The Osstem Implant is indicated for use in partially or fully edentulous mandibles and maxillae, in support of single or multiple-unit restorations including; cemented retained, or overdenture restorations, and final or temporary abutment support for fixed bridgework. It is intended for delayed loading. Ultra Wide Fixture System is intended to be used in the molar region.
Device Story
SS SA Fixture is a titanium dental implant surgically placed into the jaw bone to support prosthetic restorations. It features a sandblasted and acid-etched (SA) surface and an internal octagonal connection. The device is supplied sterile and is intended for delayed loading in the molar region. It is used by dental professionals in a clinical setting. The implant provides a stable foundation for crowns, bridges, or overdentures, restoring masticatory function and aesthetics for edentulous patients. The healthcare provider selects the appropriate fixture size based on the patient's anatomical needs and performs the surgical placement. The device's mechanical stability supports the long-term integration of the prosthetic restoration.
Clinical Evidence
No clinical studies were submitted. Substantial equivalence is supported by bench testing, including biocompatibility, sterilization validation, surface characterization, Bone to Implant Contact (BIC) analysis, surface area analysis, pullout testing, and fatigue testing conducted per ISO 14801.
Indicated for partially or fully edentulous patients requiring single or multiple-unit restorations, including cemented, screw-retained, or overdenture restorations, and abutment support for fixed bridgework. Intended for delayed loading. Ultra Wide Fixture System is indicated for use in the molar region.
Regulatory Classification
Identification
An endosseous dental implant is a prescription device made of a material such as titanium or titanium alloy that is intended to be surgically placed in the bone of the upper or lower jaw arches to provide support for prosthetic devices, such as artificial teeth, in order to restore a patient's chewing function.
Special Controls
*Classification.* (1) Class II (special controls). The device is classified as class II if it is a root-form endosseous dental implant. The root-form endosseous dental implant is characterized by four geometrically distinct types: Basket, screw, solid cylinder, and hollow cylinder. The guidance document entitled “Class II Special Controls Guidance Document: Root-Form Endosseous Dental Implants and Endosseous Dental Implant Abutments” will serve as the special control. (See § 872.1(e) for the availability of this guidance document.)(2)
*Classification.* Class II (special controls). The device is classified as class II if it is a blade-form endosseous dental implant. The special controls for this device are:(i) The design characteristics of the device must ensure that the geometry and material composition are consistent with the intended use;
(ii) Mechanical performance (fatigue) testing under simulated physiological conditions to demonstrate maximum load (endurance limit) when the device is subjected to compressive and shear loads;
(iii) Corrosion testing under simulated physiological conditions to demonstrate corrosion potential of each metal or alloy, couple potential for an assembled dissimilar metal implant system, and corrosion rate for an assembled dissimilar metal implant system;
(iv) The device must be demonstrated to be biocompatible;
(v) Sterility testing must demonstrate the sterility of the device;
(vi) Performance testing to evaluate the compatibility of the device in a magnetic resonance (MR) environment;
(vii) Labeling must include a clear description of the technological features, how the device should be used in patients, detailed surgical protocol and restoration procedures, relevant precautions and warnings based on the clinical use of the device, and qualifications and training requirements for device users including technicians and clinicians;
(viii) Patient labeling must contain a description of how the device works, how the device is placed, how the patient needs to care for the implant, possible adverse events and how to report any complications; and
(ix) Documented clinical experience must demonstrate safe and effective use and capture any adverse events observed during clinical use.
{1}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized image of a caduceus, which is a symbol often associated with healthcare and medicine. The caduceus is depicted with three intertwined strands and a wing-like shape at the top.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
May 15, 2017
Osstem Implant Co., Ltd. % David Kim Manager Hiossen Inc. 85 Ben Fairless Dr. Fairless Hills, Pennsylvania 19030
Re: K163557
Trade/Device Name: SS SA Fixture Regulation Number: 21 CFR 872.3640 Regulation Name: Endosseous Dental Implant Regulatory Class: Class II Product Code: DZE, NHA Dated: April 14, 2017 Received: April 14, 2017
Dear David Kim:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in
{2}------------------------------------------------
the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely.
Michael J. Ryan -S
for Tina Kiang, Ph.D. Acting Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
## Indications for Use
510(k) Number (if known) K163557
Device Name SS SA Fixture
Indications for Use (Describe)
The Osstem Implant is indicated for use in partially or fully edentulous mandibles and maxillae, in support of single or multiple-unit restorations including; cemented retained, or overdenture restorations, and final or temporary abutment support for fixed bridgework. It is intended for delayed loading. Ultra Wide Fixture System is intended to be used in the molar region.
Type of Use (Select one or both, as applicable)
| Prescription Use (Part 21 CFR 801 Subpart D) | Over-The-Counter Use (21 CFR 801 Subpart C) |
|----------------------------------------------|---------------------------------------------|
|----------------------------------------------|---------------------------------------------|
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
# 510(k) Summary
Date: May 11, 2017
#### 1. Company and Correspondent making the submission:
| - Submitter's Name: | OSSTEM Implant Co., Ltd. |
|-------------------------|-------------------------------------------------------------------------|
| - Address: | 66-16, Bansong-ro 513beon-gil, Haeundae-gu,<br>Busan, Republic of Korea |
| - Contact: | Mr. Hee Kwon Son |
| - Phone: | +82 51 850 2575 |
| - Correspondent's Name: | HIOSSEN Inc. |
| - Address: | 85 Ben Fairless Dr. Fairless Hills, PA 19030 |
| - Contact: | DAVID KIM |
| - Phone: | 267 759 7031 |
- 2. Device:
Trade or (Proprietary) Name : SS SA Fixture Common or usual name : Dental Implant Classification Name : Endosseous Dental Implant Regulation Number : 21CFR872.3640 Device Classification: Class II Primary Product Code: DZE Subsequent Product Code: NHA
- 3. Predicate Device:
Substantial equivalence is claimed to the following devices: Primary Predicate K120847, ET/SS Implant System, OSSTEM Implant Co., Ltd.
Reference predicate K161604, OSSTEM Implant System, OSSTEM Implant Co., Ltd.
### 4. Description:
The SS SA Fixture is a dental implant made of titanium metal intended to be surgically placed in the bone of the upper or lower jaw arches.
Fixture is made of pure titanium metal and supplied sterile. The surface is SA, Sandblasting and Acid etching, treated.
{5}------------------------------------------------
Image /page/4/Picture/1 description: The image shows the logo and contact information for OSSTEM Implant Co., Ltd. The logo is on the left side of the image, and the company name is in large, bold letters. The address is 66-16, Bansong-ro 513beon-gil, Haeundae-gu, Busan, Republic of Korea. The telephone number is +82 51 850-2500, the fax number is +82 51 850-4341, and the website address is www.osstem.com.
The SS SA Fixture is similar to other commercially available products based on the intended use, the technology used, the claims, the material composition employed and performance characteristics.
#### 1) SSIII SA Ultra-Wide Fixture
| Device<br>Description | Intended to be surgically placed in the bone of the upper or lower jaw<br>arches. Fixture is supplied sterile. Only straight type abutments are to<br>be used with the Ultra-Wide Fixture System. | |
|-----------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------|
| Material | Pure Titanium (ASTM F 67) | |
| Surface | SA surface treatment | |
| | Diameter (mm) | Length (mm) |
| Dimension | 5.96, | 6, 8.5 |
| | 6.0 | 7 |
| | 5.95 | 10 |
| | 5.92 | 11.5, 13 |
| | 6.93 | 6 |
| | 6.8 | 7, 8.5, 10, 11.5, 13 |
## - Substantial Equivalence Matrix
| | SS SA Fixture<br>(SSIII SA Ultra Wide<br>Fixture) | Primary Predicate<br>ET/SS Implant<br>System<br>(SSII/III SA Fixture) | Reference predicate<br>OSSTEM Implant<br>System | |
|------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------|
| 510(K) No. | New Device | K120847 | K161604 | |
| Manufacturer | OSSTEM Implant<br>Co., Ltd. | OSSTEM Implant<br>Co., Ltd. | OSSTEM Implant<br>Co., Ltd. | Same |
| Indication for<br>use | The Osstem Implant is<br>indicated for use in<br>partially or fully<br>edentulous mandibles<br>and maxillae, in support<br>of single or multiple-<br>unit restorations<br>including; cemented<br>retained, screw retained,<br>or overdenture<br>restorations, and final or<br>temporary abutment<br>support for fixed<br>bridgework. It is<br>intended for delayed<br>loading. Ultra Wide<br>Fixture System is<br>intended to be used in<br>the molar region. | ET/SS Implant System<br>is indicated for use in<br>partially or fully<br>edentulous mandibles<br>and maxillae, in support<br>of single or multiple-<br>unit restorations<br>including; cemented<br>retained, screw retained,<br>or overdenture<br>restorations, and final or<br>temporary abutment<br>support for fixed<br>bridgework. It is<br>intended for delayed<br>loading. The abutment<br>is intended for use with<br>a dental implant fixture<br>to provide support for<br>prosthetic restorations | The Osstem Implant<br>System is indicated for<br>use in partially or fully<br>edentulous mandibles<br>and maxillae, in support<br>of single or multiple-<br>units restorations<br>including; cemented<br>retained, screw retained,<br>or overdenture<br>restorations, and final or<br>temporary abutment<br>support for fixed<br>bridgework. It is<br>intended for delayed<br>loading. Ultra wide<br>Fixture System is<br>intended to be used in<br>the molar region.<br>Products with diameter | Same |
| | | such as crowns, bridges,<br>or overdenture. | of less than 3.25mm<br>should be used<br>exclusively for the<br>lateral incisor in the<br>maxilla and a central or<br>lateral incisor in the<br>mandible | |
| Surgery type | One stage Surgery | One stage Surgery | One or two stage<br>Surgery | Same |
| Structure | - Non Submerged<br>Fixture<br>- Self tapping<br>- Internal Octagonal<br>connection<br>- Taper Body | - Non Submerged<br>Fixture<br>- Self tapping<br>- Internal Octagonal<br>connection<br>- Straight/Taper body<br>shape | - Internal Hex-<br>connected<br>- Submerged Fixture<br>- Straight/Taper body<br>shape | Same |
| Platform<br>(D) (mm) | 6 | 4.8~6.0 | N/A | Included in<br>predicate |
| Body<br>Diameter<br>(D) (mm) | 5.92, 5.95, 5.96, 6.0,<br>6.93, 6.8 | 4.1~4.9/3.75~5.0 | 3.2, 3.5, 3.75, 3.77, 4.2,<br>4.25, 4.4, 4.45, 4.6,<br>4.63, 4.65, 4.8, 4.9,<br>5.05, 5.08, 5.1, 5.25,<br>6.2, 7.1 | Included in<br>Reference<br>predicate |
| Length (mm) | 6, 7, 8.5, 10, 11.5, 13 | 6~15.0 | 6.2, 7.0, 8.5, 10.0, 11.5,<br>13.0, 15.0, 18.0 | Included in<br>predicate |
| Material of<br>Fixture | Pure Titanium Grade 4<br>(ASTM F67) | Pure Titanium Grade 4<br>(ASTM F67) | Pure Titanium Grade 4<br>(ASTM F67) | Same |
| Surface | SA | SA | SA | Same |
| Sterilization | Radiation Sterile | Radiation Sterile | Radiation Sterile | Same |
| Shelf life | 8years | 5years | 8years | Same with<br>Reference<br>predicate |
| S E | Same<br>SS SA Fixture have same connection structure, material, manufacture process (including<br>surface treatment), principles of operation, Technological Characteristics and indication for use<br>with ET/SS Implant System (SSII/III SA Fixture)<br>And length of SS SA Fixture is also included in primary predicate ET/SS Implant System<br>(SSII/III SA Fixture)<br>Different<br>Diameter of SS SA fixture is bigger than primary predicate ET/SS Implant System (SSII/III SA<br>Fixture) but Diameter of SS SA fixture is included in range of reference predicate diameter,<br>OSSTEM Implant System.<br>Therefore SSIII SA Ultra Wide Fixture of SS SA Fixture is substantially equivalent to the<br>predicate devices | | | |
Letter(8.5 X 11in)
{6}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the logo for Osstem Implant. The logo is orange and gray. The word "OSSTEM" is in orange, and the word "IMPLANT" is in gray. There is a small superscript "6" next to the word "OSSTEM".
OSSTEM Implant Co., Ltd.
Section 007
66-16, Bansong-ro 513beon-gil, Haeundae-gu, Busan, Republic of Korea Tel: +82 51 850-2500 Fax: +82 51 850-4341 www.osstem.com
## 5. Indication for use:
The Osstem Implant is indicated for use in partially or fully edentulous mandibles and maxillae, in support of single or multiple-unit restorations including; cemented retained, screw retained, or overdenture restorations, and final or temporary abutment support for
{7}------------------------------------------------
Image /page/6/Picture/1 description: The image shows the logo and contact information for OSSTEM Implant Co., Ltd. The address is 66-16, Bansong-ro 513beon-gil, Haeundae-gu, Busan, Republic of Korea. The phone number is +82 51 850-2500, the fax number is +82 51 850-4341, and the website is www.osstem.com.
fixed bridgework. It is intended for delayed loading. Ultra Wide Fixture System is intended to be used in the molar region.
#### 6. Summary of nonclinical testing
The following nonclinical testing data were provided or relied upon in support of the Substantial equivalence determination.
#### Biocompatibility
Biocompatibility testing on the SS SA Fixture was previously conducted. As the material of construction and manufacturing processes are the same as the predicate device, ET/SS Implant System, OSSTEM Implant Co., Ltd., K120847 and OSSTEM Implant System, OSSTEM Implant Co., Ltd., K161604.
Therefore, no additional testing is required to support the biological safety of the subject devices.
#### Sterilization validation
Validation of the gamma irradiation process was previously conducted for the predicate devices, OSSTEM Implant System, OSSTEM Implant Co., Ltd., K161604. There has been no change to the manufacturing, sterilization processes since then; therefore, additional validation is not required.
#### Surface treatment characterization testing
SS SA Fixture has SA (Sandblasted and Acid etched surface) surface treatment that is exactly same with the predicate devices, ET/SS Implant System, OSSTEM Implant Co., Ltd., K120847 and OSSTEM Implant System, OSSTEM Implant Co., Ltd., K161604. There has been no change to the manufacturing or surface treatment processes since then; therefore, additional characterization testing is not required.
#### Bone to Implant Contact (BIC) analysis
BIC analysis is conducted with SS SA Fixture containing implants of lengths less than 7mm. We compared the BIC value and confirmed the substantial equivalence status with the predicate device; ET/SS implant System, OSSTEM Implant Co., Ltd., K120847.
#### Surface area analysis
Surface area analysis is conducted with SS SA Fixture containing implants of lengths less than 7mm. We compared the surface area and confirmed the substantial equivalence status with the predicate device; ET/SS implant System, OSSTEM Implant Co., Ltd., K120847.
#### Pullout test
Pullout force test was conducted with SS SA Fixture containing implants of lengths less than 7mm. We compared the pullout force value and confirmed the substantial equivalence status with the predicate device; ET/SS implant System, OSSTEM Implant Co., Ltd., K120847.
{8}------------------------------------------------
Image /page/7/Picture/1 description: The image contains two logos. The first logo on the left says "OSSTEM IMPLANT" with "OSSTEM" in orange and "IMPLANT" in gray. The second logo on the right says "OS" in large black font, followed by the number "66-16."
## Fatigue test
The Fatigue testing was considered according to the "Guidance for industry and FDA staff Class II Special Controls Guidance Document Root-form Endosseous Dental Implants and Endosseous Dental Abutment" and ISO 14801 Dentistry - Fatigue test for endosseous dental implants with the worst case scenario. SS SA Fixture has bigger diameter than the predicate device, ET/SS Implant System, OSSTEM Implant Co., Ltd., K120847. The result of SS SA Fixture static and dynamic compression - bending test will be higher than this predicate; therefore, we do not consider additional fatigue test.
- 7. Summary of clinical testing No clinical studies are submitted.
- 8. Conclusions
In accordance with the Federal Food, Drug and Cosmetic Act, 21 CFR Part 807, and based on the information provided in this premarket notification Osstem Implant Co., Ltd. concludes that the SS SA Fixture is substantially equivalent to the predicate devices as described herein.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.