← Product Code [LZO](/productcode/LZO) · K163577

# Ecofit® Hip System (K163577)

_Implantcast GmbH · LZO · Sep 25, 2017 · Orthopedic · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K163577

## Device Facts

- **Applicant:** Implantcast GmbH
- **Product Code:** [LZO](/productcode/LZO.md)
- **Decision Date:** Sep 25, 2017
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 888.3353
- **Device Class:** Class 2
- **Review Panel:** Orthopedic
- **Attributes:** Therapeutic

## Indications for Use

The EcoFit® Hip System is indicated for use as a total hip replacement in cases of: - Non-inflammatory degenerative joint disease including osteoarthritis and avascular necrosis; - Rheumatoid arthritis; - Treatment of non-union, femoral neck fracture, and trochanteric fractures of the proximal femur with head involvement that are unmanageable by other techniques; and - Revision of previously failed total hip arthroplasty. The EcoFit® Hip Stem and EcoFit® Acetabular Cup are intended for uncemented, press-fit fixation.

## Device Story

EcoFit® Hip System is a total hip replacement prosthesis. Components include femoral stems, modular Cobalt Chrome Molybdenum femoral heads, and metal-backed acetabular cups with ultra-high molecular weight polyethylene liners. Stems are collarless, straight, monoblock, flat-tapered wedge designs made of Ti6Al4V alloy with proximal plasma-sprayed commercially pure titanium (cpTi) coating. Acetabular shells are also Ti6Al4V with cpTi coating. Device is intended for uncemented, press-fit fixation. Used by orthopedic surgeons in clinical settings to replace diseased or damaged hip joints. Provides rotational stability and joint articulation. Benefits include restoration of hip function and mobility for patients with degenerative joint disease, fractures, or failed prior arthroplasty.

## Clinical Evidence

No clinical data. Bench testing only. Testing included metallic bone screw properties (ASTM F 543), liner disassembly (ASTM F 1820), taper disassembly (ASTM F 2009), impingement (ASTM F 2582), fatigue strength (ISO 7206-4, ISO 7206-6), fretting corrosion, range of motion (ISO 21535), and biocompatibility (ISO 10993). All results were acceptable or passed.

## Technological Characteristics

Materials: Ti6Al4V alloy, Cobalt Chrome Molybdenum alloy, ultra-high molecular weight polyethylene, commercially pure titanium (cpTi) plasma coating. Design: Collarless, straight, monoblock, flat-tapered wedge stem; modular acetabular cup. Fixation: Uncemented, press-fit. Standards: ASTM F 543, F 1820, F 2009, F 2582; ISO 7206-4, 7206-6, 21535, 10993.

## Regulatory Identification

A hip joint metal/ceramic/polymer semi-constrained cemented or nonporous uncemented prosthesis is a device intended to be implanted to replace a hip joint. This device limits translation and rotation in one or more planes via the geometry of its articulating surfaces. It has no linkage across-the-joint. The two-part femoral component consists of a femoral stem made of alloys to be fixed in the intramedullary canal of the femur by impaction with or without use of bone cement. The proximal end of the femoral stem is tapered with a surface that ensures positive locking with the spherical ceramic (aluminium oxide, A12 03 ) head of the femoral component. The acetabular component is made of ultra-high molecular weight polyethylene or ultra-high molecular weight polyethylene reinforced with nonporous metal alloys, and used with or without bone cement.

## Predicate Devices

- Biomet Taperloc® Complete ([K101086](/device/K101086.md), [K103755](/device/K103755.md))
- Zimmer® M/L Taper Hip Prosthesis ([K032726](/device/K032726.md))
- Biomet Taperloc® Complete Size 4mm and XR 123° ([K120030](/device/K120030.md))
- Biomet Taperloc® Complete Microplasty System ([K110400](/device/K110400.md))
- Zimmer Continuum™ and Trilogy® Integrated Taper (IT) Acetabular Systems ([K091508](/device/K091508.md))
- Aesculap Excia Total Hip System - Plasmacup SC ([K042344](/device/K042344.md)) and Plasmacup NSC ([K061699](/device/K061699.md))
- Theken Companies iNSitu Total Hip System ([K161184](/device/K161184.md))
- Smith & Nephew REFLECTION ACETABULAR COMPONENT ([K932755](/device/K932755.md), [K022556](/device/K022556.md))
- Zimmer BIOLOX® Delta Ceramic Femoral Head ([K071535](/device/K071535.md), [K130899](/device/K130899.md))
- Zimmer VerSys® Fiber Hip Prosthesis ([K061786](/device/K061786.md), [K964769](/device/K964769.md))
- Total Joint Orthopedics Klassic HD Hip System ([K143407](/device/K143407.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, with flowing lines beneath them.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

Septemeber 25, 2017

implantcast GmbH % Dave McGurl Director, Regulatory Affairs Musculoskeletal Clinical Regulatory Advisers LLC 1050 K Street NW, Suite 1000 Washington. District of Columbia 20001

Re: K163577 Trade/Device Name: Ecofit® Hip System Regulation Number: 21 CFR 888.3353 Regulation Name: Hip Joint Metal/Ceramic/Polymer Semi-Constrained Cemented Or Nonporous Uncemented Prosthesis Regulatory Class: Class II Product Code: LZO Dated: August 22, 2017 Received: August 22, 2017

Dear Mr. McGurl:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in

{1}------------------------------------------------

the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely,

# Katherine D. Kavlock -S

for Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

## Indications for Use

510(k) Number (if known)

K163577

Device Name

EcoFit® Hip System

Indications for Use (Describe)

The EcoFit® Hip System is indicated for use as a total hip replacement in cases of:

- · Non-inflammatory degenerative joint disease including osteoarthritis and avascular necrosis;
- Rheumatoid arthritis;

· Treatment of non-union, femoral neck fracture, and trochanteric fractures of the proximal femur with head involvement that are unmanageable by other techniques; and

· Revision of previously failed total hip arthroplasty

The EcoFit® Hip Stem and EcoFit® Acetabular Cup are intended for uncemented, press-fit fixation.

Type of Use (Select one or both, as applicable)X Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Druq Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement below.

{3}------------------------------------------------

## 510(k) Summary

| 510(k) SPONSOR: | implantcast GmbH<br>Lueneburger Schanze 26<br>21614 Buxtehude<br>Germany<br><br>Contact: Juliane Höppner<br>Email: j.hoeppner@implantcast.de<br>Telephone: +49 4161 744-135<br>FAX: +49 4161 744-200 |
|-----------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| TRADE NAME:     | EcoFit® Hip System                                                                                                                                                                                   |

COMMON NAME: Total Hip Replacement

| Product            | Regulation and Description | Description                                                                                   | Product Code | Device Class |
|--------------------|----------------------------|-----------------------------------------------------------------------------------------------|--------------|--------------|
| EcoFit® Hip System | 21 CFR 888.3353            | Hip joint metal/ceramic/polymer semi-constrained cemented or nonporous uncemented prosthesis. | LZO          | II           |

## PREDICATE DEVICES:

- Biomet Taperloc® Complete (K101086, K103755) .
- Zimmer® M/L Taper Hip Prosthesis (K032726) ●
- Biomet Taperloc® Complete Size 4mm and XR 123° (K120030) .
- Biomet Taperloc® Complete Microplasty System (K110400) .
- Zimmer Continuum™ and Trilogy® Integrated Taper (IT) Acetabular Systems . (K091508)
- . Aesculap Excia Total Hip System - Plasmacup SC (K042344) and Plasmacup NSC (K061699)
- . Theken Companies iNSitu Total Hip System (K161184)
- Smith & Nephew REFLECTION ACETABULAR COMPONENT (K932755, K022556) .
- Zimmer BIOLOX® Delta Ceramic Femoral Head (K071535, K130899) .
- . Zimmer VerSys® Fiber Hip Prosthesis (K061786, K964769)
- . Total Joint Orthopedics Klassic HD Hip System (K143407)

{4}------------------------------------------------

#### DEVICE DESCRIPTION:

The EcoFit® Hip System consists of EcoFit® femoral hip stems, modular Cobalt Chrome Molybdenum alloy femoral heads, and modular metal backed acetabular cups with ultra-high molecular weight polyethylene liners.

The EcoFit® femoral stem system includes three versions: the standard length stem and a Coxa Vara version. The EcoFit® femoral hip stem is collarless, straight, monoblock, flat tapered wedge design manufactured from TiAlsV4 alloy. The stem tapers from proximal to distal and is designed with a rectangular cross sectional geometry to provide rotational stability. A plasma sprayed coating of commercially pure titanium (cpTi) is applied to the proximal half of the stem. The modular femoral heads are manufactured from Cobalt Chrome Molybdenum alloy and attach to the femoral stem via a Morse style taper in a diameter of 32 mm in several neck lengths.

The EcoFit® Acetabular Cup is a two (2) piece modular cup consisting of a metal shell and a polyethylene liner. The metal shell is manufactured from TiAlsV4 alloy and a plasma sprayed coating of commercially pure titanium (cpTi) is applied to the metal shell. The polyethylene liner is manufactured from ultra-high molecular weight polyethylene and is available with an inner diameter of 32 mm for use with 32 mm outer diameter femoral head. The metal shell has a threaded insertion hole at the apex and the shell is available in two (2) versions, one with three (3) peripheral holes for the placement of bone screws as needed for adjunctive fixation and one without screw holes. A threaded apex hole plug is provided to fill the threaded insertion hole after shell placement while bone screw hole plugs are available to fill screw holes that are not used.

The EcoFit® hip stems and EcoFit® acetabular cups are intended for uncemented press-fit application.

## INTENDED USE AND INDICATIONS FOR USE:

The EcoFit® Hip System is indicated for use as a total hip replacement in cases of:

- . Non-inflammatory degenerative joint disease including osteoarthritis and avascular necrosis;
- . Rheumatoid arthritis;
- . Treatment of non-union, femoral neck fracture, and trochanteric fractures of the proximal femur with head involvement that are unmanageable by other techniques; and
- . Revision of previously failed total hip arthroplasty.

The EcoFit® Hip Stem and EcoFit® Acetabular Cup are intended for uncemented, press-fit fixation.

#### SUMMARY OF TECHNOLOGIES:

The EcoFit® Hip System is substantially equivalent to the previously cleared predicate devices based on similarities in intended use and indications, design features, materials, packaging and sterilization methods. Additionally, the EcoFit® Hip System is similar in fixation methods and comes in a similar range of sizes as the predicate devices.

#### NON-CLINICAL TESTING:

The following testing was performed to demonstrate substantial equivalency of the EcoFit® Hip System to the predicate devices. The results of the testing show that the EcoFit® Hip System is safe and effective for the proposed indications.

{5}------------------------------------------------

| Test                                                                                       | Test Method       | Results    |
|--------------------------------------------------------------------------------------------|-------------------|------------|
| Metallic Bone Screw Test<br>(Torsional Properties, Driving Torque, Pull-Out Strength)      | ASTM F 543        | Acceptable |
| Liner Disassembly Acetabular Shell - Liner<br>(Push-Out, Lever-Out, Torsionals Properties) | ASTM F 1820       | Acceptable |
| Taper disassembly<br>(Axial Disassembly Force Head - Femoral Stem)                         | ASTM F 2009       | Acceptable |
| Impingement EcoFit® Cup – Stem                                                             | ASTM F 2582       | Acceptable |
| Fatigue Strength EcoFit® Hip Stem – Distal                                                 | ISO 7206-4        | Pass       |
| Fatigue Strength EcoFit® Hip Stem - Proximal (Neck Region)                                 | ISO 7206-6        | Pass       |
| Fretting corrosion evaluation                                                              | /                 | Acceptable |
| ROM EcoFit® Hip Stem – EcoFit® Cup                                                         | ISO 21535 Annex A | Acceptable |
| Biocompatibility                                                                           | ISO 10993         | Pass       |

## CLINICAL TESTING:

Animal and clinical testing was not necessary to determine substantial equivalence between the EcoFit® Hip System and the predicate devices.

## CONCLUSIONS:

Based on the intended use, materials, design and testing provided, the 510(k) demonstrates substantial equivalence to the predicate devices cited in this summary.

---

**Source:** [https://fda.innolitics.com/device/K163577](https://fda.innolitics.com/device/K163577)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
