prescription Use (Part 21 CFR 801 Subpart D) </span> </div> | <div> <span> Over-The-Counter Use (21 CFR 801 Subpart C) </span> </div> | #### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ # SECTION 5: 510(k) Summary of Safety and Effectiveness (21 CFR § 807.92) #### Company: Advanced Cooling Therapy, Inc. Address: 3440 S. Dearborn St. #215-South Chicago, IL 60616 Phone: +1-312-725-4756 Erik Kulstad Contact Person: President Date Prepared: December 30, 2016 Device Name & Classification Esophageal Cooling Device (ECD) Trade Name: Model Number(s): ECD01-A Esophageal Thermal Regulation Device (21 CFR 870.5910) Classification Name: Product Code: PLA ll Class: ## Submitter / 510(k) Holder ## Predicate Device 1. Advanced Cooling Therapy, Inc. Esophageal Cooling Device (model #: ECD01-A); DEN140018 ## Device Description The Esophageal Cooling Device is a multi-lumen silicone tube that is placed in the esophagus in a similar manner to a standard orogastric tube to modulate a patient's temperature, while simultaneously maintaining access to the stomach to allow gastric decompression and drainage and maintain the functionality of the standard orogastric tube. Modulation and control of patient temperature is achieved by connecting the Esophageal Cooling Device to an external heat exchanger. Two lumens connect to the exchanger. A third central lumen connects to wall suction and is used for standard gastric decompression. The Esophageal Cooling Device is made of standard medical-grade silicone. It is a single-use, disposable, non-implantable device with an intended duration of use of 36 hours or less. #### Indications for Use The Esophageal Cooling Device is a thermal regulating device, intended to: - connect to a Gaymar Medi-Therm III Conductive Hyper/Hypothermia System or Stryker Altrix Precision Temperature Management System to control patient temperature, and - provide gastric decompression and suctioning. {4}------------------------------------------------ ## Technological Characteristics The indications for use have been expanded to include the Stryker Altrix Precision Temperature Management System as a compatible external heat exchanger. All other technological characteristics (including principles of operation, materials, and design) remain identical. The Stryker Altrix Precision Temperature Management System provides functionality and safety features comparable to the Gaymar Medi-Therm III Conductive Hyper/Hypothermia System. ### Non-Clinical Performance Testing Non-clinical performance testing was conducted to demonstrate the compatibility of the device with the Stryker Altrix Precision Temperature Management System. The following tests were performed: | Test performed | Discussion | |----------------|----------------------------------------------------------------------------------------------------------------------------| | Flow rate | Flow rate testing demonstrated the Esophageal Cooling Device is suitable for maintaining<br>thermal transfer capabilities. | #### Conclusion The results of all performance testing showed the subject device to be substantially equivalent to the predicate device.
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...