K163645 · Ethicon, LLC · OMD · Apr 21, 2017 · General, Plastic Surgery
Device Facts
Record ID
K163645
Device Name
DERMABOND PRINEO Skin Closure System
Applicant
Ethicon, LLC
Product Code
OMD · General, Plastic Surgery
Decision Date
Apr 21, 2017
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4011
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
DERMABOND™ PRINEO™ System is intended for topical application only to hold closed easily approximated skin edges of wounds from surgical incisions, including punctures from minimally invasive surgery, and simple, thoroughly cleansed, trauma-induced lacerations. DERMABOND™ PRINEO™ System should be used in conjunction with, but not in place of, deep dermal stitches. Additionally, the adjunct wound closure device component maintains temporary skin edge alignment along the length of the wound during application of the liquid adhesive.
Device Story
Dermabond Prineo is a sterile, liquid topical skin adhesive (2-octylcyanoacrylate) and adjunct wound closure system. Device consists of a single-use applicator with a crushable glass ampule and two self-adhering meshes. Used in clinical settings by healthcare providers to close surgical incisions up to 40 cm. Mesh is applied first to provide temporary skin edge alignment; liquid adhesive is then applied over the mesh. Adhesive polymerizes within minutes to form a microbial barrier. System acts as a topical skin closure; must be used with deep dermal stitches. Benefits include efficient skin edge approximation and microbial barrier protection for surgical wounds.
Clinical Evidence
Bench testing only. Testing included creep strength, peel strength (mesh only and mesh with adhesive), wound/tissue holding strength, set time, and set temperature. Biocompatibility testing included intracutaneous reactivity, modified ISO skin irritation, and ISO guinea pig maximization sensitization. A 14-day porcine effectiveness study was also conducted.
Technological Characteristics
Liquid 2-octylcyanoacrylate adhesive with D & C Violet No. 2 colorant. Includes two self-adhering meshes for alignment. Sterile, single-use applicator with glass ampule. Materials comply with ISO 10993-1:2009. Designed for incisions up to 40 cm.
Indications for Use
Indicated for topical closure of easily approximated surgical incisions (including minimally invasive surgery punctures) and simple, cleansed, trauma-induced lacerations in patients requiring wound closure in conjunction with deep dermal stitches.
Regulatory Classification
Identification
A tissue adhesive with adjunct wound closure device intended for the topical approximation of skin is a device indicated for topical application only to hold closed easily approximated skin edges of wounds from surgical incisions, including punctures from minimally invasive surgery, and simple, thoroughly cleansed, trauma-induced lacerations. It may be used in conjunction with, but not in place of, deep dermal stitches. Additionally, the adjunct wound closure device component maintains temporary skin edge alignment along the length of wound during application of the liquid adhesive.
Special Controls
*Classification.* Class II (special controls). The special control for this device is FDA's “Guidance for Industry and FDA Staff; Class II Special Controls Guidance Document: Tissue Adhesive with Adjunct Wound Closure Device Intended for the Topical Approximation of Skin.” See § 878.1(e) for the availability of this guidance document.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, with flowing lines connecting them.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
April 21, 2017
Ethicon, LLC Mr. Joice Pappan Regulatory Specialist II Route 22 West, P.O. Box 151 Somerville, New Jersey 08876-0151
Re: K163645 Trade/Device Name: Dermabond Prineo Skin Closure System Regulation Number: 21 CFR 878.4011 Regulation Name: Tissue adhesive with adjunct wound closure device for topical approximation of skin Regulatory Class: Class II Product Code: OMD Dated: January 20, 2017 Received: January 23, 2017
Dear Mr. Pappan:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in
{1}------------------------------------------------
the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely,
# David Krause -S
Binita S. Ashar, M.D., M.B.A., F.A.C.S. for Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K163645
Device Name DERMABOND™ PRINEO™ Skin Closure System
#### Indications for Use (Describe)
DERMABOND™ PRINEO™ System is intended for topical application only to hold closed easily approximated skin edges of wounds from surgical incisions, including punctures from minimally invasive surgery, and simple, thoroughly cleansed, trauma-induced lacerations. DERMABOND™ PRINEO™ System should be used in conjunction with, but not in place of, deep dermal stitches. Additionally, the adjunct wound closure device component maintains temporary skin edge alignment along the length of the wound during application of the liquid adhesive.
Type of Use (Select one or both, as applicable)
| <span> <span style="font-size: 100%;"> ☑ Prescription Use (Part 21 CFR 801 Subpart D) </span> </span> |
|--------------------------------------------------------------------------------------------------------------|
| <span> <span style="font-size: 100%;"> ☐ Over-The-Counter Use (21 CFR 801 Subpart C) </span> </span> |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
510(k) Pre-Market Notification Ethicon DERMABOND™ PRINEO™ Skin Closure System Ethicon, Inc.
Image /page/3/Picture/1 description: The image shows the word "ETHICON" in red, with each letter in a bold, serif font. The letters are evenly spaced and the word is centered. The color of the text is a vibrant red, which contrasts with the white background.
PART OF THE Johnson&Johnson FAMILY OF COMPANIES
# 510(k) Summar
- Submitter: Ethicon Inc. a Johnson & Johnson company P.O. Box 151 Route 22 West Somerville, NJ 08876-0151
- Joice Pappan Contact Person: Specialist II, Regulatory Affairs Ethicon, Inc. a Johnson & Johnson company Ph: (908) 218-2113 Fax: (908) 218-2595 E-mail: Jpappan@its.jnj.com
April 19, 2017 Date Prepared:
Device Trade Name: DERMABOND™ PRINEO™ Skin Closure System
Device Common Name: Topical Skin Adhesive
Class: II
Classification Name: Tissue adhesive with adjunct wound closure device intended for the topical approximation of skin (21 CFR 878.4011)
Product Code: OMD
{4}------------------------------------------------
510(k) Pre-Market Notification Ethicon DERMABOND™ PRINEO™ Skin Closure System Ethicon. Inc.
#### Predicate Devices:
| Device | Company | Product<br>Code | 510(k)<br>Number | Predicate for: |
|----------------------------------------------|---------------|-----------------|------------------|------------------------------------------------------------------------------------------------------------------------|
| DERMABOND™<br>PRINEO™ Skin Closure<br>System | Ethicon, Inc. | OMD | K133864 | Fundamental Scientific<br>Technology, Design, Intended Use,<br>Materials, Construction,<br>Performance Characteristics |
#### Device Description:
DERMABOND™ PRINEO™ Skin Closure System is a sterile, liquid topical skin adhesive containing a monomeric (2-octylcyanoacrylate) formulation and the colorant D & C Violet No. 2. It is provided in a single-use applicator packaged in a rigid blister. The applicator is composed of a crushable glass ampule contained within a pen applicator with an attached applicator tip. As applied to skin, the liquid adhesive is slightly more viscous than water and polymerizes within minutes. In vitro studies have shown that DERMABOND™ PRINEO™ System acts as a barrier to microbial penetration as long as the adhesive film remains intact. Clinical studies were not conducted to demonstrate microbial barrier properties.
DERMABOND™ PRINEOTM System also incorporates 2 self-adhering meshes that are applied to the approximated skin edges to provide temporary skin edge alignment of incisions up to 40 cm in length until the liquid adhesive is applied to achieve skin closure.
#### Indications for Use:
DERMABOND™ PRINEOTM System is intended for topical application only to hold closed easily approximated skin edges of wounds from surgical incisions, including punctures from minimally invasive surgery, and simple, thoroughly cleansed, trauma-induced lacerations. DERMABOND™ PRINEO™ System should be used in conjunction with, but not in place of, deep dermal stitches. Additionally, the adjunct wound closure device component maintains temporary skin edge alignment along the length of the wound during application of the liquid adhesive.
#### Summary of Technological Characteristics and Performance Testing:
The safety and effectiveness of the subject device DERMABOND™ PRINEO™ Skin Closure System (DP42) and the substantial equivalence to the predicate device DERMABOND™ PRINEO™ Skin Closure System (DP22) has been demonstrated via data collected in non-clinical
{5}------------------------------------------------
design verification. The results of these tests provide reasonable assurance that the proposed device has been designed and tested to assure conformance to the requirements for its intended use. No new safety or performance issues were raised during the testing. All materials used in the proposed device are the same to the predicate device and meet the requirements of ISO 10993-1:2009 Biological evaluation of medical devices - Part 1: Evaluation and Testing within a risk management process.
The following tests were completed for the DERMABOND™ PRINEO™ Skin Closure System (42cm):
| Biocompatibility/In vivoTesting | Bench Testing |
|---------------------------------------------|--------------------------------------------------|
| Irritation (Intracutaneous Reactivity) | Creep Strength |
| Irritation (Modified ISO Skin) | Peel Strength- mesh only (N/cm) |
| Sensitization (ISO Guinea Pig Maximization) | Peel adhesion Strength-Mesh with adhesive (N/cm) |
| 14 day Porcine Effectiveness Study | Wound/Tissue holding Strength (Ibf) |
| | Set Time (Seconds) |
| | Set Temperature (°C) |
#### Summary of Substantial Equivalence Comparison:
The subject DERMABOND™ PRINEO™ Skin Closure System (42 cm) is equivalent to the predicate DERMABOND™ PRINEO™ Skin Closure System (22 cm) described in K133864 with the exception of number of mesh patches. The subject device is intended to hold closed incisions up to 40 cm in length and the predicate device is for incisions up to 20 cm in length. The subject device has the same fundamental scientific technology and intended use as the current, legally marketed DERMABOND™ PRINEO™ Skin Closure System (22cm). The subject and predicate devices share the same materials, design, fundamental scientific technology (operating principle), labeling components, packaging materials and configuration and sterilization process. The additional size meets the same requirements as the current FDA cleared K133864 device.
### Conclusion:
Based on the intended use, technological characteristics, safety and performance testing, the additional size of DERMABOND™ PRINEO™ Skin Closure System (42cm) has been shown to be appropriate for its intended use and is considered to be substantially equivalent to the DERMABOND™ PRINEOTM Skin Closure System (22cm) (K133864).
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.