← Product Code [PBY](/productcode/PBY) · K163667

# RFID Localization System- Localizer Reader, RFID Localization System- Localizer Surgical Probe, RFID Localization System- Localizer Surgical Probe (5 pack), RFID Localization System- Tag Applicator, RFID Localization System -Tag Applicator (10 pack) (K163667)

_Health Beacons, Inc. · PBY · Apr 28, 2017 · General, Plastic Surgery · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K163667

## Device Facts

- **Applicant:** Health Beacons, Inc.
- **Product Code:** [PBY](/productcode/PBY.md)
- **Decision Date:** Apr 28, 2017
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 878.4300
- **Device Class:** Class 2
- **Review Panel:** General, Plastic Surgery

## Indications for Use

The Tag of the RFLS is intended for percutaneous placement in the breast to temporarily (<30 days) mark a lesion intended for surgical removal. Using image guidance (such as ultrasound or radiography) or aided by non-imaging guidance (RFLS), the RFID Tag is located and surgically removed with the target tissue. The RFLS is intended only for the non-imaging detection and localization of the Tag that has been implanted in a lesion intended for surgical removal.

## Device Story

RFID Localization System (RFLS) uses miniature RFID tags as markers for breast lesion localization. System components include a Tag, Tag Applicator, hand-held Reader, and Surgical Probe. The Tag is placed percutaneously under image guidance. During surgery, the surgeon uses the Reader and Surgical Probe to detect the RFID Tag; the system provides audible tones (increasing pitch/volume with proximity), a visual bar indicator, and a distance measurement readout. This guides the surgeon to the target tissue for excision. Used in clinical settings by breast surgeons and radiologists. Benefits include precise intra-operative localization of non-palpable lesions.

## Clinical Evidence

Bench testing only. No clinical data provided. Testing included system verification, reader/probe design verification, accuracy and repeatability evaluation, tissue marker migration, usability validation, electrosurgical tool compatibility, EMC/safety testing, biocompatibility, and magnetic resonance testing (ASTM F2119-07).

## Technological Characteristics

Materials: Tag (soda lime bioglass, polypropylene homopolymer), Applicator (304 stainless steel), Probe (ABS, 304 stainless steel, TPU, silicone elastomer), Reader (polycarbonate). Sensing: Radiofrequency wave technology. Dimensions: Tag ~2.3mm x 10.6mm. Connectivity: Standalone. Sterilization: Ethylene Oxide (Tag/Applicator), Gamma (Probe). Standards: ISO 10993, IEC 60601-1 (3rd ed), IEC 60601-1-2 (4th ed), FCC 47 CFR Part 15 subpart C, ASTM F2119-07.

## Regulatory Identification

An implantable clip is a clip-like device intended to connect internal tissues to aid healing. It is not absorbable.

## Predicate Devices

- SAVI Scout Reflector and SAVI Scout System ([K161507](/device/K161507.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized image of a caduceus, which is a symbol often associated with healthcare and medicine.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

April 28, 2017

Health Beacons, Inc. Ms. Nancy Confrey Chief Executive Officer 34 Walden Street, #753 Concord, Massachusetts 01742

Re: K163667

Trade/Device Name: RFID Localization System- Localizer Reader, RFID Localization System- Localizer Surgical Probe, RFID Localization System-Localizer Surgical Probe (5 Pack), RFID Localization System- Tag Applicator, RFIDd Localization System -tag Applicator (10 Pack) Regulation Number: 21 CFR 878.4300

Regulation Name: Implantable clip Regulatory Class: Class II Product Code: PBY Dated: April 21, 2017 Received: April 24, 2017

Dear Ms. Confrey:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device

{1}------------------------------------------------

related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely,

# David Krause -S

for Binita S. Ashar, M.D., M.B.A., F.A.C.S. Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

## Indications for Use

Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement below.

510(k) Number (if known)

K163667

Device Name RFID Localization System

Indications for Use (Describe)

The Tag of the RFLS is intended for percutaneous placement in the breast to temporarily (<30 days) mark a lesion intended for surgical removal. Using image guidance (such as ultrasound or radiography) or aided by non-imaging guidance (RFLS), the RFID Tag is located and surgically removed with the target tissue.

The RFLS is intended only for the non-imaging detection and localization of the Tag that has been implanted in a lesion intended for surgical removal.

| Type of Use (Select one or both, as applicable)              |                                                             |
|--------------------------------------------------------------|-------------------------------------------------------------|
| <div> </div>                                                 | <div> </div>                                                |
| <div> </div>                                                 | <div> </div>                                                |
| <div>   Prescription Use (Part 21 CFR 801 Subpart D)  </div> | <div>   Over-The-Counter Use (21 CFR 801 Subpart C)  </div> |

### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

# 510(k) Summary

This summary of 510(k) safety and effectiveness information is submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92.

#### l. SUBMITTER

Health Beacons, Inc. 34 Walden Street #753 Concord, MA 01742 Phone: (978) 287-4635 Fax: (978) 246-6019 Establishment Registration No.: TBD

Contact Person: Nancy Confrey Chief Executive Officer Phone: (978) 287-4635 nconfrey@healthbeacons.com

Alternate Contact: Diana DeGregorio Lincé Consulting, LLC Regulatory Affairs Consultant (925) 980-8047 ddegregorio@linceconsulting.com

Date Prepared December 23, 2016

#### ll. DEVICE

Trade Name: RFID Localization System (RFLS) Common Name: Temporary Tissue Marker Classification Name: Implantable clip Classification: 21 CFR§ 878.4300 Product Code: PBY Class II Device Class:

{4}------------------------------------------------

#### lll. PREDICATE

Cianna Medical, Inc. SAVI Scout Reflector and SAVI Scout System (K161507)

#### IV. DEVICE DESCRIPTION

The RFID Localization System (RFLS) is a marker-with-detector localization device that employs miniature RFID tags as markers and a hand-held reader that can measure distance to the tag. The RFLS is comprised of a Tag, Tag Applicator, Reader, and Surgical Probe. The Tag, when used in conjunction with the Reader and Surgical Probe, can be used as a guide for the surgeon during the excision of tissue. The RFLS is a prescription device meant only for use by trained professionals, specifically breast surgeons and diagnostic radiologists.

#### V. INDICATIONS FOR USE

The Tag of the RFLS is intended for percutaneous placement in the breast to temporarily (<30 days) mark a lesion intended for surgical removal. Using image quidance (such as ultrasound or radiography) or aided by non-imaging quidance (RFLS), the RFID Tag is located and surgically removed with the target tissue.

The RFLS is intended only for the non-imaging detection and localization of the Tag that has been implanted in a lesion intended for surgical removal.

#### VI. COMPARISON CHARACTERISTICS THE OF TECHNOLOGICAL WITH PREDICATE DEVICE

The RFLS has similar features as compared to the predicate device as shown in the following table:

| Manufacturer                                      | Cianna Medical, Inc.                                                                                              | Health Beacons, Inc.                                                    |
|---------------------------------------------------|-------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------|
| Device Name                                       | SAVI Scout Reflector and SAVI<br>Scout System                                                                     | RFID Localization System (RFLS)                                         |
| 510(k) Number                                     | K161507                                                                                                           | TBD                                                                     |
| Anatomical<br>Locations                           | Breast Tissue                                                                                                     | Same                                                                    |
| Primary Device<br>Components                      | Electromagnetic wave reflective<br>device (Reflector) pre-loaded in<br>Delivery System, Handpiece, and<br>Console | RFID Tag pre-loaded in Tag<br>Applicator, Reader, and Surgical<br>Probe |
| Breast Tissue<br>Penetration /<br>Delivery Device | Needle Implanter                                                                                                  | Same                                                                    |
| Visibility                                        | X-ray, Ultrasound                                                                                                 | Same                                                                    |
| Principles of<br>Operation                        | electromagnetic wave technology to<br>detect the tissue marker                                                    | radiofrequency wave technology to<br>detect tissue marker               |
| Localization<br>Accuracy                          | Unknown                                                                                                           | ±7 mm                                                                   |

{5}------------------------------------------------

| Manufacturer                                         | Cianna Medical, Inc.                                                                                                                                                                                                                                                              | Health Beacons, Inc.                                                                                                                                                                                                                                                     |
|------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Device Name                                          | SAVI Scout Reflector and SAVI<br>Scout System                                                                                                                                                                                                                                     | RFID Localization System (RFLS)                                                                                                                                                                                                                                          |
| 510(k) Number                                        | K161507                                                                                                                                                                                                                                                                           | TBD                                                                                                                                                                                                                                                                      |
| Indicators                                           | Visual and Audible                                                                                                                                                                                                                                                                | Same                                                                                                                                                                                                                                                                     |
| Implant<br>Delivery Device<br>Size/Working<br>Length | 16 Ga/130mm                                                                                                                                                                                                                                                                       | 12 Ga/100mm                                                                                                                                                                                                                                                              |
| Implant<br>Delivery Device<br>Needle Marking         | 1 cm marker increments                                                                                                                                                                                                                                                            | Same                                                                                                                                                                                                                                                                     |
| Implant<br>Dimensions                                | Approximately<br>0.05" (1.27 mm) x 0.49" (12.4 mm)<br>long                                                                                                                                                                                                                        | Approximately<br>0.09" (2.3 mm) diameter x 0.37"<br>(10.6 mm) long                                                                                                                                                                                                       |
| Patient Contact<br>Materials                         | Reflector: 3922 Loctite, Nitinol<br>•<br>SE508, light oxide<br>Delivery System: 304 Stainless<br>•<br>Steel<br>Handpiece: Acrylonitrile butadiene<br>•<br>styrene, Polyester film, Loctite<br>4011, NPO (COG) K20 (Dielectric<br>material)<br>• Console: N/A (no patient contact) | Tag: Soda lime bioglass,<br>polypropylene homopolymer<br>• Tag Applicator: 304 Stainless<br>Steel<br>• Surgical Probe: Acrylonitrile<br>butadiene styrene, 304 Stainless<br>•<br>Steel, TPU, Silicone Elastomer<br>Reader: Polycarbonate, Polyester<br>•<br>film (label) |
| Sterilization<br>Method                              | Single-use, sterilized components:<br>• Implanter and reflector (Ethylene<br>Oxide)<br>• Handpiece (Ethylene Oxide)                                                                                                                                                               | Single-use, sterilized components:<br>• Tag Applicator and RFID Tag<br>(Ethylene Oxide)<br>• Surgical Probe (Gamma)                                                                                                                                                      |
| Tissue Marker<br>Locator                             | Reusable, non-sterile Console                                                                                                                                                                                                                                                     | Reusable, non-sterile Reader                                                                                                                                                                                                                                             |
| Conformance<br>to Standards                          | ISO 10993<br>IEC 60601-1<br>IEC 60601-1-2<br>ASTM F2119-07                                                                                                                                                                                                                        | ISO 10993<br>IEC 60601-1 (3rd edition)<br>IEC 60601-1-2 (4th edition)<br>FCC 47 CFR Part 15 subpart C<br>ASTM F2119-07 (2013)                                                                                                                                            |

The technological characteristics and principals of operation of the RFLS is substantially equivalent to the named predicate device. The primary difference between the two systems is the Savi Scout device uses electromagnetic wave technology to detect the tissue marker, while the subject device tissue marker is located using RFID technology (radiofrequency wave). However, this minor difference in technology does not raise new or different questions of safety and effectiveness. The performance testing included in the 510k demonstrates the safety of the system for its intended use and confirms the performance of the system for use in intra-operative surgical guidance. Specifically, the performance testing demonstrates the safe use of the system to facilitate intra-operative guided surgical procedures by detecting the presence of the Tag by providing an audible tone which

{6}------------------------------------------------

increases pitch and volume with decreased distance, a visible bar indicator, and a measurement readout.

#### VII. PERFORMANCE DATA

Performance testing was conducted to evaluate and characterize the performance of the RFLS. The following performance testing was conducted on the RFLS to support a determination of substantial equivalence to the predicate device.

- RFLS System Verification ●
- Reader/Probe Design Verification
- Localizer System and Component Accuracy, Repeatability Evaluation ●
- . Test Media Evaluation
- Reader and Surgical Probe Operating Environment Evaluation ●
- Tissue Marker Migration Evaluation
- Usability Verification and Validation ●
- Electrosurgical tool compatibility
- EMC/Safety Testing
- Biocompatibility ●
- Magnetic Resonance Testing ●
- Packaging Validation ●
- Sterilization Validation

#### VIII. CONCLUSIONS

The RFLS has been carefully compared to the legally marketed predicate device with respect to intended use/indications for use, technological characteristics, anatomical sites, performance, safety characteristics, and labeling. In addition, non-clinical testing was conducted to verify and validate the performance of the device and ensure the RFLS functions as intended and meets design specifications. The comparison and non-clinical performance testing results demonstrate that the device is substantially equivalent to the predicate device for its intended use.

---

**Source:** [https://fda.innolitics.com/device/K163667](https://fda.innolitics.com/device/K163667)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
