Trovita Health Science Safe-T Feed Ready-to-Feed Pouch with ENFit Connector
K163586 · Trovrx, Inc. Dba Trovita Health Science · PIF · Mar 17, 2017 · Gastroenterology, Urology
Device Facts
Record ID
K163586
Device Name
Trovita Health Science Safe-T Feed Ready-to-Feed Pouch with ENFit Connector
Applicant
Trovrx, Inc. Dba Trovita Health Science
Product Code
PIF · Gastroenterology, Urology
Decision Date
Mar 17, 2017
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 876.5980
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Trovita Health Science Safe-T Feed Pouch with ENFit® Connector is intended to deliver liquid nutrition formula to an enteral access device (a feeding tube).
Device Story
Device is a pre-filled foil pouch for liquid nutrition delivery; features integrated female ENFit® connector and silicone valve to prevent leakage. Pouch includes hanging ring for gravity feeding or can be squeezed for bolus feeding. Used in clinical or home settings by patients or caregivers. Eliminates need to transfer formula to intermediate syringes or bags, reducing spill and contamination risks. Valve remains closed until breached by male ENFit® connector of feeding tube. Benefits include simplified administration, reduced contamination risk, and improved safety via standardized ENFit® connection.
Clinical Evidence
No clinical data. Evidence based on bench testing: risk analysis (FMEA), ENFit® dimensional analysis, flow rate analysis, valve performance, ISO 80369-3:2016 mechanical testing (leakage, stress cracking, separation resistance), biocompatibility (ISO 10993-1:2009), human factors/usability validation, and packaging/shipping testing.
Technological Characteristics
Foil pouch with integrated female ENFit® connector and silicone valve. Single-patient use, disposable. Non-sterile. Connectivity: ENFit® (ISO 80369-3:2016). Materials: Silicone valve. No software or energy source.
Indications for Use
Indicated for patients requiring enteral nutrition delivery via a feeding tube.
Regulatory Classification
Identification
A gastrointestinal tube and accessories is a device that consists of flexible or semi-rigid tubing used for instilling fluids into, withdrawing fluids from, splinting, or suppressing bleeding of the alimentary tract. This device may incorporate an integral inflatable balloon for retention or hemostasis. This generic type of device includes the hemostatic bag, irrigation and aspiration catheter (gastric, colonic, etc.), rectal catheter, sterile infant gavage set, gastrointestinal string and tubes to locate internal bleeding, double lumen tube for intestinal decompression or intubation, feeding tube, gastroenterostomy tube, Levine tube, nasogastric tube, single lumen tube with mercury weight balloon for intestinal intubation or decompression, and gastro-urological irrigation tray (for gastrological use).
Special Controls
*Classification.* (1) Class II (special controls). The barium enema retention catheter and tip with or without a bag that is a gastrointestinal tube and accessory or a gastronomy tube holder accessory is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I (general controls) for the dissolvable nasogastric feed tube guide for the nasogastric tube. The class I device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 876.9.
NeoMed NeoConnect™ Enteral Syringes with ENFit® Connector (K152857)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized caduceus symbol, which is a staff with two snakes entwined around it. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circular pattern around the symbol.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
March 17, 2017
TrovRx, Inc. dba Trovita Health Science % Julie Bodmer Regulatory Consultant Libra Medical, Inc. 8401 73rd Ave North, Suite 63 Brooklyn Park, Minnesota 55428
Re: K163586
> Trade/Device Name: Trovita Health Science Safe-T Feed™ Ready-to-Feed Pouch with ENFit® Connector Regulation Number: 21 CFR 876.5980 Regulation Name: Gastrointestinal tube and accessories Regulatory Class: Class II Product Code: PIF Dated: December 29, 2016 Received: December 30, 2016
Dear Julie Bodmer:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must
{1}------------------------------------------------
comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely,
Joyce M. Whang -S
for
Benjamin R. Fisher, Ph.D. Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known)
K163586
# Device Name
Trovita Health Science Safe-T Feed™ Ready-to-Feed Pouch with ENFit® Connector
# Indications for Use (Describe)
The Trovita Health Science Safe-T Feed Pouch with ENFit® Connector is intended to deliver liquid nutrition formula to an enteral access device (a feeding tube).
| Type of Use (Select one or both, as applicable) |
|----------------------------------------------------------------------------------------------------------------------------------|
| <div> <span> <span style="font-size: 16px;"><b> × </b> </span> Prescription Use (Part 21 CFR 801 Subpart D) </span> </div> |
| <div> <span> <span style="font-size: 16px;"> □ </span> Over-The-Counter Use (21 CFR 801 Subpart C) </span> </div> |
# CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
# *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for Trovita Health Science. The word "TROVITA" is in large, bold, blue letters. Below that, in smaller letters, is the phrase "HEALTH SCIENCE". Below that is an orange rectangle with the words "REDEFINING NUTRITION" in white letters. To the right of the word "TROVITA" is a circular design made up of orange dots.
# 510(K) SUMMARY 6.0
#### 6.1 ADMINISTRATIVE INFORMATION
Date of Summary Preparation: 16 March 2016
#### 6.1.1 CONTACT INFORMATION
| Submitter/Manufacturer: | TrovRx, Inc. dba Trovita Health Science<br>230 South Broad Street<br>17th Floor<br>Philadelphia, PA 19102<br>Tel: 908-938-7534<br>Email: info@trovitahealth.com |
|-------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Primary Submission Contact: | Julie Bodmer<br>Regulatory Consultant, Libra Medical Inc.<br>8401 73rd Ave North, Suite 63<br>Minneapolis, MN 55428<br>Phone: 612-910-3412<br>Fax: 763-477-6357<br>Email: jbodmer@libramed.com |
| Secondary Submission Contact: | Sara Petrie<br>Regulatory Consultant, Libra Medical Inc. |
8401 73rd Ave North, Suite 63 Minneapolis, MN 55428 Phone: 612-396-9849 Fax: 763-477-6357 Email: spetrie@libramed.com
#### 6.1.2 DEVICE INFORMATION
| Trade Name: | Trovita Health Science Safe-T Feed™ Ready-to-Feed Pouch<br>with ENFit® Connector |
|-----------------------------------------------------|----------------------------------------------------------------------------------|
| Common Name: | Gastrointestinal Tubes with Enteral Specific Connectors |
| Classification Name: | Gastrointestinal tubes and accessories |
| Classification Regulation: | 21 CFR 876.5980 |
| Class: | II |
| Panel: | Gastroenterology and Urology |
| Product Code: | PIF |
| Prior FDA Submissions<br>Related to Subject Device: | None |
{4}------------------------------------------------
Image /page/4/Picture/1 description: The image shows the logo for Trovita Health Science. The word "Trovita" is in large, bold, blue letters. Below that, the words "Health Science" are in smaller, gray letters. Below that, the words "Redefining Nutrition" are in a yellow box. To the right of the word "Trovita" is a cluster of orange circles.
### 6.2 PREDICATE DEVICE
The Trovita Health Science Safe-T Feed™ Ready-to-Feed Pouch with ENFit® Connector is substantially equivalent to the following devices:
- K150286 Medline Enteral Feeding Sets ●
- K152857 NeoMed NeoConnect™ Enteral Syringes with ENFit® Connector ●
### DEVICE DESCRIPTION 6.3
The Trovita Health Science Safe-T Feed™ Ready-to-Feed Pouch with ENFit® Connector consists of a foil pouch with an integrated female ENFit® connector at one end and a hanging ring at the other. A silicone valve is mounted in the inner lumen of the connector to reduce the risk of formula leakage. The outer lumen (spout) of the ENFit® connector is covered with a foil tamperevidence seal. The pouch is designed such that it can be provided pre-filled with Trovita Health Science acidified liquid nutrition formula for tube feeding.
### INDICATIONS FOR USE / INTENDED USE 6.4
The Trovita Health Science Safe-T Feed" Ready-to-Feed Pouch with ENFit® Connector is intended to deliver liquid nutrition formula to an enteral access device (a feeding tube).
#### COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH PREDICATES 6.5
The Trovita Health Science Safe-T Feed™ Ready-to-Feed Pouch with ENFit® Connector is intended to connect to male ENFit® connectors, including temporary transition sets and feeding tube administration sets. The disposable pouch is for single-patient use and is designed such that the user can connect the pouch directly to a feeding tube.
Because the device will be pre-filled with acidified liquid nutrition formula, there is no need to transfer the liquid nutrition to a syringe or feeding bag prior to feeding. The potential for spills during transfer to these intermediate devices is eliminated. The integrated valve in the connector is designed to eliminate the risk of leakage during connection to the feeding tube. The valve does not open until breached by the male ENFit® connector of the feeding tube when connected. Liquid nutrition can be administered by hanging the pouch for gravity feeding or by gently squeezing the pouch to perform a bolus feeding.
The intended use of the subject device and predicate devices are similar in that they each are single-use, disposable devices intended to provide liquid nutrition to a patient via a feeding tube. The subject device is provided non-sterile just as the Medline predicate (K150286) while the NeoMed predicate (K152857) is provided sterile. The principles of operation are the same. Each device employs a female ENFit® connector to connect directly to a male ENFit® connector on a feeding tube or extension set. The subject device differs in that the need to pre-fill the device prior to feeding is eliminated, thus reducing the risk of spills and food contamination.
{5}------------------------------------------------
Image /page/5/Picture/1 description: The image shows the logo for Trovita Health Science. The word "TROVITA" is in large, bold, blue letters. Below that, in smaller letters, are the words "HEALTH SCIENCE". Below that is a yellow rectangle with the words "REDEFINING NUTRITION" in white letters. To the right of the word "TROVITA" is a circle made of orange dots.
# 6.6 PERFORMANCE DATA
The following performance testing was conducted on the Trovita Health Science Safe-T Feed™ Ready-to-Feed Pouch with ENFit® Safety:
- Risk Analysis (FMEA) including Misconnection Assessment ●
- User Risk Analysis O
- o Risk Management Report
- ENFit® Connector Dimensional Analysis ●
- Flow Rate Analysis using commercially available enteral feed ●
- Valve Performance Testing ●
- Performance Testing per ISO 80369-3:2016 ●
- Falling Drop Positive Pressure Liquid Leakage O
- Stress Cracking O
- Resistance to Separation from Axial Load O
- Resistance to Separation from Unscrewing O
- Resistance to Overriding o
- Disconnection by Unscrewing o
- Biocompatibility Testing per ISO 10993-1:2009 ●
- Cytotoxicity o
- O Irritation / Intracutaneous Reactivity
- Sensitization o
- Human Factors and Usability Validation ●
- Packaging and Shipping Testing
### 6.7 SUBSTANTIAL EQUIVALENCE
The results of the performance testing listed above show that the subject device meets its specifications. Similar performance testing was conducted on the predicate devices. The subject device and predicate devices have similar intended use, technological characteristics, principles of operation, and treat a similar target population. This information along with the supporting documentation provided in this submission indicates that the subject device is substantially equivalent to the predicate devices.
## 6.8 CONCLUSION
The subject device and predicate devices have similar intended use and technological characteristics. The noted differences do not raise new or different questions of safety and effectiveness. Therefore, the Trovita Health Science Safe-T Feed™ Ready-to-Feed Pouch with ENFit® Connector and predicates, K150286 and K152857, are substantially equivalent.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.