K163639 · Lsi Solutions · NBY · Jun 2, 2017 · General, Plastic Surgery
Device Facts
Record ID
K163639
Device Name
LS-5 ePTFE Suture
Applicant
Lsi Solutions
Product Code
NBY · General, Plastic Surgery
Decision Date
Jun 2, 2017
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.5035
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The LS-5™ ePTFE Suture is indicated for use in all types of soft tissue approximation, including use in cardiovascular surgery.
Device Story
LS-5 ePTFE Suture is a non-absorbable, monofilament expanded polytetrafluoroethylene (ePTFE) surgical suture. Device features modified surgical stainless steel ferrules attached to both ends to facilitate loading and interfacing with LSI SOLUTIONS Suture Placement Devices. Includes detachable clear tube to prevent tangling. Used by surgeons in clinical settings for soft tissue approximation. Ferrules are removed after suturing. Device is MR safe, undyed, and contains no additives. Provides reliable suture loading for automated suturing technology; benefits patient through secure tissue approximation.
Clinical Evidence
No clinical studies were performed. Substantial equivalence was demonstrated through bench testing, including USP monograph testing for tensile strength and needle attachment, biocompatibility evaluation (ISO 10993-1), bacterial endotoxin testing, and stability testing.
Technological Characteristics
Non-absorbable monofilament ePTFE suture. Features stainless steel ferrules for device interface. Dimensions correlate to USP sizes 2-0 to 4-0. Sterilized for single-patient use. MR safe. No additives.
Indications for Use
Indicated for soft tissue approximation, including cardiovascular surgery. Contraindicated for ophthalmic surgery, microsurgery, and neural tissue.
Regulatory Classification
Identification
Nonabsorbable expanded polytetrafluoroethylene (ePTFE) surgical suture is a monofilament, nonabsorbable, sterile, flexible thread prepared from ePTFE and is intended for use in soft tissue approximation and ligation, including cardiovascular surgery. It may be undyed or dyed with an approved color additive and may be provided with or without an attached needle(s).
Special Controls
*Classification.* Class II (special controls). The special control for this device is FDA's “Class II Special Controls Guidance Document: Surgical Sutures; Guidance for Industry and FDA.” See § 878.1(e) for the availability of this guidance document.
{1}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo is circular and contains the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. In the center of the logo is an abstract symbol that resembles an eagle or a stylized human figure.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
June 2, 2017
LSI Solutions % Mr. Kevin Bentley Executive Director of Regulatory Affairs and Quality 7796 Victor-Mendon Rd Victor, New York 14564
Re: K163639
Trade/Device Name: LS-5 ePTFE Suture Regulation Number: 21 CFR 878.5035 Regulation Name: Nonabsorbable Expanded Polytetrafluoroethylene Surgical Suture Regulatory Class: Class II Product Code: NBY Dated: May 5, 2017 Received: May 9, 2017
Dear Mr. Bentley:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{2}------------------------------------------------
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely, Jennifer R. Stevenson -ਟਤ For Binita S. Ashar, M.D., M.B.A., F.A.C.S. Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
# Indications for Use
510(k) Number (if known)
K163639
Device Name LS-5TM ePTFE Suture
Indications for Use (Describe)
The LS-5™ ePTFE Suture is indicated for use in all types of soft tissue approximation, including use in cardiovascular surgery.
Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
Image /page/3/Picture/0 description: The image features the logo for LSASolutions, a company that provides innovative solutions for minimally invasive surgery. The logo is displayed in white against a blue background. The text "INNOVATIVE SOLUTIONS FOR MINIMALLY INVASIVE SURGERY" is written in a smaller font size below the company name.
#### 510(k) SUMMARY
| Submitted By: | LSI SOLUTIONS®, Inc. |
|---------------|-------------------------|
| | 7796 Victor-Mendon Road |
| | Victor, NY 14564 |
Contact Person: Jennifer Landry Regulatory Specialist Phone: (585) 760-6429 (585) 742-1039 Fax: Email: jlandry@lsisolutions.com
Date Prepared: May 17, 2017
LS-5TM ePTFE Suture Trade Name:
Common Name: Nonabsorbable Surgical Suture
Nonabsorbable Expanded Polytetrafluoroethylene Surgical Suture Classification Name: (21 CFR 878.5035, Product Code NBY)
Device Classification: Class II
## Predicate Device:
The predicate device, W. L. Gore and Associates Inc. GORE-TEX™ Suture ePTFE Nonabsorbable Monofilament (hereinafter known as GORE-TEX® Suture), was selected to
{5}------------------------------------------------
demonstrate substantial equivalence to the LS-5™ ePTFE Suture. GORE-TEX® Suture was initially approved under Premarket Approval Number P820083 on December 30, 1985 and was reclassified to Class II on September 9, 1999 (reference 21 CFR Part 878, Docket No. 94P-0347). The LS-5™ ePTFE Suture is designed and manufactured to be comparable to CV-5 GORE-TEX® Suture, which is designated to approximately correlate with a surgical suture diameter of USP size 2-0. The LS-5™ ePTFE Suture is also similar to CV-5 GORE-TEX® Suture for tensile strength and needle attachment. W. L. Gore and Associates Inc. promotes the CV-5 GORE-TEX® Suture as being comparable to USP size 4-0 monofilament polypropylene surgical suture.
LSI SOLUTIONS® Suture Placement Devices and Accessories, which was cleared under 510(k) Premarket Notification K100593 on September 24, 2010, and LSI SOLUTIONS® Suture QUICK LOAD® Products, which was cleared under 510(k) Premarket Notification K031443 on June 18, 2003 are used as reference predicates representative of LSI SOLUTIONS® Suture Placement Devices and the technological differences between the predicate and subject device.
### Device Description
Each LSI SOLUTIONS® LS-5™ ePTFE sterile surgical suture with 2 ferrules is held in a customized tray, with a suture release feature and is intended for single patient use. The LS-5™ ePTFE Suture is designed to enable the rapid, easy and reliable loading of suture into compatible LSI SOLUTIONS® Suture Placement Devices. LS-5™ ePTFE Suture is a non-absorbable, monofilament expanded polytetrafluoroethylene (ePTFE) suture. A short length of modified surgical stainless steel tubing, called a ferrule, is attached to each end of the suture for interfacing with LSI SOLUTIONS® Suture Placement Devices. The LS-5™ ePTFE Suture also includes a detachable clear suture tube to keep the suture from tangling. The LS-5™ ePTFE Suture is undyed and contains no additives. No significant loss in tensile strength retention occurs in vivo. The LS-5TM ePTFE Suture is MR safe.
### Intended Use
The LS-5™ ePTFE Suture is indicated for use in all types of soft tissue approximation, including use in cardiovascular surgery.
{6}------------------------------------------------
The predicate device, GORE-TEX® Suture, is indicated for use in all types of soft tissue approximation, including use in cardiovascular surgery, but not for use in ophthalmic surgery, microsurgery, and neural tissue. It is recommended for use where reduced suture line bleeding during cardiovascular anastomotic procedures is desired.
According to the FDA Guidance for Industry. Class II Special Controls Guidance Document: Surgical Sutures; Guidance for Industry and FDA, any surgery types, body sites, or patient populations where evidence demonstrates that the suture should not be used should be included in the contraindications of the labeling. The LS-5™ ePTFE Suture will include a contraindication that the subject device is not for use in ophthalmic surgery, microsurgery, and neural tissue. Therefore, the difference in indications do not impact the safety and effectiveness of the LS-5™ ePTFE Suture. Both the subject and predicate device are intended to be used in the approximation of all types of soft tissue approximation, including use in cardiovascular surgery, and are contraindicated for use in ophthalmic surgery, microsurgery, and neural tissue.
### Technological Characteristics (comparison to Predicate Device)
The LS-5™ ePTFE Suture is substantially equivalent in intended use and fundamental technological characteristics to the legally marketed predicate device. Both the subject and predicate devices are non-absorbable, monofilament suture manufactured from ePTFE. The surgical suture of both the subject and predicate devices is undyed and contains no additives. The LS-5™ ePTFE Suture will be offered swaged on LSI's proprietary stainless steel ferrules for use with LSI's automated suturing technology. The ferrule attachments are the only technological difference between the LS-5™ ePTFE Suture and the predicate device. The ferrules are attached to both ends of the suture in the same manner as standard needle attachments and facilitate the use of the suture with the reference predicate, LSI SOLUTIONS® Suture Placement Devices. The ferrules have identical material characteristics as the ferrules used in the reference device, specifically LSI Non-Absorbable QUICK LOAD® Suture. Similar to standard needle attachments, once suturing is complete, the ferrules are removed from the suture.
The subject device is packaged and sterilized in an equivalent manner and has equivalent labeling claims to the predicate and reference predicate devices, including indications, contraindications,
{7}------------------------------------------------
warnings, cautions and precautions. The subject device is considered substantially equivalent and has the same technological characteristics to its predicate device through comparison in design, intended use, material composition, function, and range of sizes.
## Performance Testing Summary
As recommended by the FDA Guidance for Industry, Class II Special Controls Guidance Document: Surgical Sutures; Guidance for Industry and FDA, the LS-5™ ePTFE Suture was subject to the requirements of the United States Pharmacopeia (USP) monograph for Nonabsorbable Sutures. Performance testing includes:
- Sutures Tensile Strength <881> .
- Ferrule Attachment which complies with the requirements for standard needle attachments . as defined in Sutures – Needle Attachment <871>.
The LS-5™ ePTFE Suture is offered with a surgical suture diameter comparable to the predicate device, CV-5 GORE-TEX® Suture, which is designated to approximately correlate with a surgical suture diameter of USP size 2-0. Both the LS-5IM ePTFE Suture and predicate device are manufactured from ePTFE. The internal microporous airspaces of ePTFE that result from expanding pure polytetrafluoroethylene (PTFE) produce a less dense structure that slightly exceed USP requirements for Nonabsorbable Sutures Section <861> Sutures - Diameter. LSI has performed testing to demonstrate the LS-5™ ePTFE Suture is comparable to CV-5 GORE-TEX® Suture for surgical suture diameter.
The LS-5™ ePTFE Suture is also similar to CV-5 GORE-TEX® Suture for tensile strength and needle attachment, which is promoted as being comparable to USP size 4-0 surgical suture. LSI has performed testing to demonstrate the LS-5™ ePTFE Suture is comparable to the predicate device for tensile strength and needle attachment and complies with USP requirements for Nonabsorbable Sutures Section <881> Sutures - Tensile Strength and Section <871> Sutures -Needle Attachment for 4-0 suture. In accordance with the FDA Guidance for Industry, Class II Special Controls Guidance Document: Surgical Sutures; Guidance for Industry and FDA, the labeling of the LS-5™ ePTFE Suture will include a tabular comparison of its diameter and tensile
{8}------------------------------------------------
strength to the predicate device and a tabular comparison of the USP requirements for USP size 2-0, 3-0 and 4-0.
Biocompatibility evaluation for the LS-5™ ePTFE Suture was conducted in accordance with the FDA Guidance for Industry, Use of International Standard ISO 10993-1:2009, "Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process," to demonstrate that the LS-5™ ePTFE Suture is biocompatible and substantially equivalent to the predicate device. The LS-5™ ePTFE Suture is classified as an implant device, contacting blood, for a permanent duration, which is the identical type and duration of patient contact as the predicate device. LSI has leveraged the following biocompatibility testing from the manufacturer of the ePTFE suture material:
- USP Systemic Toxicity Study .
- USP Intracutaneous Toxicity Study .
- USP Muscle Implantation Study .
Additionally, LSI has conducted the following biocompatibility testing:
- Cytotoxicity .
- . Sensitization
- . Irritation
- Rabbit Pyrogen .
- Hemocompatibility .
The ePTFE suture material used in the LS-5™ ePTFE Suture is proven to be remarkably inert and the same polymer that the legally marketed predicate device is made of. A combination of testing and risk assessment was completed to ensure the biocompatibility of the LS-5™ ePTFE Suture.
LSI has also performed testing to ensure that the LS-5™ ePTFE Suture is compatible with the LSI SOLUTIONS® Suture Placement Devices, Bacterial Endotoxins Testing (BET) to confirm that the subject device does not contain gram-negative endotoxins and stability testing to support the labeled 2 year shelf life.
{9}------------------------------------------------
### Clinical Testing
The technological characteristics, indications for use, material, manufacturing and sterilization processes are the same as the predicate and reference predicate devices, therefore, no clinical studies were deemed necessary to demonstrate the safety and effectiveness of the subject device.
### Substantial Equivalence
Both the GORE-TEX® Suture and the LS-5™ ePTFE Suture are non-absorbable, monofilament ePTFE surgical suture that are indicated for the use in all types of soft tissue approximation, including use in cardiovascular surgery. The differences between the subject and predicate device introduce no new risks and have no negative impact to the safety and effectiveness of the surgical suture. The LS-5™ ePTFE Suture is substantially equivalent to the currently marketed W. L. Gore and Associates, Inc. GORE-TEX® Suture which was approved under Premarket Approval Number P820083 on December 30, 1985 and was reclassified to Class II on September 9, 1999 (reference 21 CFR Part 878, Docket No. 94P-0347).
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.