Quantra

K163623 · Hologic, Inc. · LLZ · Oct 20, 2017 · Radiology

Device Facts

Record IDK163623
Device NameQuantra
ApplicantHologic, Inc.
Product CodeLLZ · Radiology
Decision DateOct 20, 2017
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 892.2050
Device ClassClass 2
AttributesAI/ML, Software as a Medical Device

Intended Use

The Quantra™ software application is intended for use with mammographic images acquired using digital breast x-ray systems. The Quantra software segregates breast density into categories, which may be useful in the reporting of consistent BI-RADS® breast composition categories as mandated by certain state regulations. The Quantra software reports a result for each subject, which is intended to aid radiologists in the assessment of breast tissue composition. The Quantra software produces adjunctive information; it is not a diagnostic aid.

Device Story

Quantra software analyzes digital mammographic images to estimate breast composition; inputs include distribution and texture of parenchymal tissue patterns. Device transforms these inputs to categorize breast density into BI-RADS® categories (a, b, c, d). Used in clinical settings by radiologists; output provides adjunctive information to assist in consistent reporting of breast composition. Device does not provide diagnostic aid; output affects clinical workflow by standardizing density assessment reporting.

Clinical Evidence

Bench testing only. Retrospective study of 230 screening mammography studies compared Quantra v2.2 output against 5-radiologist consensus BI-RADS 5th Edition categories. For 2D imaging, accuracy was 92.9% for fatty (a+b) and 99.1% for dense (c+d) categories. For Tomo (3D) imaging, accuracy was 92.0% for fatty and 94.0% for dense categories. No clinical or animal studies performed.

Technological Characteristics

Software-based image processing application. Operates on digital breast x-ray images. Aligned with ACR BI-RADS® 5th Edition. Complies with IEC 62304 (software life cycle) and ISO 14971 (risk management). Moderate level of concern.

Indications for Use

Indicated for use with digital mammographic images to aid radiologists in assessing breast tissue composition by segregating breast density into categories consistent with BI-RADS® 5th Edition standards.

Regulatory Classification

Identification

A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.

Special Controls

*Classification.* Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue. October 20, 2017 Hologic, Inc. Neil Kelly Principal Regulatory Affairs Specialist 36 Apple Ridge Road DANBURY, CT 06810 Re: K163623 Trade/Device Name: Quantra™ Regulation Number: 21 CFR 892.2050 Regulation Name: Picture archiving and communications system Regulatory Class: Class II Product Code: LLZ Dated: December 20, 2016 Received: December 22, 2016 Dear Neil Kelly: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. {1}------------------------------------------------ If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education (DICE) at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevicesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education (DICE) at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Sincerely. Robert Ocks Robert Ochs, Ph.D. Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ # Indications for Use 510(k) Number (if known) K163623 Device Name Quantra™M #### Indications for Use (Describe) The Quantra™ software application is intended for use with mammographic images acquired using digital breast x-ray systems. The Quantra software segregates breast density into categories, which may be useful in the reporting of consistent BI-RADS® breast composition categories as mandated by certain state regulations. The Quantra software reports a result for each subject, which is intended to aid radiologists in the assessment of breast tissue composition. The Quantra software produces adjunctive information; it is not a diagnostic aid. | Type of Use (Select one or both, as applicable) | | |-----------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------| | <div><span style="font-size:16px">☒</span> Prescription Use (Part 21 CFR 801 Subpart D)</div> | <div><span style="font-size:16px">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C)</div> | #### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ # Traditional 510(k) Summary This 510(k) Summary is submitted in accordance with the requirements of 21 CFR Part 807.92 | Date Prepared: | October 05, 2017 | |----------------------|--------------------------------------------------------------------| | Manufacturer: | Hologic, Inc.<br>36 Apple Ridge Road<br>Danbury, CT 06810 USA | | Est. Registration #: | 1220984 | | Contact Person: | Neil Kelly<br>Manager, Regulatory Affairs<br>Phone: (508) 263-8634 | #### Identification of the Device: | Proprietary/Trade Name: | Quantra™ | |-------------------------|-------------------------------------| | Software Version: | 2.2 | | Classification Name: | System, Image Processing, Radiology | | Regulatory Number: | 21 CFR 892.2050 | | Product Code: | LLZ | | Device Class: | Class II | | Review Panel: | Radiology | #### Identification of the Legally Marketed Predicate Device: | Trade name: | Quantra™ | |--------------------------|-------------------------------------| | Software Version: | 2.1 | | Classification Name: | System, Image Processing, Radiology | | Regulatory Number: | 21 CFR 892.2050 | | Product Code: | LLZ | | Device Class: | Class II | | Review Panel: | Radiology | | Submitter/510(k) Holder: | Hologic, Inc. | | Clearance: | K142037 (cleared October 10, 2014) | ## Device Description: Quantra is a software application used to produce assessments of breast composition and categorize them. A breast consists of fat and fibroglandular tissue. Fibroglandular tissue, also referred to as dense tissue, contains a mixture of fibrous connective tissue (stroma) and glandular tissue (epithelial cells), and usually appears brighter than surrounding tissue on a digital {4}------------------------------------------------ mammographic image. Abnormal lesions also appear bright on a mammogram and can be obscured or masked by fibroglandular tissue. The Quantra software is designed to estimate breast composition categories by analyzing distribution and texture of parenchymal tissue patterns which can be responsible for the masking effect during mammographic reading. ### Indications for Use: The Quantra™ software application is intended for use with mammographic images acquired using digital breast x-ray systems. The Quantra software segregates breast density into categories, which may be useful in the reporting of consistent BI-RADS® breast composition categories as mandated by certain state regulations. The Quantra software reports a result for each subject, which is intended to aid radiologists in the assessment of breast tissue composition. The Quantra software produces adjunctive information; it is not a diagnostic aid. #### Standards: IEC 62304:2006 - Medical Device Software - Software Life Cycle Process ISO 14971:2012 – Medical Devices – Application of Risk Management to Medical Devices #### Comparison with Predicate Device: The focus of this submission is to introduce improvements in the proposed Quantra software algorithm to align with ACR BI-RADS® Atlas (Fifth Edition, 2013), and to enhance the accuracy of breast composition estimation. To the end user the proposed Ouantra software is still used in the exact same way as the predicate Quantra software (K142037). The differences are inherent to the algorithm and reported calculations. #### Substantial Equivalence: | | Quantra<br>SW Version 2.1<br>Predicate (K142037) | Quantra<br>SW Version 2.2<br>(Proposed) | Comparison | |--------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------| | Indications<br>for Use | Quantra™ is a software<br>application intended for use with<br>images acquired using digital<br>breast x-ray systems. Quantra<br>calculates volumetric breast<br>density as a ratio of<br>fibroglandular tissue and total<br>breast volume estimates.<br>Quantra also provides area<br>breast density as a ratio of | The Quantra™ software<br>application is intended for use<br>with mammographic images<br>acquired using digital breast x-<br>ray systems. The Quantra<br>software segregates breast<br>density into categories, which<br>may be useful in the reporting of<br>consistent BI-RADS® breast<br>composition categories as | Similar | | | fibroglandular tissue area and<br>total breast area estimates.<br>Quantra segregates breast<br>density into categories, which<br>may be useful in the reporting of<br>consistent BI-RADS® breast<br>composition categories as<br>mandated by certain state<br>regulations. The Quantra results<br>for each image, breast, and<br>subject, are intended to aid<br>radiologists in the assessment of<br>breast tissue composition.<br>Quantra produces adjunctive<br>information; it is not an<br>interpretive or diagnostic aid. | mandated by certain state<br>regulations. The Quantra<br>software reports a result for each<br>subject, which is intended to aid<br>radiologists in the assessment of<br>breast tissue composition. The<br>Quantra software produces<br>adjunctive information; it is not<br>a diagnostic aid. | | | Level of<br>Concern | Moderate | Moderate | Same | | Input<br>Modality | Digital x-ray breast images | Digital x-ray breast images | Same | | Software<br>Product | Yes | Yes | Same | | Reportable<br>Information | Fibroglandular (dense) Tissue Volume Total Breast Volume Volumetric Breast Density Area Breast Density Volume of Fibroglandular Tissue Score Volumetric Breast Density Score Estimate of Breast Density (integer) Estimate of Breast Density (fractional) | Quantra Generated Breast Density Category | Similar. Uses<br>distribution of<br>glandular<br>tissue rather<br>than density of<br>glandular<br>tissue when<br>assessing<br>breast<br>composition. | | Statistical<br>comparison<br>to a<br>reference<br>population | Yes | Yes | Same, with the<br>proposed<br>device also<br>being<br>compared<br>against a | | | | | dataset scored to BI-RADS 5th edition. | | BI-RADS® Atlas | Yes, 4th Edition 2003 | Yes, 5th Edition 2013 | Similar, with enhancements | | Brest Composition Estimates | Yes, displayed as 1, 2, 3 and 4 | Yes, displayed as a, b, c, d | Similar, with enhanced accuracy | {5}------------------------------------------------ {6}------------------------------------------------ ## Summary of Testing: Non-clinical (Bench) testing has been performed using retrospectively collected images of 4view screening negative cases. The proposed Quantra software (Version 2.2) was tested against an independent data set of 230 studies, for which the 5 radiologists' consensus assessment of BIRADS 5th Edition categories are known. ### Quantra 2.2 2D Fatty vs. Dense | | Quantra 2.2- QDC 2D | | | | | |---------------------|---------------------|----------------|----------------|-----------------------|----------| | | | Fatty<br>(a+b) | Dense<br>(c+d) | Per category<br>total | Accuracy | | BI-RADS- 5th<br>Ed. | Fatty<br>(a+b) | 105 | 8 | 113 | 92.9% | | | Dense<br>(c+d) | 1 | 116 | 117 | 99.1% | | | | | Total | 230 | | # Quantra 2.2 Tomo Fatty vs. Dense | | Quantra2.2 – QDC Tomo (3D Center Projection) | | | | | |---------------------|----------------------------------------------|----------------|-----------------------|----------|-------| | | Fatty<br>(a+b) | Dense<br>(c+d) | Per category<br>total | Accuracy | | | BI-RADS- 5th<br>Ed. | Fatty<br>(a+b) | 104 | 9 | 113 | 92.0% | | | Dense<br>(c+d) | 7 | 110 | 117 | 94.0% | | | | Total | 230 | | | No animal studies or clinical studies were performed. ## Conclusion: Based on the information submitted in this premarket notification, and based on the indications for use, technological characteristics and performance testing, the proposed Quantra software (version 2.2) is substantially equivalent to the predicate Quantra software (version 2.1).
Innolitics
510(k) Summary
Decision Summary
Classification Order
Enter a record ID and click Load to view the document.
100%